← Product Code [LLZ](/productcode/LLZ) · K040931

# PATXFER RT (K040931)

_Brainlab AG · LLZ · Jul 22, 2004 · Radiology · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K040931

## Device Facts

- **Applicant:** Brainlab AG
- **Product Code:** [LLZ](/productcode/LLZ.md)
- **Decision Date:** Jul 22, 2004
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 892.2050
- **Device Class:** Class 2
- **Review Panel:** Radiology
- **Attributes:** Software as a Medical Device

## Indications for Use

PatXfer RT provides capabilities for the acceptance, transfer, display, storage, and digital processing of medical images and objects according to the DICOM RT standard and according to the BrainLAB proprietary data format. This includes functions for performing operations related to image manipulation, enhancement, compression, and quantification. The integrity of the original data is maintained.

## Device Story

PatXfer RT is a software system for medical image management; it interfaces between medical devices to facilitate treatment planning and verification. It accepts medical images and objects via DICOM 3.0 or BrainLAB proprietary formats from network archives, optical disks, or CD-ROMs. The system performs image manipulation, enhancement, compression, and quantification while maintaining original data integrity. It converts DICOM RT data to BrainLAB proprietary formats and vice-versa. Operated by trained professionals (physicians, nurses, technicians) via mouse or touchscreen, the system displays selected patient data, image series, and images. It maintains a history log-file of all import actions. By centralizing and processing image data, it assists clinicians in treatment planning and verification, potentially improving workflow efficiency and clinical decision-making accuracy.

## Clinical Evidence

No clinical data. Verification and validation were performed according to internal design and development procedures. Risk analysis confirmed hazards were mitigated to ALARP levels.

## Technological Characteristics

Software-based image processing system. Supports DICOM 3.0 and BrainLAB proprietary data formats. Connectivity via network, optical disk, or CD-ROM. User interface via mouse or touchscreen. Functions include image manipulation, enhancement, compression, and quantification.

## Regulatory Identification

A medical image management and processing system is a device that provides one or more capabilities relating to the review and digital processing of medical images for the purposes of interpretation by a trained practitioner of disease detection, diagnosis, or patient management. The software components may provide advanced or complex image processing functions for image manipulation, enhancement, or quantification that are intended for use in the interpretation and analysis of medical images. Advanced image manipulation functions may include image segmentation, multimodality image registration, or 3D visualization. Complex quantitative functions may include semi-automated measurements or time-series measurements.

## Special Controls

*Classification.* Class II (special controls; voluntary standards—Digital Imaging and Communications in Medicine (DICOM) Std., Joint Photographic Experts Group (JPEG) Std., Society of Motion Picture and Television Engineers (SMPTE) Test Pattern).

## Predicate Devices

- PatXfer 5.0 ([K021583](/device/K021583.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

K040961

# 510 (k) Summary of Safety and Effectiveness for PatXfer RT

#### Manufacturer:

| Address:        | BrainLAB AG<br>Ammerthalstrasse 8<br>85551 Heimstetten<br>Germany<br>Phone: +49 89 99 15 68 0<br>Fax: +49 89 99 15 68 33 |
|-----------------|--------------------------------------------------------------------------------------------------------------------------|
| Contact Person: | Mr. Rainer Birkenbach                                                                                                    |
| Summary Date:   | April 5, 2004                                                                                                            |

#### Device Name:

Trade name: Common/Classification Name: PatXfer RT System, Image Processing, Radiological

# Predicate Device:

PatXfer 5.0 (K021583)

Device Classification Name: Picture archiving and Communication System Requiatory Class: Class II

## Intended Use:

PatXfer RT provides capabilities for the acceptance, transfer, display, storage, and digital processing of medical images and objects according to the DICOM RT standard and according to the BrainLAB proprietary data format.

This includes functions for performing operations related to image manipulation, enhancement, compression, and quantification. The integrity of the original data is maintained.

## Device Description:

PatXfer RT interfaces between medical devices and provides the patient data for treatment planning or verification.

The application transfers, displays, stores, and processes data received according to the ACR-NEMA standard (DICOM 3.0) or according to the BrainLAB proprietary data format. DICOM RT data are converted into the BrainLAB proprietary data format and vice-versa for further data processing.

PatXfer RT supports Dicom RT data and BrainLAB's proprietary data formats from digital storage media, such as network archive, optical disk or CD-ROM.

To keep the application as user-friendly as possible only the necessary information for the intended procedures are displayed. The patient, image series, and images will be displayed and can be selected and processed.

{1}------------------------------------------------

PatXfer RT creates a history log-file including all the software actions to import the data and can be controlled with the mouse or by touchscreen monitor.

Typical users of the system are trained professionals, including but not limited to physicians, nurses, and technicians.

#### Substantial equivalence

A risk analysis was conducted for PatXfer RT and all hazards were mitigated to as low as reasonable possible (ALARP) and found to be acceptable. The Verification showed that PatXfer RT is safe and effective for its intended use.

PatXfer RT has been verified and validated according to BrainLAB's procedures for product design and development and found to be substantially equivalent with BrainLAB medical devices such as PatXfer 5.0 (K021583). The validation proves the safety and effectiveness of the system.

{2}------------------------------------------------

Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo consists of a stylized depiction of an eagle or bird-like figure with three curved lines forming its body and wings. The logo is encircled by the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" in a circular arrangement.

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

# JUL 22 2004

Mr. Rainer Birkenbach Executive Vice President BrainLAB AG Ammerthalstraße 8 85551 Heimstetten GERMANY

Re: K040931 Trade/Device Name: PatXfer RT Regulation Number: 21 CFR 892.2050 Regulation Name: Picture archiving and communications system Regulatory Class: II Product Code: 90 LLZ Dated: July 6, 2004 Received: July 8, 2004

Dear Mr. Birkenbach:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

{3}------------------------------------------------

This letter will allow you to begin marketing your device as described in your Section 510(k) I rab letter notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at one of the following numbers, based on the regulation number at the top of the letter:

| 8xx. 1xxx                        | (301) 594-4591 |
|----------------------------------|----------------|
| 876.2xxx, 3xxx, 4xxx, 5xxx       | (301) 594-4616 |
| 884.2xxx, 3xxx, 4xxx, 5xxx, 6xxx | (301) 594-4616 |
| 892.2xxx, 3xxx, 4xxx, 5xxx       | (301) 594-4654 |
| Other                            | (301) 594-4692 |

Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97) you may obtain. Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html.

Sincerely yours,

Nancy C. Bugden

Nancy C. Brogdon Director, Division of Reproductive, Abdominal and Radiological Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{4}------------------------------------------------

Page

| 510(k) Number (if known): | K040931 |
|---------------------------|---------|
|---------------------------|---------|

Device Name:

PatXfer RT

Indications For Use:

PatXfer RT provides capabilities for the acceptance, transfer, display, storage, and digital processing of medical images and objects according to the DICOM RT standard and according to the BrainLAB proprietary data format.

This includes functions for performing operations related to image manipulation, enhancement, compression, and quantification. The integrity of the original data is maintained.

#### (PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

Prescription Use X (Pcr 21 CFR 801.109)

ાર

Over-The-Counter Use

Nancy Broadron

(Division Sign-Off) Division of Reproductive. A and Radiological Devices 510(k) Number _

(Optional Format 1-2-96)

---

**Source:** [https://fda-staging.innolitics.com/device/K040931](https://fda-staging.innolitics.com/device/K040931)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
