← Product Code [FGE](/productcode/FGE) · K040027

# BOSTON SCIENTIFIC EXPRESS BILIARY SD PREMOUNTED STENT SYSTEM (K040027)

_Boston Scientific · FGE · Mar 30, 2004 · Gastroenterology, Urology · SESU_

**Canonical URL:** https://fda-staging.innolitics.com/device/K040027

## Device Facts

- **Applicant:** Boston Scientific
- **Product Code:** [FGE](/productcode/FGE.md)
- **Decision Date:** Mar 30, 2004
- **Decision:** SESU
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 876.5010
- **Device Class:** Class 2
- **Review Panel:** Gastroenterology, Urology
- **Attributes:** Therapeutic

## Indications for Use

The Boston Scientific Express™ Biliary SD Premounted Stent System is indicated for the palliation of malignant neoplasms in the biliary tree.

## Device Story

Express Biliary SD Premounted Stent System consists of balloon-expandable 316L stainless steel stent premounted on Monorail™ delivery catheter; used for biliary strictures caused by malignant neoplasms. Physician advances catheter over guidewire (≤0.018 in) to stricture site; balloon inflation deploys stent to maintain patency. Stent features laser-cut sinusoidal bands with reinforced proximal struts to reduce tumor in-growth. Device used in clinical setting; provides palliative relief by keeping biliary ducts open. Benefits include improved bile flow in patients with malignant biliary obstruction.

## Clinical Evidence

Bench testing only. Functional testing included bile corrosion resistance, tensile/elongation, foreshortening, recoil, expansion uniformity, radial hoop strength, over-expansion, deployment accuracy, balloon inflation/deflation time, catheter tensile strength, stent securement force, and biocompatibility. No clinical data presented.

## Technological Characteristics

316L surgical grade stainless steel stent; laser-cut sinusoidal band geometry with reinforced proximal struts. Monorail™ delivery system; dual lumen coaxial catheter. Balloon-expandable. Rated burst pressure 14 atm [1419 kPa]. Compatible with guidewires ≤0.018 in. Sterilized.

## Regulatory Identification

A biliary catheter and accessories is a tubular flexible device used for temporary or prolonged drainage of the biliary tract, for splinting of the bile duct during healing, or for preventing stricture of the bile duct. This generic type of device may include a bile collecting bag that is attached to the biliary catheter by a connector and fastened to the patient with a strap.

## Special Controls

*Classification.* Class II (special controls). The device, when it is a bile collecting bag or a surgical biliary catheter that does not include a balloon component, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 876.9.

## Predicate Devices

- Johnson and Johnson Corinthian™ IQ Transhepatic Biliary Stent and Delivery System and Ultra-soft SV Balloon Dilatation Catheter

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

| MAR 3 0 2004 |
|--------------|
|--------------|

KC40027
Page 1 of 4

## 510(k) Summary

#### General Provisions 4.1

|     | Submitter's<br>Name and<br>Address | Boston Scientific, Inc.<br>One Scimed Place<br>Maple Grove MN 55311                                                                                           |
|-----|------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------|
|     | Contact<br>Person                  | Candice R. Burns<br>Regulatory Affairs Specialist<br>763-494-2845                                                                                             |
|     | Classification<br>Name             | Biliary Catheter and Accessories<br>Product Code – 78 FGE<br>Regulation Number 21 CFR Part 876.5010                                                           |
|     | Common or<br>Usual Name            | Biliary Stent and Balloon Dilatation Catheter                                                                                                                 |
|     | Proprietary<br>Name                | Boston Scientific Corporation Express Biliary SD Premounted<br>Stent System                                                                                   |
| 4.2 | Name of Predicate<br>Device        | Johnson and Johnson Corinthian™ IQ Transhepatic Biliary<br>Stent and Delivery System and Ultra-soft SV Balloon<br>Dilatation Catheter                         |
| 4.3 | Device<br>Description              | The Express Biliary SD Premounted Stent System is available<br>in the models indicated in Table 4.3. A brief description of the<br>device components follows. |

{1}------------------------------------------------

K040027

Page 2 of 4

## 510(k) Summary

#### Device Description (continued) 4.3

| Table 4.3 Device Description |                         |                               |                         |                                      |                          |
|------------------------------|-------------------------|-------------------------------|-------------------------|--------------------------------------|--------------------------|
| Balloon<br>Diameter<br>mm    | Balloon<br>Length<br>mm | Crimped<br>Stent Length<br>mm | Stent<br>Diameter<br>mm | Rated Burst<br>Pressure<br>atm [kPa] | Catheter<br>Length<br>cm |
| 4                            | 16, 20                  | 15, 19                        | 4                       | 14 [1419]                            | 90, 150                  |
| 5                            | 16, 20                  | 15, 19                        | 5                       | 14 [1419]                            | 90, 150                  |
| 6                            | 15, 19                  | 14, 18                        | 6                       | 14 [1419]                            | 90, 150                  |
| 7                            | 16, 20                  | 15, 19                        | 7                       | 14 [1419]                            | 90, 150                  |

#### 4.3.1 Stent

The Express Biliary SD stent is a balloon expandable metallic stent intended to maintain patency of biliary strictures produced by malignant neoplasms. The stent will be available in a variety of sizes to address clinician needs.

The Express Biliary SD stent is made from 316L surgical grade stainless steel. The stent geometry is a laser-cut pattern consisting of large and small sinusoidal bands connected by axial struts. Additional struts are provided to reinforce the three connector rows of the proximal stent end, resulting in additional scaffolding designed to reduce tumor in-growth at the proximal stent end.

{2}------------------------------------------------

K040027
Page 3 of 4

## 510(k) Summary

| 4.3.2 | Delivery<br>System                             | The delivery system for the Express Biliary SD Premounted<br>Stent System consists of a Monorail™ catheter with a stent<br>crimped on a balloon near the distal tip. The balloon is<br>designed to provide an inflatable segment of known diameter<br>and length at recommended pressures. The distal segment of<br>the balloon catheter is dual lumen and coaxial. The outer<br>lumen is used for inflation of the balloon. The wire lumen<br>permits the use of guidewires (≤0.018 in / 0,46 mm) to<br>facilitate advancement of the catheter to and through the<br>stricture to be dilated. |
|-------|------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| 4.4   | Intended Use                                   | The Express Biliary SD Premounted Stent System is indicated<br>for palliation of malignant neoplasms in the biliary tree.                                                                                                                                                                                                                                                                                                                                                                                                                                                                      |
| 4.5   | Summary of<br>Technological<br>Characteristics | The Express Biliary SD Premounted Stent System will<br>incorporate a substantially equivalent design, method of<br>deployment, packaging, fundamental technology,<br>manufacturing, sterilization and intended use as those featured<br>in the predicate Johnson and Johnson Legally marketed<br>Corinthian IQ Biliary Stent and Delivery System.                                                                                                                                                                                                                                              |
| 4.6   | Non-clinical Test<br>Summary                   | Functional testing for the stent component includes:<br>• Bile corrosion resistance<br>• Tensile and elongation<br>• Foreshortening<br>• Expanded stent length<br>• Recoil<br>• Expansion uniformity<br>• Compression resistance/radial hoop strength<br>• Over-expansion<br>• Deployment pressure, accuracy and diameter<br>                                                                                                                                                                                                                                                                  |

{3}------------------------------------------------

K040027
Page 4 of 4

## 510(k) Summary

#### 4.6 Non-clinical Test Summary

### (continued)

- Functional testing for the delivery system includes:
	- Repeat balloon inflation within a stent .
	- Balloon burst within the stent .
	- Balloon/delivery system distension and compliance . labeling
	- Balloon inflation and deflation time .
	- Catheter tensile strength .
	- . Stent/balloon crossing profile
	- Stent securement force .
	- . Shelf life testing
	- Biocompatibility Assessment .

Product shelf life testing will be conducted and completed prior to the commercialization of this device. The shelf-life data must show acceptable results after 3 years of accelerated age testing in order to claim the labeled shelf life of 3 years.

Based on a comparison of intended use, design, and results of in-vitro testing, the Express Biliary SD Premounted Stent System is adequate for the intended use and is considered substantially equivalent to the legally marketed Johnson & Johnson Corinthian IQ Transhepatic Biliary Stent with Delivery Catheter.

{4}------------------------------------------------

Image /page/4/Picture/13 description: The image is a black and white logo for the Department of Health & Human Services - USA. The logo features a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is a stylized symbol consisting of three curved lines, resembling a person embracing another person.

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

### MAR 3 0 2004

Ms. Candice Burns Regulatory Affairs Specialist Boston Scientific Corporation One Scimed Place MAPLE GROVE MN 55311-1566

Re: K040027

Trade/Device Name: Boston Scientific Express™ Biliary SD Premounted Stent System Regulation Number: 21 CFR \$876.5010 Regulation Name: Biliary catheter and accessories Regulatory Class: II Product Code: 78 FGE Dated: January 7, 2004 Received: January 8, 2004

Dear Ms. Burns:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act and the limitations described below. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

The Office of Device Evaluation has determined that there is a reasonable likelihood that this device will be used for an intended use not identified in the proposed labeling and that such use could cause harm. Therefore, in accordance with Section 513(i)(1)(E) of the Act, the following limitation must appear in the Warnings section of the device's labeling:

The safety and effectiveness of this device for use in the vascular system have not been established.

Furthermore, the indication for biliary use must be prominently displayed in all labeling, including pouch, box, and carton labels, instructions for use, and other promotional materials, in close proximity to the trade name, of a similar point size, and in bold print.

{5}------------------------------------------------

#### Page 2 - Ms. Candice Burns

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and permits your device to proceed to the market. This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification if the limitation statement described above is added to your labeling.

Please note that the above labeling limitations are required by Section 513(i)(1)(E) of the Act. Therefore, a new 510(k) is required before these limitations are modified in any way or removed from the device's labeling.

If you desire specific information about the application of other labeling requirements to your device (21 CFR Part 801), please contact the Office of Compliance at (301) 594-4616. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International, and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html.

Sincerely yours,

Dániel G. Schultz, M.D.

Director Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{6}------------------------------------------------

Page 1 of 1

510(k) Number: K040027

Device Name: Boston Scientific Express™ Biliary SD Premounted Stent System

FDA's Statement of the Indications For Use for device:

The Boston Scientific Express™ Biliary SD Premounted Stent System is indicated for the palliation of malignant neoplasms in the biliary tree.

Prescription Use V (Per 21 CFR 801.109)

OR

Over-the-Counter Use

. . . . . . . . .

Nancy C. Brogdon
Division Sign-Off

(Division Sign-Off) Division of Reproductive, Abdon and Radiological Devices 510(k) Number.

---

**Source:** [https://fda-staging.innolitics.com/device/K040027](https://fda-staging.innolitics.com/device/K040027)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).