GE LOGIQ 9 DIAGNOSTIC ULTRASOUND SYSTEM, MODEL 2404587
K040251 · Ge Med. Systems Ultrasound & Prim. Care Diag.,Llc · IYN · Feb 20, 2004 · Radiology
Device Facts
Record ID
K040251
Device Name
GE LOGIQ 9 DIAGNOSTIC ULTRASOUND SYSTEM, MODEL 2404587
Applicant
Ge Med. Systems Ultrasound & Prim. Care Diag.,Llc
Product Code
IYN · Radiology
Decision Date
Feb 20, 2004
Decision
SESE
Submission Type
Special
Regulation
21 CFR 892.1550
Device Class
Class 2
Attributes
Pediatric
Indications for Use
The device is intended for use by a qualified physician for ultrasound evaluation of Fetal: Abdominal: Pediatric: Small Organ (breast, testes, thyroid); Neonatal Cephalic; Adult Cephalic; Cardiac (adult and pediatric); Peripheral Vascular; Musculo-skeletal Conventional and Superficial; Urology (including prostate); Transvaginal; and Intraoperative (abdominal, thoracic, vascular and neurosurgical).
Device Story
GE LOGIQ 9 BT04 is a mobile, general-purpose diagnostic ultrasound system; provides digital acquisition, processing, and display of ultrasound images. System includes console, keyboard, specialized controls, and CRT/LCD display. Operates via pulsed echo and Doppler imaging principles. Used by qualified physicians in clinical settings for anatomical evaluation and fluid flow analysis. Modification adds new probe options (4D3C, 4D10L, 4DE7C, P8D) and improves user interface/productivity. Output displayed to clinician for diagnostic decision-making. Benefits include enhanced diagnostic capabilities for various clinical applications.
Clinical Evidence
No clinical data required. Device safety and effectiveness supported by non-clinical bench testing, including acoustic output measurements, biocompatibility, cleaning/disinfection validation, and thermal/electrical/mechanical safety testing per applicable standards.
Technological Characteristics
Mobile console (64cm x 90cm x 140-160cm). Modalities: B-mode, M-mode, PW/CW Doppler, Color/Power Doppler, Harmonic Imaging, Coded Pulse. Connectivity: Digital acquisition/display. Transducers: 4D3C, 4D10L, 4DE7C, P8D. Conforms to 21 CFR 892.1550/1560/1570. Quality systems: 21 CFR 820, ISO 9001:2000, ISO 13485.
Indications for Use
Indicated for diagnostic ultrasound imaging or fluid flow analysis in fetal, abdominal, pediatric, small organ, neonatal/adult cephalic, cardiac, peripheral vascular, musculoskeletal, urological, transvaginal, and intraoperative applications. Prescription use only.
Regulatory Classification
Identification
An ultrasonic pulsed doppler imaging system is a device that combines the features of continuous wave doppler-effect technology with pulsed-echo effect technology and is intended to determine stationary body tissue characteristics, such as depth or location of tissue interfaces or dynamic tissue characteristics such as velocity of blood or tissue motion. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.
{0}------------------------------------------------
Koyo251
FEB 2 0 2004
Special 510(k) Premarket Notification GE Medical Systems - LOGIQ 9 Ultrasound BT04 January 30, 2004
# Attachment B:
Summary of Safety and Effectiveness Prepared in accordance with 21 CFR Part 807.92(c).
Image /page/0/Picture/5 description: The image shows the logo for General Electric (GE). The logo is a circular emblem with the letters "GE" intertwined in a stylized script. The letters are white against a black background, and the circle is outlined with a thin white border.
GE Medical Systems
General Electric Company P.O. Box 414, Milwaukee, WI 53201
#### Section a):
| 1. Submitter: | GE Medical Systems, Ultrasound and Primary Care Diagnostics, LLC<br>PO Box 414<br>Milwaukee, WI 53201 |
|---------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Contact Person: | Allen Schuh,<br>Manager, Safety and Regulatory Engineering<br>Telephone: 414-647-4385; Fax: 414-647-4090 |
| Date Prepared: | January 30, 2004 |
| 2. Device Name: | GE LOGIQ 9 Diagnostic Ultrasound BT04,<br>Ultrasonic Pulsed Echo Imaging System, 21 CFR 892.1560, 90-IYO<br>Ultrasonic Pulsed Doppler Imaging System, 21 CFR 892.1550, 90-IYN |
| 3. Marketed Device: | GE LOGIQ 9 Diagnostic Ultrasound System K011188/K030934/K032656 |
A device currently in commercial distribution.
4. Device Description: The GE LOGIQ 9 is a full featured general purpose diagnostic ultrasound system. It consists of a mobile console approximately 64 cm wide, 90 cm deep and 140-160 cm (adjustable) high that provides digital acquisition, processing and display capability. The user interface includes a computer keyboard, specialized controls and a color video CRT and LCD display. This modification will provide users with additional probe options, improved user interface and productivity.
5. Indications for Use: The device is intended for use by a qualified physician for ultrasound evaluation of Fetal: Abdominal: Pediatric: Small Organ (breast, testes, thyroid); Neonatal Cephalic; Adult Cephalic; Cardiac (adult and pediatric); Peripheral Vascular; Musculo-skeletal Conventional and Superficial; Urology (including prostate); Transvaginal; and Intraoperative (abdominal, thoracic, vascular and neurosurgical).
6. Comparison with Predicate Device: The GE LOGIQ 9 BT04 is of a comparable type and substantially equivalent to the current GE LOGIQ 9 and GE Voluson. It has the same technological characteristics, key safety and effectiveness features, physical design, construction, and materials, and has the same intended uses and basic operating modes as the predicate device.
#### Section b):
1. Non-clinical Tests: The device has been evaluated for acoustic output, biocompatibility, cleaning and disinfection effectiveness as well as thermal, electrical and mechanical safety, and has been found to conform with applicable medical device safety standards.
- 2. Clinical Tests: None required.
3. Conclusion: Intended uses and other key features are consistent with traditional clinical practice, FDA guidelines, and established methods of patient examination. The design and development process of the manufacturer conforms with 21 CFR 820, ISO 9001:2000 and ISO 13485 guality systems. The device conforms to applicable medical device safety standards and compliance is verified through independent evaluation with ongoing factory surveillance. Diagnostic ultrasound has accumulated a long history of safe and effective performance. Therefore, it is the opinion of GE Medical Systems that the GE LOGIQ 9 BT04 Diagnostic Ultrasound is substantially equivalent with respect to safety and effectiveness to devices currently cleared for market.
{1}------------------------------------------------
Image /page/1/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo is circular, with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the top half of the circle. Inside the circle is an image of an eagle with its wings spread.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
FEB 2 0 2004
Mr. Allen Schuh Manager, GE Ultrasound Safety and Regulatory Engineering GE Medical Systems, Ultrasound and Primary Care Diagnostics, LLC 4855 West Electric Avenue MILWAUKEE W1 53219
Re: K040251
Trade Name: GE LOGIQ 9 Ultrasound System with BT04 Modification Regulation Number: 21 CFR 892.1550 Regulation Name: Ultrasonic pulsed doppler imaging system Regulation Number: 21 CFR 892.1560 Regulation Name: Ultrasonic pulsed echo imaging system Regulation Number: 21 CFR 892.1570 Regulation Name: Diagnostic ultrasonic transducer Regulatory Class: Class II Product Code: 90 IYN, IYO, and ITX Dated: January 30, 2004 Received: February 3, 2004
Dear Mr. Schuh:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice. Iabeling, and prohibitions against misbranding and adulteration.
This determination of substantial equivalence applies to the following transducers intended for use with the GE LOGIQ 9 Ultrasound System, as described in your premarket notification:
Transducer Model Number
4D3C
{2}------------------------------------------------
#### 4D10L 4DE7C P8D
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
This determination of substantial equivalence is granted on the condition that prior to shipping the first device, you submit a postclearance special report. This report should contain complete information, including acoustic output measurements based on production line devices, requested in Appendix G, (enclosed) of the Center's September 30, 1997 "Information for Manufacturers Secking Marketing Clearance of Diagnostic Ultrasound Systems and Transducers." If the special report is incomplete or contains unacceptable values (e.g., acoustic output greater than approved levels), then the 510(k) clearance may not apply to the production units which as a result may be considered adulterated or misbranded.
The special report should reference the manufacturer's 510(k) number. It should be clearly and prominently marked "ADD-TO-FILE" and should be submitted in duplicate to:
> Food and Drug Administration Center for Devices and Radiological Health Document Mail Center (HFZ-401) 9200 Corporate Boulevard Rockville, Maryland 20850
This letter will allow you to begin marketing your device as described in your premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus permits your device to proceed to market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801, please contact the Office of Compliance at (301) 594-4591. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). Other general information on your responsibilities under the
{3}------------------------------------------------
Act may be obtained from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or at (301) 443-6597 or at its Internet address "http://www.fda.gov/cdrh/dsmamain.html".
If you have any questions regarding the content of this letter, please contact Rodrigo C. Perez at (301) 594-1212.
Sincerely yours,
Nancy C Brogdon
Nancy C. Brogdon Director, Division of Reproductive, Abdominal and Radiological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosures
{4}------------------------------------------------
# GE LOGIQ 9 Ultrasound System
Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:
| | Mode of Operation | | | | | | | | | | | |
|----------------------------------------------------|-------------------|---|---------------|---------------|------------------|--------------------|------------------|-------------------|---------------------|----------------|-------|--|
| Clinical Application<br>Anatomy/Region of Interest | B | M | PW<br>Doppler | CW<br>Doppler | Color<br>Doppler | Color M<br>Doppler | Power<br>Doppler | Combined<br>Modes | Harmonic<br>Imaging | Coded<br>Pulse | Other | |
| Ophthalmic | | | | | | | | | | | | |
| Fetal / Obstetrics[1] | P | P | P | P | P | P | P | P | P | P | P | |
| Abdominal[1] | P | P | P | P | P | P | P | P | P | P | P | |
| Pediatric | P | P | P | P | P | P | P | P | P | P | P | |
| Small Organ[2] | P | P | P | P | P | P | P | P | P | P | P | |
| Neonatal Cephalic | P | P | P | P | P | P | P | P | P | P | P | |
| Adult Cephalic | P | P | P | P | P | P | P | P | P | P | P | |
| Cardiac[3] | P | P | P | P | P | P | P | P | P | P | P | |
| Peripheral Vascular | P | P | P | P | P | P | P | P | P | P | P | |
| Musculo-skeletal Conventional | P | P | P | P | P | P | P | P | P | P | P | |
| Musculo-skeletal Superficial | P | P | P | P | P | P | P | P | P | P | P | |
| Other[4] | P | P | P | P | P | P | P | P | P | P | P | |
| Exam Type, Means of Access | | | | | | | | | | | | |
| Transesophageal | | | | | | | | | | | | |
| Transrectal | P | P | P | P | P | P | P | P | P | P | | |
| Transvaginal | P | P | P | P | P | P | P | P | P | P | | |
| Transuretheral | | | | | | | | | | | | |
| Intraoperative[5] | P | P | P | P | P | P | P | P | P | P | P | |
| Intraoperative Neurological | P | P | P | P | P | P | P | P | P | P | | |
| Intravascular | | | | | | | | | | | | |
| Laparoscopic | | | | | | | | | | | | |
N = new indication; P = previously cleared by FDA; E = added under Appendix E
Notes: [1] Abdominal includes renal, GYN/Pelvic
[2] Small organ includes breast, testes, thyroid.
[3] Cardiac is Adult and Pediatric.
[4] Other use includes Urology/Prostate
[5] Intraoperative includes abdominal, thoracic (cardiac), and vascular (PV).
[*] Combined modes are B/M, B/Color M, B/PWD or CWD, B/Color/PWD or CWD, B/Power/PWD.
(PIFASE DO NOT WRITE BFLOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE)
Nancy Broglon
(Division Sign-Off)
Division of Reproductive, Abdominal, and Radiological Devices
510(k) Number K040251
Prescription User (Per 21 CFR 801.109)
{5}------------------------------------------------
# GE LOGIQ 9 with 4D3C Transducer
Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:
| | | | | | | Mode of Operation | | | | | | | |
|----------------------------------------------------|---|---|---------------|---------------|------------------|--------------------|------------------|-------------------|---------------------|----------------|-------|--|--|
| Clinical Application<br>Anatomy/Region of Interest | B | M | PW<br>Doppler | CW<br>Doppler | Color<br>Doppler | Color M<br>Doppler | Power<br>Doppler | Combined<br>Modes | Harmonic<br>Imaging | Coded<br>Pulse | Other | | |
| Ophthalmic | | | | | | | | | | | | | |
| Fetal / Obstetrics | N | N | N | | N | | N | N | N | N | | | |
| Abdominal[1] | N | N | N | | N | | N | N | N | N | | | |
| Pediatric | | | | | | | | | | | | | |
| Small Organ[2] | | | | | | | | | | | | | |
| Neonatal Cephalic | | | | | | | | | | | | | |
| Adult Cephalic | | | | | | | | | | | | | |
| Cardiac[3] | | | | | | | | | | | | | |
| Peripheral Vascular | | | | | | | | | | | | | |
| Musculo-skeletal Conventional | | | | | | | | | | | | | |
| Musculo-skeletal Superficial | | | | | | | | | | | | | |
| Other | | | | | | | | | | | | | |
| Exam Type, Means of Access | | | | | | | | | | | | | |
| Transesophageal | | | | | | | | | | | | | |
| Transrectal | | | | | | | | | | | | | |
| Transvaginal | | | | | | | | | | | | | |
| Transuretheral | | | | | | | | | | | | | |
| Intraoperative | | | | | | | | | | | | | |
| Intraoperative Neurological | | | | | | | | | | | | | |
| Intravascular | | | | | | | | | | | | | |
| Laparoscopic | | | | | | | | | | | | | |
N = new indication; P = previously cleared by FDA; E = added under Appendix E
Notes: [1] Abdominal includes renal, GYN/Pelvic
[*] Combined modes are B/M, B/Color M, B/PWD or CWD, B/Color/PWD or CWD, B/Power/PWD.
| | (PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED) |
|----------------------------------------|----------------------------------------------------------------------------|
| | Concurrence of CDRH, Office of Device Evaluation (ODE) |
| | <img alt="signature" src="signature.png"/> |
| | (Division Sign-Off) |
| | Division of Reproductive, Abdominal, |
| | and Radiological Devices |
| Prescription User (Per 21 CFR 801.109) | |
| 510(k) Number | K040251 |
{6}------------------------------------------------
# GE LOGIQ 9 with 4D10L Transducer
Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:
| | Mode of Operation | | | | | | | | | | | |
|----------------------------------------------------|-------------------|---|---------------|---------------|------------------|--------------------|------------------|-------------------|---------------------|----------------|-------|--|
| Clinical Application<br>Anatomy/Region of Interest | B | M | PW<br>Doppler | CW<br>Doppler | Color<br>Doppler | Color M<br>Doppler | Power<br>Doppler | Combined<br>Modes | Harmonic<br>Imaging | Coded<br>Pulse | Other | |
| Ophthalmic | | | | | | | | | | | | |
| Fetal / Obstetrics | | | | | | | | | | | | |
| Abdominal[1] | N | N | N | | N | | N | N | N | N | | |
| Pediatric | N | N | N | | N | | N | N | N | N | | |
| Small Organ [2] | N | N | N | | N | | N | N | N | N | | |
| Neonatal Cephalic | | | | | | | | | | | | |
| Adult Cephalic | | | | | | | | | | | | |
| Cardiac[3] | | | | | | | | | | | | |
| Peripheral Vascular | N | N | N | | N | | N | N | N | N | | |
| Musculo-skeletal Conventional | N | N | N | | N | | N | N | N | N | | |
| Musculo-skeletal Superficial | | | | | | | | | | | | |
| Other | | | | | | | | | | | | |
| Exam Type, Means of Access | | | | | | | | | | | | |
| Transesophageal | | | | | | | | | | | | |
| Transrectal | | | | | | | | | | | | |
| Transvaginal | | | | | | | | | | | | |
| Transuretheral | | | | | | | | | | | | |
| Intraoperative | N | N | N | | N | | N | N | N | N | | |
| Intraoperative Neurological | N | N | N | | N | | N | N | N | N | | |
| Intravascular | | | | | | | | | | | | |
| Laparoscopic | | | | | | | | | | | | |
N = new indication; P = previously cleared by FDA; E = added under Appendix E
Notes: {2} Small organ includes breast, testes, thyroid.
[*] Combined modes are B/M, B/Color M, B/PWD or CWD, B/Color/PWD or CWD, B/Power/PWD.
| | (PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED) |
|----------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| | Concurrence of CDRH, Office of Device Evaluation (ODE) |
| | <div><img alt="signature" src="signature.png"/></div> <div>(Division Sign-Off)</div> <div>Division of Reproductive, Abdominal,</div> <div>and Radiological Devices</div> |
| Prescription User (Per 21 CFR 801.109) | 510(k) Number: K040251 |
E-4
{7}------------------------------------------------
# GE LOGIQ 9 with 4DE7C Transducer
Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:
| | Mode of Operation | | | | | | | | | | |
|----------------------------------------------------|-------------------|---|---------------|---------------|------------------|--------------------|------------------|-------------------|---------------------|----------------|-------|
| Clinical Application<br>Anatomy/Region of Interest | B | M | PW<br>Doppler | CW<br>Doppler | Color<br>Doppler | Color M<br>Doppler | Power<br>Doppler | Combined<br>Modes | Harmonic<br>Imaging | Coded<br>Pulse | Other |
| Ophthalmic | | | | | | | | | | | |
| Fetal / Obstetrics | N | N | N | | N | | N | N | N | N | |
| Abdominal[1] | | | | | | | | | | | |
| Pediatric[6] | N | N | N | | N | | N | N | N | N | |
| Small Organ[2] | | | | | | | | | | | |
| Neonatal Cephalic[6] | N | N | N | | N | | N | N | N | N | |
| Adult Cephalic | | | | | | | | | | | |
| Cardiac[3] | | | | | | | | | | | |
| Peripheral Vascular | | | | | | | | | | | |
| Musculo-skeletal Conventional | | | | | | | | | | | |
| Musculo-skeletal Superficial | | | | | | | | | | | |
| Other[4] | N | N | N | | N | | N | N | N | N | |
| Exam Type, Means of Access | | | | | | | | | | | |
| Transesophageal | | | | | | | | | | | |
| Transrectal | N | N | N | | N | | N | N | N | N | |
| Transvaginal | N | N | N | | N | | N | N | N | N | |
| Transuretheral | | | | | | | | | | | |
| Intraoperative | | | | | | | | | | | |
| Intraoperative Neurological | | | | | | | | | | | |
| Intravascular | | | | | | | | | | | |
| Laparoscopic | | | | | | | | | | | |
N = new indication; P = previously cleared by FDA; E = added under Appendix E
Notes: [4] Other use includes Urology/Prostate
[6] Probe used in surface scanning only
[*] Combined modes are B/M, B/Color M, B/PWD or CWD, B/Color/PWD or CWD, B/Power/PWD.
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Nancy Brandon
(Division Sign-Off)
Division of Reproductive, Abdominal, and Radiological Devices
Prescription User (Per 21 CFR 801.109)
510(k) Number
{8}------------------------------------------------
# GE LOGIQ 9 with P8D Transducer
Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:
| | Mode of Operation | | | | | | | | | |
|-----------------------------------------------------|-------------------|---|---------------|---------------|------------------|--------------------|------------------|-------------------|---------------------|----------------|
| Clinical Application<br>Anatomy/ Region of Interest | B | M | PW<br>Doppler | CW<br>Doppler | Color<br>Doppler | Color M<br>Doppler | Power<br>Doppler | Combined<br>Modes | Harmonic<br>Imaging | Coded<br>Pulse |
| Ophthalmic | | | | | | | | | | |
| Fetal / Obstetrics | | | | | | | | | | |
| Abdominal | | | | | | | | | | |
| Pediatric | | | | | | | | | | |
| Small Organ (specify) | | | | | | | | | | |
| Neonatal Cephalic | | | | | | | | | | |
| Adult Cephalic | | | | | | | | | | |
| Cardiac <sup>[3]</sup> | | | | | | | | | | |
| Peripheral Vascular | | | | N | | | | | | |
| Musculo-skeletal Conventional | | | | | | | | | | |
| Musculo-skeletal Superficial | | | | | | | | | | |
| Other (specify) | | | | | | | | | | |
| Exam Type, Means of Access | | | | | | | | | | |
| Transesophageal | | | | | | | | | | |
| Transrectal | | | | | | | | | | |
| Transvaginal | | | | | | | | | | |
| Transuretheral | | | | | | | | | | |
| Intraoperative (specify) | | | | | | | | | | |
| Intraoperative Neurological | | | | | | | | | | |
| Intravascular | | | | | | | | | | |
| Laparoscopic | | | | | | | | | | |
N = new indication; P = previously cleared by FDA; E = added under Appendix E Notes:
> (PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE)
Nancy C Boydon
Division Sign-Off
(Division Division of Reproductive, Abdominal, and Radiological Device
510(k) Number K040251
Prescription User (Per 21 CFR 801.109)
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.