SURGASSIST COMPUTER MEDIATED LINEAR CUTTER DIGITAL LOADING UNITS, 30 MM, 55 MM, 75 MM
K040720 · Power Medical Interventions, Inc. · GDW · Jun 3, 2004 · General, Plastic Surgery
Device Facts
Record ID
K040720
Device Name
SURGASSIST COMPUTER MEDIATED LINEAR CUTTER DIGITAL LOADING UNITS, 30 MM, 55 MM, 75 MM
Applicant
Power Medical Interventions, Inc.
Product Code
GDW · General, Plastic Surgery
Decision Date
Jun 3, 2004
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 878.4750
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The SurgASSIST® Computer Mediated Linear Cutter Digital Loading Units® have applications for general endoscopic surgery in gastrointestinal, and gynecological, general abdominal and thoracic surgical procedures for resection, transection, creation of anastomoses, and for open occlusion of the heart's left atrial appendage.
Device Story
SurgASSIST® Computer Mediated Linear Cutter (CMLC) Digital Loading Units® (DLUs) are reusable surgical stapling components used with previously cleared reloads. Device functions in general, endoscopic, gastrointestinal, gynecological, abdominal, and thoracic surgeries for tissue resection, transection, anastomosis, and left atrial appendage occlusion. DLUs are designed for multi-patient use, requiring cleaning and autoclave sterilization. Operation involves mechanical stapling and cutting; anvil clamp force increased to 100 lbs to support durability for multi-patient cycles. Surgeons operate the device during procedures to provide mechanical tissue management. Benefits include reusability and enhanced mechanical performance compared to single-use predecessors.
Clinical Evidence
No clinical data provided. Substantial equivalence is based on technological characteristics and design modifications for reusability.
Technological Characteristics
Reusable linear stapler loading units. Materials modified to withstand multiple cleaning and autoclave sterilization cycles. Mechanical actuation with increased anvil clamp force (100 lbs). Available in 30 mm, 55 mm, and 75 mm sizes. Standalone mechanical device.
Indications for Use
Indicated for patients undergoing general, endoscopic, gastrointestinal, gynecological, abdominal, or thoracic surgical procedures requiring resection, transection, or creation of anastomoses, and for open occlusion of the heart's left atrial appendage.
Regulatory Classification
Identification
An implantable staple is a staple-like device intended to connect internal tissues to aid healing. It is not absorbable.
Predicate Devices
SurgASSIST® Straight Linear Cutter Digital Loading Units® with Reloads (K020719)
SurgASSIST® Straight Linear Stapler Digital Loading Units® with Reloads and Straight Linear 4 Row No Knife Digital Loading Units® with Reloads (K040398)
{0}------------------------------------------------
K040720 Page 1/2
Power Medical Interventions, Inc. SurgASSIST® Computer Mediated Linear Cutter DLUs Traditional 510(k) Premarket Notification · March 17, 2004
JUN 0 3 2004
#### Section E
#### Traditional 510(k) - Summary
In Accordance with 21 CFR Section 807.92 Power Medical Interventions, Inc., is submitting the following 510(k) Summary:
1) Submitter Information:
Power Medical Interventions, Inc. 110 Union Square Drive New Hope, PA 18938 267-775-8151 Ph 267-775-8123 Fax
Applicant:
Barbara J. Whitman
March 17, 2004 Date of Notification:
2) Name of Device:
| Trade Name: | SurgASSIST®<br>Computer Mediated Linear Cutter (CMLC)<br>Digital Loading Units® (DLUs)<br>30 mm, 55 mm, 75 mm |
|-------------------------|---------------------------------------------------------------------------------------------------------------|
| Common Name: | Linear Staplers with Implantable Staples |
| Classification<br>Name: | Staple, Implantable, GDW |
- 3) Predicate Devices:
- a. SurgASSIST® Straight Linear Cutter Digital Loading Units® with Reloads, Power Medical Interventions, Inc., New Hope, PA. REF SLC55B, SLCR55B, SLC55G, SLCR55G, SLC30B, SLCR30B, SLC30G, SLCR30G (K020719).
- b. SurgASSIST® Straight Linear Stapler Digital Loading Units® with Reloads and Straight Linear 4 Row No Knife Digital Loading Units® with Reloads, Power Medical Interventions, Inc., New Hope, PA. |REF SLS55B, SLSR55B, SLS55G, SLSR55G, SLSR55G, SLS55B4, SLSR55B4 (K040398).
{1}------------------------------------------------
K040720 page 2/2
- c. USSC Auto Suture TA & GIA Staplers, United States Surgical, Norwalk, CT. REF TA, GIA (K032696).
- 4) Device Description:
devices described here are reusable CMLC DLUs used in The conjunction with a variety of Reloads, which were previously cleared to market under K020719 and K030653. The DLUs are designed to be cleaned and sterilized for multi-patient use.
- 5) Device Modification
The CMLC DLUs have identical technological features as the predicate devices (K020719). The CMLC DLUs underwent material changes in order for the DLUs to withstand multiple cleaning and autoclave cycles for the purpose of multi-patient use. The average anvil clamp force was increased from 50 lbs. on the predicate device to 100 lbs. on the Computer Mediated Linear Cutter DLUs.
- 6) Indications For Use
The SuraASSIST® Computer Mediated Linear Cutter Digital Loading Units® have applications for general and endoscopic surgery in gastrointestinal, gynecological, general abdominal and thoracic suraical procedures for resection, transection, creation of anastomoses, and for open occlusion of the heart's left atrial appendage.
- 7) Comparison to Predicate Devices
The CMLC DLUs have the same function as the previously cleared predicate Straight Linear Cutter Digital Loading Units® with Reloads (K020719). Material changes in the DLUs allow for cleaning, sterilizing, and multi-patient use. The average anvil clamp force was increased in the CMLC DLUs. For further details, please refer to Section J for predicate comparison chart.
The CMLC DLUs Indications For Use statement is identical to that which appears in the following two predicate devices, K040398 and K032696.
{2}------------------------------------------------
DEPARTMENT OF HEALTH & HUMAN SERVICES
Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo features a stylized caduceus symbol, which is a staff with two snakes coiled around it. The words "HEALTH & HUMAN SERVICES - USA" are arranged around the top of the symbol. The logo is black and white and has a simple, clean design.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
JUN 0 3 2004
Ms. Barbara J. Whitman Regulatory Affairs Manager Power Medical Interventions, Inc. 110 Union Square Drive New Hope, Pennsylvania 18938
Re: K040720
Trade/Device Name: SurgASSIST® Computer Mediated Linear Cutter Digital Loading Units® Regulation Number: 21 CFR 878.4750 Regulation Name: Implantable staple Regulatory Class: II Product Code: GDW Dated: March 18, 2004 Received: March 19, 2004
Dear Ms. Whitman:
We have reviewed your Section 510(k) premarket notification of intent to market the device we nave reviewed your becaon 310(t) premained is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate for use slated in the elerosure) to 1egars atment date of the Medical Device Amendments, or to commerce prior to May 20, 1976, the encordance with the provisions of the Federal Food, Drug, devices that have been recuired in accessaried of a premarket approval application (PMA). and Cosmette Act (Act) market the device, subject to the general controls provisions of the Act. The 1 ou may, ulcrefore, market the device, books requirements for annual registration, listing of general controls provisions of all rist leveling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it If your device is classified (see above) and class of regulations affecting your device can may be subject to back acated as a legulations, Title 21, Parts 800 to 898. In addition, FDA may be found in the Obas acements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean I lease be advised that i Dri brisoaanso o vour device complies with other requirements of the Act that I DA has made a aother regulations administered by other Federal agencies. You must of any I coclar statutes and regulations and limited to: registration and listing (21 Comply with an the rice brequirements)01); good manufacturing practice requirements as set CITY art 607), adomig (21 CFR Part 820); and if applicable, the electronic 101th in the quant) Bybellio (Sections 531-542 of the Act); 21 CFR 1000-1050.
{3}------------------------------------------------
## Page 2 - Ms. Barbara J. Whitman
This letter will allow you to begin marketing your device as described in your Section 510(k) This letter will anow you to begin marketing your antial equivalence of your device to a legaliy premarket notification. The FDA Inding of substantial equilies and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please If you desire specific advice for your de 110 your ac 9. Also, please note the regulation entitled, contact the Office of Complance at (301) 597 1707 (21) Part 807.97). You may obtain " Misoranding by reference to premarked nonlineaus in the Act from the Division of Small other general Informational and Consumer Assistance at its toll-free number (800) 638-2041 or
Manufacturers, International and Consumer Assistance on and also and also and Manufacturers, International and Consulter Parison Parison of Carb/dsmaldsmamain.html
Sincerely yours,
Miriam C. Provost
Celia M. Witten, Ph.D., M.D. Director Division of General, Restorative and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
### Section D
K040720
# Indications for Use
Power Medical Interventions, Inc. New Hope, PA 18938
510(k) Number (if known):
Device Name: SurgASSIST® Computer Mediated Linear Cutter Digital Loading Units®
Indications For Use:
The SurgASSIST® Computer Mediated Linear Cutter Digital Loading Units® have applications for general endoscopic surgery in gastrointestinal, and gynecological, general abdominal and thoracic surgical procedures for resection, transection, creation of anastomoses, and for open occlusion of the heart's left atrial appendage.
Prescription Use X (Part 21 CFR 801 Subpart D) AND/OR
Over-The-Counter Use (21 CFR 807 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Miriam C. Provost
(Division Sign-Off) Division of General, Restorative, and Neurological Devices
Page 1 of
000012
510(k) Number K0140720
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.