K033768 · Altus Medical, Inc. · ILY · Feb 17, 2004 · Physical Medicine
Device Facts
Record ID
K033768
Device Name
ALTUS MEDICAL OPTIONAL INFRARED HANDPIECE
Applicant
Altus Medical, Inc.
Product Code
ILY · Physical Medicine
Decision Date
Feb 17, 2004
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 890.5500
Device Class
Class 2
Attributes
Therapeutic, 3rd-Party Reviewed
Indications for Use
The Altus Medical optional infrared handpiece is intended to emit energy in the infrared spectrum to provide topical heating for the purpose of elevating tissue temperature for the temporary relief of minor muscle and joint pain and stiffness, the temporary relief of minor joint pain associated with arthritis, the temporary increase in local circulation where applied, and the relaxation of muscles. In addition, the Altus Medical optional infrared handpiece may also help muscle spasms, minor sprains and strains, and minor muscular back pain.
Device Story
Optional infrared handpiece accessory for Altus Medical CoolGlide Xeo laser system. Device emits infrared energy to provide topical heating for pain relief and muscle relaxation. Handpiece includes IR quartz lamp, thermoelectric cooler, sapphire window, and umbilical cable connecting to laser console. User or field engineer attaches handpiece to console; system provides power, drive current, and water cooling to manage heat. Device identifies itself to laser system via internal memory component. Healthcare provider applies handpiece to patient skin; output elevates tissue temperature. Benefits include temporary relief of minor musculoskeletal pain, stiffness, and spasms, and increased local circulation.
Clinical Evidence
No clinical data. Substantial equivalence is based on technological comparison and bench testing of functional characteristics against predicate devices.
Technological Characteristics
Infrared lamp (21 CFR 890.5500). Features IR quartz lamp, thermoelectric cooler, and sapphire window. Includes umbilical cable for electrical power, detector signals, and water cooling lines. Memory device identifies handpiece to host laser system. Class II device.
Indications for Use
Indicated for patients requiring topical heating for temporary relief of minor muscle/joint pain and stiffness, arthritis-related joint pain, muscle spasms, minor sprains/strains, and minor muscular back pain, or for temporary increase in local circulation and muscle relaxation.
Regulatory Classification
Identification
An infrared lamp is a device intended for medical purposes that emits energy at infrared frequencies (approximately 700 nanometers to 50,000 nanometers) to provide topical heating.
Special Controls
*Classification.* Class II (special controls). The device, when it is an infrared therapeutic heating lamp, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 890.9.
BioFlex Professional Therapy System by Meditech International (K023621)
Thermapulse by Palomar Medical Technologies, Inc. (K023621)
Submission Summary (Full Text)
{0}------------------------------------------------
FEB 17 2004.
Premarket Notification 510(k) Submission Infrared Handpiece to be used with the previously cleared CoolGlide Aesthetic Laser
K033768
1 of 3
### Attachment 17 510(k) Summary for the Optional Infrared Handpiece
#### I. General Information
| Submitter: | Altus Medical, Inc.<br>821 Cowan Road<br>Burlingame, CA 94010 |
|---------------------------|---------------------------------------------------------------|
| Contact Person: | Kathy Maynor |
| Telephone: | 650-259-5586 |
| Fax: | 650-552-9787 |
| Summary Preparation Date: | November 18, 2003 |
#### II. Names
.
Altus Medical Optional Infrared Handpiece Device Proprietary Name: Infrared Lamp -- product code 89 ILY; class II; 890.5500 Primary Classification Name:
Infrared lamp Common Name:
#### III. Predicate Devices
- K001056 Olympic Warm-up by Olympic Medical ●
- K022609 PW820 Patient Warmer by Fisher & Paykel ●
- K023621 BioFlex Professional Therapy System by Meditech International .
- K024179 Thermapulse by Palomar Medical Technologies, Inc. .
#### Product Description/Technological Characteristics IV.
The optional infrared handpiece consists of:
- an "umbilical" cable and connector, that is permanently attached to the hand piece A body and is semi-permanently attached to the laser system (detachable by positive action from the user) that houses:
- electrical cables (to support the thermoelectric coolers associated with the chilled A sapphire window, to provide power and an electrical drive current to the IR quartz lamp, to provide detector signals and to connect a memory device that identifies the hand piece);
- a supply and return water line (to remove the heat generated by the infrared lamp and A thermoelectric cooler);
Premarket Notification 510(k) Submission
Infrared Handpiece to be used with the previously cleared CoolGlide Aesthetic Laser
{1}------------------------------------------------
K033768
2 of 3
- the hand piece internals described above; and A
- the hand piece shells housing the internals and connecting to the umbilical. A
The proximal end of the umbilical cable is semi-permanently attached to the laser system console and the distal end is permanently attached to the body of the delivery hand piece. This hand piece is removable by either the user or an authorized field service engineer for replacement at the proximal end.
This optional infrared handpiece is for use specifically with the Altus Medical CoolCilide Xeo laser system previously cleared on K023954.
#### V. Statement of Intended Use
The Altus Medical optional infrared handpiece is intended to emit energy in the infrared spectrum to provide topical heating for the purpose of elevating tissue temperature for the temporary relief of minor muscle and joint pain and stiffness, the temporary relief of minor joint pain associated with arthritis, the temporary increase in local circulation where applied, and the relaxation of muscles. In addition, the Altus Medical optional infrared handpiece may also help muscle spasms, minor sprains and strains, and minor muscular back pain.
#### Rationale for Substantial Equivalence VI.
The optional infrared handpiece shares the same general indications for use, and therefore is substantially equivalent to the currently marketed infrared heat lamps. The relevant technological characteristics of the Altus Medical optional infrared handpiece compared to the predicate devices include:
Lamp type: Equivalent to the Olympic Medical Warm-Up lamp from Olympic Medical Timer: Equivalent to the Thermapulse device from Palomar Technologies Total electrical wattage: Substantially equivalent to the Olympic Medical Warm-Up Lamp
Power: Equivalent to the Thermapulse device from Palomar Technologies
Wavelength: Substantially equivalent to the BioFlex Professional Therapy System from Meditech International and the PW820 Patient Warmer from Fisher & Paykel
Protective Enclosure: Equivalent to the Thermapulse from Palomar Technologies
Equivalent to the BioFlex Professional Therapy System from Meditech Spot Size: Interntional
Skin contact: Equivalent to the BioFlex Professional Therapy System
Cooling: Substantially equivalent to the PW820 Patient Warmer
Energy: Substantially equivalent to the Bioflex Professional Therapy System and the PW820 Patient Warmer
{2}------------------------------------------------
K033768
3 of 3
Electrical Requirements: Equivalent to the PW820 Patient Warmer the BioFlex Delivery device: Substantially equivalent to the Thermapulse and Professional Therapy System
#### Safety and Effectiveness Information VII.
Technologically, the optional infrared handpiece is substantially equivalent to the listed I colliologically, the options the risks and benefits for the CoolGlide optional infrared handpiece are comparable to the predicate devices.
We therefore believe that there are no new questions of safety or effectiveness raised by the introduction of this device.
### VIII. Conclusion
The optional infrared handpiece was found to be substantially equivalent to the currently marketed infrared heat lamps. The optional infrared handpiece shares similar indications marketu mirared neat lampe. The openting functional features as, and thus are substantially equivalent to, the currently marketed predicate devices.
{3}------------------------------------------------
Image /page/3/Picture/1 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo is circular and contains the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. In the center of the circle is an abstract symbol that resembles an eagle or other bird.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
FEB 1 7 2004
Altus Medical, Inc. c/o Mr. Morten S. Christensen Underwriters Laboratories, Inc. 1655 Scott Boulevard Santa Clara, California 95050
Re: K033768
Trade/Device Name: Optional Infrarcd Handpiece for CoolGlide Lasers Regulation Number: 21 CFR 890.5500 Regulation Name: Infrared lamp Regulatory Class: II Product Code: ILY Dated: January 30, 2004 Received: February 2, 2004
Dear Mr. Christensen:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Mcdical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adultcration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has madc a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CHR 1000-1050.
{4}------------------------------------------------
## Page 2 - Mr. Morten S. Christensen
This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (301) 594-4659. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html
Sincerely yours.
Miriam C. Provost
Celia M. Witten, Ph.D., M.D. Director Division of General, Restorative and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
# Attachment 2 Indications For Use Statement as Requested by FDA
510(k) Number (if Known): K033768
Device Name: Optional Infrared Handpiece for CoolGlide Lasers
Indications For Use:
The Altus Medical optional infrared handpiece is intended to emit energy in the infrared spectrum to provide topical heating for the purpose of elevating tissue temperature for the temporary relief of minor muscle and joint pain and stiffness, the temporary relief of minor joint pain associated with arthritis, the temporary increase in local circulation where applied, and the relaxation of muscles. In addition, the Altus Medical optional infrared handpiece may also help muscle spasms, minor sprains and strains, and minor muscular back pain.
# (PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Prescription Use
(Per 21 CFR 801.109)
in A. R.
OR
Over-The-Counter Use ... .........
(Optional Format 1-2-96
Division of General, Restorative, and Neurological Devices
**510(k) Number** K0633768
Premarket Notification 510(k) Submission Infrared Handpiece to be used with the previously cleared CoolGlide Aesthetic Laser
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.