← Product Code [DGC](/productcode/DGC) · K034057

# LACRYTEST (K034057)

_Adiatec SA · DGC · Mar 9, 2004 · Immunology · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K034057

## Device Facts

- **Applicant:** Adiatec SA
- **Product Code:** [DGC](/productcode/DGC.md)
- **Decision Date:** Mar 9, 2004
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 866.5510
- **Device Class:** Class 2
- **Review Panel:** Immunology

## Indications for Use

Lacrytest is a rapid immunoassay for the total IgE detection in tears. Normal values are below 2,5 kUI/l, (3 ng/ml). This assay indicates in a qualitative way the presence of total class E immunoglobulins in tears, with level above normal value. Lacrytest offers help for ocular allergy diagnosis in a simple and fast way. Lacrytest is to be used with IgE mediated allergic ocular disorders. It allows knowing if an ocular inflammation, conjunctivitis, or dry eyes symptom could be linked to allergic ocular disorder. This test is dedicated to the physicians and clinical Laboratory.

## Device Story

Lacrytest is a rapid, single-step, immuno-chromatographic test strip for detecting total IgE in human tears. The device uses monoclonal and polyclonal antibodies to capture IgE from a single teardrop collected directly from the patient's eye. The strip is placed in contact with the eye for less than three minutes; an indicator confirms sufficient tear fluid absorption. After soaking the strip in sterile water, results appear as visible red-violet lines indicating IgE levels relative to a 2.5 kIU/L threshold. Used in physician offices, clinical laboratories, or during home visits, the test provides a semi-quantitative assessment of allergic ocular inflammation. Results are read visually by the clinician, offering a non-invasive alternative to the Schirmer assay. The test aids in the rapid diagnosis of allergic conjunctivitis, enabling faster clinical decision-making and patient treatment.

## Clinical Evidence

Clinical performance evaluated against the UniCAP reference method using 165 human serum samples. Results demonstrated a sensitivity of 91.7% and a specificity of 98.5%. Semi-quantitative accuracy was 100% except near the upper limit of intensity 2 (30 kIU/L). Precision studies compared intra-run and inter-run variance across different IgE levels. 87% of healthy subjects tested negative (<2.5 kIU/L).

## Technological Characteristics

Ready-to-use immuno-chromatographic test strip. Materials include nitrocellulose, polyester resin fibers, cellulose, and adhesive. The ocular contact membrane consists of polyester fibers and a surfactant (Pall Profile Filter/Bioinert Filter). Reagents include monoclonal/polyclonal antibodies, trihydrated tetrachloroauric acid, sodium azide, polyethylene glycol 20000, hydrolyzed polyvinyl alcohol, sucrose, and thimerosal. Detection is based on visual colorimetric change (red-violet lines). No instrumentation required.

## Regulatory Identification

An immunoglobulins A, G, M, D, and E immunological test system is a device that consists of the reagents used to measure by immunochemical techniques the immunoglobulins A, G, M, D, an E (serum antibodies) in serum. Measurement of these immunoglobulins aids in the diagnosis of abnormal protein metabolism and the body's lack of ability to resist infectious agents.

## Predicate Devices

- UNICAP TOTAL IGE ASSAY/PHARMACIA TOTAL IGE CONTROL ([K964152](/device/K964152.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}

510(k) SUBSTANTIAL EQUIVALENCE DETERMINATION
DECISION SUMMARY

A. 510(k) Number:
k034057

B. Analyte:
Total IgE

C. Type of Test:
Semi-quantitative immunoassay on test strip

D. Applicant:
Adiatec S.A.

E. Proprietary and Established Names:
Proprietary name: Lacrytest; Common name: IgE immunoassay

F. Regulatory Information:
1. Regulation section:
21 CFR § 866.5510 Immunoglobulins A, G, M, D, E immunological test system
2. Classification:
Class II
3. Product Code:
DGC – IgE antigen, antiserum, control
4. Panel:
Immunology (82)

G. Intended Use:
Lacrytest is a rapid immunoassay for the detection of total IgE in tears. Semi-quantitative detection of total IgE in tears (&lt; 2.5 kIU/liter, 2.5 - 10 kIU/liter, 10 - 40 kIU/liter, and &gt; 40 kIU/liter) indicates local IgE production associated with allergic conjunctivitis. This test is used in the physician office and professional clinical laboratories.
1. Indication(s) for use:
Local (in tears of the eye) IgE production associated with allergic conjunctivitis
2. Special condition for use statement(s):
For prescription use only.
3. Special instrument Requirements:
None

H. Device Description:
Ready-to-use test strip containing reagents to capture and visualize total IgE in tears.

{1}

Page 2 of 6

# I. Substantial Equivalence Information:

1. Predicate device name(s): Pharmacia Unicap total IgE assay
2. Predicate K number(s): k964152
3. Comparison with predicate

|  Similarities  |   |   |
| --- | --- | --- |
|  Item | Device | Predicate  |
|  Antigen Capture material | Anti-human IgE | Anti-human IgE  |
|  Differences  |   |   |
|  Item | Device | Predicate  |
|  Visualization of immobilized IgE | Gold-labeled anti-human IgE | Enzyme labeled anti-human IgE + substrate + chromogenic reagent  |
|  Sample matrix | Tears | Serum or plasma  |
|  Separation of bound and free analyte | Chromatographic separation on solid phase | Mechanical separation after absorption on solid phase  |
|  Instrument required | No | UniCap 100  |
|  Analytical assessment | Qualitative/semi-quantitative in 4 concentrations (< 2.5 kIU/liter, 2.5 - 10 kIU/liter, 10 - 40 kIU/liter, and > 40 kIU/liter | Quantitative  |

# J. Standard/Guidance Document Referenced (if applicable):

None referenced

# K. Test Principle:

The device is an immunoassay with all reagents embedded on a test strip that performs without the use of instruments. Total IgE in human tears is bound by anti-human IgE immobilized on test strips in the IgE reactive field of the test strip. Visualization of captured IgE is indicated by colloidal gold labeled anti-human IgE. Excess labeled antibodies are captured by antibodies in the control field. A control line appears in the control field and determines when enough tears have been collected. The presence of IgE in the sample is visualized by the appearance of a dark red line in the reactive field and the control field. Absence of a line in the control field, regardless of the presence of line in the reactive field, indicates an invalid assay result. The area on the support identifying the reactive field and control field are on

{2}

the back surface of the support in two colors. A sample with IgE in the normal reference range lacks a line in the reactive field but has a line in the control field. The presence of lines in the reactive and control fields indicates a sample with elevated total IgE levels. The device provides a qualitative/semi-quantitative measure of total IgE in tears when the intensity of color on the test strip is visually observed. Four possible categories of results are possible: negative;  $&lt; 2.5\mathrm{kUI / l}$ ,  $2.5 - 10\mathrm{kUI / l}$ ,  $10 - 40\mathrm{kIU / liter}$ , and  $&gt;40\mathrm{kUI / l}$ .

# L. Performance Characteristics (if/when applicable):

# 1. Analytical performance:

# a. Precision/Reproducibility:

For precision, the sponsor lists the following intra-run and inter-run precision on 3 samples

|  intra-run | Sample 1 | Sample 2 | Sample 3  |
| --- | --- | --- | --- |
|  number of determinations | 30 | 30 | 30  |
|  total IgE levels KUI/L | 5 | 20 | 50  |
|  ranges (intensities) | 0-1 | 2-3 | 3-3  |
|  mean value (intensity) | 1 | 2 | 3  |
|  variance % | 7 | 3 | 0  |
|  inter-run | Sample 1 | Sample 2 | Sample 3  |
| --- | --- | --- | --- |
|  number of determinations | 30 | 30 | 30  |
|  total IgE levels KUI/L | 5 | 20 | 50  |
|  ranges (intensities) | 0-1 | 2-3 | 3-3  |
|  mean value (intensity) | 1 | 2 | 3  |
|  variance % | 10 | 4 | 0  |

No description is given on the precision studies.

# b. Linearity/assay reportable range:

None provided

# c. Traceability (controls, calibrators, or method):

Correspondence to concentrations of the WHO Third International standard reference preparation is asserted. Artificial tear samples with defined amounts of added IgE (2, 7.5, 30 and  $50\mathrm{kIU / l}$ ) were tested in multiple replicates by three professional lab technicians. A single manufacturing lot of test strips was used for the test. The lab testers were blinded to the IgE concentration of the samples. For samples with  $2\mathrm{kIU / l}$ ,  $7.5\mathrm{kIU / l}$ , and  $50\mathrm{kIU / l}$ , the accuracy was  $100\%$ . For a sample with  $30\mathrm{kIU / l}$ , the accuracy was  $93\%$ . Since the accuracy (comparison to target value) is above  $90\%$  for each semi-

{3}

quantitative range, it can be concluded that semi-quantitative measurement using the strips corresponds with the WHO IgE reference preparation concentrations and imply correspondence with other commercially available total IgE assays using this standard.

d. Detection limit:

Three different dilutions of a control serum from an IgE assay from a different manufacturer were prepared in tears without detectable levels of IgE. One dilution contained an expected value of 2 kIU/l, one dilution contained an expected value of 2.5 kIU/l, and one dilution contained an expected value of 3 kIU/l. Additionally, a human serum sample was diluted with a pool of tears to give an expected value of 2.5 kIU/l. Twenty replicates of each dilution were tested with the assay test strips. For the sample with an expected value of 2 kIU/l, 19 of 20 replicates (95%) tested negative with the test strips. Samples with an expected value of 2.5 kIU/l or 3 kIU/l tested negative from 2 to 9 out of 20 replicates (i.e. from 10% to 45% of 20 replicates tested negative). The claimed detection limit is 2.5 kIU/l.

e. Analytical specificity:

No information provided

f. Assay cut-off:

No information provided

2. Comparison studies:

a. Method comparison with predicate device:

One hundred and sixty-five samples were tested in the proposed and predicate assay for total IgE levels using the tear matrix. A single lot of the proposed assay was used. Testers were blinded to the clinical status and results of the other test method. The following table summarizes results:

{4}

Page 5 of 6

|  Lacrytest | Pharmacia UniCAP  |   |   |
| --- | --- | --- | --- |
|   |  Positive > 2.5 kIU/l | Negative < 2.5 kIU/l | total  |
|  Positive | 89 | 1 | 90  |
|  negative | 8 | 67 | 75  |
|  Total | 97 | 68 | 165  |
|  Exact binomial 95% Conf. limit  |   |   |   |
|  observed agreement= | 0.945 | 0.899 to 0.975 |   |
|  expected agreement= | 0.508 |  |   |
|  Kappa = | 0.889 |  |   |
|  std error = | 0.0776 |  |   |
|  P(kappa=0)= p<0.001  |   |   |   |

The probability that the observed agreement is significantly higher than chance agreement is $&lt; 0.001$. This indicates significantly better agreement between test results of the proposed and predicate assay than chance agreement. The probability that observed agreement is equivalent with perfect agreement (kappa = 1) was 0.076. Since the probability is not less than 0.05, then it can be concluded that agreement is equivalent with perfect agreement.

Among 68 samples tested in both assays, agreement of negative results in both assays was $98.5\%$. Among 25 samples with values between 2.5 and $10\mathrm{kIU / l}$, agreement in both assays was $85.7\%$. Among 18 samples with values between 10 and $40\mathrm{kIU / l}$, agreement was $76\%$. Among 44 samples with values greater than $40\mathrm{kIU / l}$, agreement was $90.9\%$. Agreement of positive results in both assays is estimated to be $82\%$ (exact binomial confidence interval $73\% - 89.6\%$), the numerical average of agreement in the 4 categories.

To compare test results of physicians at the point-of-care with test results of professional lab technicians, 3 sites tested samples. At each site a physician tested 5 negative and 5 positive samples. At each site 3 lab technicians also tested the same positive and negative samples. Negative samples were randomly selected subjects visiting a physician office without allergy symptoms. Positive samples were randomly selected patients visiting a physician's office with clinical evidence of seasonal allergy, acute allergy, or periodic allergy. A quality control material prepared from a sample of the WHO standard reference material (75/502) was provided to each site. The agreement of test results among negative samples when tested by physician vs lab technician was $93\%$, $93\%$, and $87\%$ at each of the three sites. The agreement of test results among positive samples

{5}

Page 6 of 6

when tested by physician vs lab technician was 83%, 89%, and 100% at each of the test sites. The approximate agreement of physician test result with lab technician test result for all samples was 90%.

b. Matrix comparison:
No information comparing IgE concentrations from serum and tears was provided

3. Clinical studies:
a. Clinical sensitivity:
No information provided
b. Clinical specificity:
No information provided
c. Other clinical supportive data (when a and b are not applicable):

4. Clinical cut-off:
No information provided

5. Expected values/Reference range:
One hundred patient samples were randomly collected from subjects visiting a physician. Patients were absent allergy symptoms, allergy history, and ocular inflammation. Of the 100 subjects, 87 subjects had results less than 2.5 kIU/l. Thirteen patients had concentrations from 2.5 to 10 kIU/l. From this data, 85% of normal subjects had truly negative test results. The exact binomial 95% confidence interval for the normal subjects is 76.5 – 91.4%. Therefore, it can be concluded that approximately 90% of normal subjects will have negative test results.

M. Conclusion:
Based on information included in the file, the Lacrytest can be recommended as substantially equivalent to the predicate device, Pharmacia Unicap total IgE assay, regulated under 21 CFR 866.5510, Immunoglobulins A, G, M, D, E, immunoglobulin test system (class II, product code - DGC IgE antigen, antiserum, control).

---

**Source:** [https://fda-staging.innolitics.com/device/K034057](https://fda-staging.innolitics.com/device/K034057)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
