K040395 · Nihon Kohden America, Inc. · MHX · Jun 1, 2004 · Cardiovascular
Device Facts
Record ID
K040395
Device Name
CENTRAL TELEMETRY SYSTEM, MODEL WEP-4200A SERIES
Applicant
Nihon Kohden America, Inc.
Product Code
MHX · Cardiovascular
Decision Date
Jun 1, 2004
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 870.1025
Device Class
Class 2
Attributes
Pediatric
Indications for Use
The device is comprised of transmitters and a receiver communicated via WMTS radio frequency. The series of device are intended to monitor, display and record physiological data to provide cardiac and vital signs monitoring within a medical facility. The device is intended to produce a visual record of the electrical signals produced by the heart and monitor the electrocardiogram to generate audible and/or clookida organization produced by the novice is also intended to monitor patients' physiological data including heart rate, pulse rate, blood oxygen saturation (SpO2) noninvasive blood pressure (NIBP), body temperature respiratory rate and apnea. The device may generate an audible and/or visual alarm body temperature roopiratory no and limits. This device receives physiological signals via radio when a motor the transmitters. The device has the capability of communicating with external devices, such as personal computers. Vital sign data can be sent to a central monitor through the network via network card.
Device Story
Central telemetry system consisting of ZS-910P transmitter and WEP-4200A receiver; monitors ECG, SpO2, NIBP, temperature, respiration, and apnea. Inputs: physiological signals via electrodes, SpO2 probes, NIBP cuffs, and thermistors. Operation: signals transmitted via 608-614 MHz WMTS band to receiver; receiver processes data using multi-template arrhythmia detection and analysis; displays waveforms/numerics on color LCD touch-screen. Used by medical personnel in clinical settings. Output: real-time waveforms, numeric vital signs, and audible/visual alarms for arrhythmia or out-of-limit parameters. Benefits: continuous patient monitoring, automated arrhythmia detection, and centralized data management to support clinical decision-making.
Clinical Evidence
Bench testing only. No clinical data provided. Device compliance verified through electromagnetic, environmental, safety, and performance testing. Software functions and hardware operation confirmed via design validation. Complies with IEC 601-1 and 21 CFR Part 898.
Technological Characteristics
Multi-parameter monitor; color LCD touch-screen; WMTS (608-614 MHz) wireless connectivity; AC/battery power. Materials: standard medical-grade probes/thermistors. Software: multi-template arrhythmia detection, impedance-based respiration, pulse oximetry, NIBP, and temperature monitoring. Non-sterile.
Indications for Use
Indicated for cardiac and vital signs monitoring of adult, pediatric, and infant patients within a medical facility (ICU, CCU, recovery room, general ward).
Regulatory Classification
Identification
The arrhythmia detector and alarm device monitors an electrocardiogram and is designed to produce a visible or audible signal or alarm when atrial or ventricular arrhythmia, such as premature contraction or ventricular fibrillation, occurs.
Special Controls
*Classification.* Class II (special controls). The guidance document entitled “Class II Special Controls Guidance Document: Arrhythmia Detector and Alarm” will serve as the special control. See § 870.1 for the availability of this guidance document.
{0}------------------------------------------------
# SECTION 2 - 510(K) SUMMARY
Name and Address of Applicant . Nihon Kohden America, Inc. 90 Icon St. Foothill Ranch, Ca 92610
Date: 05/19/2004
# JUN = 1 2004
K040395
p. 1 of 3
Phone: (949) 580-1555 Dial 9 and then Ext. 4401 Fax: (949) 580-1550 Attn: Serrah Namini, Regulatory Affairs Assoc. Dir.
- Name of the device: WEP-4200A series Central Telemetry System .
- Trade or proprietary name: WEP-4200A series Central Telemetry System .
- The common or usual name: Central Telemetry System .
- . The classification name: Physiological Patient Monitor with Arrhytmia Detection and Alarms per 21 CFR Part 870.1025,
- The legally marketed equivalence: The predicate device is Nihon Kohden's Central . Telemetry system, WEP-8430A, 510K: K945944, dated 11/27/95. All features available with the new WEP-4200 are also available on Nihon Kohden BSM-4104A, 510K: K001693
- A description of the device: The system provides for monitoring ECG, ST levels, VPC rate, . arrhythmia events, SpO2, NIBP, temperature, respiration and apnea. The device will be offered with up to 5 multi-connectors and SpO2. The system is comprised of a receiver (WEP-4200A), and a transmitter (ZS-910P). WEP is a multi-parameter monitor consisting of a color LCD touch-screen to display waveforms and numerics of monitored parameters, multi-parameter receiver unit, visual alarm indicator, external communications port, and a removable battery pack for transmitter. Options include a thermal array recorder and a laser printer interface card.
- A summary of the technological characteristics of the device compared to the . predicate device:
The telemetry system is intended to monitor, record and display physiological data including ECG, blood oxygen saturation (SpO2), noninvasive blood pressure (NIBP), body temperature, respiration and apnea. These features are currently available in the legally marketed Nihon Kohden predicate devices, such as WEP-8430 and BSM-4100A series.
The device acquires ECG signals through electrodes placed on the patient. The ECG signals are then amplified, filtered and displayed on a color LCD. Heart rate value is determined from the ECG signal or the pulse waveform and the device may sound an alarm when the heart rate falls outside preset upper and lower limits. These features are currently available in the legally marketed Nihon Kohden predicate devices, such as WEP-8430 and BSM-4100A series.
The device monitors the ECG to detect arrhythmias. User selectable options are provided to control the alarm, storage and recording of arrhythmia events. The device is capable of generating an audible and visual alarm when an arrhythmia event is detected. The device incorporates the same multi-template arrhythmia detection and analysis of the predicate device WEP-8430 and BSM-4100A series.
The device includes a pulse oximeter, which measures the blood oxygen saturation using finger, foot, ear, and multi-site probes and displays a numeric value and a pulse plethysmograph waveform. When the SpO2 value falls outside preset upper and lower limits, the alarm is activated. The device uses the same control software as the predicate device.
The device includes a noninvasive blood pressure measurement system, which measures and displays the systolic, diastolic, and mean pressures through the use of pressure cuffs placed on the arm or leg. The NIBP measurement interval may be set to manual, continuous or selected time intervals. When NIBP measurements are set to time intervals, the user may also select an additional trigger method called PWTT as offered with the predicate device. An alarm may be generated when the systolic, diastolic or mean pressure values fall outside preset upper and lower limits.
{1}------------------------------------------------
K040395
p 2 of 3
The device measures temperature through the use of the same commercially available temperature thermistors as the predicate device WEP-8430. Temperature thermistors measures temperature through the use of the same commercially available temperature thermistors as the predicate device. The control software for the temperature function is the same in both the new device and the predicate.
The device measures respiration rate through impedance (picked up through the ECG electrodes), through the sensor or through the use of a thermistor respiration pickup transducer. Respiratory waveform is displayed on the monitor and respiration rate is calculated and displayed on the monitor. The device may generate an audible and visual alarm when the measured respiration value falls outside preset upper and lower limits. The software algorithm that determines the respiration rate count is the same in both the new device and the predicate.
A thermal array recorder is available as an option to allow the printing of waveforms and alphanumeric data. Automatic alarm recording, manual recording, and timed recording are available.
All system functions are operable from AC or battery power. A battery charging circuit is incorporated into each device.
#### A summary of the technological differences between the device compared to the . predicate devices:
The new device is an updated version of the WEP-8430A, eight patient telemetry, based on the same technology platform as the BSM-4104A bedside monitor. It uses features and capabilities from the predicate devices.
The transmitting frequency of the new device is in the 608-614 MHz (WMTS) frequency band. The predicate device transmitted in the UHF frequency band. This change was made to meet FCC and FDA recommendations for WMTS.
The new device has more storage memory, which is used to expand the storage of arrhythmia recall files and full disclosure. The basic functions of the arrhythmia recall and full disclosure remain unchanged.
#### Indications for Use: .
The device is intended to monitor, display and record physiological data to provide cardiac and vital signs monitoring. The device is intended to present a visual record of the electrical signals produced by the heart and monitor the electrocardiogram to generate visible and/or audible alarms when an arrhythmia exists. The device is also intended to monitor heart rate, pulse rate, blood oxygen saturation (SpO2), noninvasive blood pressure (NIBP), body temperature, respiration rate and apnea. The device may generate an audible and/or visual alarm when a measured rate falls outside preset limits. The system communicates in the new WMTS band as required by FCC. The device will be available for use by medical personnel on all patient populations within a medical facility, including ICU, CCU, recovery room and general ward.
### Non-clinical tests:
The device does not directly contact patients. Accessories that contact patients, such as probes and thermistors, are the same accessories as used with other legally marketed products or are comprised of the same component materials as the predicate accessories. Therefore, good laboratory practice studies were not required per 21 CFR Part 58. The device is in compliance with the FDA Special Controls for arrhythmia detector and alarm system. The device complies with IEC 601-1 sub-clause 56.3(c) implemented by 21 CFR Part 898 Performance Standard for Electrode Lead Wires and Patient Cables. The device is also in compliance with applicable sections of IEC standard as listed in this application.
### The device is not sterile.
The device was subjected to electromagnetic, environmental, safety and performance testing procedures. These tests verified the operation of the device. Design validation confirmed the operation of the software functions as well as hardware of the device.
{2}------------------------------------------------
p. 3 . F 3 Therefore, Nihon Kohden believes that the device is substantially equivalent to the Nihon Kohden
predicate devices as indicated.
:
:
1-040395
{3}------------------------------------------------
Image /page/3/Picture/11 description: The image shows the seal for the Department of Health & Human Services - USA. The seal is circular and contains the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is an abstract image of an eagle.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
JUN - 1 2004 ___
Nihon Kohden America c/o Ms. Serrah Namini RA Assoc. Dir. 90 Icon Street Foothill Ranch, CA 92610
Re: K040395
K040579
Trade Name: WEP-4200A Series Central Telemetry System Regulation Number: 21 CFR 870.1025 Regulation Name: Arrhythmia Detector and Alarm Regulatory Class: II (two) Product Code: MHX Dated: May 25, 2004 Received: May 25, 2004
# Dear Ms. Namini:
We have reviewed your Section 510(k) premarket notification of intent to market the device we have reviewed your Section 310(x) premained is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate for use stated in the encrosule) to regarly manation province Amendments, or to commence prior to May 20, 1976, are exactions of the Federal Food, Drug, devices that have occh rechasined in assere approval of a premarket approval application (PMA). and Cosment ACT (Act) that do not require approvine controls provisions of the Act. The You may, therefore, market the devices, sologie requirements for annual registration, listing of general controls provisions of the fiel labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it If your device is classified (500 a0070) ins . Existing major regulations affecting your device can thay be subject to subli dualities, Title 21, Parts 800 to 898. In addition, FDA may be found in the Outs acements concerning your device in the Federal Register.
{4}------------------------------------------------
Page 2 - Ms. Serrah Namini
Please be advised that FDA's issuance of a substantial equivalence determination does not mean I lease of advised that I DTT Unstance or our deviee complies with other requirements of the Act mar I Dri has made a sond regulations administered by other Federal agencies. You must or any I catal statutes and equirements, including, but not limited to: registration and listing (21 Comply with an the 110 - 21 CFR Part 801); good manufacturing practice requirements as set OI IT at 0077, idocime (21 OFF earlation (21 CFR Part 820); and if applicable, the electronic forth in the quality byevelisions (Sections 531-542 of the Act); 21 CFR 1000-1050. product radiation comi or to begin marketing your device as described in your Section 510(k) I ms letter will anow you'le cogni maining of substantial equivalence of your device to a legally premailed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please If you desire specific acriously 3 at (301) 594-4648. Also, please note the regulation entitled, Connect the Office of Comments to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small other general micrinational and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html
Sincerely yours,
Neil R.P. Ogden
Bram D. Zuckerman, M.D. ナレ Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
# NIHON KOHDEN AMERICA, INC.
510(k) NOTIFICATION Central Telemetry System
G. Indications for Use Statement
510(k) Number (if known): KOYO39S S
Device Name: WEP-4200A Series Telemetry system
## Indications for Use:
The device is comprised of transmitters and a receiver communicated via WMTS radio frequency. The series of device are intended to monitor, display and record physiological data to provide cardiac and vital signs monitoring within a medical facility. The device is intended to produce a visual record of the electrical signals produced by the heart and monitor the electrocardiogram to generate audible and/or clookida organization produced by the novice is also intended to monitor patients' physiological data including heart rate, pulse rate, blood oxygen saturation (SpO2) noninvasive blood pressure (NIBP), body temperature respiratory rate and apnea. The device may generate an audible and/or visual alarm body temperature roopiratory no and limits. This device receives physiological signals via radio when a motor the transmitters. The device has the capability of communicating with external devices, such as personal computers. Vital sign data can be sent to a central monitor through the network via network card.
The device will be available for use by medical personnel on patients within a medical facility including adults, children and infants. n
for BDZ
| (Division Sign-Off) | |
|------------------------------------|---------|
| Division of Cardiovascular Devices | |
| 510(k) Number | K040398 |
| Prescription Use (Part 21 CFR 801 Subpart D) | X |
|----------------------------------------------|---------------------------------------------|
| AND/OR | Over-The-Counter Use (21 CFR 807 Subpart C) |
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Page 1 of 1
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.