K033897 · Aga-Linde Health Care · ECX · Jul 19, 2004 · Anesthesiology
Device Facts
Record ID
K033897
Device Name
MEDICYL-E-LITE PORTABLE OXYGEN DELIVERY SYSTEM
Applicant
Aga-Linde Health Care
Product Code
ECX · Anesthesiology
Decision Date
Jul 19, 2004
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 868.2700
Device Class
Class 1
Attributes
Therapeutic
Indications for Use
The MEDICYL-E-Lite is an integrated portable oxygen delivery system intended to provide supplemental oxygen to adults. The device is MRI safe, MRI-compatible, and intended for use during MR imaging for MRI systems up to 3.0T. For emergency use only when administered by properly trained personnel for oxygen deficiency and resuscitation. For all other medical applications, Rx only. Compressed gas cylinders in service or in storage shall be stabilized or otherwise secured to prevent falling and rolling.
Device Story
Integrated portable oxygen delivery system; includes aluminum cylinder, valve, regulator, and flow meter. Provides supplemental oxygen to adults; MRI safe/compatible up to 3.0T. Used in clinical settings or during MR imaging; operated by trained personnel or clinicians. Device delivers oxygen via low flow settings; includes excess flow device for safety. Healthcare providers use the system to manage oxygen deficiency; benefits include safe oxygen administration in MRI environments and portability. Mechanical operation; no electronic or software components.
Clinical Evidence
Bench testing only. Evaluated for MRI safety and compatibility up to 3.0T in accordance with the CDRH Magnetic Resonance Working Group document, 'A Primer on Medical Device Interactions with Magnetic Resonance Imaging Systems' (Feb 7, 1997). Aluminum cylinders conform to 21 CFR 178.46 (Specification SAL).
Technological Characteristics
Integrated aluminum cylinder, valve, regulator, and flow meter. Pressure design 4350 psi. Features 3 filters and excess flow device. Single-stage piston-style regulator. MRI safe/compatible up to 3.0T. Dimensions: 6.75" height. Weight: 900g (E-size). No software or electronic components.
Indications for Use
Indicated for adult patients requiring supplemental oxygen. Used for emergency oxygen deficiency and resuscitation by trained personnel, or other medical applications under prescription. MRI safe/compatible up to 3.0T.
Regulatory Classification
Identification
A pressure regulator is a device, often called a pressure-reducing valve, that is intended for medical purposes and that is used to convert a medical gas pressure from a high variable pressure to a lower, more constant working pressure. This device includes mechanical oxygen regulators.
Predicate Devices
MEDICYL-E-Lite Portable Oxygen System
Praxair Grab n' Go Portable Medical Oxygen System
Submission Summary (Full Text)
{0}------------------------------------------------
# AGA Linde Healthcare
Image /page/0/Picture/2 description: The image shows the word "Linde" in a stylized, cursive font. The word is written in black ink on a white background. The "L" has a large, decorative swirl at the top, and the other letters are connected in a flowing script.
## 510(k) Summary MEDICYL-E-Lite Portable Oxygen System 510(k) Number: K033897
Submitted in accordance with the requirements of SMDA 1990 and 21CFR807.92.
| 1. APPLICANT'S INFORMATION: | |
|--------------------------------------------------|-----------------------------------------------------------------------------------|
| Edward W. Siurek III | |
| Dir., Operational Quality, IBD Global Operations | |
| PH: | (216) 328-9801 |
| FX: | (216) 328-9818 |
| Internet: | <a href="http://www.agalindehealthcare.com">http://www.agalindehealthcare.com</a> |
| Medical Establishment Registration No.: | pending |
| 2. SUBMITTER'S INFORMATION | |
| James Jochen Rogers | |
| General Manager | |
| Coastal Consulting Group, Ltd. | |
| P.O. Box 391117 | |
| Solon, OH 44139 | |
| PH/FX: | 440.546.4936 |
| Mobile: | 724.713.2298 |
| E-mail: | <a href="mailto:coastalcg@earthlink.net">coastalcg@earthlink.net</a> |
| Internet: | <a href="http://www.immersive-reality.biz">http://www.immersive-reality.biz</a> |
| 3. Date: | December 15, 2003 |
| 4. DEVICE INFORMATION | |
| Trade/Proprietary Name: | MEDICYL-E-Lite Portable Oxygen delivery system |
| Common Name: | MEDICYL-E-Lite Portable Oxygen delivery system |
| DEVICE NAME: | Cylinder, Compressed Gas, and Valve |
| Classification Panel: | Cardiovascular and Respiratory Devices |
| Classification Number: | 868.2700 |
| Product Nomenclature: | Regulator, Pressure, Gas Cylinder |
| Product Code(s): | CAN |
| Classification Number: | 868.2610 |
| Product Nomenclature: | Gauge, Gas Pressure, Cylinder/Pipeline |
| Product Code(s): | BXH |
| Classification Number: | unclassified |
| Product Nomenclature: | Cylinder, Compressed Gas, and Valve |
| Product Code(s): | ECX |
| Classification Number: | unclassified |
| Product Nomenclature: | Cylinder, Gas (Empty) |
| Product Code(s): | KGA |
Helping You Care More Effectively
6055 Rockside Woods Blvd.
Corporate Office (216) 642-6600
{1}------------------------------------------------
# AGA Linde Healthcare
Image /page/1/Picture/1 description: The image shows the word "Linde" in a stylized, cursive font. The letters are connected, and the "L" has a large, decorative swirl at the top. The word appears to be a logo or brand name.
#### 5. DEVICE CLASSIFICATION:
Empty compressed gas cylinders and compressed gas cylinder with valve assemblies are unclassified devices, and reviewed by the Anesthesiology and Respiratory Devices Branch, Division of Anesthesiology, General Hospital, Infection Control, and Dental Devices.
Gas cylinder pressure regulators and gas pressure gauges are Class I devices and exempted from pre-market notification.
## 6. PREDICATE DEVICE(s):
- MEDICYL-E-Lite Portable Oxygen System .
- . Praxair Grab n' Go Portable Medical Oxygen System
## 7. DEVICE DESCRIPTION:
The MEDICYL-E-Lite is an integrated portable oxygen delivery system intended to provide supplemental oxygen to adults. The device is MRI safe, MRI-compatible, and intended for use during MR imaging for MRI systems up to 3.0T. For emergency use only when administered by properly trained personnel for oxygen deficiency and resuscitation. For all other medical applications, Rx only. Compressed gas cylinders in service or in storage shall be stabilized or otherwise secured to prevent falling and rolling.
The intearated system includes an aluminum cylinder, valve, requlator and flow meter. The device offers low flow settings that may be clinically appropriate for certain classes of patients.
#### 8. INDICATIONS FOR USE:
The MEDICYL-E-Lite is an integrated portable oxygen delivery system intended to provide supplemental oxygen to adults. The device is MRI safe, MRI-compatible, and intended for use during MR imaging for MRI systems up to 3.0T. For emergency use only when administered by properly trained personnel for oxygen deficiency and resuscitation. For all other medical applications, Rx only. Compressed gas cylinders in service or in storage shall be stabilized or otherwise secured to prevent falling and rolling.
# 9. TECHNOLOGICAL CHARACTERISTICS:
A summary comparison of technological characteristics, including design and materials is provided in the table below:
| Parameter | MEDICYL-E-Lite<br>(MRI) | MEDICYL -E-Lite | Praxair Grab n'<br>Go |
|-----------------------|-------------------------|-----------------|-----------------------|
| Valve/Regulator | | | |
| Low Flow Settings | yes | yes | no |
| Flow Between Settings | no | no | no |
| Cylinder On/Off | yes | yes | no |
| Filling Port | active | active | active |
| Contents Gauge | non-active | non-active | active |
| Filters | 3 | 3 | 1 |
| Pressure Design | 4350psi | 4350psi | 3000psi |
| Excess Flow Device | yes | yes | no |
{2}------------------------------------------------
# AGA Linde Healthcare
Image /page/2/Picture/1 description: The image shows the word "Linde" in a stylized, cursive font. The word is written in black ink and appears to be a logo or brand name. The letters are connected, and the "L" has a decorative swirl at the top.
| Parameter | MEDICYL-E-Lite<br>(MRI) | MEDICYL -E-Lite | Praxair Grab n'<br>Go |
|------------------------------------------|---------------------------|-----------------|-----------------------|
| Single stage piston<br>style | yes | yes | yes |
| Guard | | | |
| Hand grip | 2 grip | 2 grip | 1 grip |
| Access Ports | yes | yes | no |
| Flow selector/hose<br>barb/gauge aligned | yes | yes | no |
| Color | green | green | green |
| Height | 6.75" | 6.75" | 8" |
| Cylinder | | | |
| Sizes | D, E | D, E | E |
| Weight (E) | 900gr | 900gr | 1060gr |
| Materials/construction | Aluminum | Aluminum | Steel |
| MRI Compatibility | | | |
| MRI Safe | yes; tested up to<br>3.0T | no | no |
| MRI Compatible | yes; tested up to<br>3.0T | no | no |
The manufacturer believes that the technological characteristics of the MEDICYL-E-Lite portable oxygen system is substantially similar to those of the predicate devices.
- 10. PERFORMANCE DATA:
The aluminum cylinders conform to the requirements of 21 CFR49 § 178.46, Specification SAL seamless aluminum cylinders.
The MEDICYL-E-Lite portable oxygen delivery system has been evaluated in accordance with the draft CDRH Magnetic Resonance Working Group document, A Primer on Medical Device Interactions with Magnetic Resonance Imaging Systems, dated February 7, 1997.
# 11. STATEMENT OF SUBSTANTIAL EQUIVALENCE:
Based upon the safety and performance testing and compliance with voluntary standards, the manufacturer believes that the MEDICYL-E-Lite portable oxygen delivery system is substantially equivalent to the predicate devices, and does not raise any new questions of safety or effectiveness.
{3}------------------------------------------------
DEPARTMENT OF HEALTH & HUMAN SERVICES
Image /page/3/Picture/1 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo is circular and contains the department's name around the perimeter. Inside the circle is an abstract symbol that resembles an eagle or bird in flight, composed of three curved lines.
Public Health Service
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
JUL 1 9 2004
AGA-Linde Healthcare, Incorporated C/O Mr. James Jochen Rogers General Manager Coastal Consulting Group, Limited P.O. Box 391117 Solon, Ohio 44139
Re: K033897
Trade/Device Name: Medicyl-E-Lite Portable Oxygen Delivery System Regulation Number: None Regulation Name: None Regulatory Class: Unclassified Product Code: ECX Dated: May 15, 2004 Received: May 19, 2004
Dear Mr. Rogers:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{4}------------------------------------------------
Page 2 -Mr. Rogers
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (301) 594-4646. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html
Sincerely yours,
Cluls
Chiu Lin, Ph.D. Director Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
# Indications for Use
510(k) Number (if known): K033897
Device Name: Medicyl-E-Lite Portable Oxygen Delivery System
Indications for Use:
The MEDICYL-E-Lite is an integrated portable oxygen delivery system intended to provide supplemental oxygen to adults. The device is MRI safe, MRIcompatible, and intended for use during MR imaging for MRI systems up to 3.0T.
For emergency use only when administered by properly trained personnel for oxygen deficiency and resuscitation. For all other medical applications, Rx only. Compressed gas cylinders in service or in storage shall be stabilized or otherwise secured to prevent falling and rolling.
Sta y/t
7/16/2001
Division of Anesthesiology. General Hospital. Infection Control. Dental Devices
510(k) Number:
X Prescription Use (Part 21 CFR 801 Subpart D)
AND
X Over-the-Counter Use _ (21 CFR 807 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Page 1 of ____________________________________________________________________________________________________________________________________________________________________
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.