K040020 · C.T.M. Homecare Product, Inc. · ITI · Feb 18, 2004 · Physical Medicine
Device Facts
Record ID
K040020
Device Name
C.T.M. POWER CHAIR, MODEL HS-1000
Applicant
C.T.M. Homecare Product, Inc.
Product Code
ITI · Physical Medicine
Decision Date
Feb 18, 2004
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 890.3860
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The C.T.M. Power Chair HS-1000 is an indoor/outdoor powered wheelchair that provides transportation for a disabled or elderly person.
Device Story
Battery-operated indoor/outdoor powered wheelchair; four-wheel base; adjustable padded seat with armrests; joystick controller. Rider operates joystick to control movement; speed proportional to joystick displacement; automatic electric braking on rear wheels engages upon joystick release or power-off. Disassembles into five components without tools for transport. Includes on-board battery charger and two 12V sealed lead acid batteries. Features free-wheeling mode for manual maneuvering. Controller provides self-diagnostic feedback via flashing power reserve indicator lights corresponding to specific system faults (e.g., motor, brake, battery). Benefits include increased mobility for users.
Clinical Evidence
Bench testing only. No clinical evaluations or comparative performance testing were submitted.
Technological Characteristics
Battery-powered (two 12V SLA batteries); two 420W motors; automatic electric braking system on rear wheels; 10" rear/7" front wheels; joystick controller (Dynamic Controls Shark Model); on-board 3A battery charger; 4.5 mph max forward speed; 2.1 mph max reverse speed; 30.7" turning radius; tool-free disassembly.
Indications for Use
Indicated for disabled or elderly persons requiring indoor/outdoor transportation via powered wheelchair.
Regulatory Classification
Identification
A powered wheelchair is a battery-operated device with wheels that is intended for medical purposes to provide mobility to persons restricted to a sitting position.
{0}------------------------------------------------
K040020
FEB 1 8 2004
## 510(k) SUMMARY
Submitter's Name: C.T.M. Homecare Product, Inc. 1663 lowa Ave. Riverside, CA 92507 (909) 788-8168
Date summary prepared:
December 23, 2003
Device name:
Proprietary name: Common or usual name: Classification name:
C.T.M. Power Chair HS-1000 Power chair. Powered wheelchair, Class II, 21 CFR 890.3860.
Legally marketed device for substantial equivalence comparison:
C.T.M. Power Chair HS-5600 submitted by Warepalmy Enterprise LLC (USA) and cleared for marketing under 510(k) "K002983.
Description of the device:
The C.T.M. Power Chair HS-1000 is an indoor/outdoor powered wheelchair that is battery operated. It has a base with four wheels, an adjustable padded seat with armrests, and a controller attached to one armrest which allows the rider to control the movement of the chair. It can be disassembled for transport and is provided with an on-board battery charger.
Intended use of device:
The device is an indoor/outdoor powered wheelchair that provides transportation for a disabled or elderly person.
Technological characteristics:
The device features and use parameters of the C.T.M. Power Chair HS-1000 and the C.T.M. Power Chair HS-5600 are very similar. Both are battery operated, have two motors, and have automatic braking systems. Batteries and charging recommendations are identical. Built-in battery chargers are provided with both wheelchairs. Use parameters are very similar, varying only in minor parameters such as turning radius.
Testing conducted:
Tests listed in the Guidance Document for the Preparation of Premarket Notification [510(k)] Applications for Mechanical and Powered Wheelchairs, and Motorized Three Wheeled Vehicles, July 1995, were conducted and the results included in the subject 510(k) submission.
Performance testing:
Comparative performance testing and clinical evaluations were not submitted as part of this 510(k).
{1}------------------------------------------------
## SECTION 3 - INTENDED USE
Intended use of device:
The C.T.M. Power Chair HS-1000 is an indoor/outdoor powered wheelchair that provides transportation for a disabled or elderly person.
Intended use of predicate device:
The C.T.M. Power Chair HS-5600 is an indoor/outdoor powered wheelchair that provides transportation for a disabled or elderly person.
Comparison:
The intended uses of the two products are identical. Each provides increased mobility for one individual, who is also the operator. Each device can be used in indoor and outdoor settings.
Labeling:
The intended use of the C.T.M. Power Chair HS-1000 is in the Owner's Manual found in Appendix II on the first page after the cover. The intended use of the HS-5600 is in the Owner's Manual found in Appendix V on the first page after the cover.
{2}------------------------------------------------
## SECTION 4 - DEVICE INFORMATION
#### Executive Summary
The C.T.M. Power Chair HS-1000 is an indoor/outdoor powered wheelchair that is battery operated. It has a base with four wheels, an adjustable padded seat with armrests, and a controller attached to one armrest which allows the rider to control the movement of the chair. It can be disassembled for transport and is provided with an on-board battery charger.
The C.T.M. Power Chair HS-1000 is a new device, not a modification of a previously cleared device. Numerous powered wheelchairs are currently on the market.
### Device Description
#### Wheelchair
The C.T.M. Power Chair HS-1000 is a battery-powered wheelchair with four wheels. The design of this wheelchair is quite similar to other power chairs that are already on the market. The wheelchair has a sturdy base, which contains the motors, provides spaces for the two batteries, and supports the padded seat. The seat has adjustable armrests and footrest. There is a controller with a joystick that attaches to either armrest and allows the rider to control the movement of the power chair. The rear wheels are 10" in diameter and the front wheels are 7". The diagrams in Appendix I show the wheelchair's features and dimensions. Diagram 1 shows photographs of the wheelchair with the major parts labeled. Diagram 2 shows front, rear, and side views of the C. T.M. Power Chair HS-1000 with major dimensions given.
If desired, the wheelchair can be disassembled to provide easier transport. There are 5 components: the base, the cover, the seat with attached controller, and two batteries. The base section weighs 86 lbs. The assembled chair without batteries weighs 120 lbs. With batteries, the chair weighs approximately 175 lbs. Assembly of the chair is simple and does not require any tools.
In the assembled wheelchair, the motors and batteries are located in the base unit. The controller/joystick unit is attached to the batteries with a cable that runs from the controller to the rear of the base. There are two motors in the base. They have a maximum 420w. There is a circuit breaker in the electrical line that will trip if electrical circuits are overloaded. Circuit breaker activation usually denotes a temporary event. The rider can push the circuit breaker and normal operation will usually be restored.
The braking system is automatic and electric. It is on the rear drive wheels only. The brakes are automatically "on" except when the wheelchair is turned on and the joystick has been moved away from the neutral position. When the joystick is released or moved back to neutral the brakes engage again. If the electrical brake system fails, the brakes will default to the closed or "brakes on" position, thereby
{3}------------------------------------------------
stopping the wheelchair. Braking time is 1.0 seconds and braking distance is about 10 feet. Diagram 3 in Appendix I shows the operating system which includes the brakes.
The C.T.M. Power Chair HS-1000 has a free-wheeling device located on the lefthand side of the base by the seat cushion. The normal position of the lever is "D". In this position the power wheelchair will not move without turning on the kev and having charged batteries. Moving the lever to the position marked "N" allows the wheelchair to be rolled freely without the key being turned on. This feature is very useful when maneuvering the wheelchair for battery charging or storage. The power chair is not intended to be ridden at all in this free-wheeling mode.
The power chair is controlled by the rider using the hand controls on the controller unit. It can be attached to either armrest. The controller unit is a Shark Model Controller from Dynamic Controls, Ltd. It is a relatively new controller that is being used on a variety of power chairs. It consists of a controller box, joystick, electrical cable, and imbedded software. Diagram 4 in Appendix I shows pictures of the controller unit with the parts labeled. The joystick is used to move the wheelchair. When it is in the central, neutral position, the wheelchair is stopped. If the joystick is pushed forward, the wheelchair moves forward. Pushing the joystick farther forward increases the speed at which the wheelchair moves. Pulling the joystick backward causes the wheelchair to move backwards, speed again proportional to the movement of the joystick. To turn the wheelchair, the joystick is moved in the direction of the desired turn. The minimum turning radius is 30.7". When the joystick is released, it automatically returns to the neutral position and the wheelchair comes to a stop.
The controller has an on-off switch, speed dial, power reserve indicator, and horn button. The on-off switch regulates the power to the unit. The wheelchair will not move unless the power is on and the batteries are charged. The speed dial adjusts the maximum speed that the wheelchair will go in either forward or reverse directions. The maximum forward speed is 4.5 mph. The maximum reverse speed is 2.1 mph. Turning the speed dial counterclockwise reduces the maximum speed, turning it clockwise increases the maximum speed. The horn button sounds the horn. The power reserve indicator shows the battery reserve. It consists of a series of red lights. When these lights are steady and completely lit, the batteries are fully charged. When only a few lights are lit, the batteries should be recharged as soon as possible.
The power reserve indicator also acts as a self-diagnostic warning light when it flashes to indicate selected problems with the power chair. Generally speaking, the wheelchair can not be driven when this light is flashing. The number of flashes helps determine the problem. The self-diagnostic signals are:
{4}------------------------------------------------
C.T.M. Power Chair HS-1000 510(k) Notification Page 8
| Number of flashes | Problem |
|-------------------|---------------------------------------------|
| 0 - no lights | battery is not connected correctly |
| 1 | user fault - release joystick and try again |
| 2 | battery fault |
| 3 | left motor fault |
| 4 | right motor fault |
| 5 | left park brake fault |
| 6 | right park brake fault |
| 7 | controller unit fault |
| 8 | power module fault |
| 9 | communications fault |
| 10 | unknown fault |
| 11 | incompatible controller unit |
Detailed explanations of these problems and solutions to them are given in the Troubleshooting section of the Owner's Manual for the power chair, which can be found in Appendix II.
The battery charger is on-board. It is a 3A charger. It plugs into a standard 110 volt wall socket or into a 220 volt socket.
#### Batteries
The C.T.M. Power Chair HS-1000 runs on two 12 volt SLA (sealed lead acid) batteries. These batteries are provided with the wheelchair. Replacements can be obtained from the distributor or other local sources.
#### Accessories
There are no optional accessories at this time.
#### Regulatory Status
The C.T.M. Power Chair HS-1000 is a new device that has not been previously submitted to the FDA.
{5}------------------------------------------------
DEPARTMENT OF HEALTH & HUMAN SERVICES
Image /page/5/Picture/1 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is a stylized representation of three human figures, depicted as abstract shapes, suggesting a sense of community and support.
Public Health Service
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
FEB 1 8 2004
C.T.M. Homecare Product, Inc. C/o Mr. Robert S. McQuate R.S. McQuate & Associates, Inc. 1630 Dutch Ravine Court Reno, Nevada 89521
Re: K040020
Trade/Device Name: C.T.M. Power Chair HS-1000 Regulation Number: 21 CFR 890.3860 Regulation Name: Powered wheelchair Regulatory Class: II Product Code: ITI Dated: December 29, 2003 Received: January 6, 2004
Dear Mr. McQuate:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CI'R Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the elcctronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{6}------------------------------------------------
Page 2 - Mr. Robert S. McQuate
This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (301) 594-4659. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html
Sincerely vours.
Mark A. Melleusen
Celia M. Witten, Ph.D., M.D Director Division of General, Restorative and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{7}------------------------------------------------
C.T.M. Power Chair HS-1000 510(k) Notification Page 3
# Indications for Use Statement
510(k) Number ( if known):
Device name: C.T.M. Power Chair HS-1000
Indications for Use:
The C.T.M. Power Chair HS-1000 is an indoor/outdoor powered wheelchair that provides transportation for a disabled or elderly person.
(Please do not write below this line)
-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------- is a very long line.
Concurrence of CDRH, Office of Device Evaluation (ODE)
Prescription Use
(Per 21 CFR 801,109)
OR Over-The-Counter Use
(Division Sign-Off)
Division of General, Restorative,
and Neurological Devices
510(k) Number K040020
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.