K033988 · Goldway (Us), Inc. · MWI · Feb 25, 2004 · Cardiovascular
Device Facts
Record ID
K033988
Device Name
GOLDWAY UT4000A VITAL SIGNS MONITOR
Applicant
Goldway (Us), Inc.
Product Code
MWI · Cardiovascular
Decision Date
Feb 25, 2004
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 870.2300
Device Class
Class 2
Attributes
Pediatric
Indications for Use
Goldway UT4000A Vital Signs Monitor is intended to monitor basic physiological parameters of patients from adults to neonates within any professional medical environment. The user, responsible to interpret the monitored data made available, will be a licensed healthcare practitioner. Physiological data, system alarms. and patient data analysis will be made available to the user from the monitor. Physiological data includes but is not restricted to: electrocardiograph, noninvasive blood pressure, and pulse oximetry. Goldway UT4000A is not intended for use as an apnea monitor. Goldway UT4000A is not intended for use during MRI or CT scans. The UT4000A is not designed for home use. The device is restricted to be used on one patient at a time.
Device Story
Goldway UT4000A is a patient monitor for professional medical environments; operated by licensed healthcare practitioners. Device inputs include ECG, NIBP, and SpO2 sensors. System processes physiological signals to display real-time waveforms and numerical data on a 5.6-inch color Active TFT LCD. Provides limit alarms and alerts when parameters exceed thresholds; stores data trends for retrospective review; supports serial port output for printing. Benefits include continuous patient status monitoring to assist clinical decision-making. Device is restricted to single-patient use.
Clinical Evidence
No clinical data. Bench testing performed to verify compliance with standards including ISO 9001, ISO 13485, ISO 10993-1, IEC 60601-1, IEC 60601-1-2, IEC 60601-1-4, IEC 60878, IEC 60601-2-27/30/34/49, AAMI SP10, EC 11, and EC 23.
Technological Characteristics
Physiological monitor with 5.6-inch color Active TFT LCD. Monitors ECG, NIBP, SpO2. Molded plastic housing. Standards: ISO 9001, ISO 13485, ISO 10993-1, IEC 60601-1 series, AAMI SP10, EC 11, EC 23. Provides audible/visual alarms. Connectivity via serial port for printing.
Indications for Use
Indicated for monitoring basic physiological parameters (ECG, NIBP, SpO2) in patients ranging from neonates to adults in professional medical environments. Contraindicated for use as an apnea monitor or during MRI/CT scans. Not for home use.
Regulatory Classification
Identification
A cardiac monitor (including cardiotachometer and rate alarm) is a device used to measure the heart rate from an analog signal produced by an electrocardiograph, vectorcardiograph, or blood pressure monitor. This device may sound an alarm when the heart rate falls outside preset upper and lower limits.
{0}------------------------------------------------
<033988
p 11/3
FEB 2,5 2004
#### 510(k) Summary
# 1. Submitter's Identification:
Lee & Xiao 2600 Mission Street, Suite 100 San Marino, CA 91108 Tel: (626)799-0998 Fax: (626)755-1588
Contact Person: Yingchao Xiao, Esq.
Date: December 5, 2003
## 2. Device Name:
Trade Name: Goldway UT4000A Vital Signs Monitor Common Name: Patient Monitor Classification Name: Monitor, Physiological, Patient
#### 3. Predicate Device Information:
The legally marketed device to which the submitter claims equivalence:
Goldway Model UT4000F Patient Monitor (K021154)
#### 4. Device Description:
The Goldway UT4000A monitor interprets and displays physiological data of the patient including waveforms and numerical data in real time. Goldway UT4000A can be custom configured to monitor electrocardiograph (ECG), non-invasive blood pressure (NIBP), or pulse oximetry (SpO2). For each patient vital parameter, Goldway UT4000A will be capable of providing limit alarms and alerts, storing data trends for retrospective review and sending serial port signal to printer for report printing.
#### 5. Intended Use:
Goldway UT4000A Vital Signs Monitor is intended to monitor basic physiological parameters of patients from adults to neonates within any professional medical environment. The user, responsible to interpret the monitored data made available, will be a licensed healthcare practitioner. Physiological data, system alarms. and patient data analysis will be made available to the user from the monitor.
{1}------------------------------------------------
Physiological data includes but is not restricted to: electrocardiograph, noninvasive blood pressure, and pulse oximetry.
Goldway UT4000A is not intended for use as an apnea monitor. Goldway UT4000A is not intended for use during MRI or CT scans.
The UT4000A is not designed for home use. The device is restricted to be used on one patient at a time.
# 6. Comparison with Predicate Device:
Shenzhen Goldway Industrial. Inc. has been developing and distributing physiological monitoring devices since 1995. The Substantial Equivalence of the subject device of this 510(k) notification, GoldwayUT4000A Vital Signs Monitor ("the subject device"), is claimed to Goldway Model UT4000F Patient Monitor ("the predicate device"). The predicate device is a Class II (two) device and was granted market clearance by FDA on April 11, 2003 with 510(K) control number K021154. The predicate device is designed to monitor ECG, NIBP, SpO2, IBP, ETO2, temperature, and respiration, while the subject device is for monitoring ECG, NIBP and SpO2 only. The patient data collected by the predicate device is displayed for the user on a 10.4-inch color Active TFT LCD, while the subject device is displayed on a 5.6-inch color Active TFT LCD. The packaging design of the predicate device and the subject device uses the same molded plastic. Both devices incorporate audible and visual alarm functions that are activated when set parameter limits are exceeded.
# 7. Discussion on Non-Clinical Tests Performed:
UT4000A meets the following standards:
- . ISO 9001
- ISO 13485 ●
- ISO 10993-1 .
- IEC 60601-1
- IEC 60601-1-2 ●
- IEC 60601-1-4 ●
- IEC 60878 ●
- . IEC 60601-2-27/30/34/49
- . AAMI SP10
- . EC 11
- EC 23 .
{2}------------------------------------------------
KOB,39 37 p 3/3
# 8. Discussion on Clinical Tests Performed:
Not Applicable.
#### 9. Conclusion:
:
Goldway UT4000A Vital Signs Monitor has the same intended use and similar technological characteristics as Goldway Model UT4000F Patient Monitor. Moreover, the comparisons between the subject device and the predicate device demonstrate that any of the differences in their technological characteristics do not raise any questions in terms of the subject device's safety and effectiveness. Therefore, the Goldway UT4000A Vital Signs Monitor is substantially equivalent to the predicate device.
{3}------------------------------------------------
Image /page/3/Picture/11 description: The image shows the seal of the Department of Health & Human Services (HHS). The seal features a stylized eagle with three overlapping wings, representing health, human services, and well-being. The words "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" are arranged in a circular pattern around the eagle.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
FEB 2 5 2004
Goldway (US), Inc. c/o Mr. Yingchao Xiao, Esq. Lee & Xiao Attorneys Crestmont Building 2600 Mission Street, Suite 100 San Marino, CA 91108
Rc: K033988
Trade Name: Goldway UT4000A Vital Signs Monitor Regulation Number: 21 CFR 870.2300 Regulation Name: Cardiac Monitor (including cardiotachometer and rate alarm) Regulatory Class: II (two) Product Code: MWI Dated: December 05, 2003 Received: December 23, 2003
Dear Mr. Xiao:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adultcration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{4}------------------------------------------------
Page 2 - Mr. Yingchao Xiao, Esq.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (301) 594-4648. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html
Sincerely yours,
Duna R. Vahner
A Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
# Indications for Use
510(k) Number (if known): K033988
# Device Name: Goldway UT4000A Vital Signs Monitor
Indications For Use:
Goldway UT4000A Vital Signs Monitor is intended to monitor basic physiological parameters of patients from adults to neonates within any professional medical environment. The user, responsible to interpret the monitored data made available, will be a licensed healthcare practitioner. Physiological data, system alarms, and patient data analysis will be made available to the user from the monitor.
Physiological data includes: electrocardiograph, non-invasive blood pressure, and pulse oximetry.
Goldway UT4000A is not intended for use as an apnea monitor. Goldway UT4000A is not intended for use during MRI or CT scans.
The UT4000A is not designed for home use. The device is restricted to be used on one patient at a time.
Prescription Use _____________________________________________________________________________________________________________________________________________________________ (Part 21 CFR 801 Subpart D) AND/OR
Over-The-Counter Use - - -(21 CFR 807 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Donna D. Lochner
Page 1 of ____________________________________________________________________________________________________________________________________________________________________
Division Sign-Off) Division of Cardiovascular Devices
510(K) Number_K033988
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.