K040910 · Choong Wae Medical Corp. · FMZ · Apr 21, 2004 · General Hospital
Device Facts
Record ID
K040910
Device Name
INFANT INCUBATOR, MODEL CHS-I1000
Applicant
Choong Wae Medical Corp.
Product Code
FMZ · General Hospital
Decision Date
Apr 21, 2004
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 880.5400
Device Class
Class 2
Attributes
Therapeutic, Pediatric, 3rd-Party Reviewed
Indications for Use
CHS-i1000 Infant Incubator provides a controlled thermal environment for neonates who are unable to provide their own thermoregulation. They may be used for short periods of time to facilitate the neonate's transition from the uterus to the external environment. The CHS-i1000 Infant Incubators can be used in two operating modes, such as Air Control Mode and Skin Control Mode. The CHS-i1000 incubator is not for the transport of the patient. The CHS-i1000 Infant Incubator has alarms to alert clinicians when certain patient or equipment conditions occur, such as a malfunction, or an excessive departure of the patient's temperature from the set value. The CHS-i1000 Infant Incubator incorporates other features, such as humidification of the infant environment and tilting of the bed.
Device Story
CHS-i1000 is a neonatal incubator providing controlled thermal and humidified environment; prevents external heat loss. Components: hood, base, cabinet, controller; clear acrylic hood with access ports (standard and iris for tubes/cables). Mattress adjustable to 12° incline. Operates in Air Control or Skin Control modes. Used in clinical settings by caregivers. Features alarms for malfunctions or temperature deviations. Benefits: supports thermoregulation for neonates; facilitates transition from uterus to external environment.
Clinical Evidence
Bench testing only. Performance data included electrical, mechanical, and environmental safety testing per EN/IEC 60601-1 and EN/IEC 60601-2-19; EMC testing per EN/IEC 60601-1-2; and biocompatibility testing per ISO 10993-1. All results were satisfactory.
Technological Characteristics
Acrylic hood; foam mattress; dimensions W1000 x H1367 x D634mm. Features humidification and bed tilting. Standards: EN/IEC 60601-1 (safety), EN/IEC 60601-2-19 (incubator performance), EN/IEC 60601-1-2 (EMC), ISO 10993-1 (biocompatibility).
Indications for Use
Indicated for neonates unable to provide their own thermoregulation. Used for short-term transition from uterus to external environment. Not for patient transport.
Regulatory Classification
Identification
A neonatal incubator is a device consisting of a rigid boxlike enclosure in which an infant may be kept in a controlled environment for medical care. The device may include an AC-powered heater, a fan to circulate the warmed air, a container for water to add humidity, a control valve through which oxygen may be added, and access ports for nursing care.
{0}------------------------------------------------
K040910
# APR 2 1 2004
510(k) Submission – CHS-i1000
### 510(k) Summary
This summary of 510(k) safety and effectiveness information is being submitted in accordance with requirements of 21 CFR Part 807.92.
Date: February 2, 2004
- 1. Company and Correspondent making the submission:
Name - Choong Wae Medical Corp. Address – 698, Shindaebang-Dong, Dongjack-Gu, Seoul, Korea 156-010 Telephone - +82-2-840-6806 Fax - +82-2-841-1213 Contact - Mr. Myung Hak Cha Internet - http://www.cwm.co.kr
- 2. Device :
Proprietary Name – CHS-i1000 Neonatal Incubator Common Name - Infant Incubator Classification Name - Incubator, Neonatal
### 3. Predicate Device :
Manufacturer : OHMEDA Medical Model : 4000 CARE-PLUS 510(k) No. : K974349(Decision Date - February. 12. 1998)
- 4. Classifications Names & Citations :
21CFR 880.5400, FMZ, Incubator, Neonatal, Class2 Guidance document for the preparation of premarket notifications [510(k)'s] for Neonatal and Neonatal Transport Incubators.
5. Description :
The CHS-i1000 incubator provides a controlled temperature and humidity for infants. It is designed to minimize the effect of external heat loss on infants.
{1}------------------------------------------------
Image /page/1/Picture/0 description: The image shows a sequence of handwritten characters, which appear to be "K040910". The characters are written in a cursive style, with some connections between the letters and numbers. The writing is clear and legible, although the style is informal.
000030
510(k) Submission - CHS-i1000
The incubator is composed of 4 main elements namely, the hood, base, cabinet and controller. Together they measure W1000 x H1367 x D634mm. The foam based mattress that measures W690 x D370mm positions centrally within the confines of the hood. The mattress can also be raised at the head and foot end to a maximum incline of 12°. This is achieved by turning the handles that are located on each end of the base.
The hood is made of a clear acrylic to permit clear visibility of the baby yet provide isolation for the baby. The hood comprised of access ports gives access to the baby without compromising the thermal environment. Four of the access ports(two on the front and two on the rear) allow for regular access by the caregiver. The other two access ports are in the form of iris ports with are used primarily for ventilator tubes and monitoring cables to be channeled into the hood to the baby. Pulling the mattress towards the caregiver can enhance access even further.
#### 6. Indication for use :
The CHS-i1000 Infant Incubator provides a controlled thermal environment for neonates who are unable to provide their own thermoregulation. They may be used for short periods of time to facilitate the neonate's transition from the uterus to the external environment. The CHS-i1000 Infant Incubators can be used in two operating modes, such as Air Control Mode and Skin Control Mode. The CHSi1000 incubator is not for the transport of the patient.
The CHS-i1000 Infant Incubator has alarms to alert clinicians when certain patient or equipment conditions occur, such as a malfunction, or an excessive departure of the patient's temperature from the set value.
The CHS-i1000 Infant Incubator incorporates other features, such as humidification of the infant environment and tilting of the bed.
7. Comparison with predicate device :
Choong Wae Medical Corp., believes that the CHS-i1000 Infant Incubator is substantially equivalent to the 4000 CARE-PLUS of OHMEDA Medical.
{2}------------------------------------------------
K040910
- 8. Safety, EMC and Performance Data :
Electrical, mechanical, environmental safety and performance testing according to standard EN/IEC 60601-1 and EN/IEC 60601-2-19 was performed, and EMC testing was conducted in accordance with standard EN/IEC 60601-1-2(1993), and Biocompatibility testing was conducted in accordance with standard ISO 10993-1. All test results were satisfactory.
- 9. Conclusions :
In accordance with the Federal Food, Drug and Cosmetic Act, 21 CFR Part 807 and based on the information provided in this premarket notification Choong Wae Medical Corp. concludes that CHS-i1000 is safe and effective and substantially equivalent to predicate devices as described herein.
- 10. Choong Wae Medical Corp. will update and include in this summary any other information deemed seasonably necessary by the FDA.
END
{3}------------------------------------------------
Image /page/3/Picture/1 description: The image shows the logo for the Department of Health & Human Services (HHS). The logo features a stylized eagle with three wing-like shapes, representing health, human services, and well-being. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" is arranged in a circular pattern around the eagle.
Public Health Service
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
APR 2 1 2004
Choongwae Medical Corporation C/O Mr. Chan Yo Won Responsible Third Party Official Underwriters Laboratories, Incorporation 2600 N.W. Lake Road Camas, Washington 98607-8542
Re: K040910
Trade/Device Name: CHS-i1000 Neonatal Incubator Regulation Number: 880.5400 Regulation Name: Neonatal Incubator Regulatory Class: II Product Code: FMZ Dated: March 9, 2004 Received: April 8, 2004
Dear Mr. Won:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{4}------------------------------------------------
Page 2 - Mr. Won
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807): labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (301) 594-4618. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html
Sincerely yours,
Kein Mulvey
Chiu Lin. Ph.D. Director Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
K0409/0
# Indications for Use
Cim
(Division Sign-Off) Division of Anesthesiology, General Hospital, Infection Control, Dental Devices
510(k) Number(if known):
204113410 510(k) Number:
Device Name: CHS-i1000 Neonatal Incubator
Indications for Use:
CHS-i1000 Infant Incubator provides a controlled thermal The environment for neonates who are unable to provide their own thermoregulation. They may be used for short periods of time to facilitate the neonate's transition from the uterus to the external environment. The CHS-i1000 Infant Incubators can be used in two operating modes, such as Air Control Mode and Skin Control Mode. The CHS-i1000 incubator is not for the transport of the patient.
The CHS-i1000 Infant Incubator has alarms to alert clinicians when certain patient or equipment conditions occur, such as a malfunction, or an excessive departure of the patient's temperature from the set value.
The CHS-i1000 Infant Incubator incorporates other features, such as humidification of the infant environment and tilting of the bed.
Prescription Use
(Part 21 CFR 801 Subpart D)
AND/OR
Over-The-Counter Use (Part 21 CFR 807 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation(ODE)
000033
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.