FLEXIFIL SERIES HC431 FULL FACE MASK, MODEL HC431A
Applicant
Fisher & Paykel Healthcare, Ltd.
Product Code
BZD · Anesthesiology
Decision Date
May 5, 2004
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 868.5905
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The Fisher & Paykel Healthcare Flexifit™ Series HC431 Full Face Mask is an accessory to a Noncontinuous ventilator (IPPB) as per 73 BZD, 21 CFR §868.5905. The Flexifit™ Series HC431 Full Face Mask is indicated for use by adults requiring CPAP or Bilevel ventilator treatment in home, hospital and laboratory environments for the treatment of Obstructive Sleep Apnea (OSA). It constitutes the patient to ventilator interface in a noncontinuous ventilator system. The device administers positive airway pressure via a mask which covers the nose and mouth. The HC431 Full Face Mask is a multi-patient reusable device for use on the prescription of a suitably qualified physician.
Device Story
Respiratory mask interface for CPAP/Bilevel ventilators; covers nose and mouth. Delivers positive airway pressure (3-25 cm H2O) to treat OSA. Features silicone seal for airtight fit; polycarbonate base; Glider™ mechanism for headgear attachment allowing head movement without mask displacement; quick-release cord for emergency removal. Includes non-rebreathing valve to prevent CO2 re-breathing during pressure loss and bias slots for exhaled gas purging. Used in home, hospital, or sleep lab; operated by patient or healthcare provider. Output is delivered pressure; provider monitors therapy efficacy via ventilator settings. Benefits include secure, comfortable airway pressure delivery and emergency safety features.
Clinical Evidence
Bench testing only. No clinical data provided. Testing confirmed strength, durability, pressure/flow characteristics, and connection standards compliance.
Indicated for adult patients requiring CPAP or Bilevel ventilator treatment for Obstructive Sleep Apnea (OSA) in home, hospital, or sleep lab settings.
Regulatory Classification
Identification
A noncontinuous ventilator (intermittent positive pressure breathing-IPPB) is a device intended to deliver intermittently an aerosol to a patient's lungs or to assist a patient's breathing.
{0}------------------------------------------------
K040506
# MAY = 5 2004
17 February 2004
# 510(k) Summary of Safety and Effectiveness Information
| Trade Name: | Flexifit™ Series HC431 Full Face Mask |
|----------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Classification Name: | Accessory to Noncontinuous ventilator (IPPB) - 73 BZD<br>Anesthesiology Devices, 21 CFR §868.5905 (Class II) |
| Predicate Devices: | Resmed, Full Face Mask Series 2, K023244<br>Fisher & Paykel Healthcare, Oracle Oral Mask, K033087<br>(predicate for non-rebreathing valve and biocompatibility) |
This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirements of 21 CFR §807.92:
(a)(1) - (a)(3) Refer to information above and concluding this summary.
### (a)(4) Description of the Device
The Flexifit™ Series HC431 Fuill Face Mask is an accessory to a Noncontinuous ventilator (IPPB) according to 21 CFR §868.5905. It constitutes the patient to ventilator interface in a noncontinuous ventilator system.
The Flexifit™ Series HC431 Full Face Mask is a respiratory mask which covers the mouth and nose. The mask is connected to the output breathing tube of the ventilator. The ventilator supplies air at CPAP or Bilevel pressures (typically in the range 3 - 25 cm H₂O) which is available at the mask.
The mask is positioned over the patient's nose and mouth during CPAP or Bilevel treatment. The HC431 Full Face Mask is designed to ensure the desired positive airway pressure is delivered to the patient with minimal leakage and that the mask is held securely over the nose and mouth while asleep.
The non-rebreathing valve prevents CO2 re-breathing in the event of pressure loss due to power outage or ventilator machine failure. The elbow on the front of the mask facilitates freedom of movement while maintaining circuit integrity. Bias slots on the mask provide a means to purge exhaled gases from the breathing circuit.
{1}------------------------------------------------
# 510(k) Summary of Safety and Effectiveness Information (continued)
#### (a)(5) Statement of the Intended Use
The Flexifit™ Series HC431 Full Face Mask is intended for multiple patient or single patient adult use by individuals who have been diagnosed by a physician as requiring CPAP or Bilevel ventilator treatment. A CPAP or Bilevel ventilator is typically used to treat obstructive sleep apnea (OSA) and may be used in the home, hospital or sleep laboratory.
The mask may be reprocessed and reused by healthcare facilities to allow multi patient use. The mask may be reprocessed up to 20 times.
The Flexifit™ Series HC431 Full Face Mask is designed to function as intended for up to 12 months of daily use when cared for as specified by the User Instructions by a single patient in the home.
#### (a)(6) Technological Characteristics Summary
The technological characteristics of the Flexifit™ Series HC431 Full Face Mask are equivalent to the predicate devices listed above.
The silicone seal is designed to fit around a patient's nose and mouth, and provide an airtight seal by which positive airway pressure is supplied to the patient. The seal is designed to be contoured and flexible to provide a good fit for most of the population. The seal fits under the patient's chin to provide stability to the mask and prevent the mask sliding up the face of the user. The seal is available in two sizes, both of which fit the single size mask base.
The base is manufactured from a rigid polycarbonate, and is designed to provide attachment points for the seal, headgear and breathing circuit. The upper headgear attaches to the base via the forehead rest, while the lower headgear straps attach to the base via the Glider™. The forehead rest holds the forehead cushion, and gives additional stability to the mask while preventing the seal from being over-tightened onto the bridge of the nose.
The Glider™ provides a sliding attachment between the headgear and the base, so that as the user turns their head, the Glider™ can slide through the mask base without becoming tight, as headgear does if it is rigidly attached to the base. The Glider™ has two straps, which help to maintain the mask stability. The Glider™ is held onto the mask base by the glider cover. The Glider™ is manufactured from Nylon 6, which provides good flexibility and strength, while maintaining a low coefficient of friction between the Glider™ and the polycarbonate base.
One end of the Glider™ is attached to the headgear by a clip, which incorporates a quick release cord. In case of claustrophobia, vomiting or other emergency, the patient can pull the quick release strap to release the mask from their face. The clip is manufactured from acetal, which provides good flexibility and toughness. The other end of the Glider™ is attached to the headgear via a tight hook. Both the clip and the hook allow the user to remove the mask without adjusting the headgear strap length settings.
{2}------------------------------------------------
The polycarbonate glider cover also provides the bias slots. The bias slots are thin slots, which taper gently outwards to allow the bias flow to escape from the base while remaining attached to the sidewalls of the slot. This results in minimal velocity and noise. The cover provides one sidewall of the slot, while the mask base provides the other.
The breathing circuit is attached to the mask base via the elbow and the nonrebreathing valve. The polycarbonate elbow has a free rotating swivel-snap connection on both ends. This means that the user can adjust the breathing circuit to any position they prefer. The free rotation reduces the torque transferred from the breathing circuit to the mask, so the mask disturbance is minimised.
Attached to the other end of the elbow is the non-rebreathing valve. The valve is designed to prevent carbon dioxide build up in the mask in the event of cessation of airflow from the breathing circuit. The valve consists of a thin silicon flap and two polycarbonate housings, which are ultrasonically welded together. During normal operation, the valve's silicon flaps are held open by the CPAP flow. When the flow ceases, the flaps return to their natural position, occluding the circuit and preventing back flow. This forces any exhaled air to exit the mask via the bias slots, thereby preventing CO2 build up in the mask. The valve has a standard 22mm conical connector, which the breathing circuit attaches to.
The headgear provides four attachment points to the mask. The length of each strap can be adjusted using Velcro tabs, and each strap can also be unhooked without altering or undoing the length adjustment. The headgear is made of Breath-o-prene, which helps prevent sweating. The headgear also provides an attachment point for the other end of the quick release cord.
{3}------------------------------------------------
# 510(k) Summary of Safety and Effectiveness Information (continued)
#### (b)(1) and b(2) Discussion of Non-Clinical Tests
Tests performed on the Flexifit™ Series HC431 Full Face Mask demonstrate substantial equivalence to the predicate devices.
Non-clinical tests have demonstrated effective performance in terms of strength, durability, pressure and flow characteristics, and conformance of connections to industry standards.
#### (b)(3) Conclusions Demonstrating Safety, Effectiveness and Performance
When used as intended, the Flexifit™ Series HC431 Full Face Mask has been shown to be as safe and effective as the predicate device. Specifically:
- The Flexifit™ Series HC431 Full Face Mask is a safe patient to ventilator interface . when used as an accessory to a Noncontinuous ventilator
- The Flexifit™ Series HC431 Full Face Mask is an effective means of delivering ● positive airway pressure in the treatment of OSA
- The Flexifit™ Series HC431 Full Face Mask is a reliable device when used and . maintained as specified in the Device Instructions
This information verifies that the Flexifit™ Series HC431 Full Face Mask is equivalent to the predicate device in terms of safety, effectiveness and performance.
signed:
: Ic I
te: 17/2/04
James Thompson Fisher & Paykel Healthcare Ltd
{4}------------------------------------------------
DEPARTMENT OF HEALTH & HUMAN SERVICES
Image /page/4/Picture/1 description: The image is a seal for the Department of Health & Human Services - USA. The seal is circular and contains the department's name around the perimeter. In the center of the seal is an abstract symbol that resembles an eagle or a bird in flight, composed of three curved lines.
Public Health Service
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
Mr. James Thompson Regulatory Affairs Engineer Fisher & Paykel Healthcare Limited 15 Maurice Paykel Place East Tamaki Auckland, 1701 NEW ZEALAND
Re: K040506
Trade/Device Name: Fisher & Paykel Healthcare-Flexifit™ Series HC431 Full Face Mask Regulation Number: 868.5905 Regulation Name: Noncontinuous Ventilator (IPPB) Regulatory Class: II -Product Code: BZD Dated: February 17, 2004 Received: February 27, 2004
Dear Mr. Thompson:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the cnactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into cither class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898, In addition, FDA may publish further announcements concerning your device in the Federal Register.
MAY - 5 2004
{5}------------------------------------------------
#### Page 2 - Mr. Thompson
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act): 21 CFR 1000-1050.
This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits vour device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (301) 594-4646. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807,97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html
Sincerely yours.
Ouls
Chiu Lin, Ph.D. Director Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{6}------------------------------------------------
[510(k)] Number: _ K ﮟ ﮨﮯ ﮐﮧ
17 February 2004
# Fisher & Paykel Healthcare - Flexifit™ Series HC431 Full Face Mask
### PREMARKET NOTIFICATION 510(k) INDICATIONS FOR USE STATEMENT
The Fisher & Paykel Healthcare Flexifit™ Series HC431 Full Face Mask is an accessory to a Noncontinuous ventilator (IPPB) as per 73 BZD, 21 CFR §868.5905.
The Flexifit™ Series HC431 Full Face Mask is indicated for use by adults requiring CPAP or Bilevel ventilator treatment in home, hospital and laboratory environments for the treatment of Obstructive Sleep Apnea (OSA). It constitutes the patient to ventilator interface in a noncontinuous ventilator system. The device administers positive airway pressure via a mask which covers the nose and mouth. The HC431 Full Face Mask is a multi-patient reusable device for use on the prescription of a suitably qualified physician.
Prescription Use ✓ (Per 21 CFR subpart D) and/or
Over-the-Counter Use (Per 21 CFR 807' subpart C)
(Please do not write below this line - continue on another page if needed) Concurrence of CDRH, Office of Device Evaluation (ODE)
Avery Sullivan
(Division Sign-Off Division of Anesthesiology, General Hospital, Infection Control, Dental Devices
510(k) Number: K040504
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.