MODIFICATION TO LIFESTENT SDS AND XL SDS BILIARY ENDOPROSTHESIS
K041032 · Orbus Medical Technologies · FGE · May 10, 2004 · Gastroenterology, Urology
Device Facts
Record ID
K041032
Device Name
MODIFICATION TO LIFESTENT SDS AND XL SDS BILIARY ENDOPROSTHESIS
Applicant
Orbus Medical Technologies
Product Code
FGE · Gastroenterology, Urology
Decision Date
May 10, 2004
Decision
SESU
Submission Type
Special
Regulation
21 CFR 876.5010
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The Orbus LifeStent XL SDS Biliary Endoprosthesis is intended for the palliation of malignant neoplasms in the biliary tree.
Device Story
The LifeStent XL SDS Biliary Endoprosthesis is a balloon-expandable stent designed to maintain patency of bile ducts obstructed by malignant tumors. The device consists of a laser-cut 316L stainless steel stent mounted on a percutaneous transluminal angioplasty (PTA) balloon catheter. During a clinical procedure, the system is inserted percutaneously to the diseased site; the balloon is inflated to expand the stent, which remains permanently implanted to keep the duct open. The device is intended for use by physicians in a clinical setting. It provides mechanical support to the biliary tree to alleviate obstructions caused by neoplasms, thereby improving bile flow and patient comfort.
Clinical Evidence
Bench testing only. Performance data included in vitro tests for deployment, expansion force, compression force, dimensions, corrosion susceptibility, visual appearance, balloon performance, expansion damage, and balloon tensile strength.
Technological Characteristics
Balloon-expandable stent; 316L stainless steel; laser-cut open lattice design; mounted on PTA balloon catheter; available in 100mm expanded diameter and 17, 27, 37mm lengths; permanent implant; sterile.
Indications for Use
Indicated for the palliation of malignant neoplasms in the biliary tree. Prescription use only.
Regulatory Classification
Identification
A biliary catheter and accessories is a tubular flexible device used for temporary or prolonged drainage of the biliary tract, for splinting of the bile duct during healing, or for preventing stricture of the bile duct. This generic type of device may include a bile collecting bag that is attached to the biliary catheter by a connector and fastened to the patient with a strap.
Special Controls
*Classification.* Class II (special controls). The device, when it is a bile collecting bag or a surgical biliary catheter that does not include a balloon component, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 876.9.
LifeStent XL 8 and 9mm Biliary Endoprosthesis (K040550)
Submission Summary (Full Text)
{0}------------------------------------------------
MAY 1 0 2004
K041032
Page 1 of 2
| 1. | Submitter: | Orbus Medical Technologies, Inc<br>5363 NW 35th Ave<br>Fort Lauderdale, Florida 33309<br>Phone: (954) 730-0711<br>Fax: (954) 730-7601 |
|----|------------------------|---------------------------------------------------------------------------------------------------------------------------------------|
| 2. | Contact: | Jim Clossick<br>Director of Quality Assurance and Regulatory Affairs |
| 3. | Date Prepared: | 3/1/04 |
| 4. | Device Trade Name: | LifeStent XL SDS Biliary Endoprosthesis |
| 5. | Device Common Name: | Biliary stent |
| 6. | Device Classification: | Biliary Catheter (78 FGE) |
| 7. | Predicate Devices: | LifeStent SDS Biliary Endoprosthesis |
510(k) Statement
# Description: 8.
The LifeStent XL SDS Biliary Endoprosthesis consists of a balloon expandable stent which is I ne Enehein AL SDD Dipercutaneous transluminal angioplasty (PTA) balloon catheter which acts as the delivery system. The stent is a permanently implanted device used to maintain patency of a major bile duct obstructed by tissue of an impinging tumor. The flexible, balloon expandable stent is made by laser cutting an open lattice design into a 316L stainless steel tube. The stent is supplied mounted onto a PTA balloon catheter, inserted percutaneously to the diseased site, and expanded by balloon inflation. The systems are available with expanded diameter of 100mm at stent lengths of 17, 27, 37mm.
# Intended Use: 9.
The LifeStent XL SDS Billary Endoprosthesis are intended for the palliation of malignant neoplasms in the biliary tree.
# Technological Characteristics: 10.
Comparisons of the new and predicate devices were designed to show that the technical characteristics such as materials, performance properties, biocompatibility, method of sterilization, and packaging are identical or substantially equivalent.
{1}------------------------------------------------
# Performance Data: 11.
Orbus protocols ensure that the LifeStent XL SDS Biliary Endoprosthesis perform in a manner substantially equivalent to the predicate devices during in vitro tests such as deployment, expansion force, compression force, dimensions, corrosion susceptibility, visual appearance, balloon performance, expansion damage, and balloon tensile strength.
# Conclusion 12.
Since the LifeStent XL SDS Biliary Endoprosthesis, comprised of the Ø 8-12 mm LifeStent (K023121) and the Opti-Plast XT catheter (K941706), has the same intended use, identical material properties, similar performance properties, packaging, and sterilization methods, it may be considered substantially equivalent to the predicate devices cited in (7). These devices are a line extension to the LifeStent XL 8 and 9mm Biliary Endoprosthesis previously cleared per K040550.
{2}------------------------------------------------
Image /page/2/Picture/1 description: The image shows a circular seal with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is a stylized image of an eagle with its wings spread. The eagle is facing to the right and has three wavy lines below it.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
MAY 1 0 2004
Mr. Jim Clossick Director of Quality Assurance and Regulatory Affairs Orbus Medical Technologies 5363 NW 35th Avenue Fort Lauderdale, Florida 33309
Re: K041032
Trade/Device Name: Orbus LifeStent XL SDS Biliary Endoprosthesis Regulation Number: 21 CFR §876.5010 Regulation Name: Biliary catheter and accessories Regulatory Class: II Product Code: 78 FGE Dated: April 19, 2004 Received: April 21, 2004
Dear Mr. Clossick:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the cnactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act and the limitations described below. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
The Office of Device Evaluation has determined that there is a reasonable likelihood that this device will be used for an intended use not identified in the proposed labeling and that such use could cause harm. Therefore, in accordance with Section 513(i)(1)(E) of the Act, the following limitation must appear in the Warnings section of the device's labeling:
The safety and effectiveness of this device for use in the vascular system have not been established.
Furthermore, the indication for biliary use must be prominently displayed in all labeling, including pouch, box, and carton labels, instructions for use, and other promotional materials, in close proximity to the trade name, of a similar point size, and in bold print.
{3}------------------------------------------------
Page 2 - Mr. Jim Clossick
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that I lease or acresed a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
The FDA finding of substantial equivalence of your device to a legally marketed predicate device The I Dr. Initime of vablar your device and permits your device to proceed to the market. This resure in a classis. In a base in marketing your device as described in your Section 510(k) premarket notification if the limitation statement described above is added to your labeling.
Please note that the above labeling limitations are required by Section 513(i)(1)(E) of the Act. Therefore, a new 510(k) is required before these limitations are modified in any way or removed from the device's labeling.
If you desire specific information about the application of other labeling requirements to your device (2) CFR Part 801), please contact the Office of Compliance at (301) 594-4616. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International, and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html.
Sincerely yours,
Phillips
Tillman, Ph.D. Donna-Acting Director Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
Page_ 1 of 1
510(k) Number: K041032
Device Name: Orbus LifeStent XL SDS Biliary Endoprosthesis
FDA's Statement of the Indications For Use for device:
The Orbus LifeStent XL SDS Biliary Endoprosthesis is indicated for the palliation of malignant neoplasms in the biliary tree.
Prescription Use
(Per 21 CFR 801.109) √
OR
Over-the-Counter Use _________________________________________________________________________________________________________________________________________________________
David A. Logsdon
(Division Sign-Off)
(Division of Reproductive, Abdominal, and Roon of Reproductive
and Radiological Devices 510(k) Number
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.