K040953 · Aesku, Inc. · LJM · Jun 23, 2004 · Immunology
Device Facts
Record ID
K040953
Device Name
AESKULISA ANA HEP-2
Applicant
Aesku, Inc.
Product Code
LJM · Immunology
Decision Date
Jun 23, 2004
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 866.5100
Device Class
Class 2
Indications for Use
AESKULISA ANA-Hep2 is a solid phase enzyme immunoassay for the combined qualitative detection of IgG antibodies against Hep2 cells in human serum. Each well is coated with lysed Hep2 cells and specific antigens. The test collectively detects, in one well, total ANAs against double stranded DNA (dsDNA), histones, SS-A (Ro), SS-B (La), Sm, snRNP/Sm, Scl-70, Jo-1 and centromeric antigens along with sera positive for Hep2 immunofluorescence test (IFT). The assay is a tool in the diagnosis of certain systemic rheumatic diseases and should be used in conjunction with other serological tests and clinical findings.
Device Story
AESKULISA ANA-Hep2 is a qualitative ELISA test for detecting IgG antibodies against Hep2 cells and specific nuclear antigens (dsDNA, histones, SS-A, SS-B, Sm, snRNP/Sm, Scl-70, Jo-1, centromeric) in human serum. The device uses microtiter plates coated with lysed Hep2 cells and antigens. Patient serum is incubated in the wells; if ANA antibodies are present, they bind to the antigens. After washing, anti-human IgG HRP conjugate is added, followed by TMB substrate. The resulting enzymatic colorimetric reaction (blue to yellow after acid stop) is measured at 450 nm using a microtiter plate reader. The optical density (OD) is proportional to the antibody concentration. The test is performed in a clinical laboratory setting by trained personnel. Results assist clinicians in diagnosing systemic rheumatic diseases when combined with other clinical and serological data.
Clinical Evidence
No clinical trials were performed. Substantial equivalence was established via analytical performance testing and a method comparison study. Method comparison with the predicate device (n=59) showed 100% agreement (57 positive, 2 negative). Analytical specificity was demonstrated using 57 characterized patient sera (all positive) and 91 negative control sera (11 positive for other antigens, 80 healthy donors). Precision (intra-assay and inter-assay) showed %CVs ranging from 1.5% to 3.1%. Linearity/recovery was confirmed via serial dilutions.
Technological Characteristics
Solid-phase enzyme immunoassay (EIA). Wells coated with lysed HEp-2 cells and specific nuclear antigens. Qualitative detection of IgG antibodies. Manual or automated plate processing.
Indications for Use
Indicated for the qualitative detection of IgG antibodies against Hep2 cells in human serum to aid in the diagnosis of systemic rheumatic diseases. Intended for use in conjunction with other serological tests and clinical findings. Prescription use only.
Regulatory Classification
Identification
An antinuclear antibody immunological test system is a device that consists of the reagents used to measure by immunochemical techniques the autoimmune antibodies in serum, other body fluids, and tissues that react with cellular nuclear constituents (molecules present in the nucleus of a cell, such as ribonucleic acid, deoxyribonucleic acid, or nuclear proteins). The measurements aid in the diagnosis of systemic lupus erythematosus (a multisystem autoimmune disease in which antibodies attack the victim's own tissues), hepatitis (a liver disease), rheumatoid arthritis, Sjögren's syndrome (arthritis with inflammation of the eye, eyelid, and salivary glands), and systemic sclerosis (chronic hardening and shrinking of many body tissues).
Submission Summary (Full Text)
{0}
510(k) SUBSTANTIAL EQUIVALENCE DETERMINATION
DECISION SUMMARY
DEVICE ONLY TEMPLATE
A. 510(k) Number:
k040953
B. Purpose for Submission:
New device
C. Analyte:
Antinuclear antibody
D. Type of Test:
Qualitative ELISA
E. Applicant:
AESKU INC
F. Proprietary and Established Names:
AESKULISA® ANA Hep-2 test
G. Regulatory Information:
1. Regulation section:
21CFR § 866.5100, Antinuclear Antibody Immunological Test System
2. Classification:
Class II
3. Product Code:
LKJ, Antinuclear Antibody, Antigen, Control
4. Panel:
82 Immunology
H. Intended Use:
1. Intended use(s):
AESKULISA ANA-Hep2 is a solid phase enzyme immunoassay for the combined qualitative detection of IgG antibodies against Hep2 cells in human serum. Each well is coated with lysed Hep2 cells and specific antigens. The test collectively detects, in one well, total ANAs against double stranded DNA (dsDNA), histones, SS-A (Ro), SS-B (La), Sm, snRNP/Sm, Scl-70, Jo-1 and centromeric antigens along with sera positive for Hep2 immunofluorescence test (IFT).
The assay is a tool in the diagnosis of certain systemic rheumatic diseases and should be used in conjunction with other serological tests and clinical findings.
2. Indication(s) for use:
{1}
Same as intended use.
3. Special condition for use statement(s):
The device is for prescription use only.
4. Special instrument Requirements:
Microtiter plate reader 450 nm reading filter and optional 620 nm reference filter (600-690 nm). Microplate washing device (multichannel pipette or automated system)
I. Device Description:
The assay components of the AESKULISA ANA-Hep2 include antigen coated microtiter plate, negative, positive and cut-off controls, wash buffer concentrate, sample buffer concentrate, anti-human IgG horseradish peroxidase (HRP) conjugate, 3,3',5,5' tetramethylbenzidine (TMB) substrate and 1M HCl stop solution. The positive and cut-off controls are composed of human serum of a specific dilution in standard buffer. The negative control contains standard buffer without any reactive component.
J. Substantial Equivalence Information
1. Predicate device name(s):
HELIX Diagnostics Antinuclear Antibody Screening
2. Predicate K number(s):
k954723
3. Comparison with predicate:
| Similarities | | |
| --- | --- | --- |
| Item | Device | Predicate |
| | AESKULISA ANA Hep-2 | Helix ANA Screening Test |
| Assay principle | Indirect, non-competitive qualitative ELISA test using microtiter technology. | Same |
| Indications for Use | Aid in the diagnosis of systemic rheumatic diseases and should be used in conjunction with other serological tests and clinical findings. | Aid in the diagnosis of certain systemic rheumatic diseases. |
| Sample matrix | Serum | Same |
| Analyte | Antinuclear antibodies | Same |
| Differences | | |
| Item | Device | Predicate |
| Sample buffer | To be diluted (1:5) | Ready to use |
| Wash buffer | To be diluted (1:50) | To be diluted (1:16.6) |
| Cut-off control, positive and negative controls | Ready to Use | Positive and negative controls only To be diluted (1:40) |
| Incubation times | 30, 15 plus 15 minutes | 30, 30 plus 30 minutes |
{2}
3
K. Standard/Guidance Document Referenced (if applicable):
None referenced
L. Test Principle:
Serum samples diluted 1:101 are incubated in the microplates coated with the ANA antigens. ANA antibodies, if present in the patient specimen, bind to the antigens forming antigen/antibody complexes. The unbound fraction is washed off. The anti-human IgG HRP conjugate is added and reacts with the antigen/antibody complexes in the microplates. Unbound conjugate is washed off in the following step. Addition of TMB substrate generates an enzymatic colorimetric (blue) reaction, which is stopped by diluted acid (color changes to yellow). The rate of color formation is a function of the amount of conjugate bound to the antigen/antibody complexes and is proportional to the initial concentration of the antibodies in the patient samples. The color formation is determined by measuring the OD value at 450 nm.
M. Performance Characteristics (if/when applicable)
1. Analytical performance:
a. Precision/Reproducibility:
Intra-assay
To determine intra-assay variation, 3 different sera (low, medium, high) were tested 24 times on 3 microtiter plates. The results are as follows:
| Sample No. | Overall Mean OD Ratio | Overall Mean %CV |
| --- | --- | --- |
| Low | 4.6 | 1.5 |
| Med | 2.8 | 2.0 |
| High | 1.4 | 1.8 |
Inter-assay
Inter-assay precision (%CV) was determined by running 3 different sera (low, medium and high) 18 times on three different microtiter plates for five days. The results are as follows:
| Sample No. | Overall Mean OD Ratio | Overall Mean %CV |
| --- | --- | --- |
| Low | 4.7 | 3.1 |
| Med | 3.0 | 2.5 |
| High | 1.2 | 2.4 |
b. Linearity/assay reportable range:
A dilution/recovery study was performed using two fold serial dilutions of two positive samples. The results of this study are as follows:
| Sample No. | Dilution Factor | Measured conc (OD ratio) | Expected Conc (OD ratio) | % Recovery |
| --- | --- | --- | --- | --- |
| 1 | 1/100 | 4.10 | 4.20 | 97.6 |
| | 1/200 | 2.10 | 2.10 | 100 |
{3}
4
| | 1/400 | 1.00 | 1.05 | 95.2 |
| --- | --- | --- | --- | --- |
| | 1/800 | 0.55 | 0.53 | 103.8 |
| 2 | 1/100 | 6.10 | 6.20 | 98.4 |
| | 1/200 | 3.00 | 3.10 | 96.8 |
| | 1/400 | 1.59 | 1.55 | 102.6 |
| | 1/800 | 0.79 | 0.78 | 102.0 |
c. Traceability (controls, calibrators, or method):
Positive controls are the CDC-ANA reference human sera or WHO standards for the respective antigen.
d. Detection limit:
Not applicable
e. Analytical specificity:
To demonstrate assay specificity, 57 characterized patient sera from various autoimmune diseases (SLE, MCTD, CREST and Sjogren syndrome) obtained from major hospitals which were positive on IFA Hep2 ANA were all found positive in the AESKULISA ANA-Hep2.
| disease | # of tested sera |
| --- | --- |
| SLE | 39 |
| MCTD | 3 |
| CREST | 4 |
| Sjogrens Syndrome | 4 |
| Various AI diseases | 7 |
In addition, 11 sera positive for the following antigens: tissue transglutaminase (3), Proteinase (3), thyroglobulin (2), gliadin (2) and thyroid peroxidase (2) and 80 sera from a control group were tested and found negative on the AESKULISA ANA Hep-2. The healthy donors consisted of 54% female and 46% male. Forty-one percent of the cohort was young adults (n=33), 53% middle aged (n=42) and 6% (n=5) elderly.
f. Assay cut-off:
To determine the cut-off value, serial dilutions (1:3) of an antibody specific patient serum are tested in triplicates. The OD at 450nm is determined and plotted against the dilution factor to determine a linear range. The dilution in the linear range with an OD of approximately 2.0 is defined as calibrator F and is assigned 300 U/mL. Calibrator F is diluted to a concentration of 15 U/mL and calibrated to the respective CDC ANA reference serum. The selected cut-off is equivalent to an OD of 0.5 to 0.6 of that of the reference serum. To validate the cut-off, samples from the control group used in the specificity study were tested. The cut-off value is an OD ratio of 1.0
{4}
5
2. Comparison studies:
a. Method comparison with predicate device:
Fifty nine patient sera were tested on the AESKULISA and the HELIX device. The samples consisted of 39 SLE, 7 suspected reactive Arthritis, 4 CREST Syndrome, 3 Sjögren’s Syndrome, 2 Mixed Connective Tissue Disease (MCTD), one each of SLE/ Sjögren overlap, MCTD/Sicca Symptoms, anti-Jo-1 Syndrome and Rheumatoid Arthritis. Total agreement between the two devices was 100% with 57 sera positive and two sera negative on both devices.
| | HELIX | | | |
| --- | --- | --- | --- | --- |
| | | Pos | Neg | Total |
| AeskuLisa
ANA Hep-2 | Pos | 57 | 0 | 57 |
| | Neg | 0 | 2 | 2 |
| | Total | 57 | 2 | |
b. Matrix comparison:
Serum is the only recommended matrix.
3. Clinical studies:
a. Clinical sensitivity:
Not provided
b. Clinical specificity:
Not provided
c. Other clinical supportive data (when a and b are not applicable):
Not applicable.
4. Clinical cut-off:
See assay cut-off.
5. Expected values/Reference range:
The expected value in the normal population is negative.
N. Conclusion:
The submitted material in this premarket notification is complete and supports a substantial equivalence decision.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.