BOUSSIGNAC/VYGON ENDOTRACHEAL TUBE, MODEL 6501.25/30/35 AND 6502.70/75/80
K040657 · Vygon SA · BTR · Jun 10, 2004 · Anesthesiology
Device Facts
Record ID
K040657
Device Name
BOUSSIGNAC/VYGON ENDOTRACHEAL TUBE, MODEL 6501.25/30/35 AND 6502.70/75/80
Applicant
Vygon SA
Product Code
BTR · Anesthesiology
Decision Date
Jun 10, 2004
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 868.5730
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The Boussignac/ Vygon Endotracheal Tube is indicated for airway management. The insufflation canals permit the delivery of intermittent jet ventilation and the administration of oxygen during tracheobronchial suctioning procedures. They also provide a supplemental means for administering oxygen during bronchoscopies. In these ventilatory modes, the main lumen serves as a channel for the elimination of expired gases. When the insufflation port is capped, the Boussignac/ Vygon Endotracheal Tube can function as a standard tracheal tube with ventilation occurring through the main lumen. The Boussignac/ Vygon Endotracheal Tube is intended for oral/nasal intubation. The monitoring/irrigation lumens may be used for monitoring airway pressure, for irrigation of a patient's tracheobronchial tree, to aid in the removal of accumulated secretions, for sampling tracheal gases, for anesthetizing the trachea, or for introducing suitable medication in accordance with standard medical practices.
Device Story
Boussignac/Vygon Endotracheal Tube is a multi-lumen airway device for oral/nasal intubation. Design includes Magill curve, tip, depth markings, radiopaque line, and low-pressure cuff (adult models). Device features main lumen plus eight embedded capillaries: insufflation canals deliver intermittent jet ventilation and oxygen; additional canals enable pressure monitoring, CO2 sampling, or irrigation. Used in clinical settings (e.g., OR, ICU) by physicians for airway management. When insufflation ports are capped, device functions as standard endotracheal tube. Output includes gas delivery, pressure data, and access for suctioning/medication. Benefits include simultaneous ventilation and monitoring/irrigation capabilities, facilitating airway maintenance and secretion management.
Clinical Evidence
Clinical publications were presented; however, no specific study results, sample sizes, or performance metrics were detailed in the summary document.
Technological Characteristics
Multi-lumen endotracheal tube with Magill curve. Features main lumen and 8 embedded capillaries for gas insufflation, pressure control, CO2 monitoring, or irrigation. Includes low-pressure cuff (adult), self-sealing valve, pilot balloon, and radiopaque line. Conforms to ISO standards for tracheal tubes and connectors. Non-electronic, mechanical device.
Indications for Use
Indicated for airway management in patients requiring oral/nasal intubation. Supports intermittent jet ventilation, oxygen administration during suctioning or bronchoscopy, airway pressure monitoring, tracheobronchial irrigation, secretion removal, gas sampling, and medication delivery.
Regulatory Classification
Identification
A tracheal tube is a device inserted into a patient's trachea via the nose or mouth and used to maintain an open airway.
{0}------------------------------------------------
K040657
# 510(k) Summary
Vygon. S.A
In accordance with the Safe Medical Devices Act (SMDA) of 1990 and Title of the Code of Federal in accordance with the Sale Medical Devices Fred (0.1.2.1.2017) The following summary of safety and effectiveness information is provided:
## Submitted by A.
Vygon, S.A. 5-11. Rue Adeline 95440 Ecouen France Telephone: (33) 1 39 92 63 63 Contact: Michel Hanania. Regulatory Affairs/Quality Assurance Manager Date Prepared: March 11, 2004.
#### B. Device Name
Trade or Proprietary Name: Boussignac/ Vygon Endotracheal Tube
Endotracheal Tube Common or Usual Name:
Tracheal Tube Classification Name:
#### C. Predicate Devices
The subject device is substantially equivalent to the Mallickrodt Hi-Lo® Jet Tracheal Tube (K802505). and the Vygon's Double Lumen Tracheal Tube (K960795).
{1}------------------------------------------------
## Device Description D.
I he Boussignac/ V ygon Endorracheal Tube is a standard endotracheal tube available in I he boussignat? Fygon Enaon activent and adult use. The tube design includes a Magill curve, a tip, printed depth markings (in one centimeter increments), a low-pressure cuff (adult models only) and self-sealing valve with attached pilot balloon for cuff inflation detlation. The tube also features a radiopaque line to assist in radiographic innation. The device, including featured characteristics listed above, conforms to ISO cisualization: The Gevice, motary Equipment – Tracheal Tubes and Connectors.
The tube has a main lumen and eight (8) separate canals, or "capillaries", embedded in the The ture has a main fument and eight (5) of these canals are used for gas insufflation, two (2) for pressure control. CO2 monitoring, or irrigation, and as in any endotracheal tube, one (1) for pressure control. CO2 monitoring, on the uncuffed pediatric model, six (6) of the imbedded canal is reserved for gas insufflation, while two (2) are used for pressure control, CO2 monitoring, or irrigation.
#### Intended l'se E.
The Boussignac/ Vygon Endotracheal Tube is indicated for airway management. The insullation canals permit the delivery of intermittent jet ventilation and the administration of oxygen during tracheobronchial suctioning procedures. They also provide a or oxygen daministering oxygen during bronchoscopies. In these ventilatory suppliciteman means for arres as a channel for the elimination of expired gases. When the modes, the main famen set res abussignac/ Vygon Endotracheal Tube can function as a misumation port is oupport to allation occurring through the main lumen. The Boussignac/ Vygon Endotracheal Tube is intended for oral/nasal intubation.
The monitoring/irrigation lumens may be used for monitoring airway pressure, for irrigation of a patient's tracheobronchial tree, to aid in the removal of accumulated secretions, for sampling tracheal gases, for anesthetizing the trachea, or for introducing suitable medication in accordance with standard medical practices.
## Comparison to Predicate Devices F.
As was established in this submission, the subject device is substantially equivalent to its rs was obtablished the Mallinckrodt Hi-Lo® Jet Tracheal Tube (K802505), and the Vygon's Double Lumen Tracheal Tube (K960795), cleared by the agency for commercial distribution in the United States.
The subject device has substantially equivalent indications for use as the Mallinckrodt Hi-I no Jet Tracheal Tube, being indicated for airway management via oral/nasal intubation.
The subject device is composed of the same materials as one or more of the predicate devices, all of which are established as safe for their application in the subject device.
{2}------------------------------------------------
The subject device encompasses design features which are substantially equivalent to those of The subject device encompasses design Trached Tube, including a standard 1 5mm offered by the MaillineKfodd 71-L0 - Jer 77dt inflation, low-pressure cuff adult models and connector, a pilot balloon and valve for can innelled tip, depth markings, a main luments) uncutted pediatic models, a facilopaque mic and overs and insufflation lumen(s).
for the evacuation of expired gases and mucosal secretions, and insufflation lumen(s).
The subject device provides functions equivalent to those provided by the Mallinskrodt Hi-I he subject device proviture the delivery of intermittent jet ventilation and the administration of oxygen during tracheobronchial suctioning procedures, providing a administration of oxygen during tracheroervann and ing bronchoscopies, and when the insufflation port is capped, both devices may also function as standard trached tubes with insuitiation port is capped, ood devices may and oring irrigation lumens may be ventilation occurring unough the main rains of a patient's trackeobronchial tree, to
used for monitoring airway pressure, for irrigation of a patient's tracked as an used for monitoning an way pressure, for intrigation and and eases, for ansthetizing
aid in the removal of accumulated secretions, for sampling traches modical aid in the removal of accumulated secretions, for sumpling with standard medical practices.
Further, the subject device is packaged and labeled in a manner substantially equivalent to its predicate device. the Mallinckrodt Hi-Lo® Jet Tracheal Tube.
# Summary of Non-Clinical Tests G.
Performance testing was presented.
# Summary of Clinical Tests H.
Clinical publications were presented.
# Conclusions of Non-Clinical and Clinical Tests I.
Performance testing and clinical publications were presented.
{3}------------------------------------------------
Image /page/3/Picture/1 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo consists of a circular seal with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" arranged around the perimeter. Inside the circle is an abstract symbol resembling an eagle or bird in flight, composed of three curved lines. The logo is black and white.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
JUN 1 0 2004
Ms. Laetitia Bernard Excaelia, Inc. 45900 Parsippany Court Temecula CA 92592
Re: K040657
Trade/Device Name: Boussignac Vygon Endotracheal Tube Regulation Number: 868.5730 Regulation Name: Tracheal Tube Regulatory Class: II Product Code: BTR Dated on Submission: March 11, 2004 Date Received in ODE: March 12, 2004
Dear Ms. Bernard:
We have reviewed your Section 510(k) premarket notification of intent to market the device we nave reviewed your bection be clevice is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate for use stated in the enorosale) to regary to regary the Medical Device Ameridments, or to commence prior to May 20, 1976, the exact ance with the provisions of the Federal Food, Drug, devices mat have occh rechassinou in accessfire approval of a premarket approval application (PMA). and Cosmetic Act (Fee) market the device, subject to the general controls provisions of the Act. The r ou may, therefore, market the act include requirements for annual registration, listing of general controls provisions of assisted, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it If your device is classifica (see above) me existing major regulations affecting your device can
may be subject to such additional controls. Existing major regulations FDA w may be subject to such additional controller Entisting and to 898. In addition, FDA may be found in the Code of I castal stegeming your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean Please be advised that PDA s issuated of a babbains and other requirements of the Act that FDA has made a delerimiation administered by other Federal agencies. You must of ally rederal statutes and regulations and uding, but not limited to: registration and listing (21 Comply with an the Act 3 requirements, necesses,
CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set
{4}------------------------------------------------
Page 2 – Ms. Laetitia Bernard
forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic
s (21) CFR (2000) (2005) (2005) (21) (2005-100) (21 CFR 1000-1050, This forth in the quality systems (QS) regulation (2F CJ R P & 050, 21 CFR 1000-1050. This
product radiation control provisions (Sections 531-542 of the Act), 21 CFR 1000-1050. product radiation control provisions (SCCRONS 551 e as described in your Section 510(k)
letter will allow you to begin marketing your device of your device to a letter will allow you to begin man kening your activelence of your device to a legally
premarket notification. The FDA finding of substantial equive add thus, nermits vour d premarket notification. The FDA inding of substantal equivalies of your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please
ron of the many of the many 1941 164 - Also - mass note the regulation enti If you desire specific advice for your device on our necesses note the regulation entitled,
contact the Office of Compliance at (301) 594-4646. Also, please note the regulat contact the Office of Complailed at (501) 574-10 10. Tree Part 807.97). You may obtain
"Misbranding by reference to premarket notification" (21CFR Part 807.97). "Misbranding by reference to premarket notification (C) < et Act from the Division of Small other general information on your responsibilities and its toll-free number (800) 638-2041 or
Manufacturers, International and Consumer Assistance at its toll-areanain html Manufacturers, International and Consumer Piositiatio and Consumers of the main.html.
Sincerely yours,
Ques
Chiu Lin, Ph.D. Director Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
#### Draft Labeling V.
# Indications for Use A.
510(k) Number (if known): K040657
Boussignac/ Vygon Endotracheal Tube Device Name:
Indications for Use:
The Boussignac/ Vygon Endotracheal Tube is indicated for airway management. The I he Boussignal? I ygon Endorfacile I ace 15 minutent jet ventilation and the administration of
insufflation canals permit the delivery of intermittent jet worlds a surpleme insufflation canals permit the denvery of internations. They also provide a supplemental oxygen during tracheooronicular succioining procedures. In these ventilatory modes, the main
means for administering oxygen during bronchoscopies. In these ventify institutio means for administering oxygen during of expired gases. When the insufflation port is lumen serves as a chamel for the emilitation of express gasson as a standard tracheal tube with capped. The Boulssignal - Trgon Endortached. The Boussignac/ Vygon Endotracheal Tube is intended for oral/nasal intubation.
The monitoring/irrigation lumens may be used for monitoring airway pressure, for impling I he monitoring/irrigation inflicts may be ased for the of accumulated secretions, for sampling in a patient s tracheooronemal tree, to are in the reaches, or for introducing suitable medication in accordance with standard medical practice.
Concurrence of CDRH, Office of Device Evaluation (ODE)
Prescription Use _____________________________________________________________________________________________________________________________________________________________ (Per 21 CFR 801.109) OR
Over-The-Counter Use
Aus Taylor
(Division Sign-Off
Division of Anesthesiology, General Hospital,
Infection Control, Dental Devices
**510(k) Number:**
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.