← Product Code [OTN](/productcode/OTN) · K040538

# BIOARC SP SLING KIT AND BIOARC TO SLING KIT, MODELS 72403851, 72403996 (K040538)

_American Medical Systems, Inc. · OTN · Mar 17, 2004 · General, Plastic Surgery · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K040538

## Device Facts

- **Applicant:** American Medical Systems, Inc.
- **Product Code:** [OTN](/productcode/OTN.md)
- **Decision Date:** Mar 17, 2004
- **Decision:** SESE
- **Submission Type:** Special
- **Regulation:** 21 CFR 878.3300
- **Device Class:** Class 2
- **Review Panel:** General, Plastic Surgery
- **Attributes:** Therapeutic

## Indications for Use

BioArc TO Subfascial Hammock is intended for the placement of a suburethral graft for the treatment of female stress urinary incontinence (SUI) resulting from urethral hypermobility and / or intrinsic sphincter deficiency.

## Device Story

BioArc TO Subfascial Hammock is a sterile, single-use surgical kit for treating female stress urinary incontinence. The device consists of two stainless steel helical needle passers and a mesh sling assembly with attached graft material. It utilizes a transobturator surgical approach to place a suburethral graft. The procedure is performed by a physician in a clinical/surgical setting. The device provides mechanical support to the urethra to restore continence. It combines technologies from the BioArc SP Sling Kit and the Monarc subfascial hammock.

## Clinical Evidence

Bench testing only; no clinical data provided.

## Technological Characteristics

Surgical mesh (polymeric) with stainless steel helical needle passers. Sterile, single-use kit. Transobturator surgical approach.

## Regulatory Identification

Surgical mesh is a metallic or polymeric screen intended to be implanted to reinforce soft tissue or bone where weakness exists. Examples of surgical mesh are metallic and polymeric mesh for hernia repair, and acetabular and cement restrictor mesh used during orthopedic surgery.

## Predicate Devices

- BioArc SP™ Sling Kit ([K030123](/device/K030123.md))
- MONARC™ subfascial hammock ([K023516](/device/K023516.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

MAR 1 7 2004

Kψ4ψ538

| 2.6                         | 510(k) Summary                                                                                                       |
|-----------------------------|----------------------------------------------------------------------------------------------------------------------|
| Submitter:                  | American Medical Systems<br>10700 Bren Road West<br>Minnetonka, MN 55343<br>Phone: 952-933-4666<br>Fax: 952-930-6496 |
| Contact Person:             | Denise Thompson                                                                                                      |
| Date Summary Prepared:      | February 27, 2004                                                                                                    |
| Device Common Name:         | Urethral Sling, Surgical Mesh                                                                                        |
| Device Trade Name:          | BioArc TO™ Subfascial Hammock                                                                                        |
| Device Classification Name: | Surgical Mesh, polymeric                                                                                             |
| Predicate Device:           | BioArc SP™ Sling Kit, K030123<br>MONARC™ subfascial hammock, K023516                                                 |

### Device Description:

The BioArc TO Subfascial Hammock is a suburethral sling procedure that uses a transobturator surgical approach to treat stress urinary incontinence. It is a sterile, single transobare kit consisting of two stainless steel helical shaped needle passers and a mesh sling assembly to which graft material is attached.

### Indications for Use:

BioArc TO Subfascial Hammock is intended for the placement of a suburethral graft for the treatment of female stress urinary incontinence (SUI) resulting from urethral hypermobility and / or intrinsic sphincter deficiency.

## Comparison to Predicate Device:

The BioArc TO brings two currently marketed AMS technologies, the BioArc SP Sling Kit and the Monarc subfascial hammock, together into one procedure.

The Indications for Use, fundamental scientific technology, surgical approach, sling placement, and materials are all the same as one of the two predicates.

### Supporting Information:

The risk analysis and the verification / validation activities reported in this Special 510(k) application substantiate equivalence to the predicate devices and did not raise any new questions of safety or efficacy.

### Conclusion:

The BioArc TO Subfascial Hammock is substantially equivalent to its predicates with respect to intended use, technological characteristics, and performance.

{1}------------------------------------------------

Image /page/1/Picture/1 description: The image is a black and white logo for the U.S. Department of Health and Human Services. The logo features a stylized eagle with its wings outstretched, symbolizing protection and care. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" is arranged in a circular fashion around the eagle, emphasizing the department's name and national affiliation. The overall design is simple and professional, conveying a sense of authority and trustworthiness.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002

Ms. Denise Thompson Regulatory Specialist American Medical Systems, Inc. 10700 Bren Road West MINNETONKA MN 55343

SEP 28 2022

Re: K040538 Trade/Device Name: AMS BioArc™ TO Subfascial Hammock AMS BioArc™ SP Sling Kit Regulation Number: 21 CFR 878.3300 Regulation Name: Surgical mesh Regulatory Class: II Product Code: OTN, PAG Dated: March 1, 2004 Received: March 2, 2004

Dear Ms. Thompson:

This letter corrects our substantially equivalent letter of March 17, 2004.

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must

{2}------------------------------------------------

Page 2 -

comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.

Sincerely yours.

Sincerely yours,

Benjamin R. Fuchs

Benjamin R. Fisher, Ph.D. Director Division of Reproductive. Gastro-Renal. and Urological Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{3}------------------------------------------------

1.4

# INDICATIONS FOR USE STATEMENT

<44 4538

510(k) Number (if known):

Device Name: AMS BioArc™ SP Sling Kit

Indications For Use:

The BioArc SP Sling Kit is intended for the placement of a pubourethral sling for the The BloArc St. Sling Ni. Is Intended for the pre (SUI) resulting from urethral hypermobility and / or intrinsic sphincter deficiency.

× Prescription Use_ (Per 21 CFR 801 Subpan D) AND/OR

Over-The Counter Use_ (21 CFR 807 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

Muriam C. Provost

(Division Sign-Off) Division of General, Restorative, and Neurological Devices

Page 1 of _2_

**510(k) Number** K040538

Page 12 of 28

---

**Source:** [https://fda-staging.innolitics.com/device/K040538](https://fda-staging.innolitics.com/device/K040538)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
