K040580 · D.Y. Instrument, Inc. · NRW · Aug 26, 2004 · General Hospital
Device Facts
Record ID
K040580
Device Name
AIS100 AUTOMATIC INTRAMUSCULAR STIMULATOR
Applicant
D.Y. Instrument, Inc.
Product Code
NRW · General Hospital
Decision Date
Aug 26, 2004
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 880.5580
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The D.Y. Instrument, Inc. AIS100 Automatic Intramuscular Stimulator Device is intended, as part of a treatment regimen, for 1. Intramuscular stimulation 2. Symptomatic relief of tension-type headache of myofascial origin
Device Story
The AIS100 is an electromechanical device designed to automate intramuscular stimulation (dry needling). It consists of a main control unit, a hand-held motor unit, and an AC power adapter. The device replaces manual needle stroking by providing automated forward and backward motion of a needle at a constant speed. It is operated by a clinician to reduce physical effort during treatment. The device uses a stainless-steel needle (TECA EMG needle) to perform the stimulation. By automating the stroke length and frequency, it facilitates myofascial release therapy. Clinical benefits include pain relief, improved cervical spine range of motion, and improved quality of life scores in patients with tension-type headaches.
Clinical Evidence
Prospective study of 28 individuals with tension-type headaches (duration ≥2 months). Patients received five treatments over 10 weeks, including myofascial release, AIS100 IMS (30-45 mins), hot packs, and AquaMED. Results: 71.4% experienced relief for ≥72 hours after the first treatment; 64.3% reported a 50% pain reduction after two treatments. Statistically significant improvements observed in cervical spine range of motion (flexion, extension, lateral flexion) and all seven categories of the pain disability/lifestyle index. No adverse effects reported.
Technological Characteristics
Electromechanical intramuscular stimulator. Materials: stainless-steel needle (TECA EMG). Energy source: AC power adapter. Components: main control unit, hand-held motor unit. Standards: IEC 60601-1 (safety), IEC 60601-1-2 (EMC). Operation: automated constant-speed forward/backward needle motion.
Indications for Use
Indicated for patients with chronic or recurrent pain, specifically tension-type headaches of myofascial origin.
Regulatory Classification
Identification
An acupuncture needle is a device intended to pierce the skin in the practice of acupuncture. The device consists of a solid, stainless steel needle. The device may have a handle attached to the needle to facilitate the delivery of acupuncture treatment.
Special Controls
*Classification.* Class II (special controls). The device, when it is an acupuncture point locator or a single use acupuncture needle, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 880.9. Acupuncture needles must comply with the following special controls:(1) Labeling for single use only and conformance to the requirements for prescription devices set out in 21 CFR 801.109,
(2) Device material biocompatibility, and
(3) Device sterility.
{0}------------------------------------------------
K040580 '93
# AUG 2 6 2004 510(k) SUMMARY
This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirements of SMDA 1990 and 21 CFR §807.92.
The assigned 510(k) number is: K040580
### 1. Submitter's Identification:
D.Y. Instrument, Inc. 30 Manning Lane Cherry Hill, NJ 08003
Contact: Dr. Young H. Lee, President
Date Summary Prepared: May 30, 2004
### 2. Name of the Device:
AIS100 Automatic Intramuscular Stimulator
### 3. Predicate Device Information:
K024207, Vinco Brand Acupuncture Needle
### 4. Device Description:
The AIS 100 is an intramuscular stimulator device that aids administering of intramuscular stimulation (or dry needling) for treatment. The device produces an automated forward and backward motion of a needle at a constant speed, which replaces the manual stroking of a needle when rendering the intramuscular stimulation (or dry needling) treatment. This enables the practitioner to administer the intramuscular stimulation treatment effectively with less physical effort when compared to the manual method. The device consists of a main control unit, a hand-held motor unit, and an AC power supply adapter.
### 5. Intended Use:
The DY Instrument, Inc. AIS100 Automatic Intramuscular Stimulator Device is intended, as part of a treatment regimen, for
- 1. Intramuscular stimulation
- 2. Symptomatic relief of tension-type headache of myofascial origin
{1}------------------------------------------------
Ko 40580
27
### 6. Comparison to Predicate Devices:
The needle used with the AIS100, the TECA EMG needle (K973442), is equivalent to the predicate, Vinco Acupuncture Needle (K024707), in terms of physical dimension - diameter, length, and tip shape; and material of construction - stainless-steel. Both needles are FDA-cleared, supplied in sterile packages, and have handles for manual handling. The needling technique used with the AIS100 is equivalent to one of the needling techniques used with the predicate device in the practice of acupuncture in terms of initial insertion method, the use of short multiple strokes, the shortness of stroke length, and the needling termination method.
### 7. Discussion of Non-Clinical Tests Performed for Determination of Substantial Equivalence is as follows:
Non-clinical performance testing included:
- · IEC 60601-1
- · IEC 60601-1-2
# 8. Discussion of Clinical Tests Performed:
Human clinical studies were conducted using the AIS100 Automatic Intramuscular Stimulator device to demonstrate the safety of the subject device and the intramuscular stimulation procedure in treating chronic or recurrent pain by measuring any and all adverse effects of the procedure administered using the subject device.
To demonstrate the efficacy of the intramuscular stimulation (IMS) procedure using the AIS100 in the treatment of chronic or recurrent pain, particularly the tension-headache, efficacy was measured in terms of pain recurrence and pain relief using a pain scale and range of motion of the cervical spine. The impact upon quality of life was also evaluated using the pain disability index.
A prospective study of 28 individuals with tension-type headaches of at least 2 months' duration was conducted. Patients had received various other forms of treatment or no treatment at all before entering the study. The patients received a total of five treatments over a 10 week period, with each treatment consisting of myofascial release therapy (15 mins), IMS treatment using the AIS100 (30-45 mins), hot packs (10 mins), and AquaMED (15 mins). After the first treatment, 71.4% of the patients experienced relief for 72 hrs or more, and after two treatments, 64.3% of patients experienced a 50% decrease in their level of pain. Flexion, extension, and lateral flexion range of motion improvements were also statistically significant after five treatments. Finally, patients experienced a statistically significant improvement in all seven categories of the pain disability/lifestyle index. No adverse effects have been observed.
The IMS technique using the Automatic Intramuscular Stimulator device AIS100 appears to be an effective treatment for relieving the pain associated with
{2}------------------------------------------------
tension-type headaches.
K040580\$^{373}\$
## 9. Conclusions:
The AIS100 Automatic Intramuscular Stimulator has similar intended uses and similar characteristics as the predicate device. Moreover, bench testing contained in this submission and human clinical testing supplied demonstrate that any differences in their technological characteristics or modes of operations do not raise any new questions of safety or effectiveness. Thus, the AIS100 Automatic Intramuscular Stimulator is substantially equivalent to the predicate device.
{3}------------------------------------------------
Image /page/3/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circle with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the top half of the circle. Inside the circle is an abstract symbol that resembles an eagle or bird in flight, composed of three curved lines.
Public Health Service
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
AUG 2 6 2004
Dr. Young H. Lee President D. Y. Instrument, Incorporated 30 Manning Lane, Cherry Hill, New Jersey 08003
Re: K040580
Trade/Device Name: AIS100 Automatic Intramuscular Stimulator Regulation Number: 880.5580 Regulation Name: Acupuncture Needle Regulatory Class: II Product Code: NRW Dated: August 23, 2004 Received: August 24, 2004
Dear Dr. Lee:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{4}------------------------------------------------
Page 2 - Dr. Lee
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (301) 594-4618. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html
Sincerely yours,
Siva Ruone
Chiu Lin, Ph.D.
Chiu Lin, Ph.D. Director Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
K040580
## Exhibit B
Page 1
### 510(k) Number (if known): K040580
Device Name: AIS100 Automatic Intramuscular Stimulator
### Indications For Use:
The D.Y. Instrument, Inc. AIS100 Automatic Intramuscular Stimulator Device is intended, as part of a treatment regimen, for
- 1. Intramuscular stimulation
- 2. Symptomatic relief of tension-type headache of myofascial origin
Prescription Use __X_
Over-The Counter Use
(Per 21 CFR 801 Subpart D)
OR (21 CFT 807 Subpart C)
# (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
13v (Division Sign-Off) (Division of Anesthesiology, General Hospital, Division Control, Dental Devices
510(k) Number:________________________________________________________________________________________________________________________________________________________________
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.