K041169 · Canamet Canadian Natl Medical Technologies, Inc. · DXN · Aug 24, 2004 · Cardiovascular
Device Facts
Record ID
K041169
Device Name
PIESOMETER, MODEL MK-1
Applicant
Canamet Canadian Natl Medical Technologies, Inc.
Product Code
DXN · Cardiovascular
Decision Date
Aug 24, 2004
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 870.1130
Device Class
Class 2
Indications for Use
The device is used to provide data to qualified medical personnel or trained users for the purpose of assessing the patient's cardiac health via blood pressure readings taken during daily activity for up to a 48-hour period. Measurements of systolic, diastolic, and heart rate are made and stored. The device can also be used to provide blood pressure and heart rate data in single measurement mode. The device is also intended to provide preliminary blood pressure data from ambulatory patients for non-time critical applications only. It is for use in hospitals, clinics or physicians offices by a qualified physician or trained staff member under the supervision of that physician. It is also intended for home care use by a trained patient with periodic health consultations from his or her physician. Patient diagnosis is not to be performed solely based on the results of this device.
Device Story
Piesometer MK-1 is a non-invasive automated blood pressure monitor for ambulatory use. Device utilizes electronic pressure cuff with three acoustic sensors: two sensors on biceps detect Korotkoff sounds; one sensor on triceps detects interference signals. System employs adaptive interference cancellation signal processing to filter out noise and vibration, enabling readings in high-noise environments. Device records systolic, diastolic, and heart rate data over 48-hour periods or single measurements. Used in hospitals, clinics, or home settings by physicians, staff, or trained patients. Output displayed on 32-character LCD; data stored for review. Healthcare providers use output to assess cardiac health; device intended for preliminary data collection in non-time critical applications. Patient diagnosis must not rely solely on device results.
Clinical Evidence
Bench and clinical testing performed. Results indicate device is as safe and effective as predicate devices. No specific quantitative clinical metrics (e.g., sensitivity, specificity) provided in summary.
Technological Characteristics
Auscultatory blood pressure measurement system using Korotkoff sound technique. Features three acoustic sensors (microphones) for adaptive interference cancellation. Powered by rechargeable Lithium-Ion battery. Includes 32-character LCD display. Memory capacity for 60 records. Form factor: 13.4 cm x 9.5 cm x 4.4 cm. Weight: 850 grams.
Indications for Use
Indicated for ambulatory patients requiring cardiac health assessment via blood pressure and heart rate monitoring. Suitable for use in clinical settings or home care by trained patients. Not for sole diagnostic use.
Regulatory Classification
Identification
A noninvasive blood pressure measurement system is a device that provides a signal from which systolic, diastolic, mean, or any combination of the three pressures can be derived through the use of tranducers placed on the surface of the body.
{0}------------------------------------------------
K04/169
S1/2
AUG 2 4 2004
# EXHIBIT 2 510(k) Summary
# CANAMET CANADIAN NATIONAL MEDICAL TECHNOLOGIES INC. 1120 Finch Ave West, Suite 201 Toronto ON M3J-3H7 Phone (416) 916-0469 Ext 333 Fax: (416)-916-0313 Email: Stergios@canamet.com Contact: STERGIOS STERGIOPOULOS Date: July 31, 2004
- Identification of the device 1. Proprietary-Trade Name: PIESOMETER MK-1 Classification Names: DXN, SYSTEM, MEASUREMENT, BLOOD-PRESSURE, NON-INVASIVE, Common/Usual Name: Non-invasive automated blood pressure meter.
#### 2. Equivalent legally marketed devices
This product is similar in function and design to the Spacelabs ABP (Ambulatory Blood Pressure), K031479, K941167, K904318, K855127 and, SunTech Accutracker DX, K913844.
#### Indications for Use (intended use) 3.
The device is used to provide data to qualified medical personnel or trained users for the purpose of assessing the patient's cardiac health via blood pressure readings taken during daily activity for up to a 48-hour period. Measurements of systolic, diastolic, and heart rate are made and stored. The device can also be used to provide blood pressure and heart rate data in single measurement mode. The device is also intended to provide preliminary blood pressure data from ambulatory patients for non-time critical applications only. It is for use in hospitals, clinics or physicians offices by a qualified physician or trained staff member under the supervision of that physician. It is also intended for home care use by a trained patient with periodic health consultations from his or her physician. Patient diagnosis is not to be performed solely based on the results of this device.
#### 4. Description of the Device
The Model PIESOMETER MK-1 Digital Blood Pressure Monitor measures systolic and diastolic blood pressure and heart rate using adaptive interference cancellation (signal processing) technologies. The system employs three acoustic sensors (i.e. microphones) in the electronic pressure cuff. During a blood pressurement, two acoustic sensors are placed in the traditional location on the patient's biceps. This set of sensors, operating much like a conventional stethoscope, senses the Korotkoff sounds. Unique to the PIESOMETER MK-1 is a second set of a sensor placed in the triceps area. This sensor serves to detect any interference signals that would normally compromise the reading. This allows reading to be taken even in noise and vibration intensive environment. This technology functions by identifying signals lying within the audio frequency and filtering out the interference signal. The CANAMET PIESOMETER is protected by U.S. Patent 6,520,918.
{1}------------------------------------------------
## Safety and Effectiveness, comparison to predicate device 5.
The results of bench and user/clinical testing indicate that the new device is as safe and effective as the predicate devices.
| Designation | Spacelabs ABP (Ambulatory<br>Blood Pressure), K031479,<br>K941167, K904318, K855127 | SunTech<br>Accutracker DX,<br>K913844. | PIESOMETER MK-1 |
|------------------------|---------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------|
| Operating<br>Principle | Auscultatory Technology<br>Korotkoff sound technique. | Auscultatory<br>Technology<br>Korotkoff sound<br>technique, Motion<br>Tolerant | Auscultatory<br>Technology<br>Korotkoff sound<br>technique, Motion<br>Tolerant |
| Display | 4-digit, 7-segment LCD;<br>systolic, diastolic and<br>heart rate information | 32 Character LCD | 32 Character LCD |
| Recording Time | Up to 48 hours | Up to 48 Hours | Up to 48 Hours |
| Power supply | 4 AA alkaline disposable<br>or rechargeable NiCad<br>batteries | Four 1.5 (AA) alkaline<br>batteries | Rechargeable Lithium-<br>Ion battery (Not user<br>replaceable)<br>rechargeable with<br>supplied charger. |
| External<br>dimensions | 1.1"x4.5"x3.4" (2.8 cm<br>x11.4 cm x8.6 cm)<br>Model 90217:<br>0.97"x3.94"x2.77" (2.5 cm<br>x10.0 cm x7.0 cm) | 3.25"W x 5"L x 1.30"D<br>(8.25cm x 12.7cm x<br>3.3cm) | 5.28" x 3.74" x 1.73"<br>(13.4 cm X 9.5cm X<br>4.4cm) |
| Memory | Not specified | Over 250 samples of<br>systolic, diastolic, and<br>heart rate. | 60 records of systolic,<br>diastolic, and heart rate |
| Weight | 12.2 oz. (347 g) (with<br>batteries)Model 90217: 9<br>oz. (255 g) (with<br>batteries) | 12.6oz. (357.2 grams) | 850 grams |
| Accessories | Not specified | Multiple cuff sizes | Multiple cuff sizes |
## Comparison matrix - new ys. Predicate device 6
### 7. Conclusion
After analyzing both bench and clinical testing data, it is the conclusion of Canamet that that the "PIESOMETER MK-1" Automated Blood Pressure Monitor is as safe and effective as the predicate device, has few technological differences, and has no new indications for use, thus rendering it substantially equivalent to the predicate devices.
p (_/%
{2}------------------------------------------------
Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is an abstract symbol that resembles an eagle or bird in flight, composed of three stylized lines.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
AUG 2 4 2004
Canamet Canadian National Medical Technologies, Inc. c/o Mr. Daniel Kamm, P.E. Regulatory Engineer Kamm & Associates P.O. Box 7007 Deerfield, IL 60015
Re: K041169
> Trade Name: Piesometer, Model MK-1 Regulation Number: 21 CFR 870.1130 Regulation Name: Noninvasive Blood Pressure Measurement System Regulatory Class: II (two) Product Code: DXN Dated: August 09, 2004 Received: August 10, 2004
Dear Mr. Kamm:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations. Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must
{3}------------------------------------------------
Page 2 - Mr. Daniel Kamm, P.E.
comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (301) 594-4648. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html
Sincerely yours,
Neil R.P. Ogden for
Bram D. Zuckerman, M.D.
Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
# Indications for Use
510(k) Number (if known):________________K041169
Device Name: PIESOMETER MK-1
The device is used to provide data to qualified medical personnel or trained users for the purpose of assessing the patient's cardiac health via blood pressure readings taken during daily activity for up to a 48-hour period. Measurements of systolic, diastolic, and heart rate are made and stored. The device can also be used to provide blood pressure and heart rate data in single measurement mode. The device is also intended to provide preliminary blood pressure data from ambulatory patients for non-time critical applications only. It is for use in hospitals, clinics or physicians offices by a qualified physician or trained staff member under the supervision of that physician. It is also intended for home care use by a trained patient with periodic health consultations from his or her physician.
Patient diagnosis is not to be performed solely based on the results of this device.
Prescription Use X (Part 21 CFR 801 Subpart D) AND/OR
Over-The-Counter Use (21 CFR 807 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Neil H. Golen
for BOZ
| (Division Sign-Off) Division of Cardiovascular Devices | |
|--------------------------------------------------------|---------|
| 510(k) Number | K041169 |
Page 1 of 1
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.