K033831 · Siemens AG Medical Solutions · LLZ · Dec 19, 2003 · Radiology
Device Facts
Record ID
K033831
Device Name
SIENET COSMOS
Applicant
Siemens AG Medical Solutions
Product Code
LLZ · Radiology
Decision Date
Dec 19, 2003
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 892.2050
Device Class
Class 2
Attributes
Software as a Medical Device, 3rd-Party Reviewed
Indications for Use
The SIENET Cosmos is a Picture Archiving and Communication System (PACS) intended to enhance the complete imaging workflow, i.e. Manipulating . Reading . Reporting . Viewing . Communicating / distributing . Storing / archiving of radiological softcopy images and Storing / archiving of structured (DICOM) reports. The system is a "software only" solution and is intended to assist the physician in diagnosis or treatment planning. Therefore SIENET Cosmos supports the following generic imaging workflow: - Receive scheduled exams from IS at the SIENET Cosmos archiving component SDM . - Provide relevant prior exams and reports (Structured Reports only) to the Modalities and Workplaces . - Receive and store new exams from the Modalities at the SDM . - Prepare images for reading . - Report new images, if required by comparing them with prior exams and reports . - Demonstrate exams at Radiological Demos . - Domonstante onations at Radiology (e.g. Surgery, Intensive Case Unit, wards, external referring physicians). Note: The workstation deployment syngo Viewing Studio is not intended for primary diagnosis.
Device Story
SIENET Cosmos is a software-only PACS for radiology departments and hospitals; integrates imaging workflow from scheduling to reporting and archiving. Inputs include DICOM-formatted radiological images and structured reports from imaging modalities and RIS. System comprises a backend communication/storage component (SDM) and three workplace deployments: syngo Viewing Studio (web-based, non-diagnostic), syngo Reporting Studio (basic reporting), and syngo Reporting Studio Advanced Application Bundle (advanced reporting). Operates via centralized server structure; enables data distribution to clinicians in surgery, ICU, wards, and external referring physicians. Physicians use workstations to view, manipulate, and report on images; system facilitates comparison with prior exams. Benefits include fast, continuous access to radiological data, improved workflow efficiency, and centralized storage/archiving. System installed by service engineers on customer-provided hardware.
Clinical Evidence
Bench testing only. No clinical data provided. Safety and effectiveness are supported by risk management, software development lifecycle processes, and verification/validation testing to ensure compliance with industry standards for PACS functionality.
Technological Characteristics
Software-only PACS; backend storage (SDM) runs on Solaris 8; workstations run on Windows 2000/XP. Supports DICOM-formatted images and objects. Connectivity via HL7/DICOM-compatible RIS integration. System architecture includes centralized server and multiple workplace deployments. Installed by service engineers on specified hardware.
Indications for Use
Indicated for use by radiologists and radiological technicians to manipulate, read, report, view, and distribute radiological softcopy images and structured DICOM reports to assist in diagnosis or treatment planning. Not intended for primary diagnosis when using the syngo Viewing Studio deployment.
Regulatory Classification
Identification
A medical image management and processing system is a device that provides one or more capabilities relating to the review and digital processing of medical images for the purposes of interpretation by a trained practitioner of disease detection, diagnosis, or patient management. The software components may provide advanced or complex image processing functions for image manipulation, enhancement, or quantification that are intended for use in the interpretation and analysis of medical images. Advanced image manipulation functions may include image segmentation, multimodality image registration, or 3D visualization. Complex quantitative functions may include semi-automated measurements or time-series measurements.
Special Controls
*Classification.* Class II (special controls; voluntary standards—Digital Imaging and Communications in Medicine (DICOM) Std., Joint Photographic Experts Group (JPEG) Std., Society of Motion Picture and Television Engineers (SMPTE) Test Pattern).
{0}------------------------------------------------
K033831
# 510(k) Summary
# This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirements of SMDA 1990 and 21 CFR §807.92.
#### GENERAL INFORMATION I.
#### Establishment:
| Address: | Siemens AG, Medical Solutions<br>Henkestrasse 127<br>D-91052 Erlangen<br>Germany | | |
|----------------------|-----------------------------------------------------------------------------------------------|--|--|
| Registration Number: | 3002808157 | | |
| Contact Person: | Eva Reiter<br>Quality Manager<br>Telephone: +49 (9131) 84-2680<br>Telcfax: +49 (9131) 84-8812 | | |
### Device Name and Classification:
| Trade Name: | will be first marketed as SIENET Cosmos |
|-----------------------|---------------------------------------------|
| Classification Name: | Picture Archiving and Communications System |
| Classification Panel: | Radiology |
| CFR Section: | 21 CFR §892.2050 |
| Device Class: | Class II |
| Product Code: | LLZ |
## II. SAFETY AND EFFECTIVENESS INFORMATION SUPPORTING THE SUBSTANTIAL EQUIVALENCE DETERMINATION
#### Device Description and Intended Use: .
This premarket notification covers Siemens' cnhanced PACS system SIENET Cosmos.
SIENET Cosmos is a "software only"- solution, including a backend communication and storage component and three different workplace deployments for medical imaging tasks and applications.
SIENET Cosmos is the integrated radiology suite for large radiological practices and community hospitals. Important factors are the centralized server structure, the wideranging data distribution and the overall integrated concept, ranging from scheduling the examination to reporting and archiving as well as image and report distribution.
{1}------------------------------------------------
## SIENET Cosmos Integrated Workplaces / SIENET Cosmos Image Distribution
The three SIENET Cosmos workplace deployments ...
a) syngo® Viewing Studio - for image distribution (web-based viewing application not intended for primary diagnosis!)
b) syngo® Reporting Studio - for basic reporting, inside as well as outside of the radiology (standalone workstation)
c) syngo® Reporting Studio - Advanced Application Bundle - for use inside the radiology with advanced reporting functionality
... are medical diagnostic and viewing workstations intended for manipulating, reading, reporting, viewing and communicating / distributing of radiological softcopy images and so allows radiologists and radiological technicians to receive and process all data needed.
### SIENET Cosmos Image Data Management
... ensures all authorized personnel fast and continuous access to radiological data. It's main functionality ranges from availability of images having regard to data security, open interfaces, storage media, central system administration, back-up, software distribution to providing a flexible storage hierarchy.
The main purpose is storing and archiving of radiological softcopy images and structured (DICOM) reports.
### Integration: SIENET Cosmos Imaging Workflow Management
The Workflow Management enables by integration of any HL7- / DICOM-compatible RIS (IHF Year 5) to the SIENET Cosmos PACS a consistent workflow – from patient registration to requirement scheduling to a personal worklist and supports therefore reporting, documentation or administrative tasks.
### Technological Characteristics:
SIENET Cosmos is a "software only"-system, which will be delivered on CD-ROM / DVD and installed by Siemens service engineers. Hardware-Requirements to be met arc therefore defined.
The backend communication and storage solution (SDM) is based on the Solaris 8 operating system. The workplaces are based on Windows 2000 / Windows XP operating system.
- The herewith described SIENET Cosmos supports DICOM formatted images and objects.
#### General Safety and Effectiveness Concerns: =
The device labelling contains instructions for use and any necessary cautions and warning, to provide for safe and effective use of the device.
{2}------------------------------------------------
Risk management is ensured via a risk analysis, which is used to identify potential hazards. These potential hazards are controlled via software development, verification and validation testing. To minimize electrical, mechanical, and radiation hazards, Siemens adheres to recognized and established industry practice and standards.
#### . Substantial Equivalence:
The SIENET Cosmos, addressed in this premarket notification, is substantially equivalent to the following commercially available devices:
| SIENET Diagnostic Reporting Console (DRC) | K920309 |
|------------------------------------------------------|---------|
| SIENET Diagnostic Reporting Console VA02 Upgrade | K935694 |
| SIENET Teleradiology Product Line | K955394 |
| syngo® Multimodality Workstation | K010938 |
| SIENET MagicWeb (Webserver), SIENET MagicLink I | K973131 |
| SIENET Archive Server, ISA I, ISA II, Storage Server | K920310 |
The SIENET Cosmos described in this 510(k) has the same intended use and similar technical characteristics as the devices listed above together.
In summary, Siemens is of the opinion that SIENET Cosmos does not introduce any new potential safety risks and is substantially equivalent to and performs as well as the predicate devices.
{3}------------------------------------------------
Image /page/3/Picture/1 description: The image shows the logo for the Department of Health & Human Services USA. The logo consists of a stylized eagle with three stripes forming its body, and the text "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" arranged in a circular fashion around the eagle. The eagle is facing to the right.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
DEC 1 9 2003
Siemens AG Medical Solutions % Mr. Stefan Preiss Responsible Third Party Official TÜV America, Inc. TÜV Product Service 1775 Old Highway 8 NEW BRIGHTON MN 55112-1891 Re: K033831
Trade/Device Name: SIENET Cosmos Regulation Number: 21 CFR 892.2050 Regulation Name: Picture archiving and communications systems Regulatory Class: II Product Code: 90 LLZ Dated: November 28, 2003 Received: December 10, 2003
Dear Mr. Preiss:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the clectronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{4}------------------------------------------------
This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at one of the following numbers, based on the regulation number at the top of the letter:
| 8xx.1xxx | (301) 594-4591 |
|----------------------------------|----------------|
| 876.2xxx, 3xxx, 4xxx, 5xxx | (301) 594-4616 |
| 884.2xxx, 3xxx, 4xxx, 5xxx, 6xxx | (301) 594-4616 |
| 892.2xxx, 3xxx, 4xxx, 5xxx | (301) 594-4654 |
| Other | (301) 594-4692 |
Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97) you may obtain. Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html.
Sincerely yours,
Nancy C. Bugden
Nancy C. Brogdon Director, Division of Reproductive, Abdominal and Radiological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
### INDICATIONS FOR USE
| 510(k) Number (if known): | K033831 |
|---------------------------|---------------|
| Device Name: | SIENET Cosmos |
#### Indications For Use:
The SIENET Cosmos is a Picture Archiving and Communication System (PACS) intended to enhance the complete imaging workflow, i.e.
- Manipulating .
- . Reading
- . Reporting
- Vicwing .
- Communicating / distributing .
- Storing / archiving �
- of radiological softcopy images and
- Storing / archiving of structured (DICOM) reports. �
The system is a "software only" solution and is intended to assist the physician in diagnosis or treatment planning.
Therefore SIENET Cosmos supports the following generic imaging workflow:
- Receive scheduled exams from IS at the SIENET Cosmos archiving component SDM .
- Provide relevant prior exams and reports (Structured Reports only) to the Modalities and Workplaces .
- Receive and store new exams from the Modalities at the SDM .
- Prepare images for reading .
- Report new images, if required by comparing them with prior exams and reports .
- Demonstrate exams at Radiological Demos .
- Domonstante onations at Radiology (e.g. Surgery, Intensive Case Unit, wards, external � referring physicians).
Note:
The workstation deployment syngo Viewing Studio is not intended for primary diagnosis.
(Division Sign-Off)
Division of Reproductive, Abdominal,
and Radiological Devices K033831
510(k) Number
(Please do not write below this line - continue on another page if needed)
------------------------------------------------------------------------------------------------------------------------------------------------------------------------------Concurrence of the CDRH, Office of Device Evaluation (ODE)
Over-The-Counter Use_ OR Prescription Use (Per 21 CFR 801.109)
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.