# VANGUARD PATELLA COMPONENTS (K040770)

_Biomet Manufacturing, Inc. · JWH · Apr 22, 2004 · Orthopedic · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K040770

## Device Facts

- **Applicant:** Biomet Manufacturing, Inc.
- **Product Code:** [JWH](/productcode/JWH.md)
- **Decision Date:** Apr 22, 2004
- **Decision:** SESE
- **Submission Type:** Special
- **Regulation:** 21 CFR 888.3560
- **Device Class:** Class 2
- **Review Panel:** Orthopedic
- **Attributes:** Therapeutic

## Indications for Use

The Vanguard™ Patella is intended use is for replacement of part of the knee joint in conjunction with a femoral and tibial component. Specifically: 1. Painful and disabled knee joint resulting from osteoarthritis, rheumatoid arthritis, traumatic arthritis where one or more compartments are involved. 2. Correction of varus, valqus or posttraumatic deformity. 3. Correction or revision of unsuccessful osteotomy, arthrodesis, or failure of previous joint replacement procedure.

## Device Story

Vanguard™ Patella Components are orthopedic implants designed for knee joint replacement. Device features a true dome shape, under-surface grooves, and single or 3-peg fixation options; available with or without x-ray wire. Constructed from ArCom polyethylene. Used in conjunction with femoral and tibial components to restore joint function. Intended for surgical implantation in patients with degenerative or traumatic knee conditions. Device provides a bearing surface to replace damaged patellar cartilage, aiming to reduce pain and correct joint deformity. No software or electronic components.

## Clinical Evidence

No clinical data provided. Equivalence established via engineering analysis.

## Technological Characteristics

Material: ArCom polyethylene. Design: True dome shape, single or 3-peg configuration, optional x-ray wire. Mechanical fixation. No software or electronic components.

## Regulatory Identification

A knee joint patellofemorotibial polymer/metal/polymer semi-constrained cemented prosthesis is a device intended to be implanted to replace a knee joint. The device limits translation and rotation in one or more planes via the geometry of its articulating surfaces. It has no linkage across-the-joint. This generic type of device includes prostheses that have a femoral component made of alloys, such as cobalt-chromium-molybdenum, and a tibial component or components and a retropatellar resurfacing component made of ultra-high molecular weight polyethylene. This generic type of device is limited to those prostheses intended for use with bone cement (§ 888.3027).

## Predicate Devices

- AGC® LP Patellar Button ([K912245](/device/K912245.md))
- AGC® LP Patellar Button ([K921182](/device/K921182.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

APR 2 2 2004

Image /page/0/Picture/1 description: The image shows two lines of text. The first line appears to contain the numbers "1.11", with the last "1" being slightly separated from the others. The second line contains the alphanumeric string "K040770", with each character clearly distinguishable.

Image /page/0/Picture/2 description: The image shows the word "BIOMET" in large, bold, black letters. Below the word "BIOMET" is the phrase "510(k) Summary" in a smaller font size. The text is black and the background is white.

Applicant/Sponsor:Biomet Manufacturing Corp.

Patricia Sandborn Beres Contact Person: Senior Regulatory Specialist

Proprietary Name: Vanguard™ Patella Components

Common Name: Patellar Component

Classification Name: Knee joint patellofemorotibial, cemented, polymer/metal/ polymer, semi-constrained

Legally Marketed Devices To Which Substantial Equivalence Is Claimed: AGC® LP Patellar Button - K912245 and K921182

Device Description: Features of the Vanguard™ Patella Components are as follows:

- 1", true dome shape .
- Grooves on the under surface ●
- Single or 3-peg design �
- Available with or without an x-ray wire .
- ArCom polvethvlene .

Intended Use: The intended use of the Vanguard™ Patella Components is replacement of part of the knee joint in conjunction with a femoral and tibial component. Specifically:

- 1. Painful and disabled knee joint resulting from osteoarthritis, rheumatoid arthritis, traumatic arthritis where one or more compartments are involved.
- 2. Correction of varus, valgus or posttraumatic deformity
- 3. Correction or revision of unsuccessful osteotomy, arthrodesis, or failure of previous joint replacement procedure.

Summary of Technologies: The overall design, materials and processing of the Vanguard™ Patella Components are similar to the predicate device.

Non-Clinical Testing: Engineering analysis demonstrated equivalence between the Vanguard™ Patella Components and the predicate devices.

## Clinical Testing: None provided

All trademarks are property of Biomet, Inc. MALING ADDRESS PO. BOX 587 Warsaw, IN 46581-0587

.

Image /page/0/Picture/25 description: The image shows the word "FAX" in bold, followed by the number "52" in a larger, bolder font. The number appears to be part of a phone or fax number. The rest of the number is not clear.

{1}------------------------------------------------

Image /page/1/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo is a circular seal with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is a stylized image of three human figures, represented by flowing lines, suggesting unity and collaboration.

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

APR 2 2 2004

Patricia Sandborn Beres Senior Regulatory Specialist Biomet Manufacturing Corporation 56 East Bell Drive P.O. Box 587 Warsaw, Indiana 46581-0587

Re: K040770

Trade/Device Name: Vanguard™ Patella Components Regulation Number: 21 CFR 888.3560 Regulation Name: Knee joint, patellofemorotibial, polymer/metal/polymer semi-constrained cemented prosthesis Regulatory Class: II Product Code: JWH Dated: March 24, 2004 Received: March 25, 2004

Dear Ms. Beres:

We have reviewed your Section 510(k) premarket notification of intent to market the device we nave revious and have determined the device is substantially equivalent (for the indications forcroneed a o o and in sure) to legally marketed predicate devices marketed in interstate for associated in the May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. ate nos that have boom that do not require approval of a premarket approval application (PMA). and Cosmeter Prov, 100 (110) 1107 the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can may or dayset to de of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must or any I with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

{2}------------------------------------------------

This letter will allow you to begin marketing your device as described in your Section 510(k) I mis letter will and in your your finding of substantial equivalence of your device to a legally prematicated predicated. The classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please If you attire of Compliance at (301) 594-4659. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small other general international and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html

Sincerely yours,

Miriam C. Provost

Celia M. Witten, Ph.D., M.D. Director Division of General, Restorative and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{3}------------------------------------------------

f I

## Indications for Use

510(k) Number (if known):

Device Name: Vanguard™ Patella Components

Indications For Use:

The Vanguard™ Patella is intended use is for replacement of part of the knee joint in conjunction with a femoral and tibial component. Specifically:

- 1. Painful and disabled knee joint resulting from osteoarthritis, rheumatoid arthritis, traumatic arthritis where one or more compartments are involved.
- 2. Correction of varus, valqus or posttraumatic deformity.
- 3. Correction or revision of unsuccessful osteotomy, arthrodesis, or failure of previous joint replacement procedure.

Prescription Use × (Part 21 CFR 801 Subpart D) AND/OR

Over-The-Counter Use (21 CFR 807 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

Muriam C. Provost

(Division Sign-Off) Division of General, Restorative, and Neurological Devices

Page 1 of 1

510(k) Number_K640770

---

**Source:** [https://fda-staging.innolitics.com/device/K040770](https://fda-staging.innolitics.com/device/K040770)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).