COMPPAC 200HD; CS12 POWER SUPPLY (AIRCRAFT/VECHILE); PS11 POWER SUPPLY (MAINS)
Applicant
Pneupac , Ltd.
Product Code
BTL · Anesthesiology
Decision Date
May 27, 2004
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 868.5925
Device Class
Class 2
Attributes
Therapeutic, Pediatric
Indications for Use
The Pneupac compPAC 200HD ventilator is a self contained portable gas powered automatic gas ventilator intended to provide ventilation to patients who have respiratory ditress or insufficiency. The comPAC 200HD is suitable for emergency and transport use for adult and paediatric patients greater than 20Kg in situations where convential portable ventilators may not be suitable.
Device Story
The compPAC 200HD is a portable, gas-powered automatic ventilator for emergency and transport use. It operates via an internal compressor or external gas supply (44-87 psig). Ambient air is drawn through a filter and entrained with gas to provide ventilation. The device features calibrated rotary controls for minute volume (6-14 L/min) and frequency (10-30 bpm), with a constant I:E ratio of 1:1.8. It includes a fixed pressure relief valve (40 cm H2O) and pneumatic/electronic alarms for high pressure, low pressure/disconnect, and gas supply failure. Power options include an internal NiCad battery, mains electricity (via PS11), or 24-28V DC vehicle/aircraft sources (via CS12). The device is housed in a chemically hardened case. Healthcare providers use the front panel controls and manometers to adjust and monitor ventilation. The device supports patients in remote areas or disaster relief, providing flexible power and gas input options to maintain respiratory support.
Clinical Evidence
Bench testing only. Testing conducted in accordance with the Draft Reviewer's Guidance for Ventilators (1999) and EN60601-1. Performance testing verified volume and frequency parameters against predicate devices. EMC and safety testing performed per EN 60601-1-2, EN 794-3, and various BS EN standards for immunity and emissions.
Technological Characteristics
Gas-powered ventilator with internal compressor. Materials: Chemically hardened housing. Power: Internal NiCad battery, 115/230V AC, or 12-30V DC. Connectivity: 40mm DIN NATO filter threads, Schrader gas input. Standards: EN 60601-1, EN 794-3, BS EN 61000 series. Features: Pneumatic pressure relief valve, electronic LED/audible alarms, rotary controls for frequency and volume.
Indications for Use
Indicated for adult and pediatric patients >20kg with respiratory distress or insufficiency requiring ventilation in emergency and transport settings.
Regulatory Classification
Identification
A powered emergency ventilator is a demand valve or inhalator intended to provide emergency respiratory support by means of a face mask or a tube inserted into a patient's airway.
ParaPAC Model 200 Medic Emergency Powered Ventilator (Resuscitator) (K020899)
Submission Summary (Full Text)
{0}------------------------------------------------
KO4oo9o
## MAY 2 7 2004
#### Pneupac Ltd.
Bramingham Susiness Park Liferense Was 1 .. 11:4; 2 32 .. 132 .. 152 .. ্রাম Tel - - - 44 |013681 430000 50x - - 44 100 158 1 433001 r folophelucan ny ak www.phalip... It is
### Summary of Safety and Effectiveness
| Submitter: | Pneupac Ltd. |
|------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Address: | Bramingham Business Park.<br>Enterprise Way, Luton,<br>Bedfordshire, LU3 4BU<br>United Kingdom. |
| Telephone: | (+44) (0) 1582 430000 |
| Contact: | Regulatory Affairs & Quality Assurance<br>Manager |
| Prepared: | 20 <sup>th</sup> August 2003 |
| Proprietary Name: | compPAC Ventilator Model 200, PS11<br>Mains or dc Power Supply/ Battery<br>Charger & CS12 dc Power Supply/ Battery<br>Charger |
| Common/ Classification Name: | Powered Ventilator with Power Supply/<br>Battery Charger |
| Predicate Devices: | compPAC 200 Powered Emergency Ventilator<br>& PS11 Power Supply/ Battery Charger<br>(K021841).<br>ParaPAC Model 200 Medic Emergency<br>Powered Ventilator (Resuscitator) (K020899). |
#### New Device Description:
The compPAC COM 200HD ventilator is primarily gas powered and is housed in an easily carried, chemically hardened housing is designed to accept the long endurance battery that is used to drive a small compressor, which in turn provides the inflating gas, through
X. Summary of Safety and Killestjvesses 510/kW COM200FU3
in the present case, it is impossible to continue without additional data.
we cannot proceed for the time being.
{1}------------------------------------------------
### Summary of Safety and Effectiveness
#### New Device Description (ctd.):
an 'oscillator', to the casualty. All ambient air for the ventilation of the casualty passes into the system through a fitter (NATO No: 4240-01-361-1319). About '// of the volume is compressed to system uncough in the (cre expansion in an entrainment mixing device, which entrains the other 1/3 by creating a sub-atmospheric pressure It can be operated from a range of power sources Battery! Compressed gass! Mains Electricity (via PS11 power supply/ charger unit) and any d.c. power source, ie: auxiliary aircraft/vehicle electrical source (via CS12 power supply/ charger unit) to provide maximum flexibility of operation in remote areas, in disaster relief.
The compPAC COM 200HD ventilator is a robust self-contained portable device comprising of the following standard kit parts
- Control Module -�
- Green case variant: Part No 510A2433 ◆ ાર
- Yellow case variant: Part No 510A2434 .
- Rubber Boot: Part No 510A2271 奉
- Filter: Part No W194-001 �
- Patient Valve and Hose assembly: Part No 510A1082 *
- Oz Auxiliary gas input lead, Schrader probe (BS5682): Part No 510A2600 ◆
- Support ramp: Part No 510A2372 春
- 28 volt open-ended Vehicle Supply lead to compPAC Part No 510A2564 ゃ
- compPAC C200HD User Manual Part No 504-2055A NUS HD #
And the following fitted labels
- Valance Panel Label. Part No 504-228 ◆
- Instrument Panel Label: Part No 504-227 �
- Supplementary O2 Table Label: Part No 504-224 �
- Battery fitting and removal instruction Label: Part No 504-223 �
- Alarm Bezel Label (set of 2): Part No 504-222 .
X. Sciencer) of Stately and Effectiveners 510K | COM20EMD
{2}------------------------------------------------
### Summary of Safety and Effectiveness
#### New Device Description (ctd.):
The module can be used individually as a robust self-contained portable device powered by a NiCad rechargeable battery specified for fitting inside the unit, which will provide approximately (1) hours continuous ventilation. The module weighs 187 lb with the battery installed. The ventilator has a socket to accept an external 24V d c supply from e g. an aircraft or vehicle electrical circuit
This enables the system to be used for extended periods wherever a 24-28V d.c. supply is available. The power requirement is less than 50 watts
If the Pneupac CS12 (d.c. source only) power supply) battery charger (weight 2.2 lbs) is connected to this socket, simultancous full charging of the internal battery will occar whilst the ventilator is running
If the Pneupac PSI i (inains or d.c. source) power supply / battery charger (weight. 3.86 lb) is connected to this socket, simultaneous trickle charging of internal battery will occur whilst the ventilator is running When not required as a power supply it can be switched, to fast charge the battery whilst it is in situ inside the ventilator
Additionally, specifically for aircraft use, an aeromedical EMC filter/ lead assembly (Part No 525-1015) is featured to protect the aircraft from compPAC emissions
The following diagrams illustrate the optional sources available to drive the compPAC ventilator, the electrical power source options and the battery charging options -
X: Smannary of Safety and Effect/sences: 510(13 COM200HD
Section 2/ Page 3
smiths
{3}------------------------------------------------
## Summary of Safety and Effectiveness
New Device Description (51d
New Device Description (110.):
Graphical representation of electrical power source inputs, their relevant interfaces / devices and outputs
#### To run the compPAC
Image /page/3/Figure/5 description: The image shows a diagram of an electrical power source with interfaces, devices, and outputs. The electrical power sources include a battery, 12V dc, 115V ac, 230V ac, and 24-28V dc. The interfaces and devices include PS11 for stationary use only, a dc vehicle cable, and CS12 for vehicle and aircraft use. The output is compPAC operating.
#### To charge the batters
Image /page/3/Figure/7 description: The image shows a diagram of an electrical power source with interfaces, devices, and outputs. The electrical power sources include 115V ac, 230V ac, 12V dc, and 24-28V dc. The interfaces/devices include PS11 and CS12, and the output is battery charging. PS11 provides a trickle charge while operating, and CS12 provides a standard charge while operating.
X: Summery of Saleny and Effectivenen 510(k) U M200HD
{4}------------------------------------------------
## Summary of Safety and Effectiveness
### New Device Description (ctd.);
An input gas connector is provided to allow the compPAC ventilator to be connected to a 44 to 87 psig gas supply via a legally registered high pressure regulator (not suppliated with the material battered battered hence not part of this submission) so that it can be operated independently of its its internal battery electrical supply In this way, if oxygen cylinders are available, 100% or 4.5% or allows electrical supply in this way. In usygen cymiders are are This facility also allows connection to air compressor systems
Alternatively, there is a supplementary oxygen intake connector, in paralied with the filter, enabling 21-45% oxygen to be supplied to the casualty, from e g an oxygen cylinder was a legally registered high pressure regulator (not supplied with the compPAC, hence not part of this registeres ), when the ventilator is operating on its compressor
Calibrated controls for frequency and minute volume are provided to set the required ventiation pattern. A fixed pressure relief valve is fifted to limit the peak inspiratory pressure to a maximum of 40 cm HzO and provides a pneumatically operated audible high-pressure alarm (an adjustable pressure relief device will also be available in the fittere as a retro-fittable option). In addition an pressure reach device the alarm sounds if the inflation pressure exceeds 40 cm HzO.
Most of the controls and input and output connections are mounted on the from panel, which is deeply shrouded to give maximum protection from chemical "rain"
The ventilator module comrol panel has the following features.
- Minute Volume Control. This calibrated rotary control knob gives continuous adjustment of � the minute volume delivered to the patient over the range 6 to 14 L/min.
- Frequency Control, This rotary control knob gives continuous adjussment of frequency over 兼 the range 10 to 30 breaths per minute
- I:E ratio is nominally constant at 1. i. 8 throughout the range of frequency *
X Summery of Satiety and Ellectricess 5100 - COM200800
Section 2/ Page 5
smiths · · Seat · ist finishing Bricked in
{5}------------------------------------------------
## Summary of Safety and Effectiveness
New Device Description (ctd.):
- Patient Inflation Pressure Manometer, range -10 to +100 cm HpO �
- Oxygen Concentration Control This two-position rovary control in assessed to the eas-후 100% or 45% inspired oxygen concentration when an oxygen supply is connected to the gas-100% or 43% inspired oxygen concentralion withis systed is disabled internally. In the 45% Oz position the driving gas requirement is reduced by 66% and the balance of the In the 45% U position the driving gas requirement arr which is drawn through the fifter canister
- A mechanically operated visual alarm (labelled 'Supply Gas Failure Alam') gives a . warning that the supply gas has dropped to a pressure at which the ventilator will no longer be warming that the supply gas nas artipped to a pressure it shows red, with adequate its suces the operating to specificated ( red indicates that the supply should be changed In most cases the shows will begin to oscillate from white to parial red as the supply pressure falls to the lower threshold level
The visual indication will be accompanied by an electronically generated medium priority audible warning
- Electronic aların bezel indicating *
- High Pressure Indicator Flashes Red LED with audible alarm at set relief pressure and 要 with continuous positive pressure
- Normal Cycle Indicator Flashes Green LED every time inflation pressure rises through ♪ 10 cm H2O
- Low Pressure/ Disconnect Indicator Flashes Yellow LED with audible alarm if pressure � does not rise through 10 cm H2O within ten seconds.
- Silence button silences audible alarm for 60 seconds. Flashes Orange LED to indicate to * the operator that the audible alarm is silenced.
- Electrical power indicator Flashes Yellow LED with audible alarm �
K Surners of Subery and Effectiveness (100) COM200HD
{6}------------------------------------------------
## Summary of Safety and Effectiveness
The CS12 power supply and battery charger designed to power the comply has the The CS12 power supply and Dattery from an aircraft or vehicle 24-28V dc supply has the following controls and features:
The CS12 power supply and battery charger complements the complexe with a nominal 12/24/28 vol The CS12 power supply and battery charged from any source with a nominal 12/24/28 vol.
allows the ventilator's battery to be charged from the second with annoly allows the ventilator's battery to be clarged from any only be operated simultaneously if the supply
D.C. aircraft/ vehicle supply. The ventilator may only be operated simult D.C. aircrato vemere supply. The following controls and features
- DC Outputs protected outputs to power the ventilator and supply a constant current to # charge the +28Vdc battery
- Start Button initiates a battery charging cycle If held for a few seconds, it will also ● terminate any battery charging cycle underway
- POWER Yellow Lamp 'charge only' indicates aircraft/ vehicle supply voltage is only # sufficient (10-22V) to charge the battery
- POWER Green Lamp 'compPAC & charge' indicates aircraft/ vehicle supply voltage is ◆ POWER Green Lamp compt AC of charge its battery simultator and charge its battery simultancously
- CHARGE Yellow Lamp 'Charging' indicates a battery charging cycle is underway �
- CHARGE Green Lamp 'Complete' indicates the battery charging cycle has been * successfully completed
X. Summery of Safety and Effectiversism 510(k) LOM200HD
Section 2/ Page 7
smiths Coments Shallerys of a port of the series famise
{7}------------------------------------------------
### Summary of Safety and Effectiveness
The PS11 Dower supply/ charger designed to power the compPAC 200HD ventilator and The ESTI Dower Support Chartery from a Mains (230 or 115 volt nominal AC operation) or d.c. (12 to 30 volt d c ) supply has the following controls and features
- Line Output In conjunction with the negative common, a -28V dc output for general ↓ Lifte Output = In conjunction with the negarres battle when the appropriate lead is fitted. The output is fully protected against short circuits and overload
- Trickle charge output In conjunction with the negative common, a +28V dc ourput 章 I rickle charge output - In complise ventilator This line provides a trickle charge facility for the compPAC's internal battery
- Battery Charge Output In conjunction with the negative common, this is the output used to � charge a +28V dc NiCad battery
- On/Off Switch The switch on the from panel of the PS11 selects two different operating . modes -
- 28V de & Trickle Clug. ON: This activates the 28V dc Line and Trickle Charge outputs and � open circuits the compPAC internal battery
- Output OFF & Charge Battery This activates the Battery Charge output and removes power . from the 28V dc Line and Trickle Charge outputs.
#### Aircraft Power Supply Lead:
Additionally, specifically for aircraft use, an aeromedical EMC filter/ lead assembly (Part No: 525-1015) is featured to protect the aircraft from compPAC emissions.
X Summary of Safety and Effect/temess 510 k | COM200f1)
Section 2/ Page 8
smiths Simming sessent is - 8 , so t to the fymilians for a
{8}------------------------------------------------
## Summary of Safety and Effectiveness
#### Filter:
When the compleAC is operating its imernal compresson, the incoming air is filtered. The specifications of the recommended filter are as follows -
- Fifter efficiency is at least 99.99% efficient against a 0.3um Mass Median Aerodynamic # Diameter aerosol challenge at 32 L/min
- Airflow resistance at 32 Liters per Minute is 10 to 17 mm HzO .
- Connector size is 40mm DIN NATO compatible threads �
#### Intended Use:
The Pheupac compPAC 200HD ventilator is a self contained portable gas powered aumanie I he Pheapac Compt AC 200112 ventilation to patients who have respiratory distress or insufficiency. venthator measured to provice venences and transport use for adult and pediatric patients I he compt AC 2001D is sunable for chergency and mable ventilators may not be suitable.
#### Performance Data:
The design of this ventilator uses currently available technology found in many legally marketed I de design of this ventuator these carrence that the compPAC 200HD PSII Power supply! ventuators Testing Was performics to casure were safe and would perform within the environment(s) for which they are to be marketed.
Safety testing was conducted in accordance with the Draft Reviewer's Guidance for Ventilators, Salery testing was condicted in accordance Watt Inc. Drainents for emergency and transport ventilators" 1999 and EN60601-1 "Medical Electrical Equipment - Part 1 General fransport venthators - 1727 and ES (2000)ator and power supply/ charger passes all of these tests and met all requirements of the standards
and met an requirement was performed in accordance with the following standards -
- EN 60601-1-2: 1993 *
- EN794-3: 1999 を
- BS EN 61000-3-2: 1995 'Limits for harmonic current emissions (equipment input current S. * 16A per phase)'
- BS EN 61000-4-3: 1997 "Radiated, Radio frequency, Electromagnetic field immunity test
X: Summary of Safety and Effectiveress 510(x) COM200HD
{9}------------------------------------------------
## Summary of Safety and Effectiveness
#### Performance Data (ctd.):
- BS EN 61000-4-4: 1995 Electrical fast transient/ burst immunity test.
- ◆ BS EN 61000-4-2: 1995 - Electrostatic discharge immunity test. ●
- BS EN 61000-4-5: 1995 Surge immunity test
t Generic Specification for Special Purpose Aeromedical Equipment - Issue 3, specifically -Generic Specification for Special Part 3 Issue 5, the following parts
- DRE01 Radiated E Field Emissions .
- DRS02 Radiated E Field Susceptibility .
- DCE01 Conducted Emissions �
- DCS01 Conducted Susceptibility ◆
- DCS02 Conducted Susceptibility 章
- DCS10 Electrostatic Discharge 泰
Electromagnetic compatibility (EMC), electrical mechanical durability, safety (operator and Flectromagnetic companding (EMC), electrical, and been completed The results demonstrated that the compPAC 200HD ventilator, PSI | Mains' d c Power supply charger and CSI2 d.c. Power . comply/ charger complied with the guidelines and standards and that they performed within their specifications and functional requirements.
Comparison testing of the compPAC model 200HD with its predicate counterparas the compl'AC Comparison testing of the collip AC though 2001 the performance of the delivered Volume and 200 and paral AC, 200 Medic was doile to sach. The tress were performed across the ventilators. These data Frequency paramerers are the same to tack to cach tolerances of the ventilators. These casts of the ventilators. These casts PAC 200 and catire range. All meastiements were within the species to both the compPAC 200 and paraPAC 200 Medic
The testing described above indicates that there is no functional difference between the operation The resting described above multierparts the counterparts the compense the compenses it is is not of the compl AC model 200112 with its presency parameters. Based on these results, it is our determination that the device model is safe, effective and performs as well as the legally marketed
Surnmary of Safety and Effectiveness 5100 : COM200HO
Scotion 2/ Page 10
smiths
{10}------------------------------------------------
## Summary of Safety and Effectiveness
### Performance Data (cid.):
predicate devices.
This summary of 510(k) safety and effectiveness information is being submitted in accordance
s and consisted on 2017 4 1969 wel 31 CER 807 92 This summary of 570(K) safery and enver. In CFR 807.92 with the requirements of SMDA 1990 and 21 CFR 807.92
Respectfully, Frankalo Colin Walters Comi Walter."
Regulatory Affairs and Quality Assurance Manager
X: Summery of Safety and leffectiveress 510(8 ) COM200) ID
Section 2/ Page 11
smiths 20: Sau Area
{11}------------------------------------------------
Image /page/11/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo is circular and contains the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is an abstract symbol that resembles an eagle or bird in flight, composed of three curved lines.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
MAY 27 2004
Pneupac Limited C/O Mr. Donald J. Alexander Vice President of Regulatory Affairs BCI International N7 W22025 Johnson Road Waukesha, Wisconsin 53186-1856
Re: K040090
Trade Name: compPAC 200HD with PS11 and CS12 Power Supplies Regulation Number: 21 CFR 868.5925 Regulation Name: Emergency Powcred Ventilator (Resuscitator) Regulatory Class: II Product Code: BTL Dated: April 28, 2004 Received: May 3, 2004
Dear Mr. Alexander:
We have reviewed your Section 510(k) premarket notification of intent to market the device we nave reviewed your we determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate for use stated in the encreations of the enactment date of the Medical Device Amendments, or to commerce proc to rial 2011-11-11) in accordance with the provisions of the Federal Food, Drug, ucvices that have been require approval of a premarket approval application (PMA). and coometic for ( res) ( res) ( res) ( = ) = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = 1 ou may, increrery initions of the Act include requirements for annual registration, listing of general obliable provision practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it If your device is clashines tontrols. Existing major regulations affecting your device can may be subject to sater added to the 21, Parts 800 to 898. In addition, FDA may be found in the over ments concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean r lease of actives and i promination that your device complics with other requirements of the Act that 1 1971 has made a and regulations administered by other Federal agencies. You must
{12}------------------------------------------------
Page 2 – Mr. Donald J. Alexander
comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (301) 594-4646. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html.
Sincerely vours,
Chiu Lin, Ph.D.
Chiu Lin, Ph.D. Director Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{13}------------------------------------------------
### Indications for Use
510(k) Number (if known): K040090
Device Name: compPAC 200HD Emergency Powered Ventilator, PS11 Power Supply/Charger & CS12 Power Supply/Charger
Indications for Use:
The Pneupac compPAC 200HD ventilator is a self contained portable gas powered automatic gas ventilator intended to provide ventilation to patients who have respiratory ditress or insufficiency. The comPAC 200HD is suitable for emergency and transport use for adult and paediatric patients greater than 20Kg in situations where convential portable ventilators may not be suitable
Prescription Use X (Part 21 CFR 801 Subpart D) AND/OR
Over- The- Counter Use (21 CFR 807 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Ci Suin
(Division Sign Off)
Division of Anesthesiology, General Hospital,
Infection Control, Dental Devices
510(k)
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.