The FRIALIT® PLUS Dental Implant System is indicated for use in single tooth restorations, edentulous spans restored with multiple single teeth, freestanding bridges and to retain overdentures. The implants can be used for immediate implant placement, delayed immediate or late implant placement. The XiVE® PLUS Dental Implant System is indicated for the following: Once the implant has osseointegrated, it serves to support single tooth, bridge and overdenture restorations. In the edentulous mandible, a minimum of four XiVE® dental implants (≥11mm length) are placed between the mental foramina and rigidly splinted together. In this case, bar prosthetic loading is possible immediately after implant placement. The XiVE® D3.0 PLUS Dental Implant System is indicated for single tooth restorations and splinted tooth restorations in the region of 7 and 10 and 23 to 26. The XiVE® TG PLUS Dental Implant System is indicated for single-stage implant placement, with a minimum healing phase of three months in good quality bone and four months in spongy bone, for maxillary and mandibular splinted crowns, bridges and barretained overdenture restorations. The bridge must be supported by a minimum of two transgingival threaded implants. In the edentulous maxilla, a minimum of four transgingival implants are placed in a trapezoidal distribution and rigidly splinted together. In the edentulous mandible, a minimum of four transgingival implants (≥11mm length) are placed between the mental foramina and rigidly splinted together. In this case, bar prosthetic loading is possible immediately after implant placement.
Device Story
Endosseous dental implants (FRIALIT, XiVE, XiVE TG) featuring 'PLUS' surface treatment; intended for surgical placement into jawbone by dental professionals; provides foundation for prosthetic restorations (crowns, bridges, overdentures). PLUS surface (grit-blasted, high-temperature acid-etched) designed to enhance osseointegration and removal torque compared to previous surface iterations. Used in clinical dental settings; supports immediate, delayed, or late loading protocols depending on implant type and bone quality. Benefits include improved implant stability and integration strength, facilitating functional and aesthetic tooth replacement.
Clinical Evidence
No human clinical trials presented. Evidence consists of preclinical physiochemical surface characterization and animal studies conducted at the University of Tubingen, University of Cologne, and University of São Paulo. Animal models (pigs) evaluated loaded/unloaded implants, immediate placement in infected sites, and laser therapy effects. Results indicate higher removal torque values for PLUS surface implants compared to solely acid-etched or anodic oxidized surfaces, supporting enhanced integration stability.
Technological Characteristics
Endosseous dental implants with grit-blasted and high-temperature acid-etched 'PLUS' surface. Materials and mechanical design are consistent with predicate systems. Intended for surgical implantation into maxillary or mandibular bone. No electronic components, software, or energy sources involved.
Indications for Use
Indicated for patients requiring dental restorations including single tooth, bridge, and overdenture support. Specific populations include edentulous mandible/maxilla patients requiring splinted implants. Contraindicated for sites not meeting specific bone quality or length requirements (e.g., <9.5mm or <11mm depending on specific system/region).
Regulatory Classification
Identification
An endosseous dental implant is a prescription device made of a material such as titanium or titanium alloy that is intended to be surgically placed in the bone of the upper or lower jaw arches to provide support for prosthetic devices, such as artificial teeth, in order to restore a patient's chewing function.
Special Controls
*Classification.* (1) Class II (special controls). The device is classified as class II if it is a root-form endosseous dental implant. The root-form endosseous dental implant is characterized by four geometrically distinct types: Basket, screw, solid cylinder, and hollow cylinder. The guidance document entitled “Class II Special Controls Guidance Document: Root-Form Endosseous Dental Implants and Endosseous Dental Implant Abutments” will serve as the special control. (See § 872.1(e) for the availability of this guidance document.)(2)
*Classification.* Class II (special controls). The device is classified as class II if it is a blade-form endosseous dental implant. The special controls for this device are:(i) The design characteristics of the device must ensure that the geometry and material composition are consistent with the intended use;
(ii) Mechanical performance (fatigue) testing under simulated physiological conditions to demonstrate maximum load (endurance limit) when the device is subjected to compressive and shear loads;
(iii) Corrosion testing under simulated physiological conditions to demonstrate corrosion potential of each metal or alloy, couple potential for an assembled dissimilar metal implant system, and corrosion rate for an assembled dissimilar metal implant system;
(iv) The device must be demonstrated to be biocompatible;
(v) Sterility testing must demonstrate the sterility of the device;
(vi) Performance testing to evaluate the compatibility of the device in a magnetic resonance (MR) environment;
(vii) Labeling must include a clear description of the technological features, how the device should be used in patients, detailed surgical protocol and restoration procedures, relevant precautions and warnings based on the clinical use of the device, and qualifications and training requirements for device users including technicians and clinicians;
(viii) Patient labeling must contain a description of how the device works, how the device is placed, how the patient needs to care for the implant, possible adverse events and how to report any complications; and
(ix) Documented clinical experience must demonstrate safe and effective use and capture any adverse events observed during clinical use.
{0}------------------------------------------------
FRIADENT® PLUS Dental Implant Systems Original Premarket 510(k) Notification
70
OCT 27 2004
#### SECTION 16: SUMMARY OF SAFETY AND EFFECTIVENESS
This 510(k) summary of safety and effectiveness information is being submitted in accordance with the requirements of SMDA 1990 and CFR 807.92.
#### 16.1 SUBMITTER INFORMATION
| a. | Company Name: | FRIADENT GmbH. |
|----|--------------------------------------|------------------------------------------------------|
| b. | Company Address: | Steinzeugstrasse 50<br>Mannheim D-68229<br>Germany |
| c. | Company Phone:<br>Company Facsimile: | (011) 49 06 21 4 86 1549<br>(011) 49 06 21 4 86 1866 |
| d. | Contact Person: | Heike Dietzler<br>Regulatory Affairs Manager |
| e. | Date Summary Prepared: | September 28, 2004 |
#### DEVICE IDENTIFICATION 16.2.
| a. Trade/Proprietary Name: | FRIALIT® PLUS Dental Implant System<br>XiVE® S PLUS Dental Implant System<br>XiVE® TG PLUS Dental Implant System<br>XiVE® D 3.0 PLUS Dental Implant System |
|----------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------|
| b. Classification Name: | Endosseous Dental Implants<br>21 CFR 872.3640 |
### IDENTIFICATION OF PREDICATE DEVICES 16.3
| Company | Device | 510(k) No. | Date Cleared |
|---------------|--------------------------------------|------------|--------------|
| FRIADENT GmbH | FRIALIT® Dental Implant<br>System | K031671 | 08/14/2003 |
| | XiVE® S Dental Implant<br>System | K032158 | 08/14/2003 |
| | XiVE® TG Dental Implant<br>System | K032284 | 08/14/2003 |
| | XiVE® D 3.0 Dental Implant<br>System | K032302 | 10/02/2003 |
{1}------------------------------------------------
#### DEVICE DESCRIPTION 16.4
The FRIALIT®, XiVE® and XiVE® TG dental implants with the new FRIADENT Surface have been cleared for commercial distribution. The purpose of this application is to change the name of the surface to PLUS and obtain clearance for additional marketing claims associated with the PLUS surface.
#### 16.5 SUBSTANTIAL EQUIVALENCE
The FRIADENT® PLUS Dental Implant Systems are substantially equivalent to the current FRIADENT® Dental Implant Systems in terms of design, materials, coatings, mechanical strength, prosthetic options and single tooth intended use.
#### 16.6 INTENDED USE
The FRIALIT® PLUS Dental Implant System is indicated for use in single tooth restorations, edentulous spans restored with multiple single teeth, freestanding bridges and to retain overdentures. The implants can be used for immediate implant placement, delayed immediate or late implant placement.
The XiVE®S PLUS Dental Implant System is indicated for the following: Once the implant has osseointegrated, it serves to support single tooth, bridge and overdenture restorations. In the edentulous mandible, a minimum of four XiVE® dental implants (≥9.5mm length) are placed between the mental foramina and rigidly splinted together. In this case, bar prosthetic loading is possible immediately after implant placement.
The XiVE® D3.0 PLUS Dental Implant System is indicated for single tooth restorations and splinted tooth restorations in the region of 7 to 10 and 23 to 26.
The XiVE® TG PLUS Dental Implant System is indicated for single-stage implant placement, with a minimum healing phase of three months in good quality bone and four months in spongy bone, for maxillary and mandibular splinted crowns, bridges and bar-retained overdenture restorations. The bridge must be supported by a minimum of two transgingival threaded implants. In the edentulous maxilla, a minimum of four transgingival implants are placed in a trapezoidal distribution
{2}------------------------------------------------
KC40170
FRIADENT® PLUS Dental Implant Systems Original Premarket 510(k) Notification
> and rigidly splinted together. In the edentulous mandible, a minimum of four transgingival implants (≥9.5mm length) are placed between the mental foramina and rigidly splinted together. In this case, bar prosthetic loading is possible immediately after implant placement.
#### 16.7 TECHNOLOGICAL CHARACTERISTICS
Preclinical evaluations of the FRIADENT PLUS Surface were conducted at the University of Tubingen, Germany. Several studies were conducted to determine the physiochemical surface characteristics in comparison to other commercially available surfaces.
Animal studies of the FRIADENT PLUS Surface were conducted at the University of Cologne, Germany, and the University of São Paulo, Brazil. Studies involved the evaluation of loaded and unloaded implants in pigs, immediate implant placement in periodontally infected sites, and the effect of visible and NIR low intensity laser therapy on implant osseointegration. Results of these studies support the PLUS surface characteristics, name and advertising claims for the FRIADENT Dental Implant Systems.
#### 16.8 CLASS III CERTIFICATION AND SUMMARY
This notification contains a Class III certification and summary of adverse safety and effectiveness information pursuant to 513(f) of the Federal Food, Drug, and Cosmetic Act.
### 16.9 CONCLUSIONS
This notification contains all information required by 21 CFR 807.87. A completed copy of the Premarket Notification 510(k) Reviewers Checklist is provided in this submission. Preclincal evaluations and animal studies of the FRIADENT PLUS surface have been presented to substantiate the name PLUS and associated marketing claims. Comparison of the FRIADENT dental implant systems to the predicate device shows that the device is substantially equivalent.
{3}------------------------------------------------
Image /page/3/Picture/1 description: The image shows the seal of the U.S. Department of Health & Human Services. The seal is circular and contains the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the top half of the circle. Inside the circle is a stylized image of an eagle.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
OCT 27 2004
FRIADENT GmbH C/O Ms. Carol White President Patterson Consulting Group, Incorporated 21911 Erie Lane Lake Forest, California 92630
Re: K040170
Trade/Device Name: FRIADENT PLUS Dental Implant Systems Regulation Number: 872.3640 Regulation Name: Endosseous Implant Regulatory Class: II Product Code: DZE Dated: July 28, 2004 Received: August 11, 2004
Dear Ms. White:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition. FDA may publish further announcements concerning your device in the Federal Register.
{4}------------------------------------------------
Page 2 - Ms. White
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (240) 276-0115. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html
Sincerely yours,
CluLs
Chin Lin, Ph.D.
Chiu Lin, Ph.D. Director Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
FRIADENT PLUS Dental Implant Systems Original Premarket 510(k) Notification
## INDICATION FOR USE
510(k) Number: K040170
Device Name:
FRIADENT PLUS Dental Implant Systems
Indications for Use:
The FRIALIT® PLUS Dental Implant System is indicated for use in single tooth restorations, edentulous spans restored with multiple single teeth, freestanding bridges and to retain overdentures. The implants can be used for immediate implant placement, delayed immediate or late implant placement.
The XiVE® PLUS Dental Implant System is indicated for the following: Once the implant has osseointegrated, it serves to support single tooth, bridge and overdenture restorations. In the edentulous mandible, a minimum of four XiVE® dental implants (≥11mm length) are placed between the mental foramina and rigidly splinted together. In this case, bar prosthetic loading is possible immediately after implant placement.
The XiVE® D3.0 PLUS Dental Implant System is indicated for single tooth restorations and splinted tooth restorations in the region of 7 and 10 and 23 to 26.
OR
(continue next page)
Prescription Use _____________________________________________________________________________________________________________________________________________________________ (Per 21 CFR 801.109).
Over-The-Counter Use _______________
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
| Concurrence of CDRH, Office of Device Evaluation (ODE) | <img alt="Signature" src="signature.png"/> |
|--------------------------------------------------------|------------------------------------------------------------------------------------|
| | (Division Sign-Off) |
| | Division of Anesthesiology, General Hospital,<br>Infection Control, Dental Devices |
| 510(k) Number | K040170 |
{6}------------------------------------------------
# INDICATION FOR USE (Continued)
510(k) Number:
K040170
Device Name: PLUS Dental Implant Systems
Indications for Use: (continued)
The XiVE® TG PLUS Dental Implant System is indicated for single-stage implant placement, with a minimum healing phase of three months in good quality bone and four months in spongy bone, for maxillary and mandibular splinted crowns, bridges and barretained overdenture restorations. The bridge must be supported by a minimum of two transgingival threaded implants. In the edentulous maxilla, a minimum of four transgingival implants are placed in a trapezoidal distribution and rigidly splinted together. In the edentulous mandible, a minimum of four transgingival implants (≥11mm length) are placed between the mental foramina and rigidly splinted together. In this case, bar prosthetic loading is possible immediately after implant placement.
FRIADENT PLUS grit-blasted and high temperature acid-etched implants display higher removal torque values compared to solely acid-etched implants and anodic oxidized textured implants. The high removal torque values of FRIADENT PLUS implants after a loading period of 5 months may be interpreted as an increase in the strength of endosseous implant integration and implant stability.
Prescription Use _____________________________________________________________________________________________________________________________________________________________
Over-The-Counter Use
(Per 21 CFR 801.109)
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
OR
Concurrence of CDRH, Office of Device Evaluation (ODE)
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.