K041294 · Cardiodynamics International Corp. · DSB · Dec 8, 2004 · Cardiovascular
Device Facts
Record ID
K041294
Device Name
BIOZDX HEMODYNAMIC MONITOR
Applicant
Cardiodynamics International Corp.
Product Code
DSB · Cardiovascular
Decision Date
Dec 8, 2004
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 870.2770
Device Class
Class 2
Indications for Use
The BioZDx Hemodynamic Monitor is intended to monitor and display a patient's hemodynamic parameters. These parameters include: ECG Cardiac Output Thoracic Fluid Content Left Vent. Ejection Time End Diastolic Volume Systemic Vascular Resistance PressureLeft Cardiac Work Diastolic Blood Pressure Pre-Ejection Period Heart Rate Acceleration Index Index of Contractility Mean Blood Pressure Systolic Time Ratio End diastolic Index Cardiac Index Stroke Volume Systolic Blood
Device Story
BioZDx Hemodynamic Monitor noninvasively measures hemodynamic parameters via Impedance Cardiography (ICG). Input: 8 electrodes (two dual electrodes per side of neck/thorax) inject minimal current; device reads returning voltage waveform. Processing: Proprietary DSP circuitry and software calculate hemodynamic parameters using patient inputs (gender, body frame, height, weight, age, blood pressure). Output: Hemodynamic data displayed on internal VGA screen and printed reports. Used in hospital, outpatient, and clinical settings by healthcare professionals. Replaces external simulator with built-in version; features improved patient cable with enhanced defibrillation protection and identification chips. Benefits: Provides noninvasive hemodynamic monitoring to assist clinical decision-making.
Clinical Evidence
Bench testing only. Validation included system, hardware, and signal processing testing. No clinical data presented.
Technological Characteristics
Impedance Cardiography (ICG) device. Intel 586-equivalent processor; Windows CE OS. DSP firmware (ZMARC+). Connectivity: USB serial link between PIM and computer. Integrated VGA display and printer. Patient interface: 8-electrode configuration. Defibrillation protection integrated into patient cable yoke. Software integrity checks implemented.
Indications for Use
Indicated for monitoring and displaying hemodynamic parameters in patients. Parameters include cardiac output, thoracic fluid content, systemic vascular resistance, blood pressure, and various cardiac indices. Prescription use only.
Regulatory Classification
Identification
An impedance plethysmograph is a device used to estimate peripheral blood flow by measuring electrical impedance changes in a region of the body such as the arms and legs.
Special Controls
*Classification.* Class II (special controls). The device, when it is a body composition analyzer which is not intended to diagnose or treat any medical condition, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 870.9.
{0}------------------------------------------------
# K041294
# DEC - 8 2004
### Section E - 510(k) Statement and Summary
#### Premarket Notification 510(k) Summary
#### SUBSTANTIAL EQUIVALENCE:
Identification of predicate devices, model, and manufacturer:
| Predicate device 1: | CardioDynamics BioZ.pc |
|--------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Model: | Part # BZ-500/501 |
| Manufacturer: | CardioDynamics International Corporation |
| Predicate Device 510(k): | K001081 on 6/31/2000 |
| Reason for Submission: | Repackaging and porting of functionality from an off-the-shelf PC platform to a Philips cardiograph for the user display and interface. Use of BioZDx ported software and change of operating system from Windows™ to Windows CETM. Improvements to patient cable including enhanced defib protection, and removable/replaceable left and right interconnect assemblies. Built-in rather than externally attached printer. |
| Predicate device 2: | CardioDynamics BioZ.com |
| Model: | Part # BZ-4110 |
| Manufacturer: | CardioDynamics International Corporation |
| Predicate Device 510(k): | K974725 on 3/18/1998 |
| Reason for Submission: | Repackaging and porting of functionality from an embedded PC platform to a Philips cardiograph for the user display and interface. Use of BioZDx ported software and change of operating system from Microsoft DOS 6.22 to Windows CETM. Improvements to patient cable including enhanced defib protection, and removable/replaceable left and right interconnect assemblies.. Built-in rather than externally attached printer. |
The BioZDx Hemodynamic Monitor is substantially equivalent to its predicate devices, the BioZ.pc System and the BioZ.com System currently marketed by CardioDynamics International Corporation. The justification for this substantial equivalence determination is presented below.
The BioZDx Hemodynamic Monitor is substantially equivalent to the BioZ.pc System in terms of design, intended use and principles of operation. The BioZDx Hemodynamic Monitor simply re-packages and repartitions the electronics of the product for the convenience of users who wish to have an integrated solution including a built-in printer which provides reports on standard ECG 8-1/2 x 11 grided paper.
The BioZDx has nearly identical DSP software structure to the predicate devices, with enhanced fiducial point detection techniques over those used in the predicate devices, and the addition of a statistical based modulation factor for dZ/dTmax rather than the empirical technique employed in the predicate devices. There were no unexpected test results due to these changes and the validation of these changes is included in Section J.
{1}------------------------------------------------
The user software has been simplified compared to either of the predicate devices and ported from DOS 6.22 running under Microsoft Windows to Microsoft Windows CE to take advantage of the features of the new user platform.
The new user platform is the hardware and core operating system software of the Philips PageWriter Trim Cardiograph, 510/k) K031422 on 7/3/2003, less the ECC-specific hardware and ECG application software. The Philips ECC-specific hardware and ECG application software have been replaced with their equivalent ICG counterparts.
The User Software which provides the user interface for the BioZDx system is loaded onto the PC using a software installation kit, similar to the method used for the predicate device. The predicate device used a floppy disk or an attached notebook PC for software installation. The BioZDx uses a pre-programmed PCMCIA flash memory card.. Both devices store the user software on the internal storage of the user platform component of the system. Both products use an identical software integrity check when the monitoring software is first activated, to insure that no corruption has occurred of any of the operating software used by the device.
The BioZDx provides a printed report as do the two predicate devices. In the BioZDx, the printer is internal to the device rather than attached externally. The report in the BioZDx contains the same information as the predicate devices, but is reformatted to standard grid ECG paper. Both the predicate devices and theBioZDx use the HPGL graphic language for formatting printouts and communication with the print engine.
The BioZDx product also includes an improved patient cable, with enhanced defib protection and removable/replaceable left and right distal interconnect sections. The defib protection components are identical to those used in the predicate devices, but they are moved to the patient cable yoke and closer to the patient. The distal patient cable sections also include identification chips to prevent users from operating the device with unapproved and unvalidated cable systems. All final product testing was performed with prototype versions of the improved patient cable.
CardioDynamics International Corporation has labeled the version of BioZDx software clearly on the installation kit flash memory card. The installation instructions in the kit, and the user manual instructions for use clearly specify the kit only for use with the BioZDx product.
Both BioZ.pc and the BioZ.com predicate devices as well as the new BioZDx systems are portable in design and for use in the hospital, outpatient and clinical settings. The intended use of the BioZDx (identical to the predicate device) is to noninvasively measure a patient's hemodynamic parameters using Impedance Cardiography (ICG). Monitoring is accomplished by attaching 8 electrodes to the patient (two dual electrodes on each side of the neck and thorax), injecting a minimal current through the upper electrodes, and reading the returning voltage waveform from the inner electrodes.
The BioZDx Hemodynamic Monitor utilizes CardioDynamics' proprietary DSP electronic circuitry and software incorporating formulas and algorithms to calculate the various hemodynamic parameters. The user inputs patient parameters into the user software, including patient gender, body frame size, height, weight, age and blood pressure. The Monitor then utilizes these parameters and measures the ICG signals to determine the hemodynamic properties of that particular patient interface circuity (which provides all data acquisition, processing, and isolation) is similar to the predicate devices, and tested to the same internal and Agency specifications All changes to the new product from the predicate device are verified and validated within the body of hardware and software testing. The validation reports are included in Section J for System, Hardware, and Signal Processing, and Section M for software.
{2}------------------------------------------------
K041294
The ICG electrodes (BioZTect Sensor) to be used with the BioZDx device are sold separately, and are identical to those marketed for use with both the BioZ.com predicate device products. The BioZTect Sensor received 510(k) Clearance (K011797) on 7-3-2001.
The predicate devices (the BioZ.com) and the new BioZDx Hemodynamic Monitor are IBM PC Family (Intel Pentium-class core processor-based) products, which differ primarily in partitioning and packaging. Both utilize Microsoft operating systems, either alone or within a Windows or Windows CE environment.
The predicate devices required the use of an external simulator device (the CardioDynamics BZ-4525 BioZ.sim device) to verify system performance and test patient cables. For user convenience, this function is built in to the BioZDx device.
Substantial equivalence is shown in the following table (on the next page):
{3}------------------------------------------------
.
| Attribute | BioZDx (New) | BioZ.pc (Predicate) | BioZ.com (Predicate) |
|----------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------|
| Packaging | Separate computer and<br>Patient Interface Module<br>(PIM) | Separate computer and<br>Patient Interface Module<br>(PIM) | Self-contained instrument<br>(computer and Patient<br>Interface Electronics) |
| Pt. Interface Circuitry<br>and sensors | Same as BioZ.com and<br>BioZ.pc with addition of<br>continuous lead fail. All<br>accessories except patient<br>cable same as BioZ.com | Same as new device with<br>exception of continuous<br>lead fail. All accessories<br>except patient cable same<br>as new device | Same as new device with<br>exception of continuous<br>lead fail. All accessories<br>except patient cable same<br>as new device |
| Defib Protection | Moved to yoke of patient<br>cable | Internal to BioZ.pc<br>Instrument | Internal to BioZ.com<br>Instrument |
| CPU | Intel 586 equivalent PC | Intel 586 or Equivalent PC | Intel 386EX PC |
| CPU Packaging | Separate from patient<br>interface circuitry,<br>connected with serial link | Separate from patient<br>interface circuitry,<br>connected with serial link | Internal to BioZ.com<br>Instrument |
| CPU Communications | USB Serial | RS323 Serial (External) | RS232 Serial (Internal) |
| PC Operating System | Windows CE® | Windows 98 or NT® | DOS 6.22 |
| PC Software<br>Installation Kit | Manufactured and<br>supplied per CDIC<br>Manufacturing Procedure<br>02-121 | Manufactured and<br>supplied per CDIC<br>Manufacturing Procedure<br>02-121 | Manufactured and<br>supplied per CDIC<br>Manufacturing Procedure<br>02-121 |
| Printer | Internal to Cardiograph<br>HPGL Protocol | Externally connected<br>HPGL Protocol | Externally connected<br>HPGL Protocol |
| User Interface<br>software | BioZDx ver V1.0x9 | BioZ.pc V1.52 | BioZ.com V2.26 |
| User Display | Internal VGA Screen | External PC VGA Screen | BioZ.com internal $\frac{1}{4}$<br>VGA screen |
| Blood Pressure<br>Electronics | Internal to BioZ.dx PIM<br>Suntech Module | Internal to BioZ.pc PIM<br>Suntech Module | Internal to instrument<br>CAS or Suntech Module |
| Pulse Oximeter<br>Electronics | Discontinued as a product<br>feature | Internal to PIM | Externally connected to<br>instrument via serial data<br>cable |
| DSP Packaging | Internal to PIM | Internal to PIM | Internal to BioZ.com<br>Instrument |
| DSP Firmware | CDIC ZMARC+ Ver<br>V00.04.16.10.09 | CDIC ZMARC Ver1.09 | CDIC ZMARC Ver 2.15 |
| Patient Simulator | Built-In for user<br>convenience | Separate Device (BZ-<br>4525) | Separate Device (BZ-<br>4525) |
:
### Table 1 Device Equivalency
{4}------------------------------------------------
#### Risk Analysis and Design Controls
A Risk Analysis assessment, performed on the BioZDx, revealed that the repartitioning of the circuitry present absolutely no additional safety risks. The System Hazard Analysis is included in Section J. The software Hazard Analysis is included in Section M.
CardioDynamics International Corporation is certified to EN13485, and audited by TUV Rhineland. No significant audit findings have been presented since EN14001 certification in May of 1998. All products designed and manufactured under CDIC's Quality System, which includes Design Control and Manufacturing Practices.
OSP-107 defines the following activities and deliverables related to all design activities:
- Design Input: Specific requirements for a design project, including any applicable statutory and . regulatory requirements.
- Design Output: The results of a design effort. The design output includes 1) the detailed . documentation covering all functional and performance characteristics of the device, including those related to safety, 2) the product and process documents needed to produce the device and maintain it throughout its life cycle.
- Design Verification: Confirmation that the design output meets the design input requirements. .
- Design Validation: The establishment, by objective evidence, that a device conforms to the users' . needs and intended uses.
- Design Transfer: The process of transferring knowledge (specifications) about the device from the . design group to the manufacturing group.
It is the responsibility of all members of the Engineering Department to ensure that the requirements of OSP-107 are met for all product designs and that each required deliverable is created and approved.
It is the responsibility of the representatives of the Marketing, Quality Assurance, Manufacturing and Executive departments to review, understand and approve each applicable deliverable of the Design Control process.
The result of following QSP-107 is a Device Design Record that includes design documentation associated with the design of the product. The completed Device Design Record provides a design history of the finished device and demonstrates the device design was developed, verified and validated in accordance to OSP-107. The accessories developed specifically as part of the BioZDx Hemodynamic Monitor are included in its Device Design Record.
Design Review is required, at a minimum, at the completion of each step in the design control process (i.e. Design Input, Design Output, Design Verification and Design Transfer). The design review process identifies the required deliverables for each step in the process and verifies the content and accuracy of each deliverable. The Design Control Form, OSDP-101-1, contains a list of the deliverables for each step, with the approved revision number. The signatures on both the Design Control Form and each of the approved documents constitute evidence of the design review process.
Design inputs provide the foundation for product development. All Design Inputs are as comprehensive and precise as possible without imposing design solutions. Upon completion of the design review activity
{5}------------------------------------------------
for the Design Input step, with agreement that the Design Inputs were correct and acceptable, the Design Input section of the Design Control Form is completed with all required signatures.
The Design Output deliverables are to be created by appropriate personnel to meet the specific Design Inout requirements. The Design Output deliverables demonstrate that the Design Input requirements have been met. These deliverables are listed in the Design Output section of the Design Control Form. Upon the completion of the design review activity for the Design Output step, with agreement that the listed Design Output deliverables were determined to be correct and acceptable, the Design Output section of the Design Control Form is completed with all required signatures.
Design Verification is performed by the appropriate personnel to verify the Design Outputs fulfill the Design Input requirements. The Design Verification deliverables indicate that the Design Outputs meet the specified Design Input requirements listed in the Design Verification section of the Design Control Form.
Design Validation is performed by the appropriate personnel to verify that the Design Outputs fulfills the users' needs and intended uses. The Design Validation documents indicate the Design Outputs meet the users' needs and intended uses, and are listed in the Design Validation section of the Design Control Form. Upon the completion of the design review activity for the Design Validation step, with agreement that the listed Design Validation deliverables are determined to be correct and acceptable, the Design Validation section of the Design Control Form is completed with all required signatures.
Design Transfer activity is performed by appropriate personnel to verify that the Design Output and Design Transfer deliverables are sufficient to produce a product that meets the Design Input requirements. The Design Transfer deliverables include manufacturing instruction and documentation, in addition to the Design Output deliverables that are sufficient to produce a device that meets the Design Input requirements and shall be listed in the Design Transfer section of the Design Control Form. Upon completion of the design review activity for the Design Transfer step, with agreement that the listed Design Transfer deliverables are determined to be correct and acceptable, the Design Transfer section of the Design Control Form is completed with all required signatures.
The design process can be considered complete when all sections of each Design Control Form created during the project are signed. Upon completion of the design output documents are submitted for Document Change Order (DCO).
Configuration Control and Quality Assurance are provided under CDIC Quality System Procedures. These procedures specify that all products must be DCO control, that all components used in product assembly have incoming inspection, and that all product is built only under controlled manufacturing documentation, manufacturing instruction and test procedures. Internal audits are performed to insure compliance with these procedures.
{6}------------------------------------------------
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
DEC - 8 2004
Cardiodynamics International Corp. c/o Mr. Dennis G. Hepp Chief Technology Officer 6175 Nancy Ridge Drive, #300 San Diego CA 92121
Re: K041294
Trade Name: BioZDx Hemodynamic Monitor Regulation Number: 21 CFR 870.2770 Regulation Name: Impedance Plethysmograph Regulatory Class: II (two) Product Code: DSB Dated: May 13, 2004 Received: May 14, 2004
Dear Mr. Hepp:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration. Iisting of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{7}------------------------------------------------
Page 2 - Mr. Dennis G. Hepp
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (301) 594-4648. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html
Sincerely yours.
B.Zimmerman fr
Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{8}------------------------------------------------
## Indications for Use
510(k) Number (if known):_K041294
Device Name: BioZDx HemoDynamic Monitor
Indications For Use:
The BioZDx Hemodynamic Monitor is intended to monitor and display a patient's hemodynamic parameters. These parameters include:
ECG Cardiac Output Thoracic Fluid Content Left Vent. Ejection Time End Diastolic Volume Systemic Vascular Resistance PressureLeft Cardiac Work Diastolic Blood Pressure
Pre-Ejection Period Heart Rate Acceleration Index Index of Contractility Mean Blood Pressure Systolic Time Ratio End diastolic Index Cardiac Index
Stroke Volume Systolic Blood
Prescription Use _XXX_ (Part 21 CFR 801 Subpart D)
AND/OR
Over-The-Counter Use (21 CFR 807 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
B. Simmons
n Sign-Off) cular Devices Division of Cardiov, 510(k) Number
Page 1 of ____________________________________________________________________________________________________________________________________________________________________
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.