K040340 · Personal Products Company, Div. of Mcneil-Ppc, Inc. · NUC · Apr 28, 2004 · Obstetrics/Gynecology
Device Facts
Record ID
K040340
Device Name
K-Y BRAND WARMING ULTRAGEL PERSONAL LUBRICANT
Applicant
Personal Products Company, Div. of Mcneil-Ppc, Inc.
Product Code
NUC · Obstetrics/Gynecology
Decision Date
Apr 28, 2004
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 884.5300
Device Class
Class 2
Indications for Use
K-Y® Brand WARMING UltraGEL is intended as personal lubricant to be used with or without a condom. The lubricous nature of this product helps to supplement the body's own natural lubricating fluids, thereby relieving friction to help enhance the ease and comfort of intimate sexual activity. This lubricant may be safely applied to vaginal, anal or penile tissues for purpose of lubrication, and moisturization and is compatible with latex condoms. K-Y® Brand WARMING UltraGEL has the additional benefit of imparting a warming sensation when applied to the genital area.
Device Story
K-Y® Brand WARMING UltraGEL is a non-sterile, water-soluble, clear, non-staining, non-greasy gel. Applied topically to vaginal, anal, or penile tissues by the user to reduce friction and enhance comfort during sexual activity. Imparts a gentle warming sensation upon contact with genital skin. Compatible with latex condoms. Designed for over-the-counter use by consumers. Safety established via biocompatibility testing (ISO 10993) and clinical studies evaluating sensitization and consumer perception of warmth and intimacy enhancement. No complex electronics or software involved.
Clinical Evidence
Clinical evidence includes a Human Repeated Insult Patch Test (Modified Draize Procedure) showing no contact sensitization. A consumer perception study (n=not specified) reported 80% excellent/very good rating for 'Warms on Contact' and 95% for 'Does not feel cold'. An in-home consumer use study (245 responses) reported 91.67% positive for 'Warms on Contact' and 85.01% for 'Enhances Intimacy'. Safety was monitored via gynecological examinations pre- and post-use; no serious adverse events were reported, and the product was determined not to cause irritation.
Technological Characteristics
Water-soluble gel formulation; proprietary ingredients; non-sterile. Imparts warming sensation via chemical interaction upon application. Compatible with latex condoms per ASTM D 3492. No electronic components, software, or energy sources.
Indications for Use
Indicated for use as a personal lubricant for vaginal, penile, and condom application during sexual intimacy to supplement natural lubrication and relieve friction. Suitable for over-the-counter use.
Regulatory Classification
Identification
A condom is a sheath which completely covers the penis with a closely fitting membrane. The condom is used for contraceptive and for prophylactic purposes (preventing transmission of sexually transmitted infections). The device may also be used to collect semen to aid in the diagnosis of infertility.
Special Controls
*Classification.* (1) Class II (special controls) for condoms made of materials other than natural rubber latex, including natural membrane (skin) or synthetic.(2) Class II (special controls) for natural rubber latex condoms. The guidance document entitled “Class II Special Controls Guidance Document: Labeling for Natural Rubber Latex Condoms Classified Under 21 CFR 884.5300” will serve as the special control. See § 884.1(e) for the availability of this guidance document.
Predicate Devices
K-Y® Brand Warming LIQUID Personal Lubricant
Submission Summary (Full Text)
{0}------------------------------------------------
APR 2 8 2004
### 510(k) SUMMARY
KO4034D
| Submitter | Personal Products Company Division of McNeil-PPC Inc.<br>199 Grandview Road<br>Skillman, New Jersey 08558-9418 |
|---------------------|----------------------------------------------------------------------------------------------------------------|
| Contact Person | Marylou P. Carlson<br>Mgr. Regulatory Affairs<br>(908) 904-3709 phone (908) 904-3748 fax |
| Date Prepared | February 02, 2004 |
| Proprietary Name | K-Y® Brand<br>WARMING UltraGEL Personal Lubricant |
| Common Name | Personal Lubricant |
| Classification Name | Condom: 21CFR § 884.5300 Product Code 85HIS<br>Patient Lubricant: 21CFR 880.6375 Product Code MMS |
| Predicate Device | K-Y® Brand Warming LIQUID Personal Lubricant |
# Description of Device
K-Y® Brand WARMING UltraGEL is a non-sterile, clear, non-staining, non-greasy, water soluble gel for use as a personal lubricant. This product imparts a gentle warming sensation when applied to the genitalia. This product was designed to meet a customer need for an intimate lubricant that does not feel cold when applied. K-Y® Brand Warming UltraGEL can reduce friction during sexual intercourse thereby enhancing sexual intimacy. It is compatible with latex condoms as demonstrated in Condom Compatibility Testing conducted according the standards as defined by ASTM D 3492. K-Y® Brand WARMING UltraGel is not a contraceptive nor spermicide.
# Intended Use
K-Y® Brand WARMING UltraGEL is intended as personal lubricant to be used with or without a condom.
The lubricous nature of this product helps to supplement the body's own natural lubricating fluids, thereby relieving friction to help enhance the ease and comfort of intimate sexual activity. This lubricant may be safely applied to vaginal, anal or penile tissues for purpose of lubrication, and moisturization and is compatible with latex condoms. K-Y® Brand WARMING UltraGEL has the additional benefit of imparting a warming sensation when applied to the genital area.
# Regulatory Status
Per 21CFR, 880.6375, Patient lubricant is defined as a Class I medical device intended for medical purposes that is used to lubricate a body orifice to facilitate entry of a diagnostic or therapeutic device. Patient lubricants are not exempt from 510(k) clearance. Additionally when used as an accessory to a condom, (a Class II medical device) the lubricant is considered, by FDA, as a Class II Medical Device requiring 510(k) clearance.
{1}------------------------------------------------
# 510(k) SUMMARY (continued)
# Technological Characteristics
The K-Y® Brand WARMING UltraGEL Personal Lubricant formula is proprietary. The product, has no exceptional technological characteristics and consists mainly of safe watersoluble GRAS status ingredients somewhat similar to other K-Y® Brand personal lubricants currently on the market.
#### Substantial Equivalence
K-Y® Brand WARMING UltraGEL Personal Lubricant has been shown, in laboratory tests, to be substantially equivalent to the currently marketed K-Y® Brand Warming LIQUID Personal Lubricant. Both devices have the same intended use with a variation in formula ingredients. The gentle warming technology is the special feature of the K-Y® Brand WARMING lubricant products.
### Preclinical Testing of Formulation
Biocompatibility safety studies according to International Standard ISO 10993 and General Program Memorandum G95-1 on K-Y® Brand WARMING UltraGEL were conducted by an outside laboratory, in compliance with Good Laboratory Practices (GLPs). Results form these studies, demonstrated that K-Y® Brand WARMING UltraGEL was not considered to be a contact sensitizing agent, nor was it associated with systemic toxicity.
### Human Clinical Testing
In a Human Repeated Insult Patch Test (Modified Draize Procedure), this product was compared to the currently marketed K-Y® Warming LIQUID for its potential for contact sensitization. Under the conditions of this test no evidence of contact sensitization was elicited.
A Consumer Perception Study evaluated both male and female subjects' experience of warmth with a single application of the product to their genitals during an on-site visit. Overall 80.00% of the participants rated the product as Excellent, Very Good for "Warms on Contact" and 95.00% for "Does not feel cold when applied" There were no adverse events observed or reported during the course of this study.
An In-Home Consumer Use Study was conducted to evaluate both consumer perception of warmth during sexual activity as well as the tolerability of the product through vulvo/vaginal speculum examination prior and post product use. Consented female subjects received gynecological examinations at baseline and following the last coital episode. The study was conducted in compliance with 21CFR Part 812 for Investigational Device Exemption and 21CFR Parts 50 and 56. Efficacy results for this study concluded that in 245 reported responses, 91.67% were positive for "Warms on Contact" and 85.01% were positive for "Enhances Intimacy" and 8.6% responded positively to "experienced discomfort". Additionally, after two weeks of home use, (with a minimal of two sexual intercourse encounters) there were no serious adverse events reported. Gynecological examinations detected only one instance of mottled irregular erythema of the inner thigh area at baseline, which was not present at the return visit. It was concluded that the product did not cause irritation as determined by final gynecological examination.
Preclinical and Clinical testing have provided scientific evidence that this product is safe for its intended use.
{2}------------------------------------------------
Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" around the perimeter. Inside the circle is a stylized symbol that resembles a human figure embracing another, possibly representing care and support.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
FEB 2 4 2014
Ms. Marylou (Panico) Carlson Manager, Regulatory Affairs Personal Products Company Division of McNeil-PPC, Inc. 199 Grandview Road SKILLMAN NJ 08558
Re: K040340
> Trade/Device Name: K-Y® Brand Warming UltraGel Personal Lubricant Regulation Number: 21 CFR §884.5300 Regulation Name: Condom Regulatory Class: II Product Code: NUC Dated (Date on orig SE ltr): February 6, 2004 Received (Date on orig SE ltr): February 11, 2004
Dear Ms. Carlson:
This letter corrects our substantially equivalent letter of April 28, 2004.
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be
{3}------------------------------------------------
Page 2 -
found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638 2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office
of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers. International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
**Benjamin R.Fisher -S**
Benjamin R. Fisher, Ph.D. Director Division of Reproductive, Gastro-Renal, and Urological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
# Indications for Use
510(k) Number (if known): K040340
Device Name: K-Y® Brand Warming Ultra Gel
Indications For Use:
Personal Lubricant For vaginal/penile and condom application during sexual intimacy
Prescription Use (Part 21 CFR 801 Subpart D)
AND/OR
Over-The-Counter Use _X (21 CFR 807 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
David H. Ingram
(Division Sign-Off) Division of Reproductive, Abdominal, and Radiological Devic 510(k) Number
Page 1 of ____________________________________________________________________________________________________________________________________________________________________
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.