QLAB Quantification software is a software application package. It is designed to view and quantify image data acquired on Philips Medical Systems ultrasound products.
Device Story
QLAB Quantification software (v3.0) is a PACS workstation application for viewing, manipulating, and quantifying ultrasound image data. Inputs consist of DICOM-compliant files from Philips Ultrasound systems. The device includes three plug-ins: Cardiac 3DQ (3D viewer, distance/area/volume/mass measurements via MultiPlanar Reconstruction), General Imaging 3D Viewer (volume rendering), and Microvascular Imaging (MVI) (Maximum Intensity Projection convolution of cine information). Used in clinical settings by healthcare professionals to assist in diagnostic assessment. Output includes processed images, quantitative measurements, and exported data (e.g., AVI files). The software facilitates clinical decision-making by providing standardized tools for image analysis and quantification, mirroring processing capabilities of predicate ultrasound platforms.
Clinical Evidence
No clinical data provided. The device relies on bench testing, software verification, and validation processes to demonstrate safety and effectiveness for image display and quantification.
Technological Characteristics
Software-based PACS workstation application. Complies with ISO JPEG image compression standards and Microsoft Developer's Network (MSDN) guidelines. Features include 3D volume rendering, MultiPlanar Reconstruction (MPR), and Maximum Intensity Projection (MIP) algorithms. Connectivity via DICOM-compliant file import. Designed for use with Philips Ultrasound systems.
Indications for Use
Indicated for clinicians requiring viewing and quantification of ultrasound image data acquired on Philips Medical Systems ultrasound products.
Regulatory Classification
Identification
A medical image management and processing system is a device that provides one or more capabilities relating to the review and digital processing of medical images for the purposes of interpretation by a trained practitioner of disease detection, diagnosis, or patient management. The software components may provide advanced or complex image processing functions for image manipulation, enhancement, or quantification that are intended for use in the interpretation and analysis of medical images. Advanced image manipulation functions may include image segmentation, multimodality image registration, or 3D visualization. Complex quantitative functions may include semi-automated measurements or time-series measurements.
Special Controls
*Classification.* Class II (special controls; voluntary standards—Digital Imaging and Communications in Medicine (DICOM) Std., Joint Photographic Experts Group (JPEG) Std., Society of Motion Picture and Television Engineers (SMPTE) Test Pattern).
Predicate Devices
Philips M2424 Diagnostic Ultrasound System
TomTec Cardio-View
Reference Devices
Philips HDI 5000
Submission Summary (Full Text)
{0}------------------------------------------------
# FEB 1 7 2004
K040227
page 1 of 2
# 510(k) Summary of Safety and Effectiveness
This sunmary of safety and effectiveness is provided as part of the Premarket Notification in compliance with 21CFR. Part 807, Subpart E, Section 807.92
1 ) Submitter's name, address, telephone number, contact person
Steve Singlar Regulatory Submissions Manager Philips Ultrasound 3000 Minuteman Koad Andover, MA 01810-1099 Telephone: (978) 659-1201 Facsimile: (978) 975 7324 E-mail: Steve Singlar(@Philips.com
Date prepared: 15 January 2004
2) Name of the device, including the trade or proprietary name if applicable, the common or usual name, and the classification name. if known:
| Common/Usual Name: | Picture Archiving and Communications Systems Workstation |
|----------------------|----------------------------------------------------------|
| Proprietary Name: | QLAB Quantification |
| Classification Name: | Picture Archiving and Communications System, Class II |
## 3) Device Description
QLAB version 3.0 adds a Cardiac 3DQ Plug-in (3D viewer with 3D measurements), G1 3D viewer Plug-in and an MVI Plug-in to the cleared QLAB version 2.0.
The Cardiac 3DQ plug-in provides a means of opening. displaying, manipulating and measuring 3D inage files from currently cleared Plilips Ultrasound systems The 3DQ plug-in also allows distance, area, volume and mass measurements from MultiPlanar Reconstruction (MPR) images derived the 3D data sets. The software also provides a means of exporting the data generated by the plug in module in a form accessible to the end user.
Gl (General Imaging) ND Viewer Plug-in reads DICOM compliant liles generated by currently cleared Philips Ultrasound systems. It contains tools for changing 3D volume rendering parameters. The volume rendering is done using the rendering engine shipping with the Boris Platform and Philips HDI 5000. Therefore the main volume rendering controls are the same as on the imaging system.
MVI (Microvascular Imaging) Plug-in reads DICOM compliant files generated by the Philips Boris Platform and the Philips HDI 5000 Platforms. It performs a Maximum Intensity Projection convolution of the cine' information and allows viewing of the processed information. It provides tools for export of the resulting information in a standard AVI tile formal for use in presentations. The processing is accomplished exactly as in the Predicate device (HDI 5000) with the exception that there are no user selectable processing changes possible.
{1}------------------------------------------------
K040227
Page 2 of 2
# 4) Performance Standards
No performance standards for PACS systems or components have been issued under the authority of Section 14. The QLAB software has been designed to comply with the following voluntary standards:
- MSDN Microsoft Developer's Network October 2001 .
- ISC) Joint Photographic Experts Group (JPEG) Image Compression Standard .
# 5) General Safety and Effectiveness Concerns
The device labeling contains operating instructions for the safe and effective use of the QLAB software.
## 6) Substantially Eguivalent Devices
Philips Ultrasound believes that the QLAB software is substantially equivalent to other commercially available products, specifically Philips M2424 Diaguostic Ultrasound System and the TomTec Cardio-View.
## 7) Software
Software development for the QLAB software follows documented processes for software design, verification and validation testing. A risk assessment has been completed to identify potential design hazards that could cause an error or injury based on the use of the quantification results. Appropriate steps have been taken to control all identified risks for this type of image display and quantification product.
## 8) Conclusions
The QLAB software is designed and manufactured to meet United States and international standards for the display and quantification of invages acquired on Phillips Ultrasound devices. The system is designed to incorporate components common to all image viewing systems for the display, manipulation and quantification tasks within a clinical setting. The QLAB software incorporates features of predicate devices cleared through premarket notification and no new issues of safety or effectiveness are raised.
{2}------------------------------------------------
Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo consists of a stylized eagle with three stripes forming its body and wing. The eagle is facing to the left. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular pattern around the eagle.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
Philips Ultrasound, Inc. % Ms. Laura Danielson Responsible Third Party Official TÜV Product Service 1775 Old Highway 8
NEW BRIGHTON MN 55112-1891
Re: K040227
Trade/Device Name: OLAB Quantification Software Regulation Number: 21 CFR 892.2050 Regulation Name: Picture archiving and communication system Regulatory Class: II
Product Code: 90 LLZ Dated: January 30, 2004 Received: February 2, 2004
Dear Ms. Danielson:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act), 21 CFR 1000-1050,
FEB 17 2004
{3}------------------------------------------------
This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at one of the following numbers, based on the regulation number at the top of the letter:
| 8xx.1xxx | (301) 594-4591 |
|----------------------------------|----------------|
| 876.2xxx, 3xxx, 4xxx, 5xxx | (301) 594-4616 |
| 884.2xxx, 3xxx, 4xxx, 5xxx, 6xxx | (301) 594-4616 |
| 892.2xxx, 3xxx, 4xxx, 5xxx | (301) 594-4654 |
| Other | (301) 594-4692 |
Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97) you may obtain. Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html.
Sincerely yours,
Nancy C. Bugden
Nancy C. Brogdon Director, Division of Reproductive, Abdominal and Radiological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
#### 2 FDA Indication for Use Form
510(k) Number: Unknown YO40227
Device Name: QLAB Quantification Software with Cardiac 3DQ Plug-in, 3D General Imaging Viewer Plug-in and a Microvascular Imaging Plug-in.
Indications for Use: QLAB Quantification software is a software application package. It is designed to view and quantify image data acquired on Philips Medical Systems ultrasound products.
Prescription Use: X (Part 21 CFR 801 Subpart D)
AND/OR
Over-The-Counter Use: (21 CFR 807 Subpart C)
PLEASE DO NOT WRITE BELOW THIS LINE -- CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
David L. Heymann
(Division Sign-Off) Division of Reproductive, Abdominal, and Radiological Devices 510(k) Number
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.