← Product Code [MNH](/productcode/MNH) · K040747

# ALADYN RIGID SPINE PLATE (K040747)

_Scient'X · MNH · Jun 17, 2004 · Orthopedic · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K040747

## Device Facts

- **Applicant:** Scient'X
- **Product Code:** [MNH](/productcode/MNH.md)
- **Decision Date:** Jun 17, 2004
- **Decision:** SESE
- **Submission Type:** Special
- **Regulation:** 21 CFR 888.3070
- **Device Class:** Class 2
- **Review Panel:** Orthopedic
- **Attributes:** Therapeutic

## Indications for Use

The Aladyn Rigid Spine Plate is a pedicle screw system intended to provide immobilization and stabilization of spinal segments in skeletally mature patients as an adjunct to fusion in the treatment of the following acute and chronic instabilities or deformities of the thoracic, lumbar, and sacral spine: degenerative spondylolisthesis with objective evidence of neurological impairment, fracture, dislocation, scoliosis, kyphosis, spinal tumor, and failed previous fusion (pseudarthrosis). As a pedicle screw system Aladyn Rigid Spine Plate is indicated for the treatment of severe spondylolisthesis (Grades 3 and 4) of the L5-S1 vertebra in skeletally mature patients receiving fusion by autogenous bone graft having implants attached to the lumbar and sacral spinc (L3 to sacrum) with removal of the implants after the attainment of a solid fusion.

## Device Story

Aladyn Rigid Spine Plate is a posterior pedicle screw system for spinal immobilization/stabilization. System comprises rigid plates, hemispherical pedicle screws, and polyaxial nuts. Screws are inserted into vertebrae; plate is placed over screws and secured with nuts. Used in thoracic, lumbar, and sacral spine procedures. Modification of ISOLOCK and ISOBAR systems; eliminates use of washers. Implantation performed by surgeons in clinical/OR settings. Provides rigid fixation to support fusion; implants may be removed after solid fusion attainment.

## Clinical Evidence

Bench testing only. Mechanical testing performed per ASTM F1717. Results demonstrated performance comparable to predicate ISOLOCK and ISOBAR Pedicle Screw Systems.

## Technological Characteristics

Pedicle screw spinal system consisting of rigid plates, hemispherical screws (6.2mm diameter, 38-50mm lengths), and polyaxial nuts. Materials and design are modifications of predicate systems. Mechanical testing conducted per ASTM F1717. No software or electronic components.

## Regulatory Identification

(1) Rigid pedicle screw systems are comprised of multiple components, made from a variety of materials that allow the surgeon to build an implant system to fit the patient's anatomical and physiological requirements. Such a spinal implant assembly consists of a combination of screws, longitudinal members (e.g., plates, rods including dual diameter rods, plate/rod combinations), transverse or cross connectors, and interconnection mechanisms (e.g., rod-to-rod connectors, offset connectors).(2) Semi-rigid systems are defined as systems that contain one or more of the following features (including but not limited to): Non-uniform longitudinal elements, or features that allow more motion or flexibility compared to rigid systems.

## Special Controls

*Classification.* (1) Class II (special controls), when intended to provide immobilization and stabilization of spinal segments in skeletally mature patients as an adjunct to fusion in the treatment of the following acute and chronic instabilities or deformities of the thoracic, lumbar, and sacral spine: severe spondylolisthesis (grades 3 and 4) of the L5-S1 vertebra; degenerative spondylolisthesis with objective evidence of neurologic impairment; fracture; dislocation; scoliosis; kyphosis; spinal tumor; and failed previous fusion (pseudarthrosis). These pedicle screw spinal systems must comply with the following special controls:(i) Compliance with material standards;
(ii) Compliance with mechanical testing standards;
(iii) Compliance with biocompatibility standards; and
(iv) Labeling that contains these two statements in addition to other appropriate labeling information:
 “Warning: The safety and effectiveness of pedicle screw spinal systems have been established only for spinal conditions with significant mechanical instability or deformity requiring fusion with instrumentation. These conditions are significant mechanical instability or deformity of the thoracic, lumbar, and sacral spine secondary to severe spondylolisthesis (grades 3 and 4) of the L5-S1 vertebra, degenerative spondylolisthesis with objective evidence of neurologic impairment, fracture, dislocation, scoliosis, kyphosis, spinal tumor, and failed previous fusion (pseudarthrosis). The safety and effectiveness of these devices for any other conditions are unknown.”
“Precaution: The implantation of pedicle screw spinal systems should be performed only by experienced spinal surgeons with specific training in the use of this pedicle screw spinal system because this is a technically demanding procedure presenting a risk of serious injury to the patient.”
(2) Class II (special controls), when a rigid pedicle screw system is intended to provide immobilization and stabilization of spinal segments in the thoracic, lumbar, and sacral spine as an adjunct to fusion in the treatment of degenerative disc disease and spondylolisthesis other than either severe spondylolisthesis (grades 3 and 4) at L5-S1 or degenerative spondylolisthesis with objective evidence of neurologic impairment. These pedicle screw systems must comply with the following special controls:
(i) The design characteristics of the device, including engineering schematics, must ensure that the geometry and material composition are consistent with the intended use.
(ii) Non-clinical performance testing must demonstrate the mechanical function and durability of the implant.
(iii) Device components must be demonstrated to be biocompatible.
(iv) Validation testing must demonstrate the cleanliness and sterility of, or the ability to clean and sterilize, the device components and device-specific instruments.
(v) Labeling must include the following:
(A) A clear description of the technological features of the device including identification of device materials and the principles of device operation;
(B) Intended use and indications for use, including levels of fixation;
(C) Identification of magnetic resonance (MR) compatibility status;
(D) Cleaning and sterilization instructions for devices and instruments that are provided non-sterile to the end user; and
(E) Detailed instructions of each surgical step, including device removal.
(3) Class II (special controls), when a semi-rigid system is intended to provide immobilization and stabilization of spinal segments in the thoracic, lumbar, and sacral spine as an adjunct to fusion for any indication. In addition to complying with the special controls in paragraphs (b)(2)(i) through (v) of this section, these pedicle screw systems must comply with the following special controls:
(i) Demonstration that clinical performance characteristics of the device support the intended use of the product, including assessment of fusion compared to a clinically acceptable fusion rate.
(ii) Semi-rigid systems marketed prior to the effective date of this reclassification must submit an amendment to their previously cleared premarket notification (510(k)) demonstrating compliance with the special controls in paragraphs (b)(2)(i) through (v) and paragraph (b)(3)(i) of this section.

## Predicate Devices

- ISOLOCK Pedicle Screw System ([K990721](/device/K990721.md))
- ISOBAR Pedicle Screw System ([K990118](/device/K990118.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

K040741
page 1 of 1

5100k) Summary

Date March 17, 2004

# JUN 1 7 2004

Submitter

Scient'x Batiment Calypso Parc Ariane 3 78284 Guyancourt FRANCE

## Contact person

J.D. Webb 1001 Oakwood Blvd Round Rock, TX 78681 512-388-0199

Common name Posterior pedicle screw system

### Classification name

Spondylolisthesis spinal fixation device system Pedicle screw spinal system

### Equivalent Device

The Aladyn Rigid Spine Plate is a modification to the ISOLOCK (K990721) and ISOBAR (K990118) Pedicle Screw Systems.

### Device Description

The Aladyn Rigid Spine Plates, pedicle screws and a polyaxial nut. It is a modification to the plates and nuts of the ISOLOCK Pedicle Screw System (K990721) and uses the same hemispherical screws of the ISOBAR system (K990118). Pedicle screws are inserted into the vertebrae. An Aladyn plate is placed over the screws. The rigid plates are available in six multi-hole lengths. The construct is then scurely fixed with nuts. The implantation steps of the Aladyn are the SOLOCK Plate system. The only difference is that there are no washers used with the Aladyn plate.

The hemispherical screws come in two diameters (6.2mm and in lengths ranging from 38mm to 50mm. There is also a polyaxial nut available.

#### Intended Use

The Aladyn Rigid Spine Plate is a pedicle screw system intended to provide immobilization of spinal segments in skeletally mature patients as an adjunct to fusion in the treatment of the following acute and chronic instabilities or deformities of the thoracic, lumbar, and sacral spine: degenerative spondylolisthesis with objective evidence of neurological impairment, fracture, dislocation, scoliosis, spinal tumor, and failed previous fusion (pseudarthrosis).

As a pedicle screw system Aladyn Rigid Spine Plate is indicated for the treatment of severe spondylolisthesis (Grades 3 and 4) of the LS-S1 vertebra in skeletally mature patients receiving fusion by autogenous bone graft having implants attached to the lumbar and sacral spine (L3 to sacrum) with removal of the attainment of a solid fusion.

#### Summary Nonclinical Tests

Testing was performed per ASTM F1717. Results were comparable with other devices on the ISOLOCK and ISOBAR Pedicle Screw Systems.

{1}------------------------------------------------

Image /page/1/Picture/1 description: The image shows the seal of the Department of Health & Human Services - USA. The seal is circular and contains the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. In the center of the seal is a stylized image of a caduceus, a symbol often associated with medicine and healthcare.

Public Health Service

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

JUN 1 7 2004

Scient'x c/o Mr. J.D. Webb The OrthoMedix Group, Inc. 1001 Oakwood Boulevard Round Rock, Texas 78681

Re: K040747

Trade/Device Name: Aladyn Rigid Spine Plate Regulation Number: 21 CFR 888.3070 Regulation Name: Pedicle screw spinal system Regulatory Class: II Product Code: MNH, MNI Dated: May 17, 2004 Received: May 18, 2004

Dear Mr. Webb:

We have reviewed your Section 510(k) premarket notification of intent to market the device we have reviewed your bection of the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate 101 use stated in the encreated of the enactment date of the Medical Device Amendments, or to commence prior to may 20, 1978, are exactiance with the provisions of the Federal Food, Drug. devices that have been rechasined in assess approval of a premarket approval application (PMA). alla Cosmetic Act (71ct) that do not required to the general controls provisions of the Act. The 1 ou may, merciolo, manel and not include requirements for annual registration, listing of general controls provisions of the tice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it If your device is classified (soo aboro) ins. Existing major regulations affecting your device can
may be subject to such additional controls. Existing major regulation of D thay be subject to subli additions, Title 21, Parts 800 to 898. In addition, FDA may be found in the Oous of reacts concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean Flease of advised that I DT of issuants arour device complies with other requirements of the Act mat I DA has made a dolor regulations administered by other Federal agencies. You must of any I cuttar statutes and regalaments, including, but not limited to: registration and listing (21 comply with an the Net 3 requirements, now 801); good manufacturing practice requirements as set CFK Fart 807), adocing (21 OF R Part 820); and if applicable, the electronic forth in the quality systems (Sections 531-542 of the Act); 21 CFR 1000-1050.

{2}------------------------------------------------

Page 2 - Mr. J.D. Webb

This letter will allow you to begin marketing your device as described in your Section 510(k) I his letter with anow you to ocgin mailioning your and equivalence of your device to a legally prematication. The PDA micing of easification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please if you desire specific advice for your ac 1001) 594-4659. Also, please note the regulation entitled, Colliact the Office of Company of Comment notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small other general miorination on Jour respenses its toll-free number (800) 638-2041 or Manufacturers, International and Soless http://www.fda.gov/cdrh/dsma/dsmamain.html

Sincerely yours,
Mark N Milherson

Celia M. Witten, Ph.D., M.D. Director Division of General, Restorative and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{3}------------------------------------------------

510(k) number (if known):

Device Name: Aladyn Rigid Spine Plate

Indications for Use:

## Aladyn Rigid Spine Plate Indications for Use

The Aladyn Rigid Spine Plate is a pedicle screw system intended to provide immobilization and stabilization of spinal segments in skeletally mature patients as an adjunct to fusion in the treatment of the following acute and chronic instabilities or deformities of the thoracic, lumbar, and sacral spine: degenerative spondylolisthesis with objective evidence of neurological impairment, fracture, dislocation, scoliosis, kyphosis, spinal tumor, and failed previous fusion (pseudarthrosis).

As a pedicle screw system Aladyn Rigid Spine Plate is indicated for the treatment of severe spondylolisthesis (Grades 3 and 4) of the L5-S1 vertebra in skeletally mature patients receiving fusion by autogenous bone graft having implants attached to the lumbar and sacral spinc (L3 to sacrum) with removal of the implants after the attainment of a solid fusion.

(PLEASE DO NOT WRITE BELOW THIS LINE – CONTINUE ON ANOTHER PAGE IF NEEDED)

| Concurrence of CDRH, Office of Device Evaluation (ODE)                              |                                 |
|-------------------------------------------------------------------------------------|---------------------------------|
| Prescription Use<br>(per 21 CFR 801.109)                                            | X                               |
|                                                                                     | OR                              |
|                                                                                     | Over-the-Counter Use ______     |
|                                                                                     | (Optional format 1-2-96) ______ |
| (Division Sign-off)<br>Division of General, Neurological<br>and Restorative Devices |                                 |
| 510(k) Number                                                                       |                                 |
| (Division Sign-Off)<br>Division of General, Restorative                             |                                 |

Division of General, Restorative, and Neurological Devices

Page is **510(k) Number** K040747

---

**Source:** [https://fda-staging.innolitics.com/device/K040747](https://fda-staging.innolitics.com/device/K040747)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
