BD PHOENIX AUTOMATED MICROBIOLOGY SYSTEM FOR USE WITH THE ANTIMICROBIAL AGENT CHLORAMPHENICOL 1-32 UG/ML GRAM POSITIVE
Device Facts
| Record ID | K033889 |
|---|---|
| Device Name | BD PHOENIX AUTOMATED MICROBIOLOGY SYSTEM FOR USE WITH THE ANTIMICROBIAL AGENT CHLORAMPHENICOL 1-32 UG/ML GRAM POSITIVE |
| Applicant | Becton, Dickinson & CO |
| Product Code | LON · Microbiology |
| Decision Date | Feb 3, 2004 |
| Decision | SESE |
| Submission Type | Traditional |
| Regulation | 21 CFR 866.1645 |
| Device Class | Class 2 |
Indications for Use
The BD Phoenix™ Automated Microbiology System is intended for in vitro quantitative determination of antimicrobial susceptibility by minimal inhibitory concentration (MIC) of most Gram-negative aerobic and facultative anaerobic bacteria isolates from pure culture for Enterobacteriaceae and Non-Enterobacteriaceae and most Gram-positive bacteria isolates from pure culture belonging to the genera Staphylococcus and Enterococcus. This premarket notification is for the addition of the antimicrobial agent chloramphenicol at concentrations of 1 - 32 µg/mL to Gram Positive ID/AST or AST only Phoenix panels. Chloramphenicol has been shown to be active in vitro against most strains of microorganisms listed below, as described in the FDA-approved package insert for this antimicrobial agent.
Device Story
BD Phoenix™ Automated Microbiology System performs automated antimicrobial susceptibility testing (AST) and identification (ID). System inputs: pure culture bacterial isolates suspended in broth (0.5 McFarland standard) and inoculated into sealed, multi-well polystyrene trays containing dried reagents and redox indicators. Instrument continuously incubates trays at 35°C; optical sensors monitor colorimetric oxidation-reduction changes every 20 minutes to detect bacterial growth. Growth in antibiotic-containing wells indicates resistance; lack of growth indicates susceptibility. Software-driven 'EXPERT' system interprets biochemical and AST reactions to provide MIC values and categorical interpretations (SIR) based on NCCLS standards. Used in clinical microbiology laboratories by trained technicians. Benefits include rapid, automated determination of antibiotic susceptibility, facilitating timely clinical decision-making for antimicrobial therapy.
Clinical Evidence
Performance evaluated using clinical, stock, and challenge isolates across multiple US sites. Compared against NCCLS reference broth microdilution method. Metrics included Essential Agreement (EA) and Category Agreement (CA). Reproducibility study (intra- and inter-site) demonstrated >90% intra-site and >95% inter-site reproducibility for Gram-positive isolates.
Technological Characteristics
Automated microdilution system; molded polystyrene panels with 136 micro-wells; redox indicator-based growth detection; continuous incubation at 35°C; optical turbidity and redox measurement; software-based interpretation of MIC and categorical results.
Indications for Use
Indicated for in vitro quantitative determination of antimicrobial susceptibility (MIC) of gram-positive bacteria (Staphylococcus and Enterococcus genera) to chloramphenicol (1–32 µg/mL).
Regulatory Classification
Identification
A fully automated short-term incubation cycle antimicrobial susceptibility system is a device that incorporates concentrations of antimicrobial agents into a system for the purpose of determining in vitro susceptibility of bacterial pathogens isolated from clinical specimens. Test results obtained from short-term (less than 16 hours) incubation are used to determine the antimicrobial agent of choice to treat bacterial diseases.
Special Controls
*Classification.* Class II (special controls). The special control for this device is FDA's guidance document entitled “Class II Special Controls Guidance Document: Antimicrobial Susceptibility Test (AST) Systems; Guidance for Industry and FDA.”
Predicate Devices
- VITEK® System (PMA No. N50510)
- BD Phoenix™ Automated Microbiology System with Gatifloxacin (K020321)
- BD Phoenix™ Automated Microbiology System with Ofloxacin (K020323)
- BD Phoenix™ Automated Microbiology System with Levofloxacin (K020322)