K040015 · Alm S.A. · FSY · Jan 21, 2004 · General, Plastic Surgery
Device Facts
Record ID
K040015
Device Name
MODIFICATION TO PRISMALIX SURGICAL LIGHT
Applicant
Alm S.A.
Product Code
FSY · General, Plastic Surgery
Decision Date
Jan 21, 2004
Decision
SESE
Submission Type
Special
Regulation
21 CFR 878.4580
Device Class
Class 2
Indications for Use
The PrismAlix® Surgical Light with the upgraded Energix™ Power Supply is intended to be used to provide visible illumination of the surgical area or the patient. The Energix™ Power Supply is intended for use with surgical light systems.
Device Story
Surgical light system; provides illumination for surgical area/patient. Upgraded Energix™ Power Supply incorporates software for lighthead control; features touch-pad units for on/off switching and intensity adjustment. System includes LED indicators for illumination level, bulb failure sensing, and power mode (mains/battery). Optional RS232 communication motherboard enables data transmission to remote LCD display or PC. Data includes maintenance schedules, operating parameters, hours of operation, bulb life, self-diagnosis, and voltage/current/temperature readings. Used in clinical/surgical settings; operated by surgical staff. Benefits include improved control of illumination, enhanced monitoring of system health, and simplified maintenance tracking.
Clinical Evidence
No clinical data required for this device classification submission; bench testing only.
Technological Characteristics
Surgical luminaire system. Energix™ Power Supply (WPS) includes power modules, filter PCB, and regulators. Connectivity via RS232 for remote monitoring. Software-controlled intensity regulation. Tested to UL 2601-1, CSA C22.2 No. 601.1, IEC 60601-2-41, and EN 60601-1-2. Software classified as Minor Level of Concern.
Indications for Use
Indicated for providing visible illumination of the surgical area or the patient during surgical procedures.
Regulatory Classification
Identification
A surgical lamp (including a fixture) is a device intended to be used to provide visible illumination of the surgical field or the patient.
Special Controls
*Classification.* Class II (special controls). The device, when it is an operating room lamp, a surgical instrument light, a surgical floor standing light, an endoscopic surgical light, a surgical light connector, a ceiling mounted surgical light, a surgical light carrier, surgical light accessories, a surgical lamp, a remote illuminator, or an incandescent surgical lamp, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 878.9.
Predicate Devices
ALM PrismAlix® (PRX) Series Surgical Light (K982063)
Submission Summary (Full Text)
{0}------------------------------------------------
KO40015 (pg 1 of 3)
# 510(k) SUMMARY
## ALM PrismAlix® (PRX) Surgical Light with Energix™ Power Supply
| Submitted by: | Getinge USA, Inc. (as ALM S.A.'s US Agent)<br>1777 E Henrietta Road<br>Rochester, NY 14623-3133 |
|------------------------|-------------------------------------------------------------------------------------------------|
| Contact Person: | Frederick R. Catt<br>Senior Regulatory Engineer<br>Phone: (585) 272-5013<br>Fax: (585) 272-5299 |
| Date prepared: | December 31, 2003 |
| Proprietary Name: | ALM PrismAlix® (PRX) Surgical Light |
| Common Name: | Surgical Light |
| Device Classification: | Surgical Lamp (78 FSY)<br>Class II, as listed per 21 CFR 878.4580 |
| Predicate Device: | ALM PrismAlix® (PRX) Series Surgical Light [K982063] |
#### Description of Modified Device:
The PrismAlix® (PRX) Surgical Light System is an existing product line that will allow modification via the use of an updated Energix™ Power Supply as its means to control the surgical lighthead illumination levels. The updated Energix™ Power Supply contains software that enables additional features and capabilities not available on the earlier version of Energix™ Power Supplies. These include:
- a) The power supply and regulators,
- b) one or two touch-pad units that provide the user the ability to:
- i) turn on/off one or two lightheads,
- ii) adjust illumination intensity levels and
- iii) additional Light Emitting Diodes (LEDs) are used to indicate:
- (1) the illumination level,
- (2) bulb failure sensing and
- (3) power supply mode: Mains supply: green LED or the switching to a battery back-up system results in the LED indicator changing its color to red.
{1}------------------------------------------------
K040015 (pg 2 of 3)
ALM S.A. FDA 510(k) Summary Power Surgical Light with the upgraded Energix™ Power Supply
An optional RS232 Communication Motherboard is also a new feature that allows shifting An optional ROZOE Ochinanicaa specified distance (50m, 164 ft.) away from the surgical Controls to a romote bite are provided specific user and maintenance type information/data, displayed on a Liquid Crystal Display (LCD) window located on the front information and the Energix™ power supply. The user also has an option to have data communicated to a personal computer (PC).
The LCD can show information about preventive maintenance type items, including:
- 1. programming of maintenance inspections
- 2. storage of operating parameters
- 3. hours of operation
- 4. bulb life data
- 5. self-diagnosis information
- 6. direct reading of parameters voltage, current, temperature
- 7. traceable management data serial number, software version number, production and service dates)
#### Table 1
| Energix™ Power Supply – Designations/Configurations | |
|-----------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------|
| Model | Description |
| WPS443 | Example: WPS Power Supply with two 360W Power Modules, two keypads and a RS232 Communication Motherboard |
| WPS | Base Model Designation [Transformer; plus filter PCB and regulator(s)] |
| | 2 150W Power Module and control keypad |
| | 4 360W Power Module and control keypad |
| | 0 Space holder – No #2 Power Module and 2nd control keypad (single lighthead) |
| | 2 150W Power Module #2 and 2nd control keypad |
| | 4 360W Power Module #2 and 2nd control keypad |
| | 0 Power Module(s) only |
| | 1 Power Module(s), RS232 Communication Motherboard and Battery Backup system/charger (Note: not currently planned for USA distribution) |
| | 3 Power Module(s), RS232 Communication Motherboard |
{2}------------------------------------------------
K040015 (pg 3 of 3)
ALM S.A. FDA 510(k) Summary Device: PrismAlix® Surgical Light with the upgraded Energix ™ Power Supply
Page 3 of 3 December 31, 2003
#### Intended Use:
The PrismAlix® Surgical Light with the upgraded Energix™ Power Supply is intended to be used to provide visible illumination of the surgical area or the patient. The Energix™ Power Supply is intended for use with surgical light systems.
#### Nonclinical Comparisons to Predicate Device
The PrismAlix® Surgical Light with the upgraded Energix™ Power Supply (subject device) is similar to the predicate device with the following modifications:
- Replacement of the older Energix™ design with an upgraded Energix™ Power . Supply (WPS) that includes use of software, and makes available a low energy remote touch-pad panel per lighthead.
- As the RS232 option, the Energix™ can provide the user with product condition . type information via a LCD screen or interfacing with a remote PC.
- Software is incorporated within the Energix™ design, which interfaces with input . signals sent from the dimmer control and ON/OFF lamp touch-pads. These switches provide output signals used by the intensity control regulator supplying voltage to the lamps (bulbs) within the corresponding lighthead.
- The Energix™ Power Supply can operate up to two lightheads, rated 360W each. ●
#### Test Data:
The test data supports conformance to:
- UL 2601-1 Standard for Medical Electrical Equipment, Part 1: General Requirements . for Safety
- CSA C22.2 No. 601.1 Medical Electrical Equipment, Part 1: General Requirements for . Safety
- IEC 60601-2-41 Medical electrical equipment Part 2-41: Particular requirements for . the safety of surgical luminaires and luminaires for diagnostics
- EN 60601-1-2 Medical electrical equipment Part 1-2: General requirements for safety . - Collateral standard: Electromagnetic compatibility - Requirements and tests
- Software used in the Energix™ Power Supply was tested according to the appropriate . FDA Software Guidance Documents, per its determination as a Minor Level of Concern.
#### Clinical Data:
No clinical data is required for this device classification submission.
#### Conclusion:
Based upon the information provided herein this 510(k) Premarket Notification, we conclude that the ALM PrismAlix® (PRX) Series Surgical Light System using the updated Energix™ Power Supply is substantially equivalent to the predicate device(s) and is safe and effective when used as intended.
{3}------------------------------------------------
Image /page/3/Picture/1 description: The image shows the seal of the Department of Health & Human Services - USA. The seal is circular and contains the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. In the center of the seal is an abstract image of an eagle.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
JAN 2 1 2004
ALM S.A. c/o Mr. Frederick R. Catt Sr. Regulatory Engineer Getinge USA, Inc. 1777 East Henrietta Road Rochester, New York 14623
Re: K040015
Trade/Devicc Name: PrismAlix®(PRX) Surgival Light with Energix™ Power Supply Regulation Number: 21 CFR 878.4580 Regulation Name: Surgical Lamp Regulatory Class: II Product Codes: FSY Dated: December 31, 2003 Received: January 5, 2004
Dear: Mr. Catt:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments. or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complics with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{4}------------------------------------------------
Page 2 - Mr. Frederick Catt
This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (301) 594-4659. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html
Sincerely yours,
Miriam C. Provost
A Celia M. Witten, Ph.D., M.D. Director Division of General, Restorative and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
### Indications for Use
510(k) Number (if known): K040015
Device Name: PrismAlix® (PRX) Surgical Light with Energix™ Power Supply
Indications for Use:
ALM S.A. PrismAlix® (PRX) Surgical Lights with Energix™ Power Supply are intended to be used to provide visible illumination of the surgical area or the patient. The Energix™ Power Supply is intended for use with surgical light systems.
Prescription Use (Part 21 CFR 801 Subpart D)
AND/OR
Over-The-Counter Use (21 CFR 807 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Miriam C. Provost
Page 1 of _1
Sister Sign-Ofly inion of General, Restorative . Neurological Devices
K040015
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.