ALPHA METASUL 28MM AND 32MM ACETABULAR INSERTS, STANDARD AND HOODED
K033943 · Centerpulse Orthopedics, Inc. · KWA · Oct 18, 2004 · Orthopedic
Device Facts
Record ID
K033943
Device Name
ALPHA METASUL 28MM AND 32MM ACETABULAR INSERTS, STANDARD AND HOODED
Applicant
Centerpulse Orthopedics, Inc.
Product Code
KWA · Orthopedic
Decision Date
Oct 18, 2004
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 888.3330
Device Class
Class 3
Attributes
Therapeutic
Indications for Use
The Alpha Metasul 28mm and 32mm Acetabular Inserts, standard and hooded, are intended for use in total hip arthroplasty for treatment of the following: - Patient conditions of noninflammatory degenerative joint disease (NID), e.g., avascular necrosis, osteoarthritis; - Patient conditions of inflammatory joint disease (IJD), e.g., rheumatoid arthritis; - Those patients with failed previous surgery where pain, deformity, or dysfunction persists; - Revision of previously failed hip arthroplasty.
Device Story
Alpha Metasul Acetabular Insert (28mm/32mm) is a hemispherically shaped orthopedic implant for total hip arthroplasty. Device features a 'poly-sandwich' design: UHMWPE liner thermo-mechanically bonded to a wrought hot-forged CoCr alloy metallic inlay. Metallic inlay is polished to mirror-finish; intended for use exclusively with Metasul femoral heads. Hooded versions include polyethylene overhang to reduce subluxation/instability. Device is implanted by orthopedic surgeons in clinical/hospital settings. It functions as a bearing surface within an acetabular shell to restore joint function, reduce pain, and address deformity in patients with degenerative or inflammatory joint disease or failed prior surgeries.
Clinical Evidence
No clinical data provided; substantial equivalence is supported by design comparisons and performance testing.
Technological Characteristics
Materials: UHMWPE liner and wrought hot-forged CoCr alloy inlay (ISO 5832-12, ASTM F1537). Design: Hemispherical, 'poly-sandwich' construction, standard or hooded configurations. Sizes: 32mm inserts for 52-68mm shells; 28mm inserts for 48-68mm shells. Connectivity: None. Sterilization: Not specified.
Indications for Use
Indicated for patients undergoing total hip arthroplasty suffering from noninflammatory degenerative joint disease (e.g., avascular necrosis, osteoarthritis), inflammatory joint disease (e.g., rheumatoid arthritis), failed previous surgery with persistent pain/deformity/dysfunction, or requiring revision of failed hip arthroplasty.
Regulatory Classification
Identification
A hip joint metal/metal semi-constrained, with an uncemented acetabular component, prosthesis is a two-part device intended to be implanted to replace a hip joint. The device limits translation and rotation in one or more planes via the geometry of its articulating surfaces. It has no linkage across-the-joint. This generic type of device includes prostheses that consist of a femoral and an acetabular component, both made of alloys, such as cobalt-chromium-molybdenum. The femoral component is intended to be fixed with bone cement. The acetabular component is intended for use without bone cement (§ 888.3027).
Predicate Devices
Allofit™ Acetabular System, Alpha Metasul® Acetabular Insert, 28mm Standard and Hooded (K003758)
{0}------------------------------------------------
K03 3943
## OCT 1 8 2004
510(k) Premarket Notification 510(k) Prematiet Notification
Alpha Metasul 28mm and 32mm Acetabular Inserts, Standard and Hooded
#### 510(k) SUMMARY
| MANUFACTURER: | Centerpulse Orthopedics, Ltd. (a division of Zimmer)<br>Altgasse 44<br>CH-6340<br>Baar, Switzerland |
|-----------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| SPONSOR: | Centerpulse Orthopedics, Inc. (a division of Zimmer)<br>9900 Spectrum Drive<br>Austin, TX 78717 USA |
| CONTACT: | Audrey Swearingen<br>Phone: (512) 432-9255<br>E-Mail: Audrey.Swearingen@zimmer.com |
| TRADE NAME: | Alpha Metasul® 28mm and 32mm Acetabular Inserts,<br>Standard and Hooded |
| COMMON NAME: | Total hip replacement system acetabular insert |
| CLASSIFICATION: | CFR §888.3330 (KWA) - Hip joint metal/metal<br>semiconstrained, with an uncemented acetabular<br>component, prosthesis. Metal-on-metal hip prostheses are<br>Preamendment Class III devices. |
#### PREDICATE DEVICES:
- EDICATE DEVIOEG.
Centerpulse Orthopedics Inc. Allofit™ Acetabular System, Alpha Metasul® Acetabular . Insert, 28mm Standard and Hooded (K003758)
- Insert, 202a™ Ringloc® Acetabular Liner (K002379) .
- Biomet M2a -Taper™ Acetabular System (K003363) .
#### DEVICE DESCRIPTION:
DEVICE DESCRIP TION.
The Alpha Metasul Acetabular Insert (28mm and 32mm) is a hemispherically Shaped The Alpha Metasur Acelabular Insert (Echnif and Game) polyethylene (UHMWFE)
design, composed of an outer component manufactured from polyethylene a wrought design, composed of an outer component intermo-mechanically bonded to a wrought (in compliance with TOO 503+ II2) which with ISO 5832-12 and ASTM F1537).
hot-forged CoCr alloy metallic inlay (in compliance with ISO 5832-12 and Associations not-forged CoCr alloy metalle intay (in complianter rise only with a Metasul femoral head.
The Alpha Metasul Acetabular Insert is designed for use only with a Metasul the The Alpha Metasur Acelabular Insolt is a signed for ass one on the many of the component, metal surfaces.
The Alpha Metasul 32mm Acetabular Insert, both standard and hooded, is available in The Alpha Metasur Szhm? Acetabular Shells, sizes 52mm to 68mm to 68mm (in 2mm sizes designed to mate with Allolle - Acctubular Choile, Case enning available in sizes Increments). In turn, the proposed Alpha Motabular Shells, sizes 48mm to 68mm (in 2mm designed to 'mate 'with 'Allen'' 'Allell' - 'Ale same basic design as the previously cleared Alpha increments). Dour insert sizes nate the sand hooded. They also share many Metasur 2011n1 710ctubular Though Scetabular Liner and M2a -Taper Acetabular System.
The Alpha Metasul Insert has what is commonly referred to as a "poly-sandwich" design. The Alpha Mccasur noon has mathe bearing surface of the insert, features a metallic
{1}------------------------------------------------
510(k) Premarket Notification 510(k) Premarket Notification
Alpha Metasul 28mm and 32mm Acetabular Inserts, Standard and Hooded
Metasul inlay that is polished to a mirror-finish and themo-mechanically bonded into the proven Metasul inlay that is polished to a mirror-ining the Allofit acetabular shell was aroven
polyethylene liner, which is then locked into the Alloft access of the proven polyethylene liner, which is then tocked into his blocked in snap mechanism. On the nooded inserts, the face of this por the midpoint of the incorporates an overnang of polyethylene extending ouponally would be and the subluxation and instability.
#### INTENDED USE:
INTENDED USE:
The Alpha Metasul 28mm and 32mm Acetabular Inserts, standard and hooded, are The Alpha Tivetasur Zomin' arthroplasty for treatment of the following:
- Patient conditions of noninflammatory degenerative joint disease (NID), e.g., . Patient conditions of nominiatimatory dogenorative joint disease (IJD), e.g., rheumatoid arthritis;
- Those patients with failed previous surgery where pain, deformity, or dysfunction . persists;
- Revision of previously failed hip arthroplasty. .
### BASIS OF SUBSTANTIAL EQUIVALENCE:
BASIS OF SUBSTANTIAL EQUIVALL, 102.
Performance tests, design comparisons, indications for use and labeling of the Alpha Performance lests, design compansens, indications and hooded) demonstrations Metasul 2011n and SZHIN Acelabular insorts (StanLard unes, materials, and indications for use to the predicate devices.
{2}------------------------------------------------
Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is an abstract symbol resembling an eagle or bird in flight, composed of three curved lines above three wavy lines.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
OCT 1 8 2004
Ms. Audrey Swearingen Manager, Regulatory Affairs Zimmer, Inc. 9900 Spectrum Drive Austin, Texas 78717
Re: K033943
K035743
Trade/Device Name: Alpha™ Metasul® 28mm and 32mm Acetabular Insert Regulation Number: 21 CFR 888.3330 Regulation Name: Hip joint metal/metal semi-constrained, with an uncemented acetabular component, prosthesis Regulatory Class: III Product Code: KWA Dated: September 24, 2004 Received: September 27, 2004
Dear Ms. Swearingen:
We have reviewed your Section 510(k) premarket notification of intent to market the device we nave reviewed your becaon 310(t) perceise is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate for use stated in the encrosule) to regars annount date of the Medical Device Amendments, or to commence provision to May 20, 1976, the excordance with the provisions of the Federal Food, Drug, devices that have been recure approval of a premarket approval application (PMA). and Cosmetic Acr (Free) market the device, subject to the general controls provisions of the Act. The 1 ou may, therefore, market the act include requirements for annual registration, listing of gencral controls provisions practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it If your device is classified (600 a00 royals. Existing major regulations affecting your device can may be subject to saterial regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean Frease oc advised that 1271 3 ibsuance or our device complies with other requirements of the Act that i DA has made a determinations administered by other Federal agencies. You must or any I cacial statutes and regulations, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set CFR Part 607), adomig (21 OFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{3}------------------------------------------------
Page 2 - Ms. Audrey Swearingen
This letter will allow you to begin marketing your device as described in your Section 510(k) I his letter will anow you to begin manceling your antial equivalence of your device to a legally premarket nothication. THC PDA midning of cacomman s p.
marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Patt 801), please If you desire specific advice for your act 100 case. Also, please note the regulation entitled, Conliact the Office of Complance at (3 of Critication" (21CFR Part 807.97). You may obtain "Misbranding by reference to premarket icolifical vir the Act from the Division of Small other general information on your responsible in the toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html
Sincerely yours,
Mark A. Melberer
Celia Witten, Ph.D., M.D. Director Division of General, Restorative and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
# Indications for Use
K033943 510(k) Number (if known):
Alpha™ Metasul® 28mm and 32mm Acetabular Insert Device Name:
Indications for Use:
The Alpha™ Metasul® Acetabular Insert is intended for use in total hip arthroplasty for treatment of the following:
- patient conditions of noninflammatory degenerative joint disease (NIDD), e.g., , . patient conditions of nominationatory cogstern of the matory joint disease (IJD), e.g., rheumatoid arthritis;
- rheumatolu arthritis, . persists;
- revision of previously failed hip arthroplasty. .
Prescription Use (Part 21 CFR 801 Subpart D)
AND/OR
Over-The-Counter Use (21 CFR 807 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
K033943
Mark N. Mallem
Restorative, Jeurological Devices
Page __ of____________________________________________________________________________________________________________________________________________________________________
*(Posted November 13, 2003)*
**510(k) Number**
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.