RANDOX AMPHETAMINE CLASS ASSAY
Device Facts
| Record ID | K040350 |
|---|---|
| Device Name | RANDOX AMPHETAMINE CLASS ASSAY |
| Applicant | Randox Laboratories, Ltd. |
| Product Code | DKZ · Clinical Toxicology |
| Decision Date | May 12, 2004 |
| Decision | SESE |
| Submission Type | Traditional |
| Regulation | 21 CFR 862.3100 |
| Device Class | Class 2 |
Indications for Use
The evidence® Amphetamine assays is an in vitro diagnostic test for the qualitative determination of Amphetamine Class of compounds (including amphetamines and methamphetamine), central nervous system stimulating drugs, in human urine. This is a competitive immunoassay. A primary cut-off of 1000 ng/ml has been established in accordance with SAMHSA recommendations. The evidence® Drugs of Abuse Calibrators are liquid Calibrators containing benzoylecgonine, amphetamine and methamphetamine. There are 9 levels of calibrator. They have been developed for use in calibration of the evidence® system. This assay is for use only on the automated evidence® Analyzer. Note: This test provides only a preliminary analytical result. A more specific alternative chemical method must be used to obtain a confirmed analytical result. Gas Chromatography/ Mass Spectrophotometry (GC/MS) is the preferred confirmatory method. The Amphetamine Class Assay and Calibrators must only be used by suitably qualified laboratory personnel under appropriate laboratory conditions.
Device Story
The evidence® Amphetamine Assay is an in vitro diagnostic test for use on the evidence® analyzer. It detects amphetamine compounds in urine samples at a 1000ng/ml cutoff. The system utilizes liquid calibrators (evidence® Drugs of Abuse Calibrators) to standardize the assay. Operated by qualified laboratory personnel in a clinical laboratory setting, the device provides qualitative results to assist clinicians in diagnosing and treating amphetamine use or overdose. Results are preliminary and require confirmation via GC/MS. The device benefits patients by providing rapid screening for amphetamine exposure.
Clinical Evidence
No clinical studies performed. Bench testing only. Analytical performance validated via precision/reproducibility studies (NCCLS EP5-T2), interference screening (NCCLS EP7-A), and method comparison against GC/MS and predicate (n=1336 samples). Sensitivity and specificity established using SAMHSA-recommended 1000 ng/mL cut-off.
Technological Characteristics
In vitro diagnostic assay for use on the evidence® analyzer. Employs liquid calibrators containing benzoylecgonine, amphetamine, and methamphetamine (9 levels). Qualitative detection method based on 1000ng/ml cutoff. Designed for professional laboratory use.
Indications for Use
Indicated for qualitative determination of amphetamine and methamphetamine in human urine to aid in diagnosis and treatment of amphetamine use or overdose. For use by qualified laboratory personnel. Preliminary result only; requires GC/MS confirmation.
Regulatory Classification
Identification
An amphetamine test system is a device intended to measure amphetamine, a central nervous system stimulating drug, in plasma and urine. Measurements obtained by this device are used in the diagnosis and treatment of amphetamine use or overdose and in monitoring levels of amphetamine to ensure appropriate therapy.
Special Controls
*Classification.* Class II (special controls). An amphetamine test system is not exempt if it is intended for any use other than employment or insurance testing or is intended for Federal drug testing programs. The device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 862.9, provided the test system is intended for employment and insurance testing and includes a statement in the labeling that the device is intended solely for use in employment and insurance testing, and does not include devices intended for Federal drug testing programs (*e.g.,* programs run by the Substance Abuse and Mental Health Services Administration (SAMHSA), the Department of Transportation (DOT), and the U.S. military).