← Product Code [IOR](/productcode/IOR) · K041072

# KUSCHALL CHAMPION, KUSCHALL COMPACT, KUSCHALL ULTRA-LIGHT (K041072)

_Invacare Corporation · IOR · May 12, 2004 · Physical Medicine · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K041072

## Device Facts

- **Applicant:** Invacare Corporation
- **Product Code:** [IOR](/productcode/IOR.md)
- **Decision Date:** May 12, 2004
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 890.3850
- **Device Class:** Class 1
- **Review Panel:** Physical Medicine

## Indications for Use

The Kuschall Champion, Compact and Ultra-light are mechanical wheelchairs intended to provide mobility to persons restricted to a sitting position.

## Device Story

Manually operated, user-propelled mechanical wheelchairs; intended for indoor/outdoor mobility for persons restricted to seated position. Construction includes metal frame (aluminum or carbon fiber), large rear wheels with hand rims, and front pivoting casters. Kuschall Champion features horizontal folding mechanism via hinged connecting arms; Kuschall Compact and Ultra-Light feature traditional vertical 'X' cross-brace folding mechanism. Devices are non-rigid and foldable for transport. Operated by user or caregiver; no electronic components or software.

## Clinical Evidence

Bench testing only. Compliance with EN 12183, prEN 12182, EN 1041, ISO 7176 (parts 1, 3, 8, 11, 13, 15, 16), ISO 8191 (parts 1, 2), ISO 9999, ISO 6440, and CA Bulletin 117. Testing confirmed requirements for static stability, brake efficiency, and static/impact/fatigue strength were met.

## Technological Characteristics

Manual wheelchair; aluminum or carbon fiber frame; folding mechanism (horizontal or vertical X-brace); rear wheels with hand rims; front pivoting casters. Standards: EN 12183, prEN 12182, ISO 7176 series (stability, brakes, strength, friction, labeling, ignition resistance), ISO 8191 (upholstery ignitability), ISO 9999, ISO 6440, CA Bulletin 117. No energy source; no software.

## Regulatory Identification

A mechanical wheelchair is a manually operated device with wheels that is intended for medical purposes to provide mobility to persons restricted to a sitting position.

## Predicate Devices

- Invacare Action AF-1 (now known as 'The Spyder') ([K000174](/device/K000174.md))
- Invacare MVP ([K914553](/device/K914553.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

MAY 1 2 2004

Image /page/0/Picture/1 description: The image shows the Invacare logo. The logo consists of the word "INVACARE" in bold, white, sans-serif letters, set against a black oval background. The oval has a double outline, with a thin white line separating the black fill from the outer black border.

K044072
Invacare Corporation
510(k) Submission Page 7

## 510(K) SUMMARY KUSCHALL CHAMPION, COMPACT AND ULTRA-LIGHT

This summary of 510(k) safety and effectiveness information is being supplied in accordance with the requirements of the SMDA of 1990 and 21 CFR 807.92

The assigned 510(k) number is:

Date: april 23, 2004

Submitted by: Kuschall Design AG Ringstrasse 15 Allschwill, Switzerland CH-4123

Contact Telephone: 440-329-6356 Fax: 440-326-3607

Contact Person: Carroll L. Martin, Regulatory Affairs

Trade Name: Kuschall Champion, Kuschall Compact and Kuschall Ultra-Light

Common Name: Manual wheelchair

Classification Name: Wheelchair, mechanical per 21 CFR section 890.3850

# Legally Marketed Predicate Device(s): Invacare Action AF-1 (now known as "The Spyder"), K000174, February 18, 2000 Invacare MVP, K914553, October 28, 1991

Device Description: The Kuschall Champion, Compact and Ultra-Light are manually operated, userpropelled mechanical wheelchairs. Their intended use is to provide mobility to persons restricted to a seated position.

Each wheelchair consists primarily of a metal frame that is constructed of round and oval aluminum tubing that is welded, large rear wheels with hand rims for propelling the chair and smaller front pivoting casters for steering and turning. The Kuschall Champion is also available with a carbon fiber frame. The products are designed to be lightweight, everyday wheelchairs for both indoor and outdoor use. They are all folding, non-rigid wheelchairs.

Intended Use: The intended use of the Kuschall Champion, Compact and Ultra-Light manual wheelchairs is to provide mobility to persons restricted to a seated position.

Substantial Equivalence: Product that is substantially equivalent to the Kuschall Champion is the Invacare Action AF-1, (K000174, Fcbruary 18, 2000) now known as "The Spyder". These products are both manually operated, user propelled mechanical whechchairs with the intended use of providing

INVACARE CORPORATION

{1}------------------------------------------------

Image /page/1/Picture/0 description: The image shows the Invacare logo. The logo is in black and white and features the word "INVACARE" in bold, white letters against a black oval background. The oval is slightly tilted, giving the logo a dynamic appearance.

mobility to persons restricted to a seated position. Both products consist basically of a metal frame, larger rear wheels with hand rims for propelling the wheelchair and smaller front pivoting casters for steering and turning.

The Kuschall Champion folds using two connecting arms that are hinged together in the center of the wheelchair, but are also hinged at the side frames. As such, the chair essentially folds "horizontally", the same manner in which the Invacare AF-1 folds, as opposed to the vertical folding of a cross brace wheelchair.

The folding mechanism allows the backrest and the seat to be folded simultaneously. The frame remains closed by the use of a Velcro strap while transferring the chair into the car.

The Kuschall Compact and Ultra-Light are similar to the Invacare MVP (K914553, October 28, 1991) in that all three wheelchairs are manually-operated, user propelled mechanical wheelchairs with the intended use of providing mobility to persons restricted to a seated position. All three devices consist basically of a metal frame, larger rear wheels with hand rims for propelling the wheelchair and smaller front pivoting casters for steering and turning.

The Kuschall Compact, Ultra-Light and Invacare MVP all have the ability to fold and fold in the traditional "vertical" manner. The folding mechanism is of the "X" cross brace design whereby the lower frames cross, or form an "X" in the center and are hinged at a single pivot point for folding. These chairs fold by collapsing "vertically" about the center pivot point

Additional similarities and differences are listed in the substantial equivalence matrices for each compared product. None of the differences alter the intended function and use of the devices nor do they raise any new questions of safety and effectiveness.

Software: No software is associated with these devices.

Performance Data: The following performance standards were used during the development of the Kuschall Champion, Compact and Ultra-Light and it has been determined that each wheelchair met the applicable specified requirements:

- -EN 12183: The Swedish Handicap Institute's Specification of Requirements for Manually Propelled Wheelchairs
- prEN 12182: Technical Aids for Disabled Persons General Requirements and Test -Methods
- EN 1041: Information Supplied by the Manufacturer with Medical Devices -
- -ISO 7176-1: Wheelchairs -- Part 1: Determination of Static Stability
- ISO 7176-3: Wheelchairs Part 3: Determination of Efficiency of Brakes -
- -ISO 7176-8: Wheelchairs - Part 8: Requirements and Test Methods for Static, Impact and Fatigue Strengths
- ISO 7176-11: Wheelchairs -- Part 11: Test Dummics -
- ー ISO 7176-13: Wheelchairs - Part 13: Determination of Coefficient of Friction of Test Surfaces

### INVACARE CORPORATION

{2}------------------------------------------------

Image /page/2/Picture/0 description: The image shows the Invacare logo. The logo is in black and white, with the word "INVACARE" in a bold, sans-serif font. The word is placed inside of a black oval shape. The oval shape has a white outline.

- ISO 7176-15: Wheelchairs Part 15: Requirements for Information Disclosure, -Documentation and Labeling
- ISO 7176-16: Wheelchairs Part 16: Resistance to Ignition of Upholstered Parts --Requirements and Test Methods
- ISO 8191-1: Furniture Part 1: Assessment of The Ignitability of Upholstered Furniture; -Ignition Source: Smoldering Cigarette
- ISO 8191-2: Furniture -- Part 2: Assessment of Ignitability of Upholstered Furniture; -Ignition Source: Match-Flame Equivalent
- ISO 9999: Technical Aids for Disabled Persons Classification ェ
- ISO 6440: Wheelchairs Nomenclature Terms and Definitions -
- CA Bulletin 117 Testing the Flame Retardance of Resilient Filling Materials Used in ﯿ Upholstered Furniture

#### INVACARE CORPORATION

Image /page/2/Picture/11 description: The image is a simple, minimalist composition featuring a small, dark spot against a bright white background. The spot is located in the upper-left quadrant of the image, creating a sense of asymmetry. The contrast between the dark spot and the white background is stark, drawing the viewer's attention to the small detail. The image is simple and lacks any other elements.

{3}------------------------------------------------

Image /page/3/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is a stylized image of an eagle with three stripes representing the agency's mission to protect the health of all Americans and provide essential human services.

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

MAY 1 2 2004

Carroll I.. Martin Regulatory Affairs Invacare Corporation One Invacare Way P.O. Box 4028 Elyria, Ohio 44036-2125

Re: K041072

Trade/Device Names: Kuschall Champion, Kuschall Compact and Kuschall Ultra-light Regulation Number: 21 CFR 890.3850 Regulation Name: Mechanical wheelchair Regulatory Class: I Product Code: IOR Dated: April 23, 2004 Received: April 26, 2004

Dear Mr. Martin:

We have reviewed your Scction 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adultcration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

{4}------------------------------------------------

This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (301) 594-4659. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html

Sincerely yours,

for Mark A. Melkersen

Celia M. Witten, Ph.D., M.D. Director Division of General, Restorative and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{5}------------------------------------------------

# Indications for Use

Invacare Corporation 510(k) Submission Page 5

510(k) Number (if known):

Device Name: Kuschall Champion Kuschall Compact Kuschall Ultra-light

Indications for Use:

The Kuschall Champion, Compact and Ultra-light are mechanical wheelchairs intended to provide mobility to persons restricted to a sitting position.

Prescription Use (Part 21 CFR 801 Subpart D) AND/OR

Over-The-Counter Use (21 CFR 807 Subpart C)

1

r

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of OORH, Office of Device Evaluation (ODE)

(Division Sign-Off)
Division of General, Restorative,
and Neurological Devices

510(k) Number K041072

Page 1 of

---

**Source:** [https://fda-staging.innolitics.com/device/K041072](https://fda-staging.innolitics.com/device/K041072)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).