← Product Code [FRO](/productcode/FRO) · K040683

# BIODERM SCIENCES WOUND DRESSING (K040683)

_Bioderm Sciences, Inc. · FRO · Jun 4, 2004 · SU · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K040683

## Device Facts

- **Applicant:** Bioderm Sciences, Inc.
- **Product Code:** [FRO](/productcode/FRO.md)
- **Decision Date:** Jun 4, 2004
- **Decision:** SESE
- **Submission Type:** Traditional
- **Device Class:** Class U
- **Review Panel:** SU
- **Attributes:** Therapeutic

## Indications for Use

BioDerm Sciences Wound Solution is intended to cleanse, irrigate and externally manage dermal lesions such as lacerations, post-operative (surgical) wounds, grafts, partial and full-thickness wounds, burns and ulcers (diabetic, venous stasis, pressure). It is meant to be used in conjunction with a sterile dressing that absorbs fluids (i.e. gauze, gel, alginate, foam, hydrocolloid). BioDerm Wound Solution can also be used as a wound cleanser to remove foreign matter, bacteria and tissue debris.

## Device Story

BioDerm Wound Solution is an acidic zinc-saline wet dressing for external wound management. It provides an acidic, moist environment to aid healing and can be used on skin, mucous membranes, mouth, and eyes. Supplied sterile in 50 ml injection bottles with Teflon septums; contents are withdrawn via syringe. Used by clinicians or patients to cleanse and irrigate wounds; intended to be used in conjunction with absorbent dressings like gauze, gel, alginate, foam, or hydrocolloid. The solution removes foreign matter, bacteria, and tissue debris. It is intended for sale by or on the order of a physician.

## Clinical Evidence

Bench testing and animal study provided. A study performed at the University of Miami (Department of Dermatology & Cutaneous Surgery) using a porcine model evaluated various formulations of the solution with varying metallic salt strengths as an aid to healing of incisions, grafts, burns, and partial/full-thickness wounds.

## Technological Characteristics

Acidic zinc-saline solution. Supplied sterile in 50 ml injection bottles with Teflon septums. pH ≤ 7. Intended for use with absorbent secondary dressings (gauze, gel, alginate, foam, hydrocolloid).

## Predicate Devices

- Saljet Single Dose Sterile Saline Topical Solution ([K993969](/device/K993969.md))
- 20 ml Normal Saline Topical Solution, 0.9% w/v Sodium Chloride ([K972185](/device/K972185.md))
- Hypertonic saline wet dressing - sterile ([K941999](/device/K941999.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

K040 683

Dermagran wound cleanser with zinc

#### 510(k) Summary

| Submitter:                                                     | BioDerm Sciences, Inc.<br>101 Orchard Ridge Drive, Suite 1N<br>Gaithersburg, Maryland 20878<br>Tel. (301) 216-3912<br>Fax (301) 216-3919 |
|----------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------|
| Contact Person:                                                | Ed Gubish Ph. D. <egubish@healthpathways.com><br>President, BioSciences Corp.                                                            |
| Preparation Date:                                              | May 27, 2004                                                                                                                             |
| Proprietary Device Name:                                       | BioDerm Wound Solution                                                                                                                   |
| Classification Name:                                           | Liquid Bandage                                                                                                                           |
| Predicate Devices:                                             |                                                                                                                                          |
| Saljet Single Dose Sterile Saline Topical Solution             | K993969                                                                                                                                  |
| 20 ml Normal Saline Topical Solution, 0.9% w/v Sodium Chloride | K972185                                                                                                                                  |
| Hypertonic saline wet dressing - sterile                       | K941999                                                                                                                                  |

Description:

Bio Derm Wound Solution is an acidic zinc-saline wet dressing for external wound management. Biollerm Wound Solution is an actore zinc-saline were the body in the healing process. The nonIt provides an acidic, moist environment that aids the book in the mouth aves If provides an acidic, moist environment that allow it to be used in the mouth, eyes and on mucous membranes as well as on the skin.

BioDerm Wound Solution is supplied sterile in 50 ml injection bottles with Tellon septents from the BioDeffi Wound Solution is supplied steries is applied by withdrawing the contents from the injection bottle with a syringe. The crimp caps have a pull-tal which must be removed or got The injection bothe with a sytting. The onlip supp whether the bottle has been opened or not. The contents must be used within five days after opening.

Intended Use:

BioDerm Sciences Wound Solution is intended to cleanse, irrigate and externally manage BioDerm Sciences Wound Bolution is interative (surgical) wounds, grafts, practial and full-thickness wounds, burns and ulcers (diabetic, venous stasis, pressure). It is meant to full-thickness wounds, ourils and dicers (Graoone, Toloos fluids (i.e. gauze, gel, alginate, foam, hydrocolloid).

BioDerm Wound Solution can also be used as a wound cleanser to remove foreign matter, bacteria and tissue debris.

K945802

{1}------------------------------------------------

KC40683 page 2 of 2

BioDerm Wound Solution is contraindicated for use when patients are known to have had allergic reactions to this dressing or its components. It is not suitable for use on third degree allergic reactions to this thessing of its components. We are a love a missing.

# Comparison of Technological Characteristics:

A number of wet dressings have been cleared for marketing by the FDA. All have in common A number of wel dressings nave been on sodium chloride or metallics salts with a pH less than or that they are same southous based on socialin additives such as nutrients. Dressings
equal to 7 (i.e. acidic). They may or may not contain as that as also intended for use equal to 7 (1.6. actire). They may of thay not coilant also intended for use as wound dressings are sterile. BioDerm Wound Solution is supplied sterile.

A broad range of chemical compositions is currently available. The chemical compositions of these predicate A broad range of chemical compositions is can chary of chemical compositions of these predicate BioDerm Wound Solution is willin the raige of enchiential ve " (See Table 1)

### Performance:

A study of various formulations of with varying strength levels of the metalic salts, performed at A study of various tonimations of milli rarying ology & Cutaneous Surgery) using a porcine
the University of Miami (Department of Dermatology & Cutaneous in the healing of the University of Miami (Department of BioDerm Wound Solution as an aid to healing of incision, grafts, burns, partial and full-thickness wounds. (Appendices B-E)

#### Conclusions:

When used as directed, BioDerm Wound Solution is safe and effective as a wound cleanser or wet dressing.

## Other Information:

This product is intended for sale by or on the order of a physician (or properly licensed practitioner).

{2}------------------------------------------------

Image /page/2/Picture/1 description: The image is a black and white logo for the U.S. Department of Health & Human Services. The logo features a stylized image of three human profiles facing to the right. The profiles are stacked on top of each other, with the top profile being the largest and the bottom profile being the smallest. The profiles are surrounded by a circle of text that reads "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA".

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

MAY - 1 2007

Edward R. Gubish, Ph.D. President BioDerm Sciences, Inc. 101 Orchard Ridge Drive, Suite 1N Gaithersburg, MD 20878

Re: K040683

> Trade Name: BioDerm Sciences Wound Solution Regulatory Class: Unclassified Product Code: FRO Dated: March 9, 2004 Received: March 16, 2004

Dear Mr. Gubish:

This letter corrects our substantially equivalent letter of June 4, 2004.

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{3}------------------------------------------------

Page 2 - Mr. Edward R. Gubish, Ph.D.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (sections 531-542 of the Act); 21 CFR 1000-1050.

This letter will allow you to continue marketing your device as described in your Section Fins leaser warket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (240) 276-0115. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html

Sincerely yours,

Mark N. Melkerson

Director Division of General, Restorative and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health

{4}------------------------------------------------

K040683

#### Indications for Use

510(k) Number: K040683

Device Name: BioDerm Sciences Wound Solution

Indications for Use:

BioDerm Sciences Wound Solution is intended to cleanse, irrigate and externally manage dermal lesions such as lacerations, post-operative (surgical) wounds, grafts, partial and fullthickness wounds, burns and ulcers (diabetic, venous stasis, pressure). It is meant to be used in conjunction with a sterile dressing that absorbs fluids (i.e. gauze, gel, alginate, foam, hydrocolloid).

Prescription Use (Part 21 CFR 801 Subpart D) AND/OR

Over-the-Counter Use (21 CFR 807 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence/of CDRH, Office of Device Evaluation (ODE)

(Division Sign-Off)

Division of General, Restorative,
and Neurological Devices

510(k) Number K040683

Page I of _ l

Confidential BioDerm Sciences, Inc. 4

---

**Source:** [https://fda-staging.innolitics.com/device/K040683](https://fda-staging.innolitics.com/device/K040683)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
