← Product Code [DCK](/productcode/DCK) · K041155

# WIDE RANGE C-REACTIVE PROTEIN (WR CRP) ASSAY (K041155)

_Stanbio Laboratory · DCK · Oct 14, 2004 · Immunology · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K041155

## Device Facts

- **Applicant:** Stanbio Laboratory
- **Product Code:** [DCK](/productcode/DCK.md)
- **Decision Date:** Oct 14, 2004
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 866.5270
- **Device Class:** Class 2
- **Review Panel:** Immunology

## Indications for Use

The Wide Range C-Reactive Protein (Wr CRP) Assay is an in vitro diagnostic test intended for the quantitative determination of C-reactive protein in human serum and plasma. The Stanbio Wr CRP Multi-Calibrator Set A is for use in multi-calibration of the Stanbio Wr High-Sensitivity CRP (HS-CRP) quantitative assay. The Stanbio Wr CRP Multi-Calibrator Set B is for use in multi-calibration of the Stanbio Wr CRP (Normal-CRP) quantitative assay. The Stanbio CRP (HS) Control Set A and the CRP Control Set B (Normal sensitivity) is intended to be used to validate the accuracy and precision of tests for quantitative determination of CRP on photometric systems.

## Device Story

In vitro diagnostic assay for quantitative determination of C-reactive protein (CRP) in human serum and plasma; utilizes latex particle-enhanced immunoturbidimetric technology. Device consists of two reagents (buffer and latex particles coated with anti-CRP antibodies) and two sets of calibrators/controls for high-sensitivity (HS-CRP) and normal-sensitivity ranges. Operated on automated photometric analyzers; samples and reagents are pipetted into cuvettes; immune reaction increases light scattering proportional to CRP concentration. Absorbance measured at 500 nm; concentration interpolated from a multipoint calibration curve. Used in clinical laboratory settings by trained personnel to assist clinicians in diagnosing and evaluating inflammation, infection, and tissue injury.

## Clinical Evidence

Bench testing only. Performance established via precision/reproducibility studies (CV% < 4.1%), linearity (0.005-1.5 mg/dL for HS-CRP; 0.02-30.0 mg/dL for Normal-CRP), and interference studies (no interference from bilirubin, hemoglobin, or lipids). Method comparison against predicate device showed strong correlation (r=0.986-0.999) across serum and plasma samples. No clinical sensitivity or specificity studies provided.

## Technological Characteristics

Latex particle-enhanced immunoturbidimetric assay. Reagents: Wr CRP buffer and CRP-coated latex particles. Measurement: Spectrophotometric absorbance at 500 nm. Calibration: Multipoint. Connectivity: Automated analyzer integration. Traceability: IFCC/BCR/CAP CRM 470.

## Regulatory Identification

A C-reactive protein immunological test system is a device that consists of the reagents used to measure by immunochemical techniques the C-reactive protein in serum and other body fluids. Measurement of C-reactive protein aids in evaluation of the amount of injury to body tissues.

## Predicate Devices

- Dade Behring N High Sensitivity CRP Assay (k991385)

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}

510(k) SUBSTANTIAL EQUIVALENCE DETERMINATION
DECISION SUMMARY

A. 510(k) Number:
k041155

B. Purpose for Submission:
New device

C. Analyte:
C-Reactive Protein

D. Type of Test:
Quantitative, latex particle enhanced immunoturbidimetric assay

E. Applicant:
Stanbio Laboratory

F. Proprietary and Established Names:
Wide Range C-Reactive Protein (Wr CRP) Assay

G. Regulatory Information:
1. Regulation section:
21CFR § 866.5270, C-Reactive Protein Immunological Test System
21CFR § 862.1660, Quality Control Material (Assayed and Unassayed)
21CFR § 862.1150, Calibrator
2. Classification:
Class II
Class I
Class II
3. Product Code:
DCK, C-Reactive Protein, Antigen, Antiserum and Control
JJX, Single (Specified) Analyte Controls (Assayed and Unassayed)
JIT, Calibrator, Secondary
4. Panel:
Immunology (82)
Chemistry
Chemistry

H. Intended Use:
1. Intended use(s):
The Wide Range C-Reactive Protein (Wr CRP) Assay is an in vitro diagnostic test intended for the quantitative determination of C-reactive protein in human serum and plasma.

{1}

Page 2 of 5

The Stanbio Wr CRP Multi-Calibrator Set A is for use in multi-calibration of the Stanbio Wr High-Sensitivity CRP (HS-CRP) quantitative assay.

The Stanbio Wr CRP Multi-Calibrator Set B is for use in multi-calibration of the Stanbio Wr CRP (Normal-CRP) quantitative assay.

The Stanbio CRP (HS) Control Set A and the CRP Control Set B (Normal sensitivity) is intended to be used to validate the accuracy and precision of tests for quantitative determination of CRP on photometric systems.

2. Indication(s) for use:
The measurement of C-reactive protein aids in the diagnosis and evaluation of inflammation, infection and tissue injuries.

3. Special condition for use statement(s):
The device is for prescription use only.

4. Special instrument Requirements:
The device is suitable for automated analyzers that use a multipoint calibration method. Measurements of absorbance are to be made with a spectrophotometer able to read absorbances at 570 nm.

I. Device Description:
The device test kit is comprised of two reagents, Wr CRP buffer (R1) and Wr CRP latex (R2)

There are two sets of calibrators. Set A is high sensitivity, assay range of 0.005 to 1.500 mg/dL and Set B is normal sensitivity, assay range of 0.02 to 30.00 mg/dL. Each has 5 levels. If medical condition warrants that some irritation is at hand, then the high sensitivity or Set A Calibrators would be run to determine level of inflammation. Should level exceed the top linear range for this set of calibrators, then Set B would be run to evaluate ranges above the lower level.

There are two sets of controls (Set A and Set B) with two levels each.

J. Substantial Equivalence Information:

1. Predicate device name(s):
Dade Behring N High Sensitivity CRP Assay

2. Predicate K number(s):
k991385

3. Comparison with predicate:

|  Similarities  |   |   |
| --- | --- | --- |
|  Item | Device | Predicate  |
|   | Stanbio WrCRP | Dade Behring N High Sensitivity CRP  |
|  Intended Use | Quantitative determination of CRP in human serum and plasma | Same  |
|  Sample matrix | Serum and plasma | Same  |

{2}

Page 3 of 5

|  Differences  |   |   |
| --- | --- | --- |
|  Item | Device | Predicate  |
|  Methodology | Particle enhanced turbidimetric immunoassay | Particle enhanced immunonephelometry  |
|  Calibration | Calibrated using lot independent standard | Calibrated using a lot dependent standard (included in the kit)  |

## K. Standard/Guidance Document Referenced (if applicable):

None referenced

## L. Test Principle:

The Stanbio WR™ CRP uses stabilized C - reactive protein coated latex particles which are monospecific for CRP. Calibrators, controls, and patient samples are pipetted into sample cups. Microvolume of samples and reagent buffer are automatically pipetted into individual cuvettes. Following an initial incubation, the latex enhanced anti-sera is added to the cuvettes. The immune responses created cause an increase in light scattering which correlates with the concentration of specimen CRP. Following an incubation period lasting approximately 10 minutes, the absorbance of the solution is measured at 500 nm.

A calibration curve is generated by assaying a series of calibrators with known concentrations of CRP using the instrument data reduction capability or manually plotting the changes in absorbance versus concentration. Concentration of the control and patient samples are interpolated from the calibration curve.

## M. Performance Characteristics (if/when applicable):

### 1. Analytical performance:

#### a. Precision/Reproducibility:

With Calibrator Set A (HS-CRP), within-day precision was established by running 20 assays on 3 sera. "Day-to-day" values were obtained by assaying three different sera daily for twenty days.

|  Within Day | Mean | SD | CV%  |
| --- | --- | --- | --- |
|  Sample 1 | 0.05 | 0.001 | 2.0  |
|  Sample 2 | 0.103 | 0.0009 | 0.87  |
|  Sample 3 | 1.203 | 0.007 | 0.58  |
|  Day-to- Day | Mean | SD | CV%  |
| --- | --- | --- | --- |
|  Sample 1 | 0.068 | 0.0022 | 3.2  |
|  Sample 2 | 0.237 | 0.003 | 1.27  |
|  Sample 3 | 1.074 | 0.016 | 0.99  |

With Calibrator Set B (Normal-CRP), within-day precision was established by running 20 assays on 3 sera. "Day-to-day" values were obtained by assaying two different sera daily for twenty days.

{3}

|  Within Day | Mean | SD | CV%  |
| --- | --- | --- | --- |
|  Sample 1 | 0.1024 | 0.0042 | 4.1  |
|  Sample 2 | 0.1974 | 0.0044 | 2.2  |
|  Sample 3 | 19.92 | 0.40 | 2.0  |
|  Day-to- Day | Mean | SD | CV%  |
| --- | --- | --- | --- |
|  Sample 1 | 2.459 | 0.039 | 1.6  |
|  Sample 2 | 23.26 | 0.80 | 3.4  |

# b. Linearity/assay reportable range:

Linearity of the method across the assay range for the High Sensitivity application was evaluated by serially diluting a known concentration of low CRP (2 mg/dL) with a buffer. A linear response was observed with slope of (-0.03), intercept of 1.08, and correlation coefficient of 0.9930. HS-CRP is linear from 0.-1.5 mg/dL

Linearity of the Normal CRP was evaluated by serially diluting a known concentration of high CRP (35 mg/dL) with a buffer. A linear response was observed with slope of (0.40), intercept of 1.15, and correlation coefficient of 0.9975. Normal CRP is linear from 0.02 to 30.0 mg/dL.

# c. Traceability (controls, calibrators, or method):

The calibrators and controls are verified against IFCC/BCR/CAP reference material for 15 plasma proteins CRM 470.

# d. Detection limit:

With Calibrator Set A (HS-CRP), the sensitivity of the WR® CRP is  $0.005\mathrm{mg / dL}$ .

With Calibrator Set B (Normal CRP), the sensitivity of the WR® CRP is  $0.02\mathrm{mg / dL}$ .

Sensitivity is determined by running zero calibrator a total of 20 times each for Calibrator Set A and B. This represents the lowest concentration of CRP that can be distinguished from zero.

# e. Analytical specificity:

Interfering Substances

The interference studies were conducted by taking spiked samples of serum that was assayed for their CRP content. Once the assay value was determined, the sample was split with the potential interfering substances (lipids, bilirubin and hemoglobin) added and re-assayed. A  $(\pm 10\%)$  difference was allowed for recovery from the initial assayed value.

Bilirubin up to  $40\mathrm{mg / dL}$ , hemoglobin up to  $500\mathrm{mg / dL}$  and lipemia (as triglycerides) up to  $1500\mathrm{mg / dL}$  do not interfere with the assay.

{4}

Page 5 of 5

f. Assay cut-off:
0.300 mg/dL from literature citation.

2. Comparison studies:

a. Method comparison with predicate device:
WR®CRP is compared with its predicate device, Dade Behring N High Sensitivity CRP.

With Calibrator Set A, 24 serum specimens were assayed. The correlation coefficient was 0.986 and the regression equation was $y = 1.016x + 0.125$. For plasma, 23 specimens were assayed. The correlation coefficient was 0.996 and the regression equation was $y = 1.0766x - 0.0258$.

With Calibrator Set B, 61 serum specimens were assayed. The correlation coefficient was 0.992 and the regression equation was $y = 0.9157x + 0.2224$. For plasma, 42 specimens were assayed. The correlation coefficient was 0.999 and the regression equation was $y = 1.0235x - 0.0031$

b. Matrix comparison:
Fifteen samples each of serum and plasma (in the high sensitivity range as well as the normal range) were spiked with various levels of CRP. With calibrator set A (HS-CRP), the correlation coefficient was 0.999 and the regression equation was $y = 1.0073x + 0.0004$. With calibrator set B (Normal CRP), the correlation coefficient was 0.999 and the regression equation was $y = 0.9866x - 0.2753$.

3. Clinical studies:

a. Clinical sensitivity:
Not provided.

b. Clinical specificity:
Not provided.

c. Other clinical supportive data (when a and b are not applicable):
Not applicable.

4. Clinical cut-off:
Not provided.

5. Expected values/Reference range:
Values for healthy individuals are below 0.300 mg/dL per published literature. It is recommended that each laboratory establish its own normal range.

N. Conclusion:
The submitted material in this premarket notification is complete and supports a substantial equivalence decision.

---

**Source:** [https://fda-staging.innolitics.com/device/K041155](https://fda-staging.innolitics.com/device/K041155)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
