VARELISA B2 GLYCOPROTEIN I IGA ANTIBODIES, MODELS 18948 AND 18996
K040450 · Pharmacia Deutschland GmbH · MSV · May 11, 2004 · Immunology
Device Facts
Record ID
K040450
Device Name
VARELISA B2 GLYCOPROTEIN I IGA ANTIBODIES, MODELS 18948 AND 18996
Applicant
Pharmacia Deutschland GmbH
Product Code
MSV · Immunology
Decision Date
May 11, 2004
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 866.5660
Device Class
Class 2
Indications for Use
The Varelisa B2-Glycoprotein I IgA Antibodies Kit is designed for the semiquantitative and qualitative determination of β2-glycoprotein I IgA antibodies in serum or plasma. The presence of β2-glycoprotein I antibodies can be used in conjunction with clinical findings and other laboratory tests to aid in the diagnosis of thrombotic disorders related to the primary Antiphospholipid Syndrome or occurring secondary to systemic lupus erythematosis (SLE) or other autoimmune diseases. Prescription use only.
Device Story
Varelisa B2-Glycoprotein I IgA Antibodies EIA is an indirect noncompetitive enzyme immunoassay. Input: human serum or plasma samples. Process: patient antibodies bind to human purified B2-glycoprotein I antigen coated on microplate wells; enzyme-labeled conjugate binds to antigen-antibody complex; substrate added to produce color. Output: color intensity proportional to antibody concentration. Used in clinical laboratories by technicians. Results aid physicians in diagnosing thrombotic disorders associated with primary Antiphospholipid Syndrome or secondary to SLE/autoimmune diseases. Benefits include improved diagnostic accuracy for thrombotic risk.
Clinical Evidence
No clinical studies were performed. Evidence consists of analytical performance data (precision, linearity, analytical sensitivity, specificity) and a method comparison study. Method comparison used 42 serum samples (SLE, APS, thrombosis, etc.) comparing the subject device to the predicate, showing 88.1% qualitative agreement. Analytical sensitivity was 0.03 U/mL. Interference testing showed no significant impact from bilirubin, hemoglobin, chyle, or rheumatoid factor.
Technological Characteristics
Indirect noncompetitive enzyme immunoassay (EIA). Components: microplate strips coated with human purified B2-glycoprotein I antigen, calibrators, controls, enzyme-labeled conjugate, substrate, stop solution, buffers. Manual or automated processing. No specific software algorithm class or connectivity specified.
Indications for Use
Indicated for patients suspected of having thrombotic disorders related to primary Antiphospholipid Syndrome or secondary to systemic lupus erythematosus (SLE) or other autoimmune diseases. Used for semiquantitative and qualitative determination of B2-glycoprotein I IgA antibodies in serum or plasma.
Regulatory Classification
Identification
A multiple autoantibodies immunological test system is a device that consists of the reagents used to measure by immunochemical techniques the autoantibodies (antibodies produced against the body's own tissues) in serum and other body fluids. Measurement of multiple autoantibodies aids in the diagnosis of autoimmune disorders (disease produced when the body's own tissues are injured by autoantibodies).
Predicate Devices
INOVA QUANTA Lite B2 GPI IgA (k973006)
Submission Summary (Full Text)
{0}
510(k) SUBSTANTIAL EQUIVALENCE DETERMINATION
DECISION SUMMARY
DEVICE ONLY TEMPLATE
A. 510(k) Number:
k040450
B. Purpose of Submission:
New device
C. Analyte:
$\beta 2$-Glycoprotein I IgA antibodies
D. Type of Test:
Semiquantitative and qualitative ELISA
E. Applicant:
Pharmacia Deutschland GmbH/Sweden Diagnostics (Germany) GmbH
F. Proprietary and Established Names:
Varelisa B2-Glycoprotein I IgA Antibodies EIA Kit
G. Regulatory Information:
1. Regulation section:
21 CFR 866.5660, Multiple Autoantibodies Immunological Test System
2. Classification:
Class II
3. Product Code:
MSV, System, Test, Antibodies, $\beta 2$-Glycoprotein I ($\beta 2$-CPI)
4. Panel:
Immunology (82)
H. Intended Use:
1. Intended use(s):
The Varelisa B2-Glycoprotein I IgA Antibodies Kit is designed for the semiquantitative and qualitative determination of $\beta 2$-glycoprotein I IgA antibodies in serum or plasma.
2. Indication(s) for use:
The presence of $\beta 2$-glycoprotein I antibodies can be used in conjunction with clinical findings and other laboratory tests to aid in the diagnosis of thrombotic disorders related to the primary Antiphospholipid Syndrome or occurring secondary to systemic lupus erythematosis (SLE) or other autoimmune diseases.
3. Special condition for use statement(s):
Prescription use only.
{1}
Page 2 of 6
4. Special instrument Requirements:
None.
I. Device Description:
The Varelisa B2-Glycoprotein I IgA Antibodies kit consists of an ELISA strip wells (12 strips X 8 wells) coated with purified human β2-glycoprotein I antigen. It also contains ancillary reagents such as calibrators, positive and negative controls, wash buffer, sample diluent, antibody-chromogen conjugate, enzyme substrate, stop solution for enzyme reaction, and a frame to hold ELISA strip wells. Also included are a package insert and a control certificate.
J. Substantial Equivalence Information:
1. Predicate device name(s):
INOVA QUANTA Lite B2 GPI IgA
2. Predicate K number(s):
k973006
3. Comparison with predicate:
| Similarities | | |
| --- | --- | --- |
| Item | Device | Predicate |
| Intended Use | Semiquantitative and qualitative determination of β2-glycoprotein I IgA antibodies in serum or plasma for use in conjunction with clinical findings and other laboratory tests to aid in the diagnosis of thrombotic disorders related to primary Antiphospholipid Syndrome or occurring secondary to systemic lupus erythematosis (SLE) or other autoimmune diseases. | Qualitative detection of β2-GPI IgA antibodies in serum for use in conjunction with clinical findings and other laboratory tests to aid in the diagnosis of certain autoimmune disease thrombotic disorders, such as those secondary to systemic lupus erythematosis (SLE) or other lupus-like thrombotic diseases. |
| Antigen | Human affinity-purified β2-glycoprotein I | β2-glycoprotein I, unknown source |
| Conjugate | Anti-human IgA Antibody Conjugate, Horseradish peroxidase | Same |
| Assay Principle | Indirect noncompetitive enzyme immunoassay | Same |
| Sample dilution | 1:101 | Same |
{2}
Page 3 of 6
| Differences | | |
| --- | --- | --- |
| Item | Device | Predicate |
| Specimen | Serum and plasma | Serum |
| Calibrators | Six prediluted calibrators | Five prediluted calibrators |
| Assay evaluation | OD cut-off | Decision point |
## K. Standard/Guidance Document Referenced (if applicable):
None
## L. Test Principle:
The Varelisa B2-Glycoprotein I IgA Antibodies EIA Kit is an indirect noncompetitive enzyme immunoassay. The wells of the microplate are coated with human purified $\beta 2$-glycoprotein I antigen. Antibodies specific for $\beta 2$-glycoprotein I present in the applied patient sample bind to the antigen and are detected by an enzyme labeled secondary antibody complex, which leads to the formation of an enzyme labeled conjugate-antibody-antigen complex. The enzyme labeled antigen-antibody complex converts the added substrate to form a colored solution. The rate of color formation from the chromogen (substrate) is a function of the amount of conjugate complexed with the bound antibody and thus is proportional to the initial concentration of the respective antibodies in the patient sample. The antibody concentration is determined by comparison to a calibration curve run with the assay and is reported in U/mL.
## M. Performance Characteristics (if/when applicable):
### 1. Analytical performance:
#### a. Precision/Reproducibility:
Five serum samples (low, medium, high, equivocal, negative) were selected from a serum bank. The samples were diluted 1:101 and measured in 5 runs, with 4 replicates per run. Standards and controls were analyzed in triplicate. Specified target values were: variance "within" < 12%; "between" < 15%; "total" < 15%. The specifications were met for each sample, with the following variances:
| Sample ID | | Run 1 | Run 2 | Run 3 | Run 4 | Run 5 | Mean (U/mL) | Variance | | |
| --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- |
| | | | | | | | | Within | Between | total |
| ZSB 5496 | Mean (u/mL) CV% | 16.8
1.24 | 17.1
4.19 | 18.2
3.44 | 17.5
2.85 | 17.5
6.92 | 17.4 | 4.19 | 2.18 | 4.72 |
| ZSB 6008 | Mean (u/mL) CV% | 25.6
3.28 | 24.4
4.65 | 26.2
2.08 | 25.3
4.71 | 26.1
4.39 | 25.5 | 3.93 | 2.06 | 4.43 |
| ZSB 7437 | Mean (u/mL) CV% | 53.5
3.26 | 51.5
1.97 | 54.1
1.11 | 53.3
3.48 | 57.6
2.50 | 54.0 | 4.66 | 3.42 | 5.78 |
| negative | Mean (u/mL) CV% | 3.6
5.8 | 3.5
6.6 | 4.0
5.8 | 3.5
5.6 | 4.0
6.1 | 3.7 | 6.0 | 6.7 | 9.0 |
| equiv. | Mean (u/mL) | 14.9 | 14.6 | 17.4 | 14.4 | 16.0 | 15.4 | 7.6 | 7.0 | 10.4 |
{3}
Page 4 of 6
b. Linearity/assay reportable range:
Four samples selected from a serum bank were diluted 1:1, 2:3, 1:2, 1:4, 1:8, 1:16, and 1:32 in Sample Diluent. Standards, Controls and each dilution were measured in triplicate. Specifications were that observed/expected units ratio would not differ by more than ± 20% for at least 3 successive dilutions of each tested sample.
The dilutions met the criteria and were considered linear.
Recovery of analyte was assayed using two samples from a serum bank. Each sample was tested in six replicates to determine pre-existing antibodies. Each sample was then diluted 1:101 and spiked with 1/10 volume of Standards S1 to S6 and measured in duplicate. Acceptance criteria were that recovery would be 100 ± 20%. In all cases for both samples, the acceptance criteria were met.
c. Traceability (controls, calibrators, or method):
Not traceable to external sources. The consistency of the calibration material is given by a testing procedure comprising several internal target values that must be met. Each new production of calibration material is adapted to the corresponding master calibration material.
d. Detection limit:
Sample Buffer was diluted according to Directions for Use and measured 56 times on one plate. Standards and Controls were analyzed in four replicates. Two dilutions of the first standard point, S2 (1:2, 1:4) were run in four replicates. Analytical sensitivity was calculated as the mean of the optical densities of the sample buffer plus 3SD and expressed in U/mL. The discrimination value D, proves the ability to discriminate between the lowest standard point and the background, was performed as follows:
$$
D = \frac {\eta_ {B} - \eta_ {A}}{\sqrt {\left(\sigma_ {B} ^ {2} - \sigma_ {A} ^ {2}\right)}}
$$
Where: A is Sample Buffer; B is Standard S2; ηA, ηB are mean OD's; σA, σB are SD's.
Specifications were that analytical sensitivity should be below 1 U/mL, the mean +3SD of sample buffer should be < standard point S2, and the discrimination value should be > 2.0.
The mean +3SD was 0.016 OD, which corresponded to analytical sensitivity of 0.03 U/mL. The other criteria were met.
e. Analytical specificity:
Interference was tested against potentially interfering substances found in blood: bilirubin, hemoglobin, chyle, and rheumatoid factor. Two samples from a serum bank were diluted 1:101, and another was diluted 1:201 to remain within assay range. Samples were spiked with different super-normal amounts of interferent. Spiked
{4}
Page 5 of 6
samples were analyzed in triplicate. Acceptance criteria were that spiked samples should show no more than 20% variation from unspiked sample.
| Additives (final sample concentration) | I | II | III | IV | V |
| --- | --- | --- | --- | --- | --- |
| Bilirubin F (mg/dL) | 0.0 | 4.8 | 9.5 | 14.3 | 19.0 |
| Bilirubin C (mg/dL) | 0.0 | 5.5 | 11.0 | 16.5 | 22.0 |
| Chyle (Units) | 0.0 | 357.5 | 715.0 | 1072.5 | 1430.0 |
| Hemoglobin (mg/dL) | 0.0 | 122.5 | 245.0 | 367.5 | 490.0 |
| RF (IU/mL) | 0.0 | 98.0 | 294.0 | 490.0 | - |
All samples met acceptance criteria and showed no significant influence on the test results. Additional interference tests on negative and equivocal sera with the same interferents demonstrated a negligible effect of the interferents on the equivocal samples, and a more pronounced effect on the negative samples. The negative samples, however, remained negative, so the interference was judged to be not important.
f. Assay cut-off:
The semi-quantitative cut-points were determined by measuring 432 samples from apparently healthy Caucasian donors, equally distributed by sex and age. Diluted samples, standards, and controls were analyzed in duplicate. Specifications called for the mean +2SD to be smaller than the lower limit of the equivocal range, and the mean +3SD to be smaller than the upper limit of the equivocal range. The following values were selected for negative, equivocal, and positive:
<10 U/mL = negative
10-15 U/mL = equivocal
>15 U/mL = positive
The specifications were fulfilled and no difference was detected between sexes or ages.
2. Comparison studies:
a. Method comparison with predicate device:
The following samples were used in comparing the new device with 42 samples from a serum bank at Pharmacia, representing SLE, lupus-like, thrombosis, lung inflammation, thalassemia, APS, and healthy individuals. The results were compared semiquantitatively (Varelisa) vs qualitatively (QUANTA Lite) in a 6 field analysis. 37 of 42 sera agreed qualitatively (pos/neg) for overall agreement of
{5}
Page 6 of 6
88.1%. Two sera were equivocal by the Varelisa assay and positive by the qualitative predicate. Three sera were positive by Varelisa and negative by the predicate.
| Pharmacia
Varelisa IgA | n = 42 | INOVA QUANTA Lite
Lot - | |
| --- | --- | --- | --- |
| | | Positive | Negative |
| | positive | 23 | 0 |
| | equivocal | 2 | 0 |
| | negative | 3 | 14 |
b. Matrix comparison:
Twenty blood donor samples in the positive and negative range of the assay were tested in serum, citrate plasma, heparin plasma, and EDTA plasma, in the negative range of the assay. No significant differences were noted for any of the matrices.
3. Clinical studies:
a. Clinical sensitivity:
Not applicable
b. Clinical specificity:
Not applicable
c. Other clinical supportive data (when a and b are not applicable):
Not applicable
4. Clinical cut-off:
Not applicable
5. Expected values/Reference range:
Expected value for normal healthy individuals is negative, or <10 U/mL. In a study of 432 healthy Caucasians, the following values were obtained:
N: 432
Mean: 2.7 U/mL
Mean +2SD 9.1 U/mL
Median: 2.0 U/mL
95th percentile 6.4 U/mL
N. Conclusion:
The submitted information in this premarket notification is complete, and supports a substantial equivalence decision.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.