OPUSDUO; OPUSDUO E; OPUSDUO EC; OPUS 20; OPUS SPECTRUM
K040270 · Lumenis, Ltd. · GEX · Dec 16, 2004 · General, Plastic Surgery
Device Facts
Record ID
K040270
Device Name
OPUSDUO; OPUSDUO E; OPUSDUO EC; OPUS 20; OPUS SPECTRUM
Applicant
Lumenis, Ltd.
Product Code
GEX · General, Plastic Surgery
Decision Date
Dec 16, 2004
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 878.4810
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The OpusDent Family of Er:YAG and CO2/Er:YAG Combination Laser Products (and the delivery accessories that are used with them to deliver laser energy) are intended for use in dental procedures performed in oral and maxillofacial surgery and dentistry.
Device Story
OpusDent Family of Dental Laser Systems consists of a console with control electronics/software, delivery accessories, and hand-pieces. Device utilizes Er:YAG and CO2 laser energy to perform incision, excision, cutting, ablation, vaporization, and coagulation of oral hard and soft tissues. Operated by dental professionals in clinical settings. Laser energy delivered via hand-piece to target site; clinician controls parameters via console panel. Output allows for precise tissue removal, bone contouring, and periodontal/endodontal treatment. Benefits include minimally invasive surgical access, improved hemostasis, and efficient tissue management compared to traditional mechanical instruments.
Clinical Evidence
No clinical data provided. Substantial equivalence is based on bench testing, comparison of technical characteristics, and established indications for use consistent with predicate devices.
Technological Characteristics
System comprises laser console, delivery accessories, and hand-pieces. Energy source: Er:YAG and CO2 lasers. Function: surgical laser instrument for hard/soft tissue. Connectivity: standalone console with integrated control electronics and software. Sterilization: delivery accessories are typically reusable/sterilizable. Software: embedded control system for laser parameters.
Indications for Use
Indicated for dental patients requiring hard tissue (tooth/bone) procedures, soft tissue surgery, and endodontal therapy. Hard tissue: caries removal, cavity prep, enamel etching/plasty, bone cutting/resection, apicoectomy, osseous crown lengthening, osteotomy. Soft tissue: biopsies, abscess drainage, gingivectomy/plasty, frenectomy, implant recovery, lesion removal, sulcular debridement. Endodontal: root canal access, pulpotomy, pulp extirpation, root canal debridement/prep. Contraindications not specified.
Regulatory Classification
Identification
(1) A carbon dioxide laser for use in general surgery and in dermatology is a laser device intended to cut, destroy, or remove tissue by light energy emitted by carbon dioxide.(2) An argon laser for use in dermatology is a laser device intended to destroy or coagulate tissue by light energy emitted by argon.
Predicate Devices
OpusDent Duo, OpusDent 20, and OpusDent Spectra (K021508)
Stryker Total Performance System (TPS) (K032117 and K991696)
Hoya Conbio VersaWave Dental Er:YAG Laser System (K041710)
Submission Summary (Full Text)
{0}------------------------------------------------
# DEC 1 € 2004
| A. 510(k) Summary | K040270 |
|-------------------------------|-----------------------------------------------------------------|
| General Information | |
| Submitter: | Lumenis, Inc.<br>2400 Condensa Street<br>Santa Clara, CA 950511 |
| Contact Person: | Karen L Baker/ Martha Murari, PhD |
| Summary Preparation Date: | December 2004 |
| | |
| Names | |
| Device Names: | OpusDent Family of Dental Laser Systems |
| Primary Classification Names: | Surgical Powered Laser Instrument |
#### III. Predicate Devices
I.
II.
and the contraction of the country
- OpusDent Duo, OpusDent 20, and OpusDent Spectra (K021508) .
- Stryker Total Performance System (TPS) (K032117 and K991696) .
- Hoya Conbio VersaWave Dental Er:YAG Laser System (K041710) .
### IV. Product Description
The OpusDent Family of Dental Laser Systems is comprised of the following main components:
- A light/ laser system console (including software and control electronics) with a . control and display panel;
- Delivery devices; and .
- . One or more attached hand-piece(s).
# V. Intended Use and Indications for Usc
Intended Use: The OpusDent Family of Er:YAG and CO2/Er:YAG Combination Laser Products (and the delivery accessories that are used with them to deliver laser energy) are intended for use in dental procedures performed in oral and maxillofacial surgery and dentistry.
{1}------------------------------------------------
#### VI. Rationale for Substantial Equivalence
The OpusDent Family of Er:YAG and CO2/Er:YAG Combination Laser Products (and the delivery accessories that are used with them to deliver laser energy) share the same or similar indications for use, device operation, overall technical and functional capabilities, and therefore are substantially equivalent for use in general and plastic surgery, and dentistry for surgical applications to the predicate OpusDent Duo, OpusDent 20, and OpusDent Spectra (K021508), the Stryker Total Performance System (TPS) (K03217 and K991696), and the Hoya Conbio VersaWave Dental (173) (K63217)
(K041710) (K041710).
# VII. Safety and Effectiveness Information
Review of indications for use and technical characteristics and data provided were used to support the safety and effectiveness and substantial equivalence of the OpusDent Family of Er YAG and CO2/Er.YAG Combination Laser Products for use in specific applications in the medical specialties of general and plastic surgery, and dentistry.
# VIII. Conclusion
The OpusDent Family of Er: YAG and CO2/Er: YAG Combination Laser Products (and the delivery accessories that are used with them to deliver laser energy) was found to be Spectre, (K031509) the Stephent OpusDent Duo, OpusDent 20, and OpusDent 20, and OpusDent Spectra (KO21508), the Stryke Total Performance System (TPS) (KO32117 and K991696), and the Hoya Conbio VersaWave Dental Er: YAG Laser System (K041710).
The OpusDent Family of Er: YAG and CO2/Er: YAG Combination Laser Products (and the delivery accessories that are used with them to deliver laser energy) shares the same indications for use, and similar design features, and functional features, and thus are substantially equivalent to the predicate devices.
{2}------------------------------------------------
Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized eagle with three stripes representing the agency's mission. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular pattern around the eagle. The logo is black and white.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
DEC 1 6 2004
Lumenis, Inc. c/o Ms. Ann Worden Consultant 3637 Bernal Avenue Pleasanton, California 94566
Re: K040270 Trade/Device Name: Opus Dent Family of Dental Lasers Regulation Number: 21 CFR 878.4810 Regulation Name: Laser surgical instrument for use in general and plastic surgery and in dermatology Regulatory Class: II Product Code: GEX Dated: November 16, 2004 Received: November 17, 2004
Dear Ms. Worden:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{3}------------------------------------------------
Page 2 - Ms. Ann Worden
This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (240) 276-0115. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html
Sincerely yours,
Miriam C Provost
Celia M. Witten, Ph.D., M.D. Director Division of General, Restorative and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
510(k) Number (if known): K040270
Device Name: OpusDent Family of Dental Laser Systems
Indications for Use:
The OpusDent Dental Laser System is intended to aid during dental procedures performed in oral and maxillofacial surgery and dentistry.
The Er: YAG laser component is indicated for a variety of hard tissue (tooth and bone) applications, and for the incision, excision, cutting, ablation, vaporization, and coagulation of soft tissue in oral and maxillofacial surgery and dentisty.
This includes the following:
Hard Tissue Indications of Er:YAG Laser Energy
- ♥ Caries removal
- . Cavity preparation
- Enamel etching �
- Enameloplasty, excavation of pits and fissures for placement of sealant .
Bone Indications of Er: Y AG Laser Energy
- Contact and non-contact cutting, shaving, contouring, and resection of oral osseous tissue . (bone)
- . Apicoectomy - amputation of the root end
- Osseous crown lengthening .
- Cutting bone to prepare a window access to the apex (apices) of the root(s) .
- Osseoplasty �
- Osteotomy .
*** Continued on the Following Page ***
Prescription Use > (Part 21 CFR 801 Subpart D)
AND/OR
Over-The-Counter Use (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Miriam C. Provoost
Division of General, Restorative, and Neurological Devices
Page_1 of 3
| LUMENIS, INC. | |
|-----------------------------------|----------------------------|
| MODIFICATION TO ER: 510(k) Number | K040270 |
| | MBINATION PRODUCT FAMILIES |
243 Page Revised December 2, 2004
{5}------------------------------------------------
510(k) Number (if known): K040270
Device Name: OpusDent Family of Dental Laser Systems
Indications for Use: *** Continued from Previous Page ***
Soft Tissue and Periodontal Indications of Er: YAG Laser Energy
- . Excisional and incisional biopsies
- Exposure of unerrupted teeth .
- Incision and drainage of abscesses .
- Gingival incision and excision .
- t Gingivoplasties
- Gingivectomies, Gingivectomy in case of hyperplasias of the gingival or excision of t hyperplasias
- Gingival troughing for crown impressions ●
- Hemostasis ●
- Implant recovery .
- Frenectomies and frenotomies .
- � Fibromatosis (fibroma removal)
- Benign and malignant lesion removal .
- � Operculectomy
- . Oral papillectomies
- Reduction of gingival hypertrophy ●
- . Soft tissue crown lengthening
- Preprosthetic surgery; flabby alveolar ridge, vestibuloplasty, exposure of implants, � hyperplasia, epulides, papilomas, fibromatoses, benign growths
- . Vestibuloplasty
- Sucular debridement (removal of diseased or inflamed soft tissue in the periodontal pocket to . improve clinical indices including gingival index, gingival bleeding index, probe depth, attachment loss and tooth mobility)
*** Continued on the Following Page ***
Prescription Use > (Part 21 CFR 801 Subpart D)
AND/OR
Over-The-Counter Use (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Page 2 of 3
LUMENIS, INC. MODIFICATION TO ER: YAG AND CO₂/ER:Y AG COMBINATION PRODUCT FAMILIES
244 Page Revised December 2, 2004
{6}------------------------------------------------
510(k) Number (if known): K040270
Device Name: OpusDent Family of Dental Laser Systems
Indications for Use: *** Continued from Previous Page ***
Endodontal Applications of Er: YAG Laser Energy:
- Tooth preparation to obtain access to root canal .
- Pulpotomy, Pulpotomy as an adjunct to root canal therapy t
- . Pulp extirpation
- Root canal debridement and cleaning .
- . Root canal preparation including enlargement
The CO2 laser component is indicated for the incision, excision, cutting, ablation, vaporization, and coagulation of soft tissue in a variety of soft tissue applications, and in oral and maxillofacial surgery and dentistry.
Periodontal Applications of CO2 Laser Energy:
- Removal of diseased or inflamed soft tissue in the periodontal pocket (sulcular debridement) �
- Biopsies .
- Frenectomy, Frenum release .
- . Gingivectomy
- Gingivoplasty
- Papillectomy #
- . Vestibuloplasty
- Hyperplasia ●
- Operculectorny �
- Drainage (abscess) .
- Flap surgery �
- Fibroma (nonmalignant tumor; mucosa, tongue) .
- Epulis (tumor of the gum) ●
- Aphthous ulcers ●
- Removal of soft tissue, cysts, and tumors .
Prescription Use (Part 21 CFR 801 Subpart D)
AND/OR
Over-The-Counter Use (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Page 3 of
LUMENIS, INC. MODIFICATION TO ER: YAG AND CO₂/ER: Y AG COMBINATION PRODUCT FAMILIES
245 Page Revised December 2, 2004
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.