← Product Code [JGS](/productcode/JGS) · K040977

# GENCHEM ELECTROLYTE BUFFER (K040977)

_Genchem, Inc. · JGS · Dec 27, 2004 · Clinical Chemistry · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K040977

## Device Facts

- **Applicant:** Genchem, Inc.
- **Product Code:** [JGS](/productcode/JGS.md)
- **Decision Date:** Dec 27, 2004
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 862.1665
- **Device Class:** Class 2
- **Review Panel:** Clinical Chemistry

## Indications for Use

GenChem ISE Electrolyte Buffer, when used in conjunction with the GenChem ISE Electrolyte Reference, GenChem CO2 Acid Reagent, GenChem CO2 Alkaline Buffer, GenChem Wash Concentrate, and appropriate Calibrators or Calibration Standards, is intended for the quantitative determination of sodium, potassium, chloride, and total CO2 in serum and plasma, and sodium, potassium and chloride in urine, and Reference will also determine calcium in serum, plasma and urine. Sodium results are for the diagnosis and treatment of aldosteronism (excessive secretion of the hormone aldosterone), diabetes insipidus (chronic excretion of large amounts of dilute urine, accompanied by extreme thirst), adrenal hypertension, Addison's disease (caused by the destruction of the adrenal glands), dehvdration, inappropriate antidiuretic hormone secretion, or other diseases involving electrolyte imbalance. Potassium results are used to monitor electrolyte imbalance in the diagnosis and treatment of diseases and conditions characterized by low or high blood potassium levels. Chloride results are used in the treatment of electrolyte and metabolic disorders such as cystic fibrosis and diabetic acidosis. Carbon dioxide results are used in the diagnosis and treatment of numerous and potentially serious disorders associated with changes in the body's acid-base balance. Calcium results are used in the diagnosis and treatment of parathyroid diseases, chronic renal disease and tetany (intermittent muscular contractions or spasm).

## Device Story

The GenChem ISE Electrolyte Buffer Kit is a liquid reagent system used on the Beckman SYNCHRON CX3 clinical chemistry analyzer. It facilitates the quantitative measurement of electrolytes (Na, K, Cl, CO2, Ca) in biological fluids (serum, plasma, urine, CSF). The device operates via Ion Specific Electrodes (ISE); samples are mixed with high ionic strength buffer to stabilize activity coefficients. For CO2, a secondary reaction with acid reagent converts bicarbonate to CO2 gas, which diffuses through a silicone membrane to alter the pH of an alkaline buffer; the rate of pH change is proportional to total CO2 concentration. The system is intended for use by laboratory professionals in clinical settings. Results are used by clinicians to diagnose and manage electrolyte imbalances, acid-base disorders, and endocrine conditions. The kit provides standardized, automated analysis, enabling rapid clinical decision-making regarding patient hydration, metabolic status, and renal function.

## Clinical Evidence

Bench testing only. Precision/reproducibility evaluated per NCCLS EP5-A (N=60). Linearity evaluated per NCCLS EP6-A using Least Squares method (R2=1.000 for all analytes). Analytical specificity evaluated per NCCLS EP7-A; no interference observed from hemoglobin (500 mg/dL), bilirubin (20 mg/dL), or lipemia (1800 mg/dL). Method comparison study performed on 80 adult patient samples (serum, plasma, urine, CSF) comparing GenChem reagents to Beckman reagents via least squares linear regression; results showed high correlation (r=0.935–1.000).

## Technological Characteristics

Liquid reagent (Tris-Phosphate 1.5 M). Sensing principle: Potentiometric Ion Specific Electrodes (ISE) and pCO2 electrode. Dimensions/form factor: Kit components for use on Beckman SYNCHRON CX3 analyzer. Storage: Room temperature. Analytical method: Nernst equation-based potential measurement.

## Regulatory Identification

A sodium test system is a device intended to measure sodium in serum, plasma, and urine. Measurements obtained by this device are used in the diagnosis and treatment of aldosteronism (excessive secretion of the hormone aldosterone), diabetes insipidus (chronic excretion of large amounts of dilute urine, accompanied by extreme thirst), adrenal hypertension, Addison's disease (caused by destruction of the adrenal glands), dehydration, inappropriate antidiuretic hormone secretion, or other diseases involving electrolyte imbalance.

## Predicate Devices

- Electrode, Ion Specific, Sodium, Potassium, Chloride, CO2, Calcium ([K801896](/device/K801896.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}

Page 1 of 8

# 510(k) SUBSTANTIAL EQUIVALENCE DETERMINATION DECISION SUMMARY DEVICE ONLY TEMPLATE

A. 510(k) Number: k040977

B. Purpose For Submission:

Premarket Notification 510(k) for GenChem, Inc. intentions to manufacture and market the GenChem ISE ELECTROLYTE BUFFER Kit.

C. Analyte:

Sodium, Potassium, Chloride Total CO₂ and Calcium

D. Type of Test: Quantitative ISE

E. Applicant: GenChem, Inc.

F. Proprietary and Established Names:

GenChem, Inc., ISE Electrolyte Buffer Kit

G. Regulatory Information:

Regulation section:

1. Regulation section:

21 CFR §862.1665 - Sodium test system.
21 CFR §862.1600 - Potassium test system.
21 CFR §862.1170 - Chloride test system.
21 CFR §862.1160 - Bicarbonate/carbon dioxide test system.
21 CFR §862.1145 - Calcium test system.

2. Classification:

Class II

3. Product Code:

JGS (sodium)
CEM (potassium)
CGZ (chloride)
JFL (bicarbonate/ carbon dioxide)
JFP (calcium)

{1}

Page 2 of 8

4. Panel:

75 (Chemistry)

H. Intended use(s):

1. Intended use(s)

GenChem ISE ELECTROLYTE BUFFER, when used in conjunction with the GenChem ISE Electrolyte Reference, GenChem CO₂ Acid Reagent, GenChem CO₂ Alkaline Buffer, GenChem Wash Concentrate, and appropriate Calibrators or Calibration Standards, is intended for the quantitative determination of Sodium, Potassium, Chloride and Total CO₂ in serum and plasma, Sodium, Potassium and Chloride in urine, and Chloride in cerebrospinal fluid on the Beckman SYNCHRON CX3 System. On appropriately configured SYNCHRON CX3 System, GenChem ISE Electrolyte Reference will also determine Calcium in serum, plasma and urine.

Sodium results are for the diagnosis and treatment of aldosteronism (excessive secretion of the hormone aldosterone), diabetes insipidus (chronic excretion of large amounts of dilute urine, accompanied by extreme thirst), adrenal hypertension, Addison’s disease (caused by the destruction of the adrenal glands), dehydration, inappropriate antidiuretic hormone secretion, or other diseases involving electrolyte imbalance. Potassium results are used to monitor electrolyte imbalance in the diagnosis and treatment of diseases and conditions characterized by low or high blood potassium levels. Chloride results are used in the treatment of electrolyte and metabolic disorders such as cystic fibrosis and diabetic acidosis. Carbon dioxide results are used in the diagnosis and treatment of numerous and potentially serious disorders associated with changes in the body’s acid-base balance. Calcium results are used in the diagnosis and treatment of parathyroid diseases, chronic renal disease and tetany (intermittent muscular contractions or spasm).

2. Indication(s) for use:

GenChem ISE ELECTROLYTE BUFFER, when used in conjunction with the GenChem ISE Electrolyte Reference, GenChem CO₂ Acid Reagent, GenChem CO₂ Alkaline Buffer, GenChem Wash Concentrate, and appropriate Calibrators or Calibration Standards, is intended for the quantitative determination of Sodium, Potassium, Chloride and Total CO₂ in serum and plasma, Sodium, Potassium and Chloride in urine, and Chloride in cerebrospinal fluid on the Beckman SYNCHRON CX3 System. On appropriately configured SYNCHRON CX3 System, GenChem ISE Electrolyte Reference will also determine Calcium in serum, plasma and urine.

Sodium results are for the diagnosis and treatment of aldosteronism (excessive secretion of the hormone aldosterone), diabetes insipidus (chronic excretion of large amounts of dilute urine, accompanied by extreme thirst), adrenal hypertension, Addison’s disease (caused by the destruction of the adrenal glands), dehydration, inappropriate antidiuretic hormone secretion, or other diseases involving electrolyte imbalance. Potassium results

{2}

are used to monitor electrolyte imbalance in the diagnosis and treatment of diseases and conditions characterized by low or high blood potassium levels. Chloride results are used in the treatment of electrolyte and metabolic disorders such as cystic fibrosis and diabetic acidosis. Carbon dioxide results are used in the diagnosis and treatment of numerous and potentially serious disorders associated with changes in the body's acid-base balance. Calcium results are used in the diagnosis and treatment of parathyroid diseases, chronic renal disease and tetany (intermittent muscular contractions or spasm).

3. Special condition for use statement(s): For Prescription Use.
4. Special instrument Requirements: Beckman SYNCHRON CX3 System.

# I. Device Description:

Sodium, Potassium, Chloride,  $\mathrm{CO}_{2}$  and Calcium are determined by the use of Ion Specific Electrodes, the conductivity of which is proportional to the concentration of electrolyte in the sample which is mixed with high ionic strength buffer using the ISE Solution as the reference.

# J. Substantial Equivalence Information:

1. Predicate device name(s): Beckman ISE Buffer
2. Predicate K number(s): (k801896)
3. Comparison with Predicate:

|  Device Name | GenChem ISE Electrolyte Buffer Kit | Predicate Device Beckman ISE Buffer  |
| --- | --- | --- |
|  510(k) Number | (k040977) | (k801896)  |
|  Chemical Principle | Measures electrolyte activity by use of ion specific electrode | Measures electrolyte activity by use of ion specific electrode  |
|  Intended Use | For the quantitative determination of Na, K, CL, CO2and Calcium in serum, plasma or urine. | For the quantitative determination of Na, K, CL, CO2and Calcium in serum, plasma or urine.  |
|  Format | Liquid ready to use | Liquid ready to use  |
|  Composition | Tris-Phosphate 1.5 M | Tris-Phosphate 1.5 M  |
|  Linearity | Sodium 0-200 mmol/L Potassium 0.9-15.2 mmol/L Chloride 0-150 mmol/L CO2 0-40 mmol/L Calcium 0.8-14.3 mg/dL | Sodium 0-200 mmol/L Potassium 0.9-15.2 mmol/L Chloride 0-150 mmol/L CO2 0-40 mmol/L Calcium 0.8-14.3 mg/dL  |
|  Storage | Room Temperature | Room Temperature  |

{3}

# K. Standard/Guidance Document Referenced (if applicable):

Within-Day and Day-to-Day precision was determined according to NCCLS EP5-A. Linearity was performed according to NCCLS EP6-A Guideline. Analytical specificity Determined according to NCCLS EP7-A.

# L. Test Principle:

The sample is mixed with the high ionic strength ISE Electrolyte Buffer. This dilution minimizes the variation in the activity coefficients of the analytes to be measured. As the sample passes through the flow cell, a potential is generated at the surface of the ion selective electrodes. The calcium, chloride, potassium and sodium concentrations of the sample can then be determined from these potentials using the Nerst equation.

Before the sample leaves the flow cell, it is further diluted with the CO2 Acid Reagent. The acid in this reagent converts the bicarbonate in the sample to carbonic acid, which escapes the solution in the form of carbon dioxide. Some of this carbon dioxide transverses the silicone membrane of the CO2 electrode and lowers the pH of the CO2 Alkaline Buffer. The rate of this pH exchange is proportional to the amount of total CO2 in the sample.

# M. Performance Characteristics (if/when applicable):

# 1. Analytical performance:

# a. Precision/Reproducibility:

Serum and CSF controls, and urine pools were each assayed for calcium, chloride, potassium, sodium, and total CO2 twice per day in triplicate on a SYNCHRON CX3 System. Data were collected on ten different days over a thirty day period.

|  Analyte |   |   | Within Run |   | Total  |   |
| --- | --- | --- | --- | --- | --- | --- |
|  Sample | n | mean | SD | %CV | SD | %CV  |
|  Calcium in mg/dL  |   |   |   |   |   |   |
|  Serum 1 | 60 | 7.9 | 0.16 | 2.1 | 0.15 | 1.9  |
|  Serum 2 | 60 | 11.1 | 0.09 | 0.8 | 0.11 | 1.0  |
|  Serum 3 | 60 | 14.2 | 0.13 | 0.9 | 0.16 | 1.1  |
|  Urine 1 | 59 | 3.4 | 0.12 | 3.5 | 0.13 | 3.9  |
|  Urine 2 | 60 | 10.9 | 0.33 | 3.1 | 0.30 | 2.8  |
|  Chloride in mmol/L  |   |   |   |   |   |   |
|  Serum 1 | 60 | 84 | 1.16 | 1.4 | 1.21 | 1.4  |
|  Serum 2 | 60 | 103 | 0.60 | 0.6 | 1.04 | 1.0  |
|  Serum 3 | 60 | 122 | 1.01 | 0.8 | 1.34 | 1.1  |
|  Urine 1 | 59 | 64 | 1.16 | 1.8 | 1.21 | 1.9  |
|  Urine 2 | 60 | 235 | 2.33 | 1.0 | 5.67 | 2.4  |
|  CSF 1 | 58 | 117 | 1.37 | 1.2 | 1.65 | 1.4  |

{4}

CSF 2 58 99 1.00 1.0 1.45 1.5

# Potassium in mmol/L

|  Serum 1 | 60 | 2.6 | 0.021 | 0.8 | 0.034 | 1.3  |
| --- | --- | --- | --- | --- | --- | --- |
|  Serum 2 | 60 | 5.2 | 0.031 | 0.6 | 0.040 | 0.8  |
|  Serum 3 | 60 | 7.8 | 0.074 | 0.9 | 0.083 | 1.1  |
|  Urine 1 | 59 | 27.1 | 0.23 | 0.9 | 0.27 | 1.0  |
|  Urine 2 | 60 | 123.7 | 1.67 | 1.4 | 1.67 | 1.4  |

# Sodium in mmol/L

|  Serum 1 | 60 | 114 | 1.71 | 1.5 | 1.51 | 1.3  |
| --- | --- | --- | --- | --- | --- | --- |
|  Serum 2 | 60 | 144 | 0.87 | 0.6 | 0.94 | 0.7  |
|  Serum 3 | 60 | 173 | 1.54 | 0.9 | 1.59 | 0.9  |
|  Urine 1 | 59 | 45 | 2.02 | 4.4 | 1.78 | 3.9  |
|  Urine 2 | 60 | 159 | 2.19 | 1.4 | 1.98 | 1.2  |

# Total  $\mathbf{CO}_2$  in mmol/L

|  Serum 1 | 60 | 13 | 0.40 | 3.2 | 0.38 | 3.0  |
| --- | --- | --- | --- | --- | --- | --- |
|  Serum 2 | 60 | 22 | 0.32 | 1.4 | 0.36 | 1.6  |
|  Serum 3 | 60 | 31 | 0.53 | 1.7 | 0.66 | 2.1  |

# b. Linearity/assay reportable range:

Linearity was performed according to NCCLS Guideline EP6-A. Commercially available linearity standards were analyzed in triplicate on the Beckman CX3 and the results analyzed by the Least Squares method. The results are shown below. Specimens exceeding these limits should be diluted with normal saline and reanalyzed. Multiply the result by the appropriate dilution factor.

|   | Intercept | Slope | R2 | Sey | Range  |
| --- | --- | --- | --- | --- | --- |
|  |   |   |   |   |   |
|  Calcium | 0.34 | 0.911 | 1.000 | 0.35 | 0.8 – 14.3 mg/dL  |
|  |   |   |   |   |   |
|  Na | 2.37 | 0.968 | 1.000 | 2.61 | 0 – 200 mmol/L  |
|  |   |   |   |   |   |
|  K | -0.09 | 1.017 | 1.000 | 0.03 | 0.9 – 15.2 mmol/L  |
|  |   |   |   |   |   |
|  CL | -0.71 | 0.999 | 1.000 | 0.99 | 0 – 197 mmol/L  |
|  |   |   |   |   |   |
|  CO2 | 0.15 | 1.000 | 1.000 | 0.41 | 0 – 40 mmol/L  |

{5}

Page 6 of 8

c. Traceability (controls, calibrators, or method):

The use of ion selective electrodes in the potentiometric determination of sodium and potassium ions was described over seventy years ago. However, practical methods for their measurement were not reported until recently. A sodium electrode was developed with a selectivity over potassium of about 300 to 1. Simon and Proda described a valinomycin potassium electrode with a selectivity over sodium of about 1000 to 1. Recent advances in calcium electrode technology have made potentiometric calcium measurements both accurate and practical. Chloride electrodes have been routinely used for potentiometric titrations of chloride. However, recent improvements in selectivity and sensitivity have made possible the direct potentiometric analysis of chloride in biological specimens. Stow and Randall describe a procedure for the direct measurement of the partial pressure of  $\mathrm{CO}_{2}$  using a  $\mathrm{pCO}_{2}$  electrode in 1954. This electrode has since been made more sensitive.

d. Detection limit:

The sensitivity of this methodology was documented through the repetitive assay of a serum control first with a known concentration and then diluting the sample until the minimum result was obtained and then run in replicates of 10 on the SYNCHRON CX3 System. Under the conditions described the following limits of detection were established:

|  Analyte | Limit of Detection  |
| --- | --- |
|  Calcium | 1.5 mg/dL  |
|  Sodium | 10 mmol/L  |
|  Potassium | 1.0 mmol/L  |
|  Chloride | 15 mmol/L  |
|  Total CO2 | 5.0 mmol/L  |

e. Analytical specificity:

Determined according to NCCLS EP7-A. Hemoglobin levels up to  $500\mathrm{mg / dL}$ , Bilirubin levels up to  $20\mathrm{mg / dL}$ , and Lipemia levels up to  $1800\mathrm{mg / dL}$  were tested and did not show any adverse effect on a stock sample with levels of sodium at  $148\mathrm{mmol / L}$ , potassium at  $5.2\mathrm{mmol / L}$ , chloride at  $119\mathrm{mmol / L}$ ,  $\mathrm{CO}_{2}$  at  $19\mathrm{mmol / L}$ , and calcium at  $9.4\mathrm{mg / dL}$ . Stock solutions of the substance to be tested were prepared at  $20\mathrm{x}$  concentrations and  $0.5\mathrm{ml}$  of this stock was placed in a  $10\mathrm{ml}$  volumetric flask and made up to volume with the base pool. The control stock was prepared similarly but with water as the diluent. Only Sodium Heparin, Lithium Heparin and Ammonium Heparin up to 45 Units/mL are acceptable anticoagulants.

f. Assay cut-off:

Not applicable for this type of device.

{6}

Page 7 of 8

2. Comparison studies:

a. Method comparison with predicate device:

Serum, plasma, cerebrospinal fluid and urine specimens, collected from adult patients, were assayed for calcium, chloride, potassium, sodium, and total $\mathrm{CO}_{2}$ on a SYNCHRON CX3 System using GenChem and Beckman flow cell reagents, wash solutions and calibrators. Results were compared by least squares linear regression where $X =$ Beckman Results and $Y =$ GenChem Results.

|  AnalyteSpecimen | Regression Statistics |   |   | Summary Statistics  |   |   |
| --- | --- | --- | --- | --- | --- | --- |
|   |  Unit | n | m | b | r | range  |
|  Calcium |  |  |  |  |  |   |
|  Serum | mg/dL | 80 | 0.989 | 0.0 | 0.985 | 7.1 - 10.6  |
|  Plasma | mg/dL | 80 | 0.990 | 0.0 | 0.995 | 7.1 - 10.7  |
|  Urine | mg/dL | 74 | 1.007 | -0.2 | 0.998 | 2.4 - 15.2  |
|  Chloride |  |  |  |  |  |   |
|  Serum | mmol/L | 80 | 0.988 | 1.0 | 0.935 | 98 - 127  |
|  Plasma | mmol/L | 80 | 0.998 | 0.8 | 0.985 | 98 - 127  |
|  Urine | mmol/L | 78 | 1.049 | -5.1 | 0.999 | 22 - 289  |
|  CSF | mmol/L | 44 | 1.024 | -3.4 | 0.985 | 113 - 152  |
|  Potassium |  |  |  |  |  |   |
|  Serum | mmol/L | 80 | 0.969 | 0.13 | 1.000 | 3.2 - 10.8  |
|  Plasma | mmol/L | 80 | 0.987 | 0.15 | 1.000 | 3.2 - 10.8  |
|  Urine | mmol/L | 80 | 0.993 | 0.01 | 1.000 | 3.5 - 136  |
|  Sodium |  |  |  |  |  |   |
|  Serum | mmol/L | 80 | 0.930 | 9.1 | 0.938 | 132 - 159  |
|  Urine | mmol/L | 78 | 1.000 | -0.3 | 1.000 | 17 - 288  |
|  Total CO2 |  |  |  |  |  |   |
|  Serum | mmol/L | 80 | 0.949 | 1.2 | 0.953 | 9.5 - 29  |
|  Plasma | mmol/L | 80 | 0.965 | 0.9 | 0.960 | 9.5 - 29  |

b. Matrix Comparison

See above method comparison studies.

3. Clinical studies:

a. Clinical sensitivity:

Clinical studies are not typically submitted for this device type.

b. Clinical specificity:

{7}

Page 8 of 8

Clinical studies are not typically submitted for this device type.
c. Other clinical supportive data (when a and b are not applicable):

Not applicable

4. Clinical cut-off:

Not applicable

5. Expected values/Reference range:

The expected values for calcium, chloride, potassium, sodium, and total CO₂ are listed below. Use these ranges only as guides. Each laboratory should establish its own reference ranges.

Reference Ranges¹

|  Analyte | Specimen | Conventional Units | SI Units  |
| --- | --- | --- | --- |
|  Calcium | Serum/Plasma | 8.4 - 10.2 mg/dL | 2.10 - 2.55 mmol/L  |
|   |  Urine | 100 - 300 mg/day | 2.5 - 7.5 mmol/day  |
|  Chloride | Serum/Plasma | 101 - 111 mmol/L | same  |
|   |  Urine | 110 - 250 mmol/day | same  |
|   |  CSF | 118 - 132 mmol/L | same  |
|  Potassium | Serum/Plasma | 3.5 - 5.1 mmol/L | same  |
|   |  Urine | 25 - 125 mmol/day | same  |
|  Sodium | Serum/Plasma | 136 - 145 mmol/L | same  |
|   |  Urine | 40 - 220 mmol/day | same  |
|  Total CO₂ | Serum/Plasma | 22 - 28 mmol/L | same  |

¹ Burtis, C.A., Ashwood, E.R. (eds.). Tietz Textbook of Clinical Chemistry. W.B. Saunders Company. Philadelphia, PA. (1994).

N. Conclusion:

The submitted material in this premarket notification is complete and supports a substantial equivalence decision.

---

**Source:** [https://fda-staging.innolitics.com/device/K040977](https://fda-staging.innolitics.com/device/K040977)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
