S-42 UV, T-42 UV TORIC AND M-42 UV MULTIFOCAL (HEFILCON A) SOFT (HYDROPHILIC) CONTACT LENS FOR DAILY WEAR
K040900 · Optical Connection, Inc. · LPL · Jun 15, 2004 · Ophthalmic
Device Facts
Record ID
K040900
Device Name
S-42 UV, T-42 UV TORIC AND M-42 UV MULTIFOCAL (HEFILCON A) SOFT (HYDROPHILIC) CONTACT LENS FOR DAILY WEAR
Applicant
Optical Connection, Inc.
Product Code
LPL · Ophthalmic
Decision Date
Jun 15, 2004
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 886.5925
Device Class
Class 2
Indications for Use
The S 42 UV Single Vision (hefilcon A) Soft (Hydrophilic) Contact Lens is indicated for daily wear for the correction of refractive ametropia (myopia and hyperopia) in aphakic or not-aphakic persons with non-diseased eyes that may exhibit refractive and/or corneal astigmatism up to 2.00 diopters that does not interfere with visual acuity. The T 42 UV Toric (hefilcon A) Soft (Hydrophilic) Contact Lens is indicated for daily wear for the correction of refractive ametropia (myopia, hyperopia and astigmatism) in aphakic and/or not aphakic persons with non-diseased eyes that may exhibit refractive and/or corneal astigmatism up to 5.00 Diopters. The M 42 UV Multifocal (hefilcon A) Soft (Hydrophilic) Contact Lens is indicated for daily wear for the correction of refractive ametropia (myopia and hyperopia) and presbyopia in aphakic or not-aphakic persons with non-diseased eyes that may exhibit refractive and/or corneal astigmatism up to 2.00 diopters that does not interfere with visual acuity. The lenses may be disinfected using chemical (not heat) or hydrogen peroxide disinfecting systems. Eyecare practitioners may prescribe the lenses for daily wear and/or frequent replacement. When prescribed for a Frequent Replacement Program, the lenses may be disinfected using chemical (not heat), or hydrogen peroxide disinfecting systems.
Device Story
Hemispherical flexible soft contact lenses; made of hefilcon A (HEMA/NVP copolymer); 42% water content; tinted with Piqment Blue 15; incorporates UV-absorbing compound 2-(benzoyl-3-hydroxyphenoxy)ethyl acrylate. Available in single vision, toric (back surface design), and multifocal (aspheric design) configurations. Prescribed by eyecare practitioners for daily wear or frequent replacement. Patient self-inserts and removes lenses; cleans/disinfects using chemical or hydrogen peroxide systems. Provides refractive correction for ametropia and presbyopia; UV-blocking feature reduces ocular UV exposure.
Clinical Evidence
Bench testing only. Physicochemical testing demonstrated equivalency to predicates. Biocompatibility testing (systemic injection, primary ocular irritation, cytotoxicity) confirmed non-toxic and non-irritating nature. UV-blocking compound extraction testing showed no unsafe levels.
Technological Characteristics
Material: hefilcon A (HEMA/NVP copolymer cross-linked with EGDMA). Water content: 42%. UV-absorbing compound: 2-(benzoyl-3-hydroxyphenoxy)ethyl acrylate. Tint: Piqment Blue 15. Manufacturing: Molded. Refractive index: 1.416. Specific gravity: 1.039. Dk (35°C): 13.375 x 10^-11. Visible light transmittance: 98%. UV transmittance: <10%.
Indications for Use
Indicated for daily wear correction of myopia, hyperopia, astigmatism, and presbyopia in aphakic or non-aphakic patients with non-diseased eyes. Astigmatism correction limits: 2.00D for Single Vision/Multifocal, 5.00D for Toric.
Regulatory Classification
Identification
A soft (hydrophilic) contact lens is a device intended to be worn directly against the cornea and adjacent limbal and scleral areas of the eye to correct vision conditions or act as a therapeutic bandage. The device is made of various polymer materials the main polymer molecules of which absorb or attract a certain volume (percentage) of water.
{0}------------------------------------------------
K040900
## Optical Connection Inc.
# 510(k) Summary
## Submitter Information:
Optical Connection, Inc. a subsidiary of St. Shine Optical Co., Ltd. 3315 Almaden Expressway, Suite 25 San Jose, CA, 95125
| Contact Person: | Garold L. Edwards, O.D., F.A.A.C<br>Regulatory Consultant |
|-----------------|-----------------------------------------------------------|
| Telephone: | (408) 221-3860 |
| Fax: | (831) 335-0166 |
- Date Prepared: March 23, 2004
## Device Name:
| Common Name: | Soft (Hydrophilic) Contact Lens |
|--------------------------|--------------------------------------------------------------------------------------|
| Trade/Proprietary Names: | S 42 UV Single Vision (hefilcon A) Soft<br>(Hydrophilic) Contact Lens for Daily Wear |
| | T 42 UV Toric (hefilcon A) Soft (Hydrophilic)<br>Contact Lens for Daily Wear |
| | M 42 UV Multifocal (hefilcon A) Soft (Hydrophilic)<br>Contact Lens for Daily Wear |
| Classification Name: | Soft (Hydrophilic) Contact Lens |
| Device Classification: | Class II (21 CFR 886.5925) |
{1}------------------------------------------------
### Predicate Devices:
The molded Specialty 42 UV, Specialty 42 UV Toric (hefilcon A) Soft (Hydrophilic) Contact Lenses and Saview 55 Multifocal (methafilcon A) Soft (Hydrophilic Contact Lens were selected as the predicate devices.
Specialty 42 UV, Specialty 42 UV Toric (hefilcon A) and Saview 55 Multifocal (methafilcon A) devices are manufactured in the same facility, under the same quality system, using the same molding, tinting, packaging and sterilization processes.
### Description of Devices:
S 42 UV Single Vision, T 42 UV Toric, and M 42 UV Multifocal (hefilcon A) Soft (Hydrophilic) Daily Wear Contact Lenses are hemispherical flexible shells that cover the cornea and a portion of the adjacent sclera. The 42 UV Contact Lens is available in a single vision lens design, the 42 UV Toric Contact Lens is available in a back surface design, and the 42 UV Multifocal lens is available in an aspheric lens design. The lens material (hefilcon A) is a hydrophilic polymer of 2-hydroxyethyl methacrylate (HEMA) and n-vinyl-2-pyrrolidone (NVP) cross-linked with ethyleneglycol dimethacrylate (EGDMA), using AIBN as the initiator. The lens contains 42% water by weight. 42 UV, 42 UV Toric, and 42 UV Multifocal (hefilcon A) Soft (Hydrophilic) Contact Lenses for Daily Wear are tinted using Piqment Blue 15, ([Phthalocyaninato(2-)] copper) which is approved for coloring contact lenses under 21 CFR 5 74.3045 and a UV absorbing compound, 2-(benzoyl-3-hydroxyphenoxy)ethyl acrylate, has been incorporated into the lens polymer.
{2}------------------------------------------------
# Comparison to Predicate Devices
| Lens type<br>Predicate<br>device 510(k) | S 42 UV Single<br>Vision, T 42 UV<br>Toric and<br>M 42 UV Multifocal<br>(hefilcon A) Soft<br>(Hydrophilic)<br>Contact Lenses for<br>Daily Wear<br>(molded) | S 42 UV Single<br>Vision, T 42 UV<br>Toric and<br>M 42 UV Multifocal<br>(hefilcon A) Soft<br>(Hydrophilic)<br>Contact Lenses for<br>Daily Wear<br>(lathe-cut) | Specialty 42 and<br>Specialty 42 Toric<br>(hefilcon A) Soft<br>(Hydrophilic)<br>Contact Lenses for<br>Daily Wear | Specialty 42 UV and<br>Specialty 42 UV<br>Toric (hefilcon A)<br>Soft (Hydrophilic)<br>Contact Lenses for<br>Daily Wear | Saview 55<br>Multifocal<br>(methafilcon A) Soft<br>(Hydrophilic)<br>Contact Lens for<br>Daily Wear |
|-----------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------|
| | | | K000376 | K011089 | |
| | | | | | K030548 |
| Material | hefilcon A | hefilcon A | hefilcon A | hefilcon A | methafilcon A |
| Material<br>classification | Hydrophilic Lens<br>Group 1 | Hydrophilic Lens<br>Group 1 | Hydrophilic Lens Group<br>1 | Hydrophilic Lens Group<br>1 | Hydrophilic Lens Group<br>4 |
| Indication for<br>use | myopia, hyperopia,<br>astigmatism, and<br>presbyopia | myopia, hyperopia,<br>astigmatism, and<br>presbyopia | myopia, hyperopia and<br>astigmatism | myopia, hyperopia and<br>astigmatism | myopia, hyperopia and<br>presbyopia |
| Water<br>content | 42% | 42% | 42% | 42% | 55% |
| Visible light<br>transmittance | 98% | 98% | 98% | 98% | 97.6% |
| UV<br>transmittance | < 10% | < 10% | N/A | < 10% | N/A |
| Dk (35°C) | 13.375 x 10 -11 | 13.375 x 10 -11 | 13.250 x 10 -11 | 13.375 x 10 -11 | 19.5 x 10 -11 |
| Powers | +20.00 to -20.00<br>Diopters | +20.00 to -20.00<br>Diopters | +20.00 to -20.00<br>Diopters | +20.00 to -20.00<br>Diopters | +12.00 to -20.00<br>Diopters; continuous<br>add power to +3.25 |
| Color | blue visibility | blue visibility | clear or blue visibility | blue visibility | blue visibility |
| Refractive<br>index | 1.416 | 1.416 | 1.417 | 1.416 | 1.3974 (wet) |
| Specific<br>gravity | 1.039 | 1.039 | 1.031 | 1.039 | N/A |
| Method of<br>manufacture | Molded | Lathe-cut | Moulded | Molded | Molded |
{3}------------------------------------------------
## Indications for Use:
The S 42 UV Single Vision (hefilcon A) Soft (Hydrophilic) Contact Lens is indicated for daily wear for the correction of refractive ametropia (myopia and hyperopia) in aphakic or not-aphakic persons with non-diseased eyes that may exhibit refractive and/or corneal astigmatism up to 2.00 diopters that does not interfere with visual acuity.
The T 42 UV Toric (hefilcon A) Soft (Hydrophilic) Contact Lens is indicated for daily wear for the correction of refractive ametropia (myopia, hyperopia and astigmatism) in aphakic and/or not aphakic persons with non-diseased eyes that may exhibit refractive and/or corneal astigmatism up to 5.00 Diopters.
The M 42 UV Multifocal (hefilcon A) Soft (Hydrophilic) Contact Lens is indicated for daily wear for the correction of refractive ametropia (myopia and hyperopia) and presbyopia in aphakic or not-aphakic persons with non-diseased eyes that may exhibit refractive and/or corneal astigmatism up to 2.00 diopters that does not interfere with visual acuity.
The lenses may be disinfected using chemical (not heat) or hydrogen peroxide disinfecting systems. Eyecare practitioners may prescribe the lenses for daily wear and/or frequent replacement. When prescribed for a Frequent Replacement Program, the lenses may be disinfected using chemical (not heat), or hydrogen peroxide disinfecting systems.
## Description of Safety and Substantial Equivalence:
A series of pre-clinical tests were performed to demonstrate the safety and effectiveness of S 42 UV Single Vision, T 42 UV Toric, and M 42 UV Multifocal (hefilcon A) Soft (Hydrophilic) Contact Lenses for Daily Wear, and to establish substantial equivalence to the predicate devices.
Results of Systemic Injection, Primary Ocular Irritation and Cytotoxicity Tests show the lenses to be non-toxic and non-irritating. 42 UV lenses were extracted and evaluated for presence of the UV blocking compound. Results showed no evidence of unsafe amounts of this compound in the extracts. Physicochemical testing of 42 UV lenses demonstrated equivalency to the predicate devices.
## Conclusion:
Information submitted in the 510(k) establishes that S 42 UV Single Vision, T 42 UV Toric, and M 42 UV Multifocal (hefilcon A) contact lenses have comparable physicochemical properties to the predicate devices and do not raise questions of safety and effectiveness. Shelf life testing has shown the lenses remain sterile and that lens properties do not change before the expiration date. Therefore, the devices are substantially equivalent to the predicate devices.
{4}------------------------------------------------
DEPARTMENT OF HEALTH & HUMAN SERVICES • USA
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
JUN 1 5 2004
Optical Connection, Inc. c/o Garold L. Edwards, O.D., F.A.A.O. Regulatory Consultant 3315 Almaden Expressway, Suite 25 San Jose, CA 95118
Re: K040900
Trade/Device Name: rade/Device Name:
S 42 UV Single Vision (hefilcon A) Soft (Hydrophilic) Contact Lens for Daily Wear (Molded) S 42 UV Single Vision (nemoon A) Soft (Hydrophilic) Contact Lens for Daily Wear (Molded) 1 42 UV Multifocal (hefileon A) Soft (Hydrophilic) Contact Lens for Daily Wear (Molded) Regulation Number: 21 CFR 886.5925 Regulation Name: Soft (hydrophilic) contact lens Regulatory Class: Class II Product Code: LPL Dated: March 23, 2004 Received: April 6, 2004
Dear Dr. Edwards:
We have reviewed your Section 510(k) premarket notification of intent to market the device we nave leviewed your boother 5. only programially equivalent (for the indications for releved above and have acceminate marketed predicate devices marketed in interstate commerce use stated in the eneround date of the Medical Device Amendments, or to devices that provision in way 20, 1970, the onactions with the provisions of the Federal Food, Drug, and Cosmetic have been receasined in are are a proval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general therefore, market the act include requirements for annual registration, listing of devices, good oonarols provisions or allabeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it IT your device is olassinod (600 above) and Existing major regulations affecting your device can be inay oc subject to subli additions, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that I lease be advisod that i Dr is issual.org ur device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all I cacration and regulations annument limited to: registration and listing (21 CFR Part 807); the Her s requirements, metading, vanufacturing practice requirements as set forth in the quality labeling (21 OF RT art 807), good nort 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{5}------------------------------------------------
This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (301) 594-4613. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html
Sincerely yours,
A. Kely C. Rosenthal
A. Ralph Rosenthal, M.D Director Division of Ophthalmic and Ear, Nose and Throat Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{6}------------------------------------------------
# Indications for Use
510(k) Number: K040900
Device Names:
S 42 UV Single Vision (hefilcon A) Soft (Hydrophilic) Contact Lens for Daily Wear T 42 UV Toric (hefilcon A) Soft (Hydrophilic) Contact Lens for Daily Wear M 42 UV Multifocal (hefilcon A) Soft (Hydrophilic) Contact Lens for Daily Wear
Indications for Use:
S 42 UV Single Vision (hefilcon A) Soft (Hydrophilic) Contact Lens for Daily Wear is indicated for daily wear for the correction of refractive ametropia (myopia and hyperopia) in aphakic or not aphakic persons with non-diseased eyes that may exhibit refractive and/or corneal astigmatism up to 2.00 Diopters that does not interfere with visual acuity.
T 42 UV Toric (hefilcon A) Soft (Hydrophilic) Contact Lens for Daily Wear is indicated for daily wear for the correction of refractive ametropia (myopia, hyperopia and astigmatism) in aphakic and not aphakic persons with non-diseased eyes that may exhibit refractive and/or corneal astigmatism up to 5.00 Diopters.
M 42 UV Multifocal (hefilcon A) Soft (Hydrophilic) Contact Lens for Daily Wear is indicated for daily wear for the correction of refractive ametropia (myopia and hyperopia) and presbyopia in aphakic or not aphakic persons with non-diseased eyes that may exhibit refractive and/or corneal astigmatism up to 2.00 Diopters that does not interfere with visual acuity.
The lenses may be disinfected using chemical (not heat) or hydrogen peroxide, disinfecting systems. Eyecare practitioners may prescribe the lenses for daily wear and/or frequent replacement.
Prescription Use _____________________________________________________________________________________________________________________________________________________________ (Part 21 CFR 801 Subpart D)
AND/OR -
Over-The-Counter Use (21 CFR 807 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
```
(Division Sign-Off)
Division of Ophthalmic Evg
510(k) Number K040900
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.