ACS:180 AND ADVIA CENTAUR B-TYPE NATRIURETIC PEPTIDE (BNP) ASSAYS
Applicant
Bayer Healthcare, LLC
Product Code
NBC · Clinical Chemistry
Decision Date
Jun 14, 2004
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 862.1117
Device Class
Class 2
Attributes
Real-World Evidence
Real-World Evidence
Submission
Device
Sponsor
RWD Sources
RWE Use Summary
Key Tags
K040425 · Jun 14, 2004
ACS:180 AND ADVIA CENTAUR B-TYPE NATRIURETIC PEPTIDE (BNP) ASSAYS
Bayer Healthcare, LLC
ENTIRE-TIMI 23 (Thrombolysis in Myocardial Infarction) clinical trial data
Retrospective analysis of baseline BNP concentrations from a multi-national clinical trial to determine the clinical cut-off (80 pg/mL) for predicting 30-day survival in patients following myocardial infarction.
ENTIRE-TIMI 23 (Thrombolysis in Myocardial Infarction) study; Retrospective analysis of a multi-national clinical trial; Follow-up/Duration: 30 days; Study Period: February 2000 to September 2001
438 patients with acute myocardial infarction (AMI); Sample Size: 438
Not applicable for this study
30-day mortality and composite endpoints of death and heart failure
Indications for Use
For in vitro diagnostic use in the quantitative determination of B-type natriuretic peptide (BNP) in human plasma using the ADVIA Centaur® System. This assay is indicated for the measurement of plasma BNP as an aid in the diagnosis and assessment of severity of heart failure. This test, in conjunction with other known risk factors, can also be used to predict survival in patients after myocardial infarction. This assay is not intended for use on any other system. For in vitro diagnostic use in the quantitative determination of B-type natriuretic peptide (BNP) in human plasma using the ACS:180® System. This assay is indicated for the measurement of plasma BNP as an aid in the diagnosis and assessment of severity of heart failure. This test, in conjunction with other known risk factors, can also be used to predict survival in patients after myocardial infarction. This assay is not intended for use on any other system.
Device Story
Fully automated two-site sandwich immunoassay; uses direct chemiluminescent technology. Input: human plasma sample (EDTA anticoagulant). Process: sample incubated with acridinium ester-labeled monoclonal mouse anti-human BNP antibody (Lite Reagent) and biotinylated monoclonal mouse anti-human antibody coupled to streptavidin magnetic particles (Solid Phase). Immuno-complex forms; unbound conjugates washed away. Output: relative light units (RLUs) measured via luminometer; signal proportional to BNP concentration. Used in clinical laboratories; operated by trained technicians. Results assist clinicians in diagnosing heart failure, assessing severity, and predicting post-myocardial infarction survival.
Clinical Evidence
Retrospective analysis of 438 patients from ENTIRE-TIMI 23 study. Primary endpoint: 30-day mortality post-myocardial infarction. BNP >80 pg/mL associated with higher risk of death (17.4% vs 1.8%, p<0.0001). Odds ratio for death: 11.5 (unadjusted), 8.3 (adjusted). Sensitivity/specificity data provided across age/gender groups for heart failure diagnosis using 100 pg/mL threshold.
Technological Characteristics
Chemiluminescent immunoassay; uses acridinium ester-labeled monoclonal mouse anti-human BNP F(ab')2 fragment and biotinylated monoclonal mouse anti-human antibody on streptavidin magnetic particles. Fully automated, random-access analyzer. Sample volume 100 µL. EDTA plasma matrix. Measuring range <2 to 5000 pg/mL.
Indications for Use
Indicated for quantitative measurement of plasma BNP in patients to aid in diagnosis and severity assessment of heart failure and to predict survival post-myocardial infarction in conjunction with other risk factors.
Regulatory Classification
Identification
The B-type natriuretic peptide (BNP) test system is an in vitro diagnostic device intended to measure BNP in whole blood and plasma. Measurements of BNP are used as an aid in the diagnosis of patients with congestive heart failure.
Special Controls
*Classification.* Class II (special controls). The special control is “Class II Special Control Guidance Document for B-Type Natriuretic Peptide Premarket Notifications; Final Guidance for Industry and FDA Reviewers.”
{0}
# 510(k) SUBSTANTIAL EQUIVALENCE DETERMINATION DECISION SUMMARY DEVICE ONLY TEMPLATE
A. 510(k) Number:
k040425
B. Purpose for Submission:
Labeling changes, Indications for Use modified to include: assessment of severity of heart failure and, in conjunction with other known risk factors, to predict survival in patients after myocardial infarction.
C. Analyte:
B-type natriuretic peptide test system (BNP)
D. Type of Test:
Quantitative
E. Applicant:
Bayer HealthCare LLC
F. Proprietary and Established Names:
Bayer Diagnostics ADVIA® Centaur® BNP Assay, Bayer Diagnostics ACS:180® BNP Assay
G. Regulatory Information:
1. Regulation section:
21 CFR 862.1117 B-type natriuretic peptide test system
2. Classification:
Class II
3. Product Code:
NBC
4. Panel:
75 Chemistry
H. Intended Use:
1. Intended use(s):
For in vitro diagnostic use in the quantitative determination of B-type natriuretic peptide (BNP) in human plasma using the ADVIA Centaur® System. This assay is indicated for the measurement of plasma BNP as an aid in the diagnosis and assessment of severity of heart failure. This test, in conjunction with other known risk factors, can also be used to predict survival in patients after myocardial infarction. This assay is not intended for use on any other system.
{1}
For in vitro diagnostic use in the quantitative determination of B-type natriuretic peptide (BNP) in human plasma using the ACS:180® System. This assay is indicated for the measurement of plasma BNP as an aid in the diagnosis and assessment of severity of heart failure. This test, in conjunction with other known risk factors, can also be used to predict survival in patients after myocardial infarction. This assay is not intended for use on any other system.
2. Indication(s) for use: See Intended Use above
3. Special condition for use statement(s): Prescription use
4. Special instrument Requirements: Bayer Diagnostics ADVIA® Centaur®, Bayer Diagnostics ACS:180®
# I. Device Description:
5 ReadyPack ® primary reagent packs containing ADVIA Centaur® BNP Lite Reagent (monoclonal mouse anti-human BNP fragment antibody labeled with acridium ester in buffer with bovine gamma globulin, mouse gamma globulin and preservatives) and Solid Phase (monoclonal mouse anti-human BNP antibody in bovine gamma globulin, mouse gamma globulin and preservatives), ADVIA Centaur® Master Curve card for 500 tests or 1 ReadyPack ® primary reagent pack containing ADVIA Centaur® BNP Lite Reagent and Solid Phase, ADVIA Centaur® Master Curve card for 100 tests
Or
6 vials of ACS:180® BNP Lite Reagent, 6 vials of ACS:180® Solid Phase, ACS:180® BNP Master Curve Card or 1 vial of ACS:180® BNP Lite Reagent, 1 vial of ACS:180® Solid Phase, ACS:180® BNP Master Curve Card
# J. Substantial Equivalence Information:
1. Predicate device name(s): Bayer Diagnostics ADVIA® Centaur® BNP Assay
2. Predicate K number(s): k031038
3. Comparison with predicate:
| Similarities | | |
| --- | --- | --- |
| Item | Device | Predicate |
| Assay principle | Chemiluminescence immunoassay | Same |
| Traceability | Synthetic human BNP in buffer based matrix | Same |
| Sample type | EDTA plasma | Same |
{2}
Page 3 of 9
| Differences | | |
| --- | --- | --- |
| Item | Device | Predicate |
| Indications for Use | For in vitro diagnostic use in the quantitative determination of B-type natriuretic peptide (BNP) in human plasma using the ADVIA Centaur® and ACS:180® Systems. This assay is indicated for the measurement of plasma BNP as an aid in the diagnosis and assessment of severity of heart failure. This test, in conjunction with other known risk factors, can also be used to predict survival in patients after myocardial infarction. This assay is not intended for use on any other system. | For in vitro diagnostic use in the quantitative determination of B-type natriuretic peptide (BNP) in human plasma using the ADVIA® Centaur® System. This assay is indicated for the measurement of plasma BNP as an aid in the diagnosis of heart failure. This assay is not intended for use on any other system |
| Assay cut-off | Age and gender matched descriptive statistics provided. Decision threshold of 100 pg/mL recommended for diagnosis of heart failure. Decision threshold of 80 pg/mL recommended for prediction of survival after myocardial infarction. | Age and gender matched descriptive statistics provided. Decision threshold of 100 pg/mL recommended. |
K. Standard/Guidance Document Referenced (if applicable):
NCCLS EP5, NCCLS EP7-P
L. Test Principle:
The ADVIA Centaur and ACS:180 BNP assays are fully automated two-site sandwich immunoassays using direct chemiluminescent technology, which uses constant amounts of two monoclonal antibodies. The first antibody, in the Lite Reagent, is an acridinium ester labeled monoclonal mouse anti-human BNP F(ab')₂ fragment specific to the ring structure of BNP. The second antibody, in the Solid Phase, is a biotinylated monoclonal mouse anti-human antibody specific to the C-terminal portion of BNP, which is coupled to streptavidin magnetic particles. A direct relationship exists between the amount of BNP present in the patient sample and the amount of relative light units (RLUs) detected by the system.
{3}
Page 4 of 9
# M. Performance Characteristics (if/when applicable):
## 1. Analytical performance:
### a. Precision/Reproducibility:
ADVIA Centaur: Six samples were assayed 4 times in 20 runs on 2 systems (n = 160 for each sample), over a period of 20 days. Two separate studies were performed, each with a separate lot of reagents. Within run precision % CV ranged from 4.3 % at 29.4 pg/mL to 2.1% at 1736 pg/mL, with total precision of 4.7 % and 2.9 %, respectively.
ACS:180: Six samples were assayed 3 times in 20 runs on 2 systems (n = 120 for each sample) over a period of 20 days. Within run precision % CV ranged from 7.9 % at 51.5 pg/mL to 2.5% at 1783 pg/mL, with total precision of 9.9 % and 3.8 %, respectively.
### b. Linearity/assay reportable range:
Reportable range: ADVIA Centaur <2 to 5000 pg/mL, ACS 180 < 7.5 to 5000 pg/mL. Patient samples with high BNP levels were mixed in various proportions with patient samples containing low levels of BNP. When compared to the expected value, the measured (recovered) values of BNP averaged 103% with a range of 93 to 113% for the ADVIA Centaur and averaged 98% with a range of 96 to 102% for the ACS:180.
### c. Traceability (controls, calibrators, or method):
The ADVIA Centaur and ACS:180 BNP assays are traceable to an internal standard manufactured using synthetic human BNP (amino acid 77-108). Assigned calibrator doses and ranges for quality control material are traceable to this standardization.
### d. Detection limit:
The functional sensitivity is defined as the lowest BNP concentration determined at a coefficient of variation of 20%. The functional sensitivity was determined to be 2.5 pg/mL for the ADVIA Centaur and 11.0 pg/mL for the ACS:180. The minimum detectable concentration (MDC) of the assay is < 2 pg/mL for the ADVIA Centaur and < 7.5 pg/mL for the ACS:180. The MDC is defined as the concentration of BNP that corresponds to the relative light units (RLU's) that are 2 SD > the mean RLU's of 20 replicate determinations of the BNP zero standard. This response is an estimate of the MDC with 95% confidence.
### e. Analytical specificity:
| Specimens that are ... | Demonstrate ≤ 5% change in results up to ... |
| --- | --- |
| hemolyzed | 1000 mg/dL of hemoglobin |
| lipemic | 800 mg/dL of triglycerides |
| | 1000 mg/dL of cholesterol |
| uremic | 200 mg/dL of urea |
{4}
Page 5 of 9
| Specimens that are . . . | Demonstrate ≤ 5% change in results up to . . . |
| --- | --- |
| | 2.5 mg/dL of creatinine |
| icteric | 25 mg/dL of unconjugated bilirubin |
| Specimens that are . . . | Demonstrate ≤ 7% change in results up to . . . |
| icteric | 25 mg/dL of conjugated bilirubin |
| proteinemic | 5.3 g/dL of human IgG |
A total of 55 commonly used pharmaceutical drugs were added to human plasma-based samples at two times the maximum therapeutic dosage and evaluated for potential interference. The results demonstrated $\leq 10\%$ interference from each drug.
The drug Neseritide is a synthetic form of BNP-32 which is thought to be virtually identical to the endogenous active hormone BNP.
The Bayer BNP assays measure Neseritide as BNP. The following statement in the labeling addresses the use of the test with patients receiving Neseritide:
It has been reported that patients with acute decompensated heart failure who are candidates for nesiritide (recombinant BNP) infusion should have a baseline BNP measurement taken prior to initiation of therapy. Measurements taken during infusion are reflective of the dose of nesiritide. Because of the short half-life of BNP (20 minutes), measurements taken 2 hours after the cessation of treatment again reflect the level of endogenous BNP. It has also been reported that following infusion, endogenous BNP levels return to baseline by 1-2 hours and continue to drop at 6 hours to about $80\%$ of preinfusion levels, suggesting a resetting of the neuro-hormonal axis and improvement in ventricular wall tension as a result of treatment. The ADVIA Centaur and ACS:180 BNP assays are not approved for nesiritide monitoring.
f. Assay cut-off:
See clinical cut-off below
2. Comparison studies:
a. Method comparison with predicate device:
The method comparison was previously established for the ADVIA Centaur in k031038. For 949 samples in the range of 0 to 1290 pg/mL, the relationship between the ADVIA Centaur BNP assay and a predicate device is described by the Passing & Bablok regression equation:
$$
\text{ADVIA Centaur BNP} = 0.74 \times (\text{Predicate Device}) - 0.60 \, \text{pg/mL}
$$
$$
\text{Correlation coefficient} \, (r) = 0.90
$$
For 730 samples in the range of $< 7.5$ to $4135 \, \text{pg/mL}$, the relationship between the ACS:180 BNP assay and the ADVIA Centaur BNP assay is described by the Passing & Bablok regression equation:
$$
\text{ACS:180 BNP} = 1.03 \times (\text{ADVIA Centaur BNP}) - 0.10 \, \text{pg/mL}
$$
$$
\text{Correlation coefficient} \, (r) = 1.00
$$
{5}
Page 6 of 9
b. Matrix comparison:
This test has been evaluated with plasma using EDTA as the anticoagulant. Serum, sodium citrate, lithium heparin and sodium fluoride sample tubes have also been tested and are not recommended. This information was provided in k031038.
3. Clinical studies:
a. Clinical sensitivity:
The clinical sensitivity and specificity of the ADVIA Centaur BNP assay using a decision threshold of 100 pg/mL for various age groups within each gender are presented in the following tables:
Clinical Sensitivity and Specificity vs. Age and Gender
| Males | | | | | |
| --- | --- | --- | --- | --- | --- |
| | Age Group | | | | |
| | <45 years | 45-54 years | 55-64 years | 65-74 years | 75 + years |
| % Sensitivity | 58.7 | 49.2 | 69.9 | 83.7 | 88.6 |
| 95% Confidence Interval | 40.4 - 71.0 | 36.4 - 62.1 | 61.0 - 77.9 | 75.1 - 90.2 | 80.9 - 93.9 |
| % Specificity | 100 | 100 | 99.5 | 96.8 | 94.6 |
| 95% Confidence Interval | 97.2 - 100 | 97.4 - 100 | 97.6 - 100 | 93.2 - 98.8 | 85.1 - 98.9 |
| Females | | | | | |
| --- | --- | --- | --- | --- | --- |
| | Age Group | | | | |
| | <45 years | 45-54 years | 55-64 years | 65-74 years | 75 + years |
| % Sensitivity | 45.5 | 56.3 | 60.4 | 68.9 | 87.2 |
| 95% Confidence Interval | 24.4 - 67.8 | 37.7 - 73.7 | 45.3 - 74.2 | 53.4 - 81.8 | 79.7 - 92.6 |
| % Specificity | 99.5 | 99.3 | 97.8 | 97.2 | 79.8 |
| 95% Confidence Interval | 97.1 - 100 | 96.4 - 100 | 94.4 - 99.4 | 93.5 - 99.0 | 69.9 - 87.6 |
b. Clinical specificity:
see above
c. Other clinical supportive data (when a and b are not applicable):
4. Clinical cut-off:
The decision threshold for diagnosing heart failure was determined based on the BNP level at the 95th percentile of the Reference Group. The most appropriate decision threshold for diagnosing heart failure apparent from these distributions is 100 pg/mL. This BNP value translates into a general specificity of the test of greater than 97 %.
The decision threshold for predicting survival in patients with myocardial infarction is 80 pg/mL. Retrospective analysis of BNP was performed on 438 patients with AMI from the ENTIRE-TIMI 23 (Thrombolysis in Myocardial Infarction) study, a multi-national trial conducted between February 2000 and
{6}
September 2001. The study design has been published previously. Patients were eligible for inclusion if they had an episode of ischemic discomfort of at least 30 minutes duration within the prior 6 hours, and exhibited positive ECG changes. The baseline BNP concentration was significantly higher in patients who died within 30 days (n = 15, 89 pg/mL, 25th - 75th: 40-192) compared to survivors (n = 423, 15 pg/mL, 25th - 75th: 8.8 - 32, p<0.0001). BNP levels greater than 80 pg/mL were associated with a substantially higher risk of death through 30 days of follow-up (17.4% vs. 1.8%, p<0.0001). The odds ratio for death within 30 days for patients with BNP levels greater than 80 pg/mL was 11.5. The odds ratio for death within 30 days for patients with BNP levels greater than 80 pg/mL, adjusted for age, history of hypertension, and prior angina, was 8.3 with a 95% confidence interval of 2.7 to 25.8. Patients with elevated BNP levels also had an increased risk of composite end points for death and heart failure combined (23.9% vs. 5.1%, p<0.0001). The odds ratio for death within 30 days for patients with BNP levels greater than 80 pg/mL was 5.8. The odds ratio for death or CHF within 30 days for patients with BNP levels greater than 80 pg/mL, adjusted for age, history of CHF, history of hypertension, and prior angina, was 3.6 with a 95% confidence interval of 1.5 to 8.8. Elevated levels of BNP at initial presentation are associated with an increased risk of mortality in patients with myocardial infarction.
# 5. Expected values/Reference range:
The expected results for the ADVIA Centaur BNP assay were previously established (k031038). These results were confirmed for the ACS:180 by analyzing 730 samples (see method comparison). The circulating BNP concentration was determined from 1521 individuals without heart failure (785 women and 736 men). This population included apparently healthy individuals and individuals with hypertension, diabetes, renal insufficiency, and chronic obstructive pulmonary disease. The descriptive statistics for BNP concentrations in the population without heart failure are shown in the following tables. These values are representative of the results obtained from clinical studies. Clinical studies indicate that BNP levels increase with age in the general population with the highest values seen in individuals greater than 75 years of age. In this subgroup of patients, age needs to be taken into consideration for accurate interpretation of test results.
| All | | | | | | |
| --- | --- | --- | --- | --- | --- | --- |
| | Age Group | | | | | |
| | All | <45 years | 45-54 years | 55-64 years | 65-74 years | 75 + years |
| Mean, pg/mL | 23.2 | 11.9 | 15.6 | 19.5 | 28.3 | 60.3 |
| SD, pg/mL | 32.5 | 12.9 | 15.9 | 22.6 | 25.4 | 73.0 |
| Median, pg/mL | 14.5 | 8.6 | 10.4 | 13.8 | 22.1 | 43.7 |
| 95thPercentile, pg/mL | 70.8 | 33.3 | 46.7 | 53.2 | 72.3 | 176 |
| % < 100 pg/mL | 97.4 | 99.7 | 99.7 | 98.8 | 97.0 | 85.5 |
| Minimum, pg/mL | <2 | <2 | <2 | <2 | <2 | <2 |
| Maximum, pg/mL | 576 | 128 | 119 | 286 | 164 | 576 |
| N | 1521 | 317 | 291 | 403 | 365 | 145 |
{7}
Page 8 of 9
| Males | Age Group | | | | | |
| --- | --- | --- | --- | --- | --- | --- |
| | All | <45 years | 45-54 years | 55-64 years | 65-74 years | 75 + years |
| Mean, pg/mL | 17.9 | 9.1 | 11.2 | 14.5 | 25.8 | 41.9 |
| SD, pg/mL | 22.9 | 9.4 | 11.8 | 13.9 | 25.1 | 48.8 |
| Median, pg/mL | 11.3 | 5.9 | 7.6 | 11.9 | 17.8 | 26.1 |
| 95th Percentile, pg/mL | 54.3 | 29.4 | 32.8 | 38.8 | 67.6 | 121 |
| % < 100 pg/mL | 98.6 | 100 | 100 | 99.5 | 96.8 | 94.6 |
| Minimum, pg/mL | <2 | <2 | <2 | <2 | <2 | <2 |
| Maximum, pg/mL | 250 | 56.6 | 88.9 | 132 | 151 | 250 |
| N | 736 | 129 | 140 | 223 | 188 | 56 |
| Females | Age Group | | | | | |
| --- | --- | --- | --- | --- | --- | --- |
| | All | <45 years | 45-54 years | 55-64 years | 65-74 years | 75 + years |
| Mean, pg/mL | 28.1 | 13.8 | 19.8 | 25.6 | 31.0 | 71.9 |
| SD, pg/mL | 38.8 | 14.6 | 18.0 | 29.0 | 25.5 | 82.9 |
| Median, pg/mL | 18.5 | 10.4 | 14.8 | 19.4 | 25.7 | 54.3 |
| 95th Percentile, pg/mL | 86.1 | 35.9 | 56.7 | 75.5 | 72.9 | 167 |
| % < 100 pg/mL | 96.3 | 99.5 | 99.3 | 97.8 | 97.1 | 79.8 |
| Minimum, pg/mL | <2 | <2 | <2 | <2 | <2 | <2 |
| Maximum, pg/mL | 576 | 128 | 119 | 286 | 164 | 576 |
| N | 785 | 188 | 151 | 180 | 177 | 89 |
## Patients with Heart Failure
To establish the expected results for the ADVIA Centaur BNP assay in individuals with heart failure, plasma samples were obtained from 722 patients diagnosed with heart failure (264 women and 458 men). The descriptive statistics for BNP concentrations in patients with heart failure are presented in the following tables. These values are representative of the results obtained from clinical studies. In addition, laboratories should be aware of their respective institution's current practice for the evaluation of heart failure.
| Heart Failure Population – All | | NYHA Functional Class | | | |
| --- | --- | --- | --- | --- | --- |
| | All | NYHA I | NYHA II | NYHA III | NYHA IV |
| Mean, pg/mL | 505 | 178 | 270 | 525 | 1134 |
| SD, pg/mL | 711 | 347 | 402 | 576 | 1141 |
| Median, pg/mL | 262 | 64.3 | 130 | 355 | 843 |
| 5th percentile, pg/mL | 10.8 | 1.6 | 5.4 | 21.1 | 109 |
| 95th percentile, pg/mL | 1873 | 772 | 999 | 1696 | 3157 |
| % ≥ 100 pg/mL | 72.6 | 43.1 | 58.7 | 82.0 | 95.8 |
| Minimum, pg/mL | <2 | <2 | <2 | <2 | 4.0 |
| Maximum, pg/mL | 6989 | 2310 | 3107 | 4052 | 6989 |
| N | 722 | 72 | 242 | 289 | 119 |
{8}
| Heart Failure Population - Males | | | NYHA Functional Class | | |
| --- | --- | --- | --- | --- | --- |
| | All | NYHA I | NYHA II | NYHA III | NYHA IV |
| Mean, pg/mL | 518 | 121 | 308 | 542 | 1214 |
| SD, pg/mL | 726 | 135 | 475 | 588 | 1200 |
| Median, pg/mL | 245 | 77.7 | 135 | 339 | 950 |
| 5th percentile, pg/mL | 10.7 | 3.9 | 4.4 | 23.2 | 71.5 |
| 95th percentile, pg/mL | 1946 | 400 | 1280 | 1852 | 3157 |
| % ≥ 100 pg/mL | 72.9 | 44.7 | 61.3 | 81.4 | 93.9 |
| Minimum, pg/mL | <2 | <2 | <2 | <2 | 33.7 |
| Maximum, pg/mL | 6989 | 552 | 3107 | 3503 | 6989 |
| N | 458 | 47 | 150 | 194 | 66 |
| Heart Failure Population - Females | | | NYHA Functional Class | | |
| --- | --- | --- | --- | --- | --- |
| | All | NYHA I | NYHA II | NYHA III | NYHA IV |
| Mean, pg/mL | 482 | 285 | 207 | 492 | 1034 |
| SD, pg/mL | 687 | 551 | 228 | 556 | 1068 |
| Median, pg/mL | 291 | 62.5 | 117 | 355 | 779 |
| 5th percentile, pg/mL | 11.0 | 0 | 9.5 | 15.9 | 115 |
| 95th percentile, pg/mL | 1575 | 1447 | 552 | 1518 | 2970 |
| % > 100 pg/mL | 72.0 | 40.0 | 54.3 | 83.2 | 98.1 |
| Minimum, pg/mL | <2 | <2 | <2 | 4.8 | 4.0 |
| Maximum, pg/mL | 5845 | 2310 | 1231 | 4052 | 5845 |
| N | 264 | 25 | 92 | 94 | 53 |
These results show that there is a relationship between the severity of the clinical signs and symptoms of heart failure and the median BNP concentrations of each NYHA functional class.
# N. Conclusion:
The submitted material in this premarket notification is complete and supports a substantial equivalence decision.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.