MYTECH/HAPPY LIFE BLOOD PRESSURE MONITOR, MODEL HPL-200
Applicant
Mytech Technology Co., Ltd.
Product Code
DXN · Cardiovascular
Decision Date
May 17, 2004
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 870.1130
Device Class
Class 2
Indications for Use
The device is a noninvasive blood pressure measurement system intended to measure the systolic and diastolic blood pressures and pulse rate of an adult individual, over age 18, at home by using a non-invasive technique in which an inflatable cuff is wrapped around the wrist. The cuff circumference is limited to be 5.25" – 7.75".
Device Story
Wrist-worn non-invasive blood pressure monitor; measures systolic/diastolic blood pressure and pulse rate. Operates via inflatable cuff wrapped around wrist; intended for home use by adult patients. Device features LCD display for output; provides automated readings to user. Substantially equivalent to HPL-100; shares core technological characteristics, power supply, and measurement accuracy. Safety and EMC testing performed per EN 60601-1 and EN 60601-1-2 standards.
Clinical Evidence
Bench testing only. Device performance evaluated against AAMI/ANSI SP10 standards. Electric safety and EMC testing conducted per EN 60601-1 and EN 60601-1-2. No clinical trial data provided.
Technological Characteristics
Non-invasive wrist-cuff blood pressure monitor. Features LCD display, internal memory, and battery power supply. Cuff circumference range: 5.25"–7.75". Complies with EN 60601-1 (safety) and EN 60601-1-2 (EMC).
Indications for Use
Indicated for adult individuals over age 18 for non-invasive measurement of systolic/diastolic blood pressure and pulse rate at home. Cuff circumference limited to 5.25"–7.75".
Regulatory Classification
Identification
A noninvasive blood pressure measurement system is a device that provides a signal from which systolic, diastolic, mean, or any combination of the three pressures can be derived through the use of tranducers placed on the surface of the body.
{0}------------------------------------------------
#### MYTECH TECHNOLOGY CO., LTD
MAY 1 7 2004
5F.6, Alley 2, Lane 222, Lien Cheng Road, Chung Ho City, Taipei Hsien, 235, Taiwan, ROC Telephone: 886-2-2247 4816 Fax: 886-2-2247 7024 Email: sale@mytech.com.tw
### "510(k) Summary"
| Submitter's Name: | MYTECH TECHNOLOGY CO., LTD. |
|-------------------------------|--------------------------------------------------------------------------------------------|
| Address: | 5F.6, Alley 2, Lane 222, Lien Cheng Road,<br>Chung Ho City, Taipei Hsien, 235, Taiwan, ROC |
| Telephone: | 886-2-2247 4816 |
| FAX: | 886-2-2247 7024 |
| Contact Person: | Dr. Jen, Ke-Min |
| Date Summary Prepared: | 5/7/2004 |
| Proprietary Name: | MYTECH WRIST BLOOD PRESSURE<br>MONITOR, HPL-200 |
| Common Name: | BLOOD PRESSURE MONITOR |
| Classification Name: | NON-INVASIVE BLOOD-PRESSURE<br>MEASUREMENT SYSTEM<br>(per 21CFR section 870.1130) |
| Device Class: | Class II (performance standards) |
| Specialty: | CARDIOVASCULAR |
| Product code: | DXN |
| Legally Marketed (Predicate): | MYTECH WRIST BLOOD PRESSURE<br>MONITOR, HPL-100 |
| Device: | 510(k) No. K030561 |
510(k) No: K030561
{1}------------------------------------------------
#### MYTECH TECHNOLOGY CO., LTD
5F.6, Alley 2, Lane 222, Lien Cheng Road, Chung Ho City, Taipei Hsien, 235, Taiwan, ROC Telephone: 886-2-2247 4816 Fax: 886-2-2247 7024 Email: sale@mytech.com.tw
### Description of the new device:
Basically, the subject device and the predicated device are the same in intended use, technological characteristics, power supply, LCD display, range, accuracy, and operating environments, memory, dimensions, weight, and storage environments. The only difference is the display layout. We present two photos for HPL-100 and HPL-200 respectively in the following page for comparison. Since we also did the relevant electric safety and EMC testing for HPL-200, the safety and effectiveness aspects are not raised.
They are decided to be substantially equivalent.
# Technological Characteristics of our new device compared to the predicate device:
The technological characteristic of MYTECH HPL-100 is substantially equivalent to HPL-200. I MYTECH HPL-100 is of generally the same form and intended to be used in the same manner as the substantially equivalent product; and is passed the relevant EMC and Safety standards. Thus there are substantially equivalent.
{2}------------------------------------------------
## MYTECH TECHNOLOGY CO., LTD.
5F.6, Alley 2, Lane 222, Lien Cheng Road, Chung Ho City, Taipei Hsien, 235, Taiwan, ROC Telephone: 886-2-2247 4816 Fax: 886-2-2247 7024 Email: sale@mytech.com.tw
#### Test Summary:
#### ELECTRIC SAFETY and EMC test reports, 1. General safety EN 60601-1:1990+A1+A2+A11+A12+A13 PASS EMC conformity EN 60601-1-2: 1993 PASS
#### 2. WOVEN COTTON SHEETING
Uses the 510K Blood-Pressure Cuff
- PERFORMANCE & CLINICAL TEST 3. AAMI / ANSI SP10
MYTECH TECHNOLOGY CO., LTD. believes this information and referred document to be sufficient for the FDA to find our proposed device substantially equivalent to the predicate product and other products currently in distribution.
JAMES WU
Submitter, 5/7/2004 General Manager MYTECH TECHNOLOGY CO., LTD.
{3}------------------------------------------------
Image /page/3/Picture/1 description: The image contains the text "Public Health Service". The text is in a bold, sans-serif font. The words are stacked on top of each other, with "Public" on the top line, "Health" on the second line, and "Service" on the third line.
Image /page/3/Picture/2 description: The image shows the seal of the Department of Health & Human Services - USA. The seal is circular and contains the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. In the center of the seal is an abstract image of an eagle with its wings spread.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
## MAY 1 7 2004
Mytech Technology Co., Ltd. c/o Dr. Ke-Min Jen Cro Dr. its its in mean Industrial Research Society No. 58, Fu-Chiun St. Hsin-Chu City, China (Taiwan) 300
Re: K041228
K041220
Trade Name: Mytcch/Happy Life Blood Pressure Monitor, HPL-200 Regulation Number: 21 CFR 870.1130 Regulation Name: Noninvasive Blood Pressure Measurement System Regulatory Class: II (two) Product Code: DXN Dated: May 07, 2004 Received: May 10, 2004
Dear Dr. Jen:
We have reviewed your Section 510(k) premarket notification of intent to market the devices in we nave reviewed your Section 910(t) premained is substantially equivalent (for the indications reletenced above and nave determined by marketed predicate devices marketed in interstate for use stated in the encrosure to regars) at regally and the Medical Device Amendments, or to commerce prior to May 20, 1976, the encordance with the provisions of the Federal Food, Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). and Costlience Act (Act) that do not require appt to the general controls provisions of the Act. The You may, therefore, market the dovices, sale, of the more
general controls provisions of the Act include requirements for annual registration, listing of general controls provisions of the research and prohibitions against misbranding and adulteration.
If your device is classified (see above) into cither class II (Special Controls) or class III (PMA), it If your device is classified (500 abs. Existing major regulations affecting your device can
may be subject to such additional controls. Existing major regulations of the con Inay be subject to suen additional vehicler , Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean Frease oc advised that I Dri 3 modalso wrough with other requirements of the Act
{4}------------------------------------------------
Page 2 – Dr. Ke-Min Jen
or any Federal statutes and regulations administered by other Federal agencies. You must of any i cacal statutes and regaranents, including, but not limited to: registration and listing (21 Comply with an the Free brequirements, while and manufacturing practice requirements as set CITY art 607), adoling (21 CFR Part 820); and if applicable, the electronic forth in the quality by sections (Sections 531-542 of the Act); 21 CFR 1000-1050. product radiation oonal of to begin marketing your device as described in your Section 510(k) I ins icher witi anow you to begin maing of substantial equivalence of your device to a legally prematics notification: "The President on for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please If you desire specific acripliance at (301) 594-4648. Also, please note the regulation entitled, Contact the Office of Course to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small other general information on your responser Assistance at its toll-free number (800) 638-2041 or Manufacturers, Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html
Sincerely yours,
Dimna R. Vochner
Image /page/4/Picture/5 description: The image shows a close-up of a signature. The signature is illegible, but it appears to be written in cursive. The signature is made up of a series of loops and curves. The signature is written in black ink.
Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
## MYTECH TECHNOLOGY CO., LTD.
5F.6, Alley 2, Lane 222, Lien Cheng Road, Chung Ho City, Taipei Hsien, 235, Taiwan, ROC Fax: 886-2-2247 7024 Telephone: 886-2-2247 4816 Email: sale(@mytech.com.tw
510(k) Number ( if known): ___________________________________________________________________________________________________________________________________________________
Device Name: MYTECH / HAPPY LIFE BLOOD PRESSURE MONITOR_ HPL-200
Indications for use:
The device is a noninvasive blood pressure measurement system intended to measure the systolic and diastolic blood pressures and pulse rate of an adult individual, over age 18, at home by using a non-invasive technique in which an inflatable cuff is wrapped around The cuff circumference is limited to be 5.25" – 7.75". the wrist.
Prescription Use (Part 21 CFR 801 Subpart D)
AND/OR
Over-The-Counter Use V (21 CFR 807 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Dring R. Volumes
(Division Sign-Off) Division of Cardiovascular Devices
510(k) Number上04/228
Page 1 of 1 -
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.