SMITH & NEPHEW ELECTROTHERMAL 20S SPINE GENERATOR, MODEL 7209975

K033981 · Smith & Nephew, Inc. · GEI · Feb 25, 2004 · General, Plastic Surgery

Device Facts

Record IDK033981
Device NameSMITH & NEPHEW ELECTROTHERMAL 20S SPINE GENERATOR, MODEL 7209975
ApplicantSmith & Nephew, Inc.
Product CodeGEI · General, Plastic Surgery
Decision DateFeb 25, 2004
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 878.4400
Device ClassClass 2
AttributesTherapeutic

Indications for Use

The Smith & Nephew ElectroThermal® 20S Spine System is intended to create lesions in nervous tissue, and to coagulate and decompress disc material when used in combination with Smith & Nephew thermal/coagulating probes. The generator and accessories are intended for use by qualified medical personnel trained in the use of electrosurgical equipment.

Device Story

20-watt electrothermal generator; provides RF energy for thermal heating of specialized catheters (SpineCATH, Decompression Catheter, RF Denervation Probe). Inputs: temperature and impedance monitoring signals from probes. Operation: delivers temperature-controlled RF energy; uses biphasic square wave for stimulation voltage; maintains effective tissue heating via preset time and temperature settings. Used in clinical settings by trained medical personnel. Output: controlled thermal energy for lesion creation, coagulation, and decompression. Benefits: enables precise, temperature-controlled minimally invasive spine procedures.

Clinical Evidence

Bench testing only. Performance testing demonstrated substantial equivalence to predicate devices regarding the delivery of temperature and time-controlled RF energy.

Technological Characteristics

20-watt electrosurgical generator. Sensing/actuation: temperature and impedance monitoring. Energy source: RF energy; biphasic square wave for stimulation. Form factor: standalone generator unit. Connectivity: compatible with specific Smith & Nephew thermal/coagulating probes. Control: temperature-controlled feedback loop.

Indications for Use

Indicated for patients requiring lesion creation in nervous tissue, or coagulation and decompression of disc material. Intended for use by qualified medical personnel trained in electrosurgical equipment. Contraindicated when, in the physician's judgment, an electrosurgical procedure is not in the patient's best interest.

Regulatory Classification

Identification

An electrosurgical cutting and coagulation device and accessories is a device intended to remove tissue and control bleeding by use of high-frequency electrical current.

Predicate Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Endosconv contin & Nephew, in 150 Minuteman Road Andover MA 01810 ww smith-nephew com KO33981 ## SECTION V 510(k) SUMMARY OF SAFETY AND EFFECTIVENESS INFORMATION as required by the Safe Medical Devices Act of 1990 and codified in 21 CFR 807.92 upon which the substantial equivalence is based. # Smith & Nephew ElectroThermal® 20S Spine Generator ### Date Prepared: December 22, 2003 #### A. Submitter's Name: Smith & Nephew, Inc., Endoscopy Division 150 Minuteman Road Andover, MA 01810 #### B. Company Contact Janice Haselton Regulatory Affairs Specialist II Phone: (978) 749-1494 (978) 749-1443 Fax: #### C. Device Name | Trade Name: | Smith & Nephew ElectroThermal® 20S Spine Generator | |----------------------|-------------------------------------------------------| | Common Name: | Electrosurgical Spine Generator | | Classification Name: | Electrosurgical Cutting and Coagulation & Accessories | #### D. Predicate Devices The Smith & Nephew Smith & Nephew ElectroThermal® 20S Spine Generator is substantially equivalent in Intended Use and Fundamental Scientific Technology to the following legally marketed device in commercial distribution: Smith & Nephew ORA-50 S, cleared in K993854 and Radionic's RFG-3C Plus Lesion Generator, cleared in K982489. {1}------------------------------------------------ K033981 (pg 2 of 3) #### Description of Device E. The proposed Smith & Nephew ElectroThermal® 20S Spine generator is a 20-watt electrothermal generator. It is intended to be used to create lesions in nervous tissue and to coagulate and decompress material when used in combination with Smith & Nephew thermal/coagulating probes. The generator provides temperature and impedance monitoring of energy to maintain effective tissue heating during temperature controlled applications. Smith & Nephew ElectroThermal® 20S Spine generator is designed to be used in conjunction with Smith and Nephew Spine products. These products include the Smith & Nephew SpineCATH® Intradiscal Catheter, the Smith & Nephew Decompression Catheter and the Smith & Nephew RF Denervation Probe and Smith & Nephew RF Probe. #### Intended Use F. The Smith & Nephew ElectroThermal® 20S Spine System is intended to create lesions in nervous tissue, and to coagulate and decompress disc material when used in combination with Smith & Nephew thermal/coagulating probes. The generator and accessories are intended for use by qualified medical personnel trained in the use of electrosurgical equipment. Contraindications for Use: There are no known absolute contraindications to the use of electrosurgery. The Smith & Nephew ElectroThermal® 20S Spine System is contraindicated, when in the judgement of the physician, an electrosurgical procedure would be contrary to the best interest of the patient. #### G. Comparison of Technological Characteristics The Smith & Nephew ElectroThermal® 20S Spine generator is substantially equivalent to the Smith & Nephew ORA-50 S, cleared in K993854, based on the following: - · Both control and monitor temperature - · Both monitor impedance - · Both use RF energy to thermally heat the SpineCATH® and Decompression catheters - · Both provide preset settings for time and temperature to deliver RF. The Smith & Nephew ElectroThermal® 20S Spine generator is substantially equivalent to the Radionic's RFG-3C Plus Lesion Generator cleared in K982489 based on the following. - · Both control and monitor temperature - · Both monitor impedance - · Both use a biphasic square wave to deliver stimulation voltage {2}------------------------------------------------ $$ \begin{pmatrix} \xi \ \beta \ \xi \end{pmatrix} \quad \begin{pmatrix} \xi \ \xi \ \xi \end{pmatrix} \quad \begin{pmatrix} \xi \ \beta \ \xi \ \xi \end{pmatrix} $$ - · Both use temperature controlled RF energy to create lesions in nervous tissue ### H. Summary Performance Data The performance testing conducted on the Smith & Nephew ElectroThermal® 20S Spine generator demonstrates substantial equivalents to the Radionic's RFG-3C Plus Lesion Generator cleared in K982489 and the Smith & Nephew ORA-50 S cleared in K993854 based on deliverance of temperature and time of RF controlled energy. {3}------------------------------------------------ DEPARTMENT OF HEALTH & HUMAN SERVICES Image /page/3/Picture/1 description: The image shows the logo for the Department of Health & Human Services (HHS). The logo features a stylized image of an eagle or bird with three wing-like shapes, suggesting movement or flight. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" is arranged in a circular pattern around the bird, indicating the department's name and the country it represents. Public Health Service FEB 2 5 2004 Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850 Ms. Janice Haselton Regulatory Affairs Specialist II Smith & Nephew, Inc. Endoscopy Division 150 Minuteman Road Andover, Massachusctts 01810 Re: K033981 Trade/Device Name: The Smith & Nephew ElectroThermal® 20S Spine Generator Regulation Number: 21 CFR 878.4400 Regulation Name: Electrosurgical cutting and coagulation device and accessories Regulatory Class: II Product Code: GEI Dated: Deccmber 22, 2003 Received: December 23, 2003 Dear Ms. Haselton: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complics with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencics. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. {4}------------------------------------------------ Page 2 - Ms. Janice Haselton This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (301) 594-4659. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html Sincerely yours, Mark A. Milliken Cclia M. Witten, Ph.D., M.D. Director Division of General, Restorative and Neurological Devices ()ffice of Device Evaluation Center for Devices and Radiological Health Enclosure {5}------------------------------------------------ ## Indications for Use 510(k) Number (if known):__K033981 Device Name: The Smith & Nephew ElectroThermal® 20S Spine System Indications For Use: The Smith & Nephew ElectroThermal® 20S Spine System is intended to create lesions in nervous tissue, and to coagulate and decompress disc material when used in in nervous tious, and to our ew thermal/coagulating probes. The generator and oombination will be mare by qualified medical personnel trained in the use of electrosurgical equipment. Prescription Use × (Part 21 CFR 801 Subpart D) AND/OR Over-The-Counter Use (21 CFR 807 Subpart C) (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED) for Concurrence of CDRH, Office of Device Evaluation (ODE) (Division Sign-Off) Division of General, Restorative, and Neurological Devices 510 Number K033981 Page 1 of
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...