DSM as part of a standard x-ray mammography unit is part of a stereotactic lesion localization system that has the application of localizing, and then giving a physician the capability of performing fine needle aspiration or core biopsy of lesions determined to be suspicious through prior mammographic examination
Device Story
Digital Spot Mammography (DSM) system functions as an add-on to standard x-ray mammography units and stereotactic breast biopsy systems. Input consists of x-ray signals captured by a Charge-Coupled Device (CCD) digital image receptor. The system processes these signals to produce near real-time digital images on a high-resolution flat-panel monitor. Operated by clinicians in a clinical setting, the system allows physicians to visualize suspicious breast lesions to guide fine needle aspiration or core biopsy procedures. Output includes digital images for display and archiving on CD/DVD media. DICOM connectivity enables query/retrieval operations. The device assists in precise lesion targeting, potentially improving biopsy accuracy and patient outcomes.
Clinical Evidence
No clinical data provided; bench testing only.
Technological Characteristics
System includes workstation, monitor, trackball, keyboard, computer, and CCD digital image receptor. Imaging area 5.0 cm x 5.0 cm; 512 or 1024 pixel resolution. Intel-based microprocessor. Connectivity via DICOM. Compliance with IEC-601-1, IEC 601-1-x, IEC 601-2-xx, UL 1950, FCC-A, DHHS, CSA-950, and IEC 950 standards.
Indications for Use
Indicated for patients requiring stereotactic lesion localization, fine needle aspiration, or core biopsy of breast lesions identified as suspicious via prior mammographic examination.
Regulatory Classification
Identification
A mammographic x-ray system is a device intended to be used to produce radiographs of the breast. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.
{0}------------------------------------------------
| Summary of Safety and Effectiveness | | |
|-------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------|
| K040884 | | APR 2 3 2004 |
| Submitter: | Hologic, Inc.<br>35 Crosby Drive<br>Bedford, MA 01730 | |
| Contact Person: | Gail Yaeker-Daunis<br>Sr. Regulatory Specialist | |
| Date Prepared: | April 1, 2004 | |
| Common Name: | DSM | |
| Proprietary Name: | Digital Spot Mammography System | |
| Predicate Devices: | Lorad DSM K921962A & K030666 | |
| Intended Use: | DSM as part of a standard x-ray mammography unit is part of a<br>stereotactic lesion localization system that has the application of<br>localizing, and then giving a physician the capability of performing<br>fine needle aspiration or core biopsy of lesions determined to be<br>suspicious through prior mammographic examination | |
## Compliance Statement:
The DSM has been designed for certification to International Electrotechnical Commission Standard IEC-601-1, IEC 601-1-x, IEC 601-2-xx
cUL 187, UL 1950, FCC-A, DHHS, CSA-950, IEC 950
A comprehensive Operator's Manual provided with each system is user friendly and comprehensive, thus ensuring safe and effective operation of the DSM.
This information is provided pursuant to the requirements of the Safe Medical Devices Act of 1990 (SMDA).
{1}------------------------------------------------
#### General Information Section 1.0
Ko4o884
## Introduction 1.1
The purpose of this Special 510(k) is to modify the Lorad Digital Spot Mammography System which was originally cleared for market as 510 (k) # K921962A At that time, DSM was cleared for use on the Lorad Mammography systems and the Lorad StereoGuide Stereotactic Breast Biopsy System for use in localization and stereotactic procedures.
DSM was last modified as part of the Lorad MultiCare Platinum System (formerly StereoGuide) with Digital Spot Mammography as 510 (k) # K030666.
At this time, Lorad wishes to modify DSM to include DICOM capabilities and to provide capability for use with both the MultiCare Platinum Breast Biopsy System and Lorad Mammography Systems.
#### 1.2 Indications for Use
The indications for use of the DSM are the same as submitted in the Lorad MultiCare Platinum System, 510 (k) # K030666, and repeated below:
DSM as part of a standard x-ray mammography unit is part of a stereotactic lession localization system that has the application of localizing, and then giving a physician the capability of performing fine needle aspiration or core biopsy of lesions determined to be suspicious through prior mammographic examination.
### 1.3 Substantial Equivalence
The DSM and the predicate device DSM as cleared in MultiCare Platinum 510(k) # K030666 have the same intended use, the same general configuration, the same principles of operation, and similar operating parameters. The DSM is now available for use with both Lorad MultiCare Platinum Breast Biopsy Table System and M-IV Mammography System. The addition of DICOM capability is considered the only significant change to the device.
{2}------------------------------------------------
DSM is comprised of a workstation, monitor, trackball, keyboard, computer and digital image receptor.
Image /page/2/Picture/3 description: In the image, a computer is sitting on a desk. The computer has a monitor, keyboard, and tower. The desk is made of wood and has a simple design. The computer is likely used for work or school.
Table 1 below compares the current DSM to the predicate DSM cleared on February 27, 2003 as 510(k) # K030666.
| DSM Console | DSM (K030666) Lorad Multicare | DSM |
|---------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------|
| | Platinum | |
| Tissue Imaging Area | 5.0 cm x 5.0 cm<br>Operator control of Image contrast &<br>luminance: Choice of 512 or 1024<br>pixel resolution<br>Large 18 in., high resolution, gray<br>scale flat screen monitor | Same |
| Near Real-Time<br>Image Display | 512 Mode - approx. 4 seconds<br>1024 Mode - approx 8 seconds | Same |
| Image Device | Charge-Coupled Device (CCD) | Same |
| Monitor | Flat Panel | Same |
| Display Area | 18 in. diagonal (minimum) 160°<br>viewing angle (minimum) | Same |
| Computer<br>Microprocessor | Intel CPU | Same |
| Archive Media | DVD + R/RW, CD - R/RW | Same |
| Operating System | Operating system will only recognize<br>disks, DVD/CD formatted with DOS,<br>Windows 9X/NT/2000 | Same |
| User Controls | Power, Brightness, Contrast, Picture<br>Tilt, Height Vertical Position,<br>Horizontal Position | Same |
{3}------------------------------------------------
Section 1.0
| DSM Console | DSM (K030666) Lorad Multicare<br>Platinum | DSM |
|-------------|-------------------------------------------|------|
| Approvals | UL 1950, FCC-A, DHHS, CSA-950,<br>IEC 950 | Same |
| DICOM | DICOM Query/Retrieval Operations | Same |
{4}------------------------------------------------
Image /page/4/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized depiction of an eagle or bird-like figure with three curved lines representing its wings or body. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" is arranged in a circular fashion around the bird symbol.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
# APR 2 3 2004
Ms. Gail Yaeker-Daunis Sr. Regulatory Specialist Hologic, Inc. 36 Apple Ridge Road DANBURY CT 06810
Re: K040884 Trade/Device Name: Lorad Digital Spot Mammography System (DSM) Regulation Number: 21 CFR 892.1710 Regulation Name: Mammographic x-ray system Regulatory Class: II Product Code: 90 IZH Dated: April 2, 2004 Received: April 6, 2004
Dear Ms. Yaeker-Daunis:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA), You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050
{5}------------------------------------------------
This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at one of the following numbers, based on the regulation number at the top of the letter;
| 8xx. 1xxx | (301) 594-4591 |
|----------------------------------|----------------|
| 876.2xxx, 3xxx, 4xxx, 5xxx | (301) 594-4616 |
| 884.2xxx, 3xxx, 4xxx, 5xxx, 6xxx | (301) 594-4616 |
| 892.2xxx, 3xxx, 4xxx, 5xxx | (301) 594-4654 |
| Other | (301) 594-4692 |
Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97) you may obtain. Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html.
Sincerely yours,
Nancy C. Brogdon
Nancy C. Brogdon Director, Division of Reproductive, Abdominal and Radiological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{6}------------------------------------------------
510(k) Number (if known):
Device Name: Lorad Digital Spot Mammography System (DSM)
Indications for Use:
DSM as part of a standard x-ray mammography unit is part of a stereotactic lesion Down as part of a standad as the application of localizing, and then giving a physician the location 37 stem that has the uppers aspiration or core biopsy of lesions determined to be suspicious through prior mammographic examination
Concurrence of CDRH, Office of Device Evaluation (ODE)
Prescription Use V 21 CFR 801.109
OR
Over-the-Counter Use
David h. leppmm
(Division Sign-Off) Division of Reproductive, Abdominal, and Radiological Devices 14008 510(k) Number
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.