## Indications for Use

The intended use of the Mobile Single Circular Camera (MSC) and the Mobile Digital Cardiac Camera (MDCC) is to detect the location and distribution of gamma ray emitting radionuclides in the body and store data for analysis. This device includes accessories such as signal analysis and display equipment, patient and equipment supports, radionuclide anatomical markers, component parts and accessories.

## Device Story

Mobile gamma camera system (MSC/MDCC) designed for mobile environment deployment; utilizes detector head to capture gamma ray emissions from radionuclides within patient body; transforms inputs into digital data for analysis and display; operated by clinical staff in mobile truck settings; incorporates 'Dome Mounting' pads on gantry for high load isolation during transport; provides diagnostic imaging data to assist healthcare providers in clinical decision-making; benefits patients by enabling nuclear medicine imaging services at remote or varied locations.

## Clinical Evidence

No clinical data provided. Substantial equivalence is based on technological comparison and third-party safety certification.

## Technological Characteristics

Emission computed tomography system; detector head hardware identical to predicate; gantry optimized for mobile platform mounting; includes 'Dome Mounting' pads for load isolation; software identical to predicate; regulatory class II; product code 90-KPS.

## Regulatory Identification

An emission computed tomography system is a device intended to detect the location and distribution of gamma ray- and positron-emitting radionuclides in the body and produce cross-sectional images through computer reconstruction of the data. This generic type of device may include signal analysis and display equipment, patient and equipment supports, radionuclide anatomical markers, component parts, and accessories.

## Predicate Devices

- Mobile Single Rectangular Camera ([K032779](/device/K032779.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

K033956

# 501(k) SUMMARY

'JAN 2 0 2004

Labelling:

### SUBMITTERS IDENTIFICATION

| Applicant's Name and Street Address:        | IS2 Medical Systems Inc.<br>20 Gurdwara Rd., Units 3-10<br>Ottawa, Ontario, Canada<br>K2E 8B3 |
|---------------------------------------------|-----------------------------------------------------------------------------------------------|
| Manufacturing Site:                         | IS2 Medical Systems Inc.<br>20 Gurdwara Rd., Units 3-10<br>Ottawa, Ontario, Canada<br>K2E 8B3 |
| FDA Registration #                          | 9615403                                                                                       |
| Contact Person:                             | Victor Woodburn, Manager Quality and Regulatory                                               |
| Contact Person E-mail address:              | vwoodburn@is2medical.com                                                                      |
| Telephone and Fax Number of Contact Person: | T-(613) 228-8755, F-(613) 228-8228                                                            |
| Date of Submission:                         | December 3, 2003                                                                              |
|                                             | DEVICE NAME                                                                                   |
| Device Name (Common):                       | Gamma Camera                                                                                  |
| Proprietary Name:                           | Mobile Single Circular Camera (MSC)<br>& Mobile Digital Cardiac Camera (MDCC)                 |
| Classification Name:                        | Emission Computed Tomography System                                                           |
| Product Code:                               | 90-KPS                                                                                        |
| CFR:                                        | 21CFR 892.1200                                                                                |
| Device Class:                               | II                                                                                            |
| Predicate Device:                           | Mobile Single Rectangular Camera                                                              |
| (Predicate) 510(k) No.:                     | K032779                                                                                       |

Labels and Instructions for Use can be found in Attachment # 1 no changes to the labels or Instructions for Use have occurred except for adding information labels and instructions specific to the application of mounting and using the Camera systems in a Mobile Environment.

{1}------------------------------------------------

#### INTRODUCTION

This 510(k) Premarket Notification has been prepared to demonstrate that the, Mobile Single Circular Camera (MSC) and the Mobile Digital Cardiac Camera (MDCC) manufactured by IS2 Medical Systems Inc., is substantially equivalent to the Mobile Single Rectangular Camera (MSR) which has previously been through the 510(k) equivalent in the Movie Barge nessant of the systems include the mounting of the gantry to the mobile platform incorporating the use of "Dome Mounting" pads to facilitate high load isolation.

#### INTENDED USE

The intended use of the Mobile Single Circular Camera (MSC) and the Mobile Digital Cardiac Camera (MDCC) is to detect the location and distribution of gamma ray emitting radionuclides in the body and store data for andysis. This device includes accessories such as signal analysis and display equipment, patient and equipment supports, radionuclide anatomical markers, component parts and accessories. The Indications for Use found in Attachment 2.

### DETERMINATION OF SUBSTANTIAL EQUIVALENCE

The intended use of the Mobile Single Circular Camera (MSC) and the Mobile Digital Cardiac Camera (MDCC) is the same range of studies to that of the Mobile Single Rectangular Camera (MSR). The detector head is identical in hardware and software. The gantry of the Mobile Single Circular Camera (MSC) and the Mobile Digital Cardiac Camera (MDCC) are optimised for being attached to a mobile platform which is transported from location to location in a truck and have the same range of automatic clinical motions of Mobile Single Rectangular Cumera (MSR).

The Mobile Single Circular Camera (MSC) and the Mobile Digital Camera (MDCC)) has been deemed safe and effective and is certified to the same safety standards as the predicate device by a third party organization prior to use on patients. A matrix was constructed comparing the features and intended use of the Mobile Single Circular Camera (MSC) and the Mobile Digital Cardiac Camera (MDCC) with the predicate device. We conclude that the Mobile Single Circular Camera (MSC) and the Mobile Digital Cardiac Camera (MDCC) is substantially equivalent to the predicate device and that no new safety or effectiveness concerns are raised.

{2}------------------------------------------------

Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" around the perimeter. Inside the circle is an abstract symbol that resembles three stylized human profiles facing to the right.

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

JAN 2 0 2004

Mr. Victor Woodburn Manager Quality and Regulatory IS2 Medical Systems, Inc. 20 Gurdwara Rd., Units 3-10 Ottawa, Ontario, K2E 8B3 CANADA

Re: K033956

Trade/Device Name: Mobile Single Circular Camera (MSC & Mobile Digital Cardiac Camera (MDCC) Regulation Number: 21 CFR 892.1200 Regulation Name: Emission computed tomography system Regulatory Class: II Product Code: 90 KPS Dated: December 18, 2003

Received: December 22, 2003

Dear Mr. Woodburn:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820), and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

{3}------------------------------------------------

Page 2

This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at one of the following numbers, based on the regulation number at the top of the letter:

| 8xx.1xxx                         | (301) 594-4591 |
|----------------------------------|----------------|
| 876.2xxx, 3xxx, 4xxx, 5xxx       | (301) 594-4616 |
| 884.2xxx, 3xxx, 4xxx, 5xxx, 6xxx | (301) 594-4616 |
| 892.2xxx, 3xxx, 4xxx, 5xxx       | (301) 594-4654 |
| Other                            | (301) 594-4692 |

Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97) you may obtain. Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html.

Sincerely yours,

Nancy C. Brogdon

Nancy C. Brogdon Director, Division of Reproductive, Abdominal and Radiological Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{4}------------------------------------------------

# 510(k) Number (K033956)

Device Name: Mobile Single Circular (MSC) and the Mobile Digital Cardiac Camera (MDCC)

Indications for Use: The intended use of the Mobile Single Circular (MSC) and the Mobile Digital Cardiac Camera (MDCC) is to detect the location and distribution of gamma ray emitting radionuclides in the body and store data for analysis. This device gamma ray entiting factoriable in ades accessories such as signal analysis and display restics on a moone pratent and equipment supports, radionuclide anatomical markers, component parts and accessories.

# (PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)

|  | Concurrence of CDRH, Office of Device Evaluation (ODE) |  |
|--|--------------------------------------------------------|--|
|  | (Division Sign-Off)                                    |  |
|  | Division of Reproduction, Abdomin                      |  |
|  | a ogical Devices K03386                                |  |
|  | 510(k) Number                                          |  |

| Prescription Use (Per 21 CFR 801.109) | YES |
|---------------------------------------|-----|
|---------------------------------------|-----|

OR

| Over-The-Counter Use | NO                       |
|----------------------|--------------------------|
|                      | (Optional Format 1-2-96) |

RECEIVED DATE : 01/07 14:40'04 FROM : 3014804224

---

**Source:** [https://fda-staging.innolitics.com/device/K033956](https://fda-staging.innolitics.com/device/K033956)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).