K041088 · C.R. Bard, Inc. · LJS · Nov 1, 2004 · General Hospital
Device Facts
Record ID
K041088
Device Name
TRIFUSION, MODEL 0609190/0659350
Applicant
C.R. Bard, Inc.
Product Code
LJS · General Hospital
Decision Date
Nov 1, 2004
Decision
SESE
Submission Type
Abbreviated
Regulation
21 CFR 880.5970
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The Hickman® TriFusion™ Triple Lumen Long-Term Central Venous Catheter is indicated for use in attaining short term or long term vascular access for intravenous infusion therapy and blood sampling via the internal jugular vein, external jugular vein, and subclavian vein. All Hickman® TriFusion™ catheters are designed for apheresis, and the administration of I.V. fluids, blood products, drugs, and parenteral nutrition solutions, as well as blood withdrawal. The Hickman® TriFusion™ catheter incorporates three large, equal size lumens appropriate for apheresis procedures.
Device Story
Hickman® TriFusion™ Catheter is a 12 Fr, triple-lumen, radiopaque polyurethane long-term central venous catheter. Device features three equal-sized lumens with distal lumen extending beyond proximal lumens; proximal end includes luer connectors, occlusion clamps, and priming volume ID tags. Used by clinicians for apheresis, blood withdrawal, and administration of fluids/drugs/nutrition. Catheter is implanted to provide vascular access; healthcare providers monitor infusion/withdrawal via external luer connections. Benefits include multi-lumen capability for simultaneous procedures and apheresis compatibility.
Clinical Evidence
Bench testing only. Performance testing included dimensions, flow rate, tensile/elongation/stiffness, assembly/attachment strength, leakage, burst/collapse pressure, flexural fatigue, hemolysis, priming volume, PET cuff bond strength, radiopacity, tunneler security, and recirculation.
Technological Characteristics
12 Fr triple-lumen radiopaque polyurethane catheter. Features three equal-sized lumens. Includes luer connectors, occlusion clamps, and PET cuff. Complies with ISO 10555-1, ISO 10555-3, ISO 594-1, and ISO 594-2. Provided sterile.
Indications for Use
Indicated for patients requiring short or long-term vascular access for intravenous infusion therapy, blood sampling, apheresis, administration of I.V. fluids, blood products, drugs, and parenteral nutrition via internal jugular, external jugular, or subclavian veins.
Regulatory Classification
Identification
A percutaneous, implanted, long-term intravascular catheter is a device that consists of a slender tube and any necessary connecting fittings, such as luer hubs, and accessories that facilitate the placement of the device. The device allows for repeated access to the vascular system for long-term use of 30 days or more, and it is intended for administration of fluids, medications, and nutrients; the sampling of blood; and monitoring blood pressure and temperature. The device may be constructed of metal, rubber, plastic, composite materials, or any combination of these materials and may be of single or multiple lumen design.
Special Controls
*Classification.* Class II (special controls) Guidance Document: “Guidance on Premarket Notification [510(k)] Submission for Short-Term and Long-Term Intravascular Catheters.”
Predicate Devices
Hickman® Long-Term Central Venous Catheter (K830233)
Submission Summary (Full Text)
{0}------------------------------------------------
NOV - 1 2004
·
:
K041088 Additional Information
# Section 5 510(k) Summary of Safety and Effectiveness Information
## Hickman® TriFusion™ Catheters
### 5.1 Submitter Information:
| Submitter Name: | Bard Access Systems, Inc. (BAS)<br>(Wholly owned Subsidiary of C.R. Bard, Inc.) |
|----------------------|---------------------------------------------------------------------------------|
| Address: | 5425 W. Amelia Earhart Drive<br>Salt Lake City, UT 84116 |
| Telephone Number: | (801) 595-0700, Ext. 7136 |
| Fax Number: | (801) 595-5425 |
| Contact Person: | Lynn M. Kirchoff |
| Date of Preparation: | August 17, 2004 |
### 5.2 Device Name:
| Device Name: | Hickman® TriFusion™ Catheter |
|-----------------------|----------------------------------------------------------------------------------------------------------|
| Trade Name: | TriFusion™ |
| Common/Usual Name: | Long-Term Intravascular Catheter |
| Classification Name: | 80 LJS – Long-Term Intravascular Catheter<br>21 CFR 880.5970 - Class II<br>Implanted Blood Access Device |
| Classification Panel: | General Hospital |
#### 5.3 Predicate Device:
| Device Name: | Hickman® Long-Term Central Venous Catheter |
|-------------------------|----------------------------------------------------------------------------------------------------------|
| Trade Name: | Hickman® |
| Common/Usual Name: | Long-Term Intravascular Catheter |
| Classification Name: | 80 LJS – Long-Term Intravascular Catheter<br>21 CFR 880.5970 – Class II<br>Implanted Blood Access Device |
| Classification Panel: | General Hospital |
| Premarket Notification: | K830233, cleared for marketing on February 28, 1983 |
### 5.4 Device Description:
The device description of the subject Hickman TriFusion™ Catheter is as follows:
- The Hickman TriFusion™ Catheters are open-ended triple lumen radiopaque polyurethane . catheters.
- The Hickman TriFusion™ Catheters are 12 Fr triple lumen with up to 27 cm insertion length. .
- The Hickman TiFusion™ Catheters have three equal sized lumens with the distal lumen extending . beyond the proximal lumens.
- The proximal end of the TriFusion™ Catheter consists of three luer connectors, occlusion . clamps, and priming volume ID tags.
- Catheters are provided sterile in two kit configurations, an Intermediate Tray and a . Microintroducer (MI) Tray.
{1}------------------------------------------------
#### ર્સ્ડ Intended Use:
5.6
The Hickman® TriFusion™ Triple Lumen Long-Term Central Venous Catheter is indicated for use in attaining short term or long term vascular access for introvenons influsion them blood sampling va the internal jugular vent external pagare vent and subclani van
All Hickman® TriFusion™ catheters are designed for apheresis, and the administration of I.V. fluids, blood products, drugs, and parenteral nuritions, as well as blood withdrawal The Hickman® Trif usion™ catheter incorporates three large, equal size lumens appropriate for apheresis procedures.
This is the same intended use as the currently marketed 12.5 Fr T/L Hickman catheter covered by K830233, concurrence date 02/28/1983.
Technological Characteristics Summary:
510(k) Substantial Equivalence Decision Tree:
New device is compared to Marked Device?
Yes.
Does the new device have the same indication statement as the predicate?
Yes, with minor modifications to the indication verbiage.
Does the new device have the same technological characteristics, e.g. design, material, etc.?
Not in all regards. The TriFusion™ Catheters have some minor differences from the predicate T/L Hickman Catheters. However, the basic fundamental scientific technology of the catheter has not changed.
Could the new characteristics affect safety or effectiveness?
Yes. The new characteristics could affect the safety or effectiveness of the device.
Do the new characteristics raise new types of safety and effectiveness questions?
No. There are no new types of safety and effectiveness questions.
Do accepted scientific methods exist for assessing effects of the new characteristics?
Yes. Reliance was placed on recognized standards to evaluate the device's performance.
- Guidance on Premarket Notification [510(k)] Submission for Short-Term and Long-. Term Intravascular Catheters, dated 3/16/95
- ISO 10555-1:1997, Sterile, single-use intravascular catheters, Part 1. General ● requirements
- ISO 10555-3:1997, Sterile, single-use intravascular catheters, Part 3. Central . venous catheters
- ISO 594-1:1986, Conical fittings with a 6% (Luer) taper for syringes, needles and certain other medical equipment-Part 1: General Requirements
- ISO594-2:1986, Conical fittings with a 6% (Luer) taper for syringes, needles and certain other medical equipment-Part 2:Luer Fittings
{2}------------------------------------------------
Image /page/2/Picture/0 description: The image shows handwritten text on a white background. The text at the top appears to be a code or identifier, possibly "KC2/1088". Below this, there's the word "Kocijuss" and "p.3". The phrase "Additional Information" is at the bottom of the image.
Are performance data available to assess effects of new characteristics"
Yes. Ventication testing was preformed according to the referenced standards, as well as in accordance with in-house protocols. The test results met the requirements and were compared to the predicate devices.
### Do performance date demonstrate equivalence?
Yes. Performance data demonstrate that the TriFusion™ Catheters are substantially equivalent to the predicate 12.5 Fr T/L Hickman® Catheters.
### 5.7 Nonclinical Performance Testing
Testing was performed per FDA's Guidance on Premarket Notification [510(k)] Submission for Short-Term and Long-Term Intravascular Catheters, dated 3/16/95:
### Biocompatibility
### Guidance Testing Performed
- 1) Dimensions
- 2) Flow rate
- Tensile, Elongation and Stiffness (modulus) of catheter body 3)
- 4) Tensile strength of catheter body to hub [connector] attachment (Assembly)
- 5) Catheter tip (distal) attachment strength
- Leakage at hub 6)
- Catheter burst pressure (positive internal pressure) 7)
- 8) Catheter collapse (negative internal pressure)
- 9) Catheter flexural fatigue tolerance (Body & Tip)
### In-House Protocols
Additional Tests to Establish Safety & Effectiveness:
- 10) Hemolysis
- 11) Priming volume
- 12) PET cuff bond tensile strength
- 13) Radiopacity
- 14) Tunneler security test
- 15) Recirculation
#### ર-8 Conclusion:
The Hickman TriFusion™ catheter is substantially equivalent to the legally marketed predicate device, the 12.5 Fr T/L Hickman Long-Term Catheter, covered by K830233, concurrence date 02/28/1983.
{3}------------------------------------------------
DEPARTMENT OF HEALTH & HUMAN SERVICES
Image /page/3/Picture/1 description: The image shows the seal of the U.S. Department of Health & Human Services. The seal features a stylized eagle with three tail feathers, representing the three levels of government: federal, state, and local. The eagle is surrounded by the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" in a circular arrangement.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
NOV - 1 2004
Ms. Lynn M. Kirchoff Regulatory Affairs Specialist Bard Access Systems, Incorporated 5425 West Amelia Earhart Drive Salt Lake City, Utah 84116
Re: K041088
Trade/Device Name: Hickman® TriFusion™ Catheter Regulation Number: 880.5970 Regulation Name: Percutaneous Implanted Long-Term Intravascular Catheter Regulatory Class: II Product Code: LJS Dated: August 17, 2004 Received: August 18, 2004
Dear Ms. Kirchoff:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{4}------------------------------------------------
Page 2 - Ms. Kirchoff
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (240) 276-0115. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html
Sincerely vours,
Carl
Chin-Lin, Ph.D.
Chiu Lin, Ph.D. Director Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
# Indications for Use
510(k) Number (if known): ____K041088_________________________________________________________________________________________________________________________________________
Device Name: _Hickman® TriFusion™ Catheter____________________________________________________________________________________________________________________________________
Indications for Use:
The Hickman® TriFusion™ Triple Lumen Long-Term Central Venous Catheter is indicated for use in attaining short term or long term vascular access for intravenous infusion therapy and blood sampling via the internal jugular vein, external jugular vein, mader morap your . All Hickman® TriFusion™ catheters are designed for apheresis, and the administration of I.V. fluids, blood products, drugs, and parenteral nutrition solutions, as well as blood withdrawal. The Hickman® TriFusion™ catheter incorporates three large, equal size lumens appropriate for apheresis procedures.
Prescription Use \/
(Part 21 CFR 801 Subpart D)
AND/OR
Over-The-Counter Use (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Antony V. me
(Division Sign-Off) (Division Sign-On)
Division of Anesthesiology, General Hospital, Infection Control, Dental Devices
510(k) Number: K414988
Page of
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.