← Product Code [CDS](/productcode/CDS) · K040973

# BUN REAGENT (K040973)

_Genchem, Inc. · CDS · Dec 27, 2004 · Clinical Chemistry · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K040973

## Device Facts

- **Applicant:** Genchem, Inc.
- **Product Code:** [CDS](/productcode/CDS.md)
- **Decision Date:** Dec 27, 2004
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 862.1770
- **Device Class:** Class 2
- **Review Panel:** Clinical Chemistry

## Indications for Use

The BUN Reagent is to be used for the quantitative determination of urea nitrogen in serum, plasma and urine on the Beckman SYNCHRON CX3® System to aid in the diagnosis of renal function and pre renal disease states, such as cardiac decompensation and others.

## Device Story

Reagent kit containing urease and surfactants; designed for use on Beckman SYNCHRON CX3® Analyzer. Input: serum, plasma, or urine samples. Principle: enzymatic reaction (urease) followed by conductivity measurement on the CX3 system. Output: quantitative urea nitrogen concentration (mg/dL). Used in clinical laboratory settings by trained technicians. Results assist physicians in assessing renal function and diagnosing pre-renal conditions. Benefits include standardized diagnostic testing for kidney health and related metabolic states.

## Clinical Evidence

Bench testing only. Precision/reproducibility evaluated over 30 days (n=60 per sample); CVs ranged 0.7%–9.4%. Linearity confirmed 2–300 mg/dL (R=1.00). Sensitivity limit 1.43 mg/dL. Analytical specificity tested against hemoglobin, bilirubin, and lipemia with no adverse effects. Method comparison (n=80 serum/plasma, n=79 urine) against predicate showed high correlation (R² 0.998–1.000, slopes 0.979–0.995).

## Technological Characteristics

Reagent formulation containing urease and surfactants. Operates via conductivity measurement on Beckman SYNCHRON CX3® Analyzer. Analytical range 2–300 mg/dL. Compatible with serum, plasma (heparin/EDTA), and urine. No specific materials of construction or software algorithms described beyond standard analyzer integration.

## Regulatory Identification

A urea nitrogen test system is a device intended to measure urea nitrogen (an end-product of nitrogen metabolism) in whole blood, serum, plasma, and urine. Measurements obtained by this device are used in the diagnosis and treatment of certain renal and metabolic diseases.

## Predicate Devices

- Beckman Bun reagent for the CX3 (k761061)

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}

# 510(k) SUBSTANTIAL EQUIVALENCE DETERMINATION DECISION SUMMARY DEVICE ONLY TEMPLATE

A. 510(k) Number: k040973
B. Purpose For Submission:

Premarket Notification 510(k) for GenChem, Inc. intentions to manufacture and market the GenChem Bun Reagent Kit.

C. Analyte: Bun (Blood Urea Nitrogen)
D. Type of Test:

Quantitative, Photometric End-Point

E. Applicant: GenChem, Inc.
F. Proprietary and Established Names:

GenChem, Inc, BUN Reagent

G. Regulatory Information:

Regulation section:

1. Regulation section:

21 CFR §862.1770 - Urea nitrogen test system.

2. Classification:

Class II

3. Product Code:

CDS

4. Panel:

75 (Chemistry)

H. Intended use(s):

1. Intended use(s)

{1}

Page 2 of 6

GenChem BUN Reagent is to be used for the quantitative determination of urea nitrogen in serum, plasma and urine on the Beckman SYNCHRON CX3 System to aid in the diagnosis of renal function and pre renal disease states, such as cardiac decompensation and others.

2. **Indication(s) for use:**
GenChem BUN Reagent is to be used for the quantitative determination of urea nitrogen in serum, plasma and urine on the Beckman SYNCHRON CX3 System to aid in the diagnosis of renal function and pre renal disease states, such as cardiac decompensation and others.

3. **Special condition for use statement(s):** For Prescription Use.

4. **Special instrument Requirements:** Beckman CX3 Systems.

I. **Device Description:**
The device is a reagent containing sufficient Urease, surfactants and other ingredients necessary for optimum system operation on the Beckman SYNCHRON CX3 System.

J. **Substantial Equivalence Information:**
GenChem claims substantial equivalence to the Beckman BUN reagent for the CX3.

1. **Predicate device name(s):** Beckman Bun reagent for the CX3
2. **Predicate K number(s):** (k761061)
3. **Comparison with Predicate:**

|  Device Name | GenChem Bun Reagent Kit | Predicate Device Beckman Electrode, Ion Specific, Urea Nitrogen  |
| --- | --- | --- |
|  510(k) Number | (k040973) | (k761061)  |
|  Chemical Principle | Urease, Conductivity | Urease, Conductivity  |
|  Intended Use | For the quantitative determination of urea nitrogen in serum, plasma, and urine | For the quantitative determination of urea nitrogen in serum, plasma, and urine  |
|  Format | Liquid, ready to use | Liquid, ready to use  |
|  Composition | Urease, buffer | Urease, buffer  |
|  Linearity | 0-150 mg/dL | 0-150 mg/dL  |
|  Storage | 2-8 °C | 2-8 °C  |

{2}

Page 3 of 6

# K. Standard/Guidance Document Referenced (if applicable):

Within-Day and Day-to-Day precision was determined according to NCCLS EP5-A. Linearity was performed according to NCCLS EP6-A Guideline. Analytical specificity Determined according to NCCLS EP7-A.

# L. Test Principle:

This test utilizes a conductivity reaction method. In 1961, Chin and Kroontje reported a procedure for urea determination that measured the conductivity of the ammonium bicarbonate end product. This method was later modified to measure the rate of increase in the reagent conductivity.

Principle of Procedure

$$
\mathrm {U r e a} + 3 \mathrm {H} _ {2} 0 \rightarrow 2 \mathrm {N H} _ {4} ^ {+} + \mathrm {H C O} _ {3} ^ {-} + \mathrm {O H} ^ {-}
$$

Urea from the sample is converted to ionic ammonium bicarbonate by the urease reagent. The rate of increase in conductivity, measured 11.5 seconds after sample addition, is proportional to the concentration of urea in the sample.

# M. Performance Characteristics (if/when applicable):

1. Analytical performance:

a. Precision/Reproducibility:

Control sera and diluted urine pools were each assayed twice per day in triplicate on a SYNCHRON CX3 System. Data were collected on ten different days over a thirty day period.

Precision of BUN Recoveries in (mgN/dL)

|  Sample | n | Within Run |   | Total Imprecision  |   |   |
| --- | --- | --- | --- | --- | --- | --- |
|   |   |  mean | 1SD | %CV | 1SD | %CV  |
|  Serum 1 | 60 | 7.1 | 0.65 | 9.1 | 0.66 | 9.4  |
|  Serum 2 | 60 | 35.4 | 0.62 | 1.8 | 0.66 | 1.9  |
|  Serum 3 | 60 | 63.8 | 0.50 | 0.8 | 0.80 | 1.3  |
|  Urine 1 | 60 | 21.7 | 0.89 | 4.1 | 0.82 | 3.8  |
|  Urine 2 | 60 | 112.2 | 0.75 | 0.7 | 1.25 | 1.1  |

b. Linearity/assay reportable range:

Linearity was performed according to NCCLS Guideline EP6-A. Commercially available linearity standards ranging from 0 to $158\mathrm{mg / dl}$ were analyzed in triplicate on the

{3}

Page 4 of 6

Beckman CX3 and the results analyzed by the Least Squares method. The results gave a slope of 0.995 with an intercept of -0.12, a standard error of estimate of 0.49 and $r^2 = 1.00$ and is shown below. Specimens exceeding these limits should be diluted with normal saline and reanalyzed. Multiply the result by the appropriate dilution factor.

|   |  | Usable Ranges  |   |
| --- | --- | --- | --- |
|  Specimens | Range | Conventional Units | SI Units  |
|  All | Normal | 2 - 150 mgN/dL | 2-53.6mmol/L  |
|  All | ORDAC* | 150 - 300 mgN/dL | 53-107.2mmol/L  |

c. Traceability (controls, calibrators, or method):

Beckman Calibration Standards 1 and 2 for the CX3 System

d. Detection limit:

The sensitivity of this method was investigated by assaying serum first with a known concentration and then diluting the sample until the minimum result obtained and then run in replicates of 10 on the SYNCHRON CX3 System. Under the conditions described the limit of detection for this method was found to be $2.0\mathrm{mg/dL}$.

Analyte  Limit of Detection

BUN  $2.0\mathrm{mg/dL}$

e. Analytical specificity:

Determined according to NCCLS EP7-A. Hemoglobin levels up to $500\mathrm{mg/dL}$, Bilirubin levels up to $20\mathrm{mg/dL}$, and Lipemia levels up to $1800\mathrm{mg/dL}$ were tested and did not show any adverse effect on a stock sample with a BUN level of $15\mathrm{mg/dL}$. Stock solutions of the substance to be tested were prepared at $20\mathrm{x}$ concentrations and $0.5\mathrm{ml}$ of this stock was placed in a $10\mathrm{ml}$ volumetric flask and made up to volume with the base pool. The control stock was prepared similarly but with water as the diluent. Heparin, Lithium Heparin Ammonium Heparin, and EDTA are acceptable anticoagulants.

f. Assay cut-off:

Not applicable for this type of device.

2. Comparison studies:

a. Method comparison with predicate device:

Serum and plasma specimens, and urine specimens diluted with 9 parts normal saline, ranging from 4 to $300\mathrm{mg/dL}$ were collected from adult patients and assayed for urea nitrogen on a SYNCHRON CX3 System using GenChem and Beckman BUN reagents.

{4}

Page 5 of 6

Results were compared by least squares linear regression and the following statistics were obtained:

|  VALUE | SERUM | PLASMA | URINE  |
| --- | --- | --- | --- |
|  Intercept | -0.3 | -0.2 | 0.9  |
|  Slope | 0.995 | 0.989 | 0.979  |
|  R² Value | 0.999 | 0.998 | 1.000  |
|  N | 80 | 80 | 79  |
|  Range | 4-300 | 4-300 | 6-142  |

b. Matrix Comparison

See above method comparison studies.

3. Clinical studies:

a. Clinical sensitivity:

Clinical studies are not typically submitted for this device type.

b. Clinical specificity:

Clinical studies are not typically submitted for this device type.

c. Other clinical supportive data (when a and b are not applicable):

Not applicable

4. Clinical cut-off:

Not applicable

5. Expected values/Reference range:

The expected values for urea are listed below. Use these ranges only as guides. Each laboratory should establish its own normal ranges.

{5}

Page 6 of 6

|  Specimens | Conventional Units | SI Units  |
| --- | --- | --- |
|  Serum/Plasma | 7 - 18 mgN/dL | 2.5 – 6.4 mmol/L  |
|  Urine | 12 - 20 gN/day | 428 - 714 mmol/day  |

1. Burtis, C.A., Ashwood, E.R. (eds.). Tietz Textbook of Clinical Chemistry. W.B. Saunders Company. Philadelphia, PA. (1994).

N. Conclusion:
The submitted material in this premarket notification is complete and supports a substantial equivalence decision.

---

**Source:** [https://fda-staging.innolitics.com/device/K040973](https://fda-staging.innolitics.com/device/K040973)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
