FSH MENOPAUSE PREDICTOR TEST

K041165 · ACON Laboratories, Inc. · CGJ · Jun 17, 2004 · Clinical Chemistry

Device Facts

Record IDK041165
Device NameFSH MENOPAUSE PREDICTOR TEST
ApplicantACON Laboratories, Inc.
Product CodeCGJ · Clinical Chemistry
Decision DateJun 17, 2004
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 862.1300
Device ClassClass 1

Indications for Use

The FSH Menopause Predictor Test is a qualitative, one-step, midstream assay for the detection of Follicle Stimulating Hormone (FSH) in urine to be used as an aid in predicting menopause. The FSH Menopause Predictor Test is intended for over-the-counter use by the lay consumer.

Device Story

The FSH Menopause Predictor Test is a lateral flow immunoassay for home use. It utilizes a midstream urine sample as input. The device employs mouse monoclonal-dye conjugate and goat polyclonal-solid phase antibodies to detect FSH. As urine wicks through the device, antibody-dye conjugates bind to FSH, forming a complex that binds to anti-FSH antibodies in the reaction zone. This produces a pink-rose color band (T Line) if FSH concentration is ≥ 25 mIU/mL. A reference line confirms adequate sample volume and wicking. The user visually compares the T line intensity to the reference line within 3-8 minutes. The result aids the consumer in predicting menopause status. The device is intended for self-testing by lay consumers in a home environment.

Clinical Evidence

Clinical trial with 70 female participants. Comparison of ACON FSH Menopause Predictor Test vs. InstaCheck® predicate. Participants performed testing at home; samples were also tested by a study coordinator. Results: ACON test (trained lab tech) vs. InstaCheck® accuracy >99% (99%-99% CI). ACON test (consumer) vs. ACON test (trained lab tech) accuracy 94% (92.1%-95.6% CI). ACON test (consumer) vs. InstaCheck® accuracy 94% (92.1%-95.6% CI).

Technological Characteristics

Two-site sandwich immunochromatographic assay. Materials: mouse monoclonal dye conjugate, goat polyclonal solid-phase antibodies. Form factor: midstream test wand. Qualitative colorimetric readout. Sensitivity: 25 mIU/mL. Traceability: WHO Urinary FSH 1st International Standard.

Indications for Use

Indicated for lay consumers to qualitatively detect FSH in urine as an aid in predicting menopause and confirming hormone changes associated with menopausal symptoms.

Regulatory Classification

Identification

A follicle-stimulating hormone test system is a device intended to measure follicle-stimulating hormone (FSH) in plasma, serum, and urine. FSH measurements are used in the diagnosis and treatment of pituitary gland and gonadal disorders.

Predicate Devices

Submission Summary (Full Text)

{0} # 510(k) SUBSTANTIAL EQUIVALENCE DETERMINATION DECISION SUMMARY DEVICE ONLY TEMPLATE A. 510(k) Number: k041165 B. Purpose for Submission: New menopause predictor device C. Analyte: Follicle Stimulating Hormone (FSH) D. Type of Test: Qualitative E. Applicant: ACON Laboratories, Inc. F. Proprietary and Established Names: FSH Menopause Predictor Test G. Regulatory Information: 1. Regulation section: 21 CFR 862.1300 2. Classification: Class I, meets the limitations of exemptions 862.9 (c) (9) 3. Product Code: CGJ 4. Panel: 75 Chemistry H. Intended Use: 1. Intended use(s): The FSH Menopause Predictor Test is a qualitative, one-step, midstream assay for the detection of Follicle Stimulating Hormone (FSH) in urine as a confirmation of body hormone changes related to the symptoms associated with the changes of menopause. The FHS Menopause Predictor Test is intended for use by the lay consumer. 2. Indication(s) for use: The FSH Menopause Predictor Test is a qualitative, one-step, midstream assay for the detection of Follicle Stimulating Hormone (FSH) in urine to be used as an aid in predicting menopause. The FSH Menopause Predictor Test is intended for over-the-counter use by the lay consumer. {1} Page 2 of 4 3. Special condition for use statement(s): This is intended for over-the-counter (OTC) sales to lay consumers. 4. Special instrument Requirements: Not Applicable I. Device Description: The FSH Menopause Predictor Test is supplied with two wands that employ a combination of mouse monoclonal dye conjugate and goat polyclonal-solid phase antibodies on the test region and a package insert. J. Substantial Equivalence Information: 1. Predicate device name(s): Instacheck® Menopause Predictor Test 2. Predicate K number(s): k023408 3. Comparison with predicate: | Similarities | | | | --- | --- | --- | | Item | Device | Predicate | | Intended Use | Detection of FSH in urine | Detection of FSH in urine | | | Qualitative test for use with urine. | Qualitative test for use with urine. | | Principle/Methodology | Uses colorimetric visual interpretation | Uses colorimetric visual interpretation | | | Require urine addition to the device | Require urine addition to the device | | Sensitivity | 25 mIU/mL | 25 mIU/mL | | Differences | | | | Item | Device | Predicate | | Principle/Methodology | Rapid chromatographic immunoassay using combination of mouse monoclonal dye conjugate and goat polyclonal solid phase antibodies. | Rapid chromatographic immunoassay using anti-FSH monoclonal antibody-colloidal gold conjugate and gold-labeled donkey antibodies. | K. Standard/Guidance Document Referenced (if applicable): None referenced L. Test Principle: The device is a two-site sandwich immunochromatographic assay. {2} Page 3 of 4 # M. Performance Characteristics (if/when applicable): ## 1. Analytical performance: a. Precision/Reproducibility: Five batches of normal human urine demonstrated to be negative for FSH were spiked to levels of 0, 12.5, 25.0, 37.5 and 75.0 mIU/mL. Thirty-five consumers blindly tested each of the 5 batches. The results showed an excellent repeatability for all of the positive and negative blind urine samples. Seventy-nine consumers were given 2 urine samples from women experiencing perimenopausal symptoms and/or irregular menstrual cycle. Seventy-five of the participants were able to correctly interpret their testing results as directed in the package insert. b. Linearity/assay reportable range: N/A c. Traceability (controls, calibrators, or method): WHO Urinary FHS 1st International Standard d. Detection limit: The sensitivity of the test is 25 mIU/mL. A panel of FSH was prepared by spiking human urine specimens with FSH to concentrations of 0, 12.5, 25.0, 37.5 and 75 mIU/mL. 35 devices per standard level were tested. Samples with concentrations of FSH equal to or greater than 25 mIU/mL were identified as positive for all samples. Samples lower than 25 mIU/mL were identified as negative for all samples. e. Analytical specificity: The specificity of ACON FSH Menopause Predictor test was tested with similar hormones and compounds found in normal human urine. Luteinizing hormones (1000 mIU/mL), human chorionic gonadotropin (100,000 mIU/mL) and thyroid stimulation hormone (1000 mIU/mL) were added to positive and negative urine and did not interfere with the results. The specificity of ACON FSH Menopause Predictor test was tested with various OTC drugs, chemical and biological analytes. The analytes were added to a urine sample with negative FSH levels and a urine sample with 25 mIU/ML FSH level. None of the 22 substances tested interfered with the test results. pH adjusted urine was tested in duplicate by adding FSH and did not interfere with the performance of the ACON FSH Menopause Predictor test. f. Assay cut-off: See Detection Limit above. {3} Page 4 of 4 2. Comparison studies: a. Method comparison with predicate device: The ACON FSH Menopause Predictor test was compared to the InstaCheck® FSH Menopause test (K023408) using 70 women. 70 females, ages 40-58 tested their own urine at home using the ACON FSH Menopause Predictor test, followed by a trained laboratory technician. The technician tested each sample using the ACON FSH Menopause Predictor test and the InstaCheck® FSH Menopause. This process was repeated a week later. There was an accuracy of 94% between the ACON FSH Menopause Predictor test used by the consumer and the ACON FSH Menopause Predictor test used by a trained laboratory technician. There was a >99% agreement between the ACON FSH Menopause Predictor test and the InstaCheck® when conducted by a trained laboratory technician. There was a 94% accuracy between the consumer results and the professional InstaCheck® results. 69 tests were done with midstream test method and the others were conducted using the dip method. b. Matrix comparison: N/A 3. Clinical studies: a. Clinical sensitivity: N/A b. Clinical specificity: N/A c. Other clinical supportive data (when a and b are not applicable): N/A 4. Clinical cut-off: N/A 5. Expected values/Reference range: The expected values are based on literature and in previous sensitivity studies that demonstrated adequate performance at the cutoff of 25 mIU/mL. N. Conclusion: The submitted information in this premarket notification is complete and supports a substantial equivalence decision.
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...