K040249 · Leonhard Lang GmbH · DRX · Feb 24, 2004 · Cardiovascular
Device Facts
Record ID
K040249
Device Name
SKINTACT ECG ELECTRODE
Applicant
Leonhard Lang GmbH
Product Code
DRX · Cardiovascular
Decision Date
Feb 24, 2004
Decision
SESE
Submission Type
Special
Regulation
21 CFR 870.2360
Device Class
Class 2
Indications for Use
Skintact® radiolucent and MRI-compatible ECG electrodes are designed for use in general electrocardiographic procedures where ECG monitoring is deemed necessary and is ordered by a physician. Such procedures include in particular patient ECG surveillance and ECG diagnosis recording. Skintact radiolucent and MRI-compatible ECG electrodes are single-use, non-sterile, disposable and are to be used on intact (uninjured) skin.
Device Story
Skintact® ECG electrodes are disposable, self-adhesive snap electrodes used for patient ECG surveillance and diagnostic recording. This version features radiolucent and MRI-compatible properties, achieved by replacing the standard snap with a carbon snap. The device consists of a carbon ABS sensor element coated with silver and a silver chloride layer, applied to the skin to capture electrical signals for transmission to an ECG monitor. Used in clinical settings by healthcare professionals; the output is a visual ECG trace on a monitor, assisting clinicians in cardiac rhythm assessment and diagnosis. The device benefits patients by enabling continuous monitoring or diagnostic testing while remaining compatible with imaging environments (X-ray/MRI).
Clinical Evidence
Bench testing only. Electrical performance verified against ANSI/AAMI EC12:2000 (AC impedance, DC offset, defibrillation recovery, noise, bias current). Biocompatibility testing (cytotoxicity, irritation, sensitization) confirmed safety. Radiolucency tested per ASTM F640-79. MRI compatibility tested per ASTM F2052-00, F2182-02a, and F2119-01. Comparison of ECG traces between carbon electrodes and predicate silver-snap electrodes demonstrated equivalent signal performance.
Technological Characteristics
Disposable, self-adhesive snap electrode. Sensor: Carbon ABS element coated with silver/silver chloride. Radiolucent and MRI-compatible design. Electrical performance meets ANSI/AAMI EC12:2000. Packaging: Water-vapour-proof, heat-sealed, aluminized pouches. No software or active electronic components.
Indications for Use
Indicated for patients requiring ECG monitoring or diagnostic recording during general electrocardiographic procedures. For use on intact skin only.
Regulatory Classification
Identification
An electrocardiograph electrode is the electrical conductor which is applied to the surface of the body to transmit the electrical signal at the body surface to a processor that produces an electrocardiogram or vectorcardiogram.
Special Controls
*Classification.* Class II (special controls). The device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 870.9. The special control for this device is the FDA guidance document entitled “Class II Special Controls Guidance Document: Electrocardiograph Electrodes.” See § 870.1(e) for availability information of guidance documents.
{0}------------------------------------------------
FEB 2 4 2004
# 510(k) Summary
# SUBMITTED ON BEHALF OF:
Company Name: Leonhard Lang GmbH Archenweg 56 Address: 6020 Innsbruck Austria
| Telephone: | ++ 43 / 512 / 33 4 25 7 |
|------------|-------------------------|
| Fax: | ++ 43 / 512 / 39 22 10 |
by:
Elaine Duncan, MS.M.E., RAC President, Paladin Medical, Inc. PO Box 560 Stillwater, MN 55082 715-549-6035 715-549-5380
Fax:
Telephone:
Elaine Duncan
DATE PREPARED:
CONTACT PERSON:
January 30, 2004
| Trade Name: | Skintact® ECG Electrode |
|----------------------|------------------------------------|
| Common Name: | Disposable ECG Electrodes |
| Classification Name: | Electrocardiograph (ECG) electrode |
SUBSTANTIALLY EQUIVALENT TO: Skintact® radiolucent and MRI-compatible ECG Electrodes Skintact radiolucent and MRI-compatible ECG Electrodes are substantially equivalent to the Skintact® ECG Electrodes (the manufacturer's predicate devices) and equivalent to predicate ECG electrodes cleared under K971444 and K991105.
DESCRIPTION of the DEVICE: Skintact® ECG Electrodes (and as also to be offered for sale under various private label tradenames) will now also be offered radiolucent and MRI-compatible. Just like the predicate electrodes, radiolucent and MRI-compatible ECG electrodes are self-adhesive, nonsterile, single-use disposable snap electrodes. The radiolucent and MRI-compatible ECG electrodes are identical in size, shape and configuration to the Skintact ECG electrodes currently marketed by Leonhard Lang, GmbH.
The Skintact radiolucent and MRI-compatible ECG Electrodes are composed of the same materials as the predicate devices except the snap, which is made of carbon snap guarantees an unimpaired performance during the shelf-life of the product. All electrodes include an carbon ABS sensor element coated with silver. The silver layer is either completely or partially (in the areas in contact with the conductive gel) covered with a silver chloride layer.
## INDICATIONS FOR USE:
Skintact® radiolucent and MRI-compatible ECG electrodes are designed for use in general electrocardiographic procedures where ECG monitoring is deemed necessary and is ordered by a physician. Such procedures include in particular patient ECG surveillance and ECG diagnosis recording. Skintact radiolucent and MRI-compatible ECG electrodes are single-use, non-sterile, disposable and are to be used on intact (uninjured) skin. (NO CHANGE to ORIGINAL INDICATION for USE.)
{1}------------------------------------------------
#### SUMMARY of TESTING:
All materials used in this new device have been cleared in the following 510(k)s: K 982521 Skintact® ECG Electrodes, S&W ECG Electrodes K 023503 Leonhard Lang Skintact® ECG Electrodes with solid adhesive gel Leonhard Lang Skintact® ECG Electrodes with KS 01 solid wet gel K 024264 Leonhard Lang Skintact® ECG Electrodes with KL 02 liquid gel K 024247
Biocompatibility testing of all materials cleared by 510(k)s confirms the materials are biocompatible and does not introduce any risks. The following testing of the different materials showed no adverse results: Cytotoxicity; Skin Irritation; Sensitization.
The ANSI/AAMI EC 12:2000 "Disposable ECG electrodes" was used to define the requirements for Skintact ECG Electrodes. All electrical tests are according to ANSI/AAMI EC 12:2000. A certification to conformance EC12:2000 with this standard has been provided. The testing conducted was: AC impedance; DC offset voltage; Defibrillation overload recovery; Combined offset instability and internal noise; Bias current tolerance.
The shelf life of radiolucent and MRI-compatible electrodes - was tested in real-time aging. All Skintact® ECG Electrodes (and as also to be offered for sale under various private label tradenames) are packaged in water-vapour-proofed, heat-sealed, non-transparent, aluminized pouches. Leonhard Lang has about 20 years of experience with this packaging and has met requirements for the 24 months shelf-life. No differences were required for packaging the Skintact radiolucent and MRI-compatible ECG electrodes compared to the predicate devices.
The results of these tests confirm that the shelf-life of Leonhard Lang Skintact radiolucent and MRIcompatible ECG Electrodes is well inside the limits defined in ANSI/AAMI EC12-2000. Thus the conclusion that the electrical performance of the electrodes within the limits during their shelf-life of 24 months. The comparison with the predicate devices and the data from radiolucent and MRIcompatible electrodes shows similar results. The difference is negligible in the limits defined in ANSI/AAMI FC12-2000. Therefore electrical performance of the predicate devices and Skintacte radiolucent and MRI-compatible ECG electrodes is equivalent.
Clinical data: Repeating EC12:2000 testing confirms equivalent data and the change in snap can not affect adhesive performance. So material change -- in this case the change in snap - does not affect wear tests and adhesive performance. Therefore these tests were not repeated. The potential effect of material change to the conducting signal was evaluated by repeating the trace testing on the carbon electrodes with different approved gels and determined that the carbon electrodes perform the same. Comparing the ECG traces between carbon electrodes with different approved gels and predicate device ECG electrodes demonstrate that carbon electrodes are equivalent to the predicate device ECG electrodes.
#### Radiolucency:
Radiolucency was tested according standard ASTM F640-79 "Standard Test Methods for Radiopacity of Plastics for Medical Use". Reference: "Radiopacity of ECG-Electrodes by Leonhard Lang GmbH - A study by Dr.Recheis Wolfgang, Dr.Verius Michael, Dr.Huttary Ralf, Mag.Torbica Pavle''.
#### MRI-Compatibility:
MRI companbility was tested according ASTM F 2052-00 "Standard Test Method for Magnetically Induced Displacement Force on Passive Implants in the Magnetic Resonance Environment", ASTM F 2182-02a "Standard Test Method for Measurement of Radio Frequency Induced Heating Near Passive Implants During Magnetic Resonance Imaging", ASTM F 2119-01 "Standard Test Method for Evaluation of MR Image Artifacts from Passive Implants". Reference: "MR-compatibility of ECG-Electrodes by Leonhard Lang GmbIl - A study by Dr.Recheis Wolfgang, Dr.Verius Michael, Dr.Huttary Ralf, Mag.Torbica Pavle".
{2}------------------------------------------------
DEPARTMENT OF HEALTH & HUMAN SERVICES
Public Health Service
Image /page/2/Picture/2 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circle with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is an abstract symbol that resembles an eagle or bird in flight, composed of three curved lines.
FEB 2 4 2004
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
Leonard Lang GmbH c/o Ms. Elaine Duncan, M.S.M.E., RAC President Paladin Medical, Inc. P.O. Box 560 Stillwater, MN 55082-0560
Re: K040249
Trade Name: Skintack ECG Electrode Regulation Number: 21 CFR 870.2360 Regulation Name: Electrograph Electrode Regulatory Class: II (two) Product Code: DRX Dated: January 30, 2004 Received: February 03, 2004
## Dear Ms. Duncan:
We have reviewed your Scction 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The r ou may) dierer orovisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{3}------------------------------------------------
Page 2 - Ms. Elaine Duncan, M.S.M.E., RAC
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please n Joan the Office of Compliance at (301) 594-4648. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html
Sincerely yours,
Dinna R. Lochner
MBram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
# Indications for Use
510(k) Number (if known): K040249
Device Name: Skintact® radiolucent and MRI-Compatible ECG Electrodes
Indications For Use:
Skintact radiolucent and MRI-compatible ECG electrodes are designed for use in general electrocardiographic procedures where ECG monitoring is deemed necessary and is ordered by a physician. Such procedures include in particular patient ECG surveillance and ECG diagnosis recording.
Skintact radiolucent and MRI-compatible ECG electrodes are single use, non-sterile and disposable and are to be used on intact (uninjured) skin.
Prescription Use J (Part 21 CFR 801 Subpart D) AND/OR
Over-The-Counter Use (21 CFR 807 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
R. Sochner
Jivision Sign-Off) ാത്രീംഗ് Olgh Silverscular Devices
10(k) Number K040249
Page 1 of /
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.