K040735 · Alm S.A. · FSY · May 21, 2004 · General, Plastic Surgery
Device Facts
Record ID
K040735
Device Name
ALM X'TEN (X10) SURGICAL LIGHT SYSTEM
Applicant
Alm S.A.
Product Code
FSY · General, Plastic Surgery
Decision Date
May 21, 2004
Decision
SESE
Submission Type
Special
Regulation
21 CFR 878.4580
Device Class
Class 2
Indications for Use
ALM X'Ten™ (X10) Surgical Light Systems with Energix™ Power Supply are intended to be used to provide visible illumination of the surgical area or the patient. An optional "LEDinside" feature is available and produces lower intensity levels of light intended for minimally invasive surgeries, procedures and examinations. The Energix™ Power Supply is intended for use with surgical light systems.
Device Story
The ALM X'Ten™ Surgical Light System is a ceiling-mounted surgical lighting system used in operating rooms. It provides high-intensity illumination for surgical sites and a lower-intensity "LEDinside" ambient light mode for minimally invasive procedures. The system is powered by the Energix™ Power Supply, which allows for adjustable light intensity via a yoke-mounted keypad or remote RS232 communication. Some configurations include integrated video camera pre-wiring and flat-screen display supports. The device is operated by surgical staff to provide clear visualization of the patient. The system includes bulb failure indicators and an optional LCD for maintenance and operational status. By providing adjustable, reliable illumination, the device assists clinicians in performing surgical tasks, potentially improving procedural accuracy and patient outcomes.
Clinical Evidence
No clinical data is required for this device classification submission; bench testing only.
Technological Characteristics
Ceiling-mounted surgical luminaire; Energix™ Power Supply; adjustable intensity (30, 70, 110, 150 Lux @ 1m); RS232 connectivity; yoke-mounted keypad controls; video camera/flat-screen integration. Conforms to UL 60601-1, UL 60601-2-41, CSA C22.2 No. 601.1, IEC 60601-2-41, EN 60601-1-2, and FCC Part 15.
Indications for Use
Indicated for providing visible illumination of the surgical area or patient during surgical procedures, examinations, and minimally invasive surgeries.
Regulatory Classification
Identification
A surgical lamp (including a fixture) is a device intended to be used to provide visible illumination of the surgical field or the patient.
Special Controls
*Classification.* Class II (special controls). The device, when it is an operating room lamp, a surgical instrument light, a surgical floor standing light, an endoscopic surgical light, a surgical light connector, a ceiling mounted surgical light, a surgical light carrier, surgical light accessories, a surgical lamp, a remote illuminator, or an incandescent surgical lamp, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 878.9.
Predicate Devices
ALM Angenieux® (AX) Series Surgical Light (K904965)
Reference Devices
ALM PrismAlix® (PRX) Series Surgical Lights
Submission Summary (Full Text)
{0}------------------------------------------------
MAY 21 2004
K040735
## 510(k) SUMMARY
### ALM X'Ten™ Surgical Light System
| Submitted by: | Getinge USA, Inc. (as ALM S.A.'s US Agent)<br>1777 E Henrietta Road<br>Rochester, NY 14623-3133 |
|------------------------|--------------------------------------------------------------------------------------------------|
| Contact Person: | Frederick R. Catt<br>Senior, Regulatory Engineer<br>Phone: (585) 272-5013<br>Fax: (585) 272-5299 |
| Date prepared: | March 19, 2004 |
| Proprietary Name: | ALM X'TenTM Surgical Light System |
| Common Name: | Surgical Light |
| Device Classification: | Surgical Lamp (78 FSY)<br>Class II, as listed per 21 CFR 878.4580 |
| Predicate Device: | ALM Angenieux® (AX) Series Surgical Light [K904965] |
#### Description of Device:
The ALM X'Ten'M Surgical Light System is a new product designation intended to identify a family of surgical lights that will use a similar set of design principals as the Angenieux (AX) Series Surgical Light Systems. The primary predicate device focuses on comparisons to the Angenieux® AX10, which is the medium sized lighthead within the Angenieux® (AX) Series.
The ALM X'Ten™ Surgical Light provides a broadened set of features and options that include video camera and flat screen display capabilities, bulb failure indicator(s), and an ambient light mode ("LEDinside") as results from this redesign effort. The user can toggie from the Surgical Light mode to the Ambient Light mode via use of a keypad switch
{1}------------------------------------------------
mounted on the yoke of the lighthead. The "LEDinside" mode presents a ring of light mouthed from the bottom surface of the light, centered on the outer periphery of the sterile handle. Intensity can be adjusted to four levels: 30, 70, 110 and 150 Lux (@ 1m) and is sufficient to work on the patient in a dim or darkened operating room.
The ALM X'Ten™ Product Family currently has four configurations available, as shown within Table 1.
#### Table 1
| ALM X'Ten TM Surgical Light and Systems –<br>Designations/Configurations | |
|---------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Model | Description |
| X10 DF | X'TenTM SURGICAL LIGHT, CEILING MOUNTED, ONE X10 LIGHTHEAD |
| X10 duo DF | X'TenTM SURGICAL LIGHT, CEILING MOUNTED, TWO X10 LIGHTHEADS |
| X10 duo DF V | X'TenTM SURGICAL LIGHT, CEILING MOUNTED, TWO X10 LIGHTHEADS<br>WITH ONE OF THOSE INCLUDING VIDEO PREWIRING AND AMBIENT<br>LIGHT ("LEDinside" FUNCTION) |
| X10 duo DF V FS | X'TenTM SURGICAL LIGHT, CEILING MOUNTED, TWO X10 LIGHTHEADS<br>WITH ONE OF THOSE INCLUDING VIDEO PREWIRING AND AMBIENT<br>LIGHT ("LEDinside" FUNCTION), AND ONE FLAT SCREEN SUPPORT |
Note: Additional model configurations are planned for development. These will be incorporating use of the ALM X'Ten™ lighthead subassemblies.
#### Intended Use:
The ALM X'Ten™ Surgical Lights are intended to be used to provide visible illumination of the surgical area or the patient. The "LEDinside" ambient light mode is intended for minimally invasive surgeries, procedures and examinations.
#### Nonclinical Comparisons to Predicate Device
The ALM X'Ten'™ Surgical Light (subject device) is similar to the predicate device with the following modifications:
- The ALM X'Ten™ Surgical Light uses the Energix™ Power Supply (WPS) to . provide the electrical power to the light source. Surgical Light intensity levels are adjustable from the Energix keypad or remotely with RS232 communication port option. Bulb failure indicators provided. Optional LCD provides operation and maintenance information about the surgical light system, lightheads and bulbs.
- The ALM X'Ten™ introduces the "LED inside" option feature. This is an ambient . light used to illuminate a larger patient area and is intended for minimally invasive surgery applications. Controls are located on the yoke keypad for switchover from Surgical Light to Ambient Light. "LED inside" light level adjustments are at 30, 70, 110 and 150 Lux.
{2}------------------------------------------------
- Added Capabilities of Video Camera (fixed focus or zoom), Multi-Media and Flat . Screen Displays (Note: ALM X'Ten™ with "LED inside" feature is included) similar to the ALM PrismAlix® (PRX) Series Surgical Lights.
- Added an optional adapter for single use sterile sleeves/gloves on a standard . handle system.
- Modified lighthead design, updating its appearance and suspension means. .
#### Test Data:
The test data supports conformance to:
- UL 60601-1 Standard for Medical Electrical Equipment, Part 1: General Requirements . for Safety
- UL60601-2-41 Medical electrical equipment Part 2-41: Particular requirements for . the safety of surgical luminaires and luminaires for diagnostics
- CSA C22.2 No. 601.1 Medical Electrical Equipment, Part 1: General Requirements for . Safety
- CSA C.22.2 No. 60601-2-41 Medical electrical equipment Part 2-41: Particular . requirements for the safety of surgical luminaires and luminaires for diagnostics
- IEC 60601-2-41 Medical electrical equipment Part 2-41: Particular requirements for . the safety of surgical luminaires and luminaires for diagnostics
- EN 60601-1-2 Medical electrical equipment Part 1-2: General requirements for safety . - Collateral standard: Electromagnetic compatibility - Requirements and tests
- . FCC Part 15
- Software used in the ALM X'Ten™ Surgical Light was tested according to the . appropriate FDA Software Guidance Documents, per its determination as a Minor Level of Concern.
#### Clinical Data:
No clinical data is required for this device classification submission.
#### Conclusion:
The modifications incorporated into the ALM X'Ten™ Surgical Light System designs use those desired design features from both the ALM PrismAlix® (PRX) Series and Angenieux® (AX10) Surgical Light Systems. Based upon the information provided herein this 510(k) Premarket Notification, we conclude that ALM X'Ten' Surgical Light Systems are substantially equivalent to the predicate device(s) and is safe and effective when used as intended.
{3}------------------------------------------------
Image /page/3/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized graphic of an eagle or bird-like figure with three curved lines representing its wings or body. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular fashion around the graphic.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
MAY 2 1 2004
ALM S.A. c/o Mr. Frederick R. Catt Senior Regulatory Engineer Getinge USA, Inc 1777 East Henrietta Road Rochester, New York 14623
Re: K040735
K040733
Trade/Device Name: ALM X Ten™ Surgical Light System with Energix™ Power Supply Regulation Number: 21 CFR 878.4580 Regulation Name: Surgical lamp Regulatory Class: II Product Code: TSY Dated: April 30, 2004 Received: May 3, 2004
Dear Mr. Catt:
We have reviewed your Section 510(k) premarket notification of intent to market the device we nave leviewed your becamed the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate for use stated in the encrease) to regarry the Medical Device Amendments, or to commerce prior to May 20, 1978, the ensordance with the provisions of the Federal Food, Drug, devices that have been require approval of a premarket approval application (PMA). alle Cosment Act (71ct) that do not requent to the general controls provisions of the Act. The You may, therefore, manxet the act include requirements for annual registration, listing of general controls provisions of use tice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it If your device is classified (600 a00 vyrols. Existing major regulations affecting your device can may be subject to such additional controlio visitions, Title 21, Parts 800 to 898. In addition, FDA may be found in the overning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean r least of advised that I DTT Pressan that your device complies with other requirements of the Act that IDA has Inade a dotermissions administered by other Federal agencies. You must or any recetal statutes and regulanents annouding, but not limited to: registration and listing (21 Comply with an the Hec oreense than and the start \$01); good manufacturing practice requirements as set CFN in the quality systems (QS) regulation (21 CFR Part 820), and if applicable, the electronic forth in the quality systems (Sections 531-542 of the Act); 21 CFR 1000-1050.
{4}------------------------------------------------
Page 2 - Mr. Frederick R. Catt
This letter will allow you to begin marketing your device as described in your Section 510(k) This letter will anow you to begin marketing of substantial equivalence of your device to a legally premaired notification. The PDF mailing of cation for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please If you desire specific advice for your as (301) 594-4659. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small other gelleral information on Jour response Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html
Sincerely yours,
Miriam C. Provost
* Celia M. Witten, Ph.D., M.D. Director Division of General, Restorative and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
### Indications for Use
510(k) Number (if known): K 040735
# Device Name: ALM X'Ten™ Surgical Light System with Energix™ Power Supply
Indications for Use:
ALM X'Ten™ (X10) Surgical Light Systems with Energix™ Power Supply are intended ALM X Tell "" (X10) Surgical Elght Systems hands in the surgical area or the patient.
An optional "LEDinside" feature is available and produces lower intensity levels of An optional LEDIffside Teature is a railable and procedures and examinations.
The Energix™ Power Supply is intended for use with surgical light systems.
V Prescription Use (Part 21 CFR 801 Subpart D)
AND/OR
Over-The-Counter Use _ (21 CFR 807 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Miriam C. Provost
(Division Sign-Off) Division of General, Restorative, and Neurological Devices
Page 1 of 1
**510(k) Number** K040735
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.