← Product Code [LRG](/productcode/LRG) · K041150

# MICROSCAN SYNERGIES PLUS GRAM-NEGATIVE MIC/COMBO PANELS (K041150)

_Dade Behring, Inc. · LRG · Jul 22, 2004 · Microbiology · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K041150

## Device Facts

- **Applicant:** Dade Behring, Inc.
- **Product Code:** [LRG](/productcode/LRG.md)
- **Decision Date:** Jul 22, 2004
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 866.1640
- **Device Class:** Class 2
- **Review Panel:** Microbiology

## Indications for Use

The MicroScan® Synergies plus™ Gram-Negative MIC/Combo Panel is used to determine quantitative and/or qualitative antimicrobial agent susceptibility of colonies grown on solid media of rapidly growing aerobic and facultative anaerobic gram-negative bacilli. After inoculation, panels are incubated for 4.5 - 18 hours at 35℃ +/- 1℃, in a WalkAway SI or equivalent, and read by the MicroScan® Instrumentation. Additionally, the panels may be incubated in a non-CO2 incubator and the AST portions can be read visually, according to the Package Insert. This particular submission is for the addition of the antimicrobial Cefazolin, at concentrations of 0.5 to 32 ug/ml, to the test panel.

## Device Story

MicroScan® Synergies plus™ Gram-Negative MIC/Combo Panels utilize dehydrated microdilutions of antimicrobials in Mueller Hinton Broth. System inputs: isolated bacterial colonies from solid media, prepared via turbidity method (0.5 McFarland standard). Device operates on WalkAway® -SI System using colorimetric optics (color wheel/lamp/photosensor). System performs initial read at 2.5 hours, followed by automated reads at 4.5–18 hours based on organism growth rate. Algorithm subtracts initial read from final read to minimize variation; thresholds adjust for fermenter/non-fermenter growth rates. Output: quantitative MIC values and qualitative susceptibility interpretations. Used in clinical microbiology laboratories by technicians to determine antimicrobial susceptibility. Results inform physician antibiotic selection for gram-negative infections.

## Clinical Evidence

Bench testing only. Performance evaluated using fresh and stock clinical isolates and challenge strains. Compared against NCCLS frozen reference panels. Essential Agreement for Cefazolin was >97%. Reproducibility and precision testing performed using the Turbidity inoculum method and WalkAway® S7 System. Quality control testing confirmed acceptable performance.

## Technological Characteristics

Miniaturized broth dilution susceptibility test; dehydrated Mueller-Hinton broth; 96-well panel format; automated incubation (4.5-18 hours) and optical reading via WalkAway® S7 System; manual visual reading option; utilizes fluorogenic identification and antimicrobial susceptibility testing technologies.

## Regulatory Identification

An antimicrobial susceptibility test powder is a device that consists of an antimicrobial drug powder packaged in vials in specified amounts and intended for use in clinical laboratories for determining in vitro susceptibility of bacterial pathogens to these therapeutic agents. Test results are used to determine the antimicrobial agent of choice in the treatment of bacterial diseases.

## Predicate Devices

- MicroScan® rapID/S plus Gram Negative MIC/Combo Panels – gentamicin ([K020185](/device/K020185.md))
- MicroScan® Dried Gram-Negative MIC/Combo Panels ([K862140](/device/K862140.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}

510(k) SUBSTANTIAL EQUIVALENCE DETERMINATION
DECISION SUMMARY
DEVICE ONLY TEMPLATE

A. 510(k) Number:
K041150

B. Purpose for Submission:
To add Cefazolin to the Gram-Negative MIC/Combo Panels on the Synergies plus™ System

C. Analyte:
Cefazolin at 0.5 – 32 ug/mL

D. Type of Test:
Quantitative growth based AST test

E. Applicant:
Dade Behring – Dade MicroScan, Inc.

F. Proprietary and Established Names:
MicroScan® Synergies plus™ Gram-Negative MIC/Combo Panels

G. Regulatory Information:

1. Regulation section:
866.1645-Fully automated short-term incubation cycle antimicrobial susceptibility system
866.1640 – Antimicrobial Susceptibility test powder

2. Classification:
Class II

3. Product Code:
LON – Automated AST system short incubation
LRG-Instrument for Auto Reader &amp; Interpretation of Overnight Antimicrobial Susceptibility Systems
JWY – Manual Antimicrobial Susceptibility Test Systems

4. Panel:
83 Microbiology

H. Intended Use:

1. Intended use(s):
For use with MicroScan® Synergies plus™ panels read on the
WalkAway® -SI System (including upgraded WalkAway® -40 or
WalkAway® -96 to meet WalkAway® SI equivalence). MicroScan®
panels are designed for use in determining quantitative and/or qualitative
antimicrobial agent susceptibility and/or identification to the species level
of colonies, grown on solid media, of rapidly growing aerobic and
facultative anaerobic gram-negative bacilli. (Enterobacteriaceae, glucose
non-fermenters, and non-Enterobacteriaceae glucose fermenters).

2. Indication(s) for use:

{1}

Page 2 of 6

The testing of cefazolin at concentrations of 0.5-32 ug/ml to the gram negative test panel for testing Enterobacteriaceae at 4.5-18 hours or 16-20 hours for a overnight reading

3. Special condition for use statement(s):
- Do not report results for Klebsiella oxytoca.
- All Enterobacter spp., Citrobacter freundii, Morganella morganii, Proteus vulgaris, Providencia spp.. Serratia spp. or Yersinia enterolitica will not be reported in the Software or patient reports because of natural resistance.
- Turbidity method of inoculum preparation only.
- For prescription use only.

4. Special instrument Requirements:

I. Device Description:
The MicroScan® rapID/S plus™ Panel contains microdilutions of each antimicrobic in various concentrations on dehydrated and dried panels with Mueller Hinton Broth and various nutrients. Each panel contains two control wells: a no-growth control well (contains water only/no nutrients or broth), and a growth control well (contains test medium without antibiotic). The panel is rehydrated and inoculated at the same time with 0.1 ml of suspension prepared by the turbidity method (inoculum equated to a McFarland 0.5 standard, transferred to inoculum water containing pluronic-D/F-a wetting solution).

J. Substantial Equivalence Information:

1. Predicate device name(s):
MicroScan® rapID/S plus Gram Negative MIC/Combo Panels – gentamicin
MicroScan® Dried Gram-Negative MIC/Combo Panels

2. Predicate K number(s):
K020185
K862140

3. Comparison with predicate:

|  Similarities  |   |   |
| --- | --- | --- |
|  Item | Device | Predicate  |
|  Intended use | MicroScan® panels are designed for use in determining quantitative and/or qualitative antimicrobial agent susceptibility and/or identification to the species level of colonies, grown on solid media, of rapidly growing aerobic and facultative anaerobic gram-negative bacilli. (Enterobacteriaceae, glucose non-fermenters, and non- | same  |

{2}

Page 3 of 6

|   | Enterobacteriaceae glucose fermenters). |   |
| --- | --- | --- |
|  Specimen | Isolated colonies from culture used | Isolated colonies from culture used  |
|  Inoculum | Inoculum density to 0.5 McFarland standard | Inoculum density to 0.5 McFarland standard  |
|  Incubation | <16 hours
16 – 18 hours | < 16 hours
16-18 hours  |
|  Results | Quantitative with qualitative interpretations | Quantitative with qualitative interpretations  |
|  Technology | Growth based | Growth based  |
|  Differences  |   |   |
|  Item | Device | Predicate  |
|  Panels | Dried cefazolin at 0.5-32 ug/ml | Dried gentamicin at 0.12-32 ug/ml  |
|  Reading | Uses both a early read and overnight methods in the same system | Overnight system uses only the overnight reading methods.  |
|  Inoculum preparation | Turbidity method of inoculation only. | Inoculum prepared from isolated colonies using either the Turbidity method or Prompt® system  |
|  Instrument | WalkAway® -SI System or equivalent | autoScan® -4 or WalkAway®  |
|  Antibiotic | Cefazolin at 0.5-32 ug/mL | Different concentrations depending on the antibiotic  |

## K. Standard/Guidance Document Referenced (if applicable):

“Class II Special Controls Guidance Document: Antimicrobial Susceptibility Test (AST) Systems; Guidance for Industry and FDA”; NCCLS M7 (M100-S14) “Methods for Dilution Antimicrobial Susceptibility Tests for Bacteria That Grow Aerobically; Approved Standard”

## L. Test Principle:

The WalkAway® SI uses a Colorimetric Optics System consisting of a color wheel/lamp assembly and a Photosensor. There is an initial read at 2.5 hours with a possible final read at 4.5, 5.5, 6.5, 8, 10, 12, 16, or 18 hours (overnight instrument readings, manual readings) depending on the growth rate of the organism being tested. The initial read is subtracted from the final read to minimize variations from all components of the system. The time of final read is dependent on the growth rate of the organism and the sensitivity of the automatic reader since cell densities below $2 \times 10^{7}$ cells/ml are not detected. Reading considerations are built into the reading for faster growing and slower growing organisms. Organisms that do not reach a specific threshold at 4.5 hours have the minimum threshold raised at 5.5 hours. This allows for fermenters (faster grower organisms) to be read at 4.5, 5.5 or 6.5 hours and delay the reading of non-fermenters

{3}

Page 4 of 6

(slow growing) to 8, 10, 12 and up to 18 hours. Less than 16 hour readings use data from one read to the next to determine when the appropriate time for reading the MIC is. MICs that are clearly susceptible or clearly resistant can be reported earlier than those with MIC closer to the interpretative breakpoints. Indicator wells are also used to detect delayed mechanisms of resistance.

## M. Performance Characteristics (if/when applicable):

### 1. Analytical performance:

#### a. Precision/Reproducibility:

Reproducibility was demonstrated using 25 isolates tested at 3 sites on 3 separate days in triplicate. All isolates had a mode that was on scale. The study included the testing on the WalkAway® SI read at 4.5 to 18 hours if necessary, -20 hour readings and manual readings at 16-18 hours incubation. All results were &gt;95% reproducible. An additional study was performed on readings performed at &lt;16 hours on the WalkAway®SI with results that were &gt;95% reproducible.

#### b. Linearity/assay reportable range:

Not applicable

#### c. Traceability, Stability, Expected values (controls, calibrators, or method):

The recommended QC isolate was tested &gt;90 times with acceptable results on all testing days with the reference method. The percent that did not grow in the 4.5-16 hour window for the E. coli was 12.5%. Quality control results demonstrated the ability of the different reading parameters (manual and instrument) to produce acceptable results. There does appear to be a slight trend for the E. coli ATCC 25922 when read by the instrument in the 4.5 to 18 hour read times to be slightly more resistant.

|  Results  |   |   |   |   |   |   |
| --- | --- | --- | --- | --- | --- | --- |
|  Organism | Conc in ug/ml | # reference | MicroScan®  |   |   |   |
|   |  |  | Strictly >16 hour incubation readings |   | Read when ready (4.5-18 hours)  |   |
|   |  |  | Manual overnight | Instrument overnight | 4.5-15h instrument | 16-18 h instrument  |
|  E. coli ATCC 25922 Range 1-4 | ≤ 0.5 |  |  |  |  |   |
|   |  1 | 45 | 96 | 100 | 12 | 4  |
|   |  2 | 64 | 16 | 12 | 2 | 94  |
|   |  4 |  |  |  |  |   |
|   |  8 |  |  |  |  |   |

{4}

There is a trend when reading the read when ready feature for the results of the E. coli at the 16-18 hour read time to be more resistant than the other MicroScan® readings.

Inoculum density control: A turbidity meter was used for the turbidity inoculation method. Colony counts were also performed using the turbidity method when inoculating both the dried MicroScan® panels and the frozen reference panels. The turbidity method of inoculation for test and all QC strains tested had an average inoculum that was in the range of  $2.4 \times 10^{5}$  to  $5.9 \times 10^{5}$

# d. Detection limit:

Not applicable

# e. Analytical specificity:

Not applicable

# f. Assay cut-off:

Not applicable

# 2. Comparison studies:

# a. Method comparison with predicate device:

Clinical testing was performed at three sites using fresh isolates supplemented with stock isolates of Enterobacteriaceae. A comparison of the MicroScan® Dried Gram-Negative test panel results was made to the reference method conducted as recommended in the NCCLS standard M7-A6. Testing of the reference method and the MicroScan panels was performed at the same time. A challenge set was also tested at one site and compared to the reference broth dilution result mode that was determined by previous testing of each isolate multiple times in the recommended reference panel. The chart below demonstrates the performance when comparing the Synergies plus™ and the reference method using the NCCLS interpretive criteria. The Synergies plus™ contains readings that were performed at times between 4.5 and 18 hours of incubation with  $61\%$  providing those results in  $&lt; 16$  hours. An additional comparison was done with readings on the instrument after overnight incubation and also read manually when incubated 16-18 hours. Performance by these alternate reading methods was also acceptable with no apparent differences or trends.

|   | total | EA | %EA | Total evaluable | EA of evaluable | %EA | CA | %CA | #R | min | maj | vmj  |
| --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- |
|  Clinical | 444 | 435 | 98% | 308 | 301 | 97.7 | 414 | 93.2% | 142 | 29 | 0 | 1  |
|  Challenge | 75 | 73 | 97.3 | 68 | 66 | 97.1 | 68 | 90.7 | 11 | 5 | 1 | 1  |
|  Combined | 519 | 508 | 97.9 | 376 | 367 | 97.6 | 482 | 92.9 | 153 | 31 | 1 | 2  |

EA-Essential Agreement

CA-Category Agreement

R-resistant isolates

maj-major discrepancies

vmj-very major discrepancies

min-minor discrepancies

{5}

Page 6 of 6

Evaluations were also performed using the FDA approved interpretations (one well shift to less conservative) with acceptable performance also.

**b. Matrix comparison:**
Not applicable

**3. Clinical studies:**

**a. Clinical sensitivity:**
Not applicable

**b. Clinical specificity:**
Not applicable

**c. Other clinical supportive data (when a and b are not applicable):**
Not applicable

**4. Clinical cut-off:**
Not applicable

**5. Expected values/Reference range:**
$\leq 8$ (S), 16 (I), $\geq 32$ (R)
The interpretative criteria and QC are the same as recommended in NCCLS. All values will be included in the package insert.

**N. Conclusion:**
The submitted information in this premarket notification is complete and supports a substantial equivalence decision.

---

**Source:** [https://fda-staging.innolitics.com/device/K041150](https://fda-staging.innolitics.com/device/K041150)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
