K033767 · Jeil Medical Corporation · DZE · Feb 24, 2004 · Dental
Device Facts
Record ID
K033767
Device Name
DUAL TOP ANCHOR SYSTEM SCREWS
Applicant
Jeil Medical Corporation
Product Code
DZE · Dental
Decision Date
Feb 24, 2004
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 872.3640
Device Class
Class 2
Indications for Use
This device is intended to provide a fixed anchorage point for attachment of orthodontic appliances to facilitate the orthodontic movement of teeth. It is used temporarily and is removed after orthodontic treatment has been completed. Screws are intended for single use only. For use in adults over the age of 12.
Device Story
Dual Top Anchor System Screws are titanium alloy bone screws used in orthodontics to provide temporary fixed anchorage for appliances like brackets, wires, and elastic bands. The device features a dual-head design; the upper head is hex-shaped with a hole for wire attachment. The tip is designed for self-drilling and self-tapping at the surgeon's option. Screws range from 1.4mm to 2.0mm in diameter and 6.0mm to 12mm in length. The device is provided non-sterile and requires steam sterilization before use. It is intended for single-use only in adult patients. The clinician implants the screw into the jawbone to serve as a stable anchor point, allowing for controlled orthodontic tooth movement. Once treatment is complete, the screw is removed. The device benefits patients by providing a stable, temporary skeletal anchorage point, potentially improving the efficiency and outcomes of orthodontic tooth movement.
Clinical Evidence
No clinical data provided. Substantial equivalence is based on bench testing and comparison of technological characteristics and intended use to legally marketed predicate devices.
Technological Characteristics
Titanium alloy (ASTM F 136-98) bone screws. Dimensions: 1.4mm-2.0mm diameter, 6.0mm-12mm length. Dual-head design with hex interface and wire hole. Self-drilling/self-tapping tip. Non-sterile, requires steam sterilization. Mechanical device; no software or electronics.
Indications for Use
Indicated for adults over age 12 requiring fixed anchorage points for orthodontic appliances to facilitate tooth movement. Contraindicated for patients where temporary anchorage is not required or who are under 12 years of age.
Regulatory Classification
Identification
An endosseous dental implant is a prescription device made of a material such as titanium or titanium alloy that is intended to be surgically placed in the bone of the upper or lower jaw arches to provide support for prosthetic devices, such as artificial teeth, in order to restore a patient's chewing function.
Special Controls
*Classification.* (1) Class II (special controls). The device is classified as class II if it is a root-form endosseous dental implant. The root-form endosseous dental implant is characterized by four geometrically distinct types: Basket, screw, solid cylinder, and hollow cylinder. The guidance document entitled “Class II Special Controls Guidance Document: Root-Form Endosseous Dental Implants and Endosseous Dental Implant Abutments” will serve as the special control. (See § 872.1(e) for the availability of this guidance document.)(2)
*Classification.* Class II (special controls). The device is classified as class II if it is a blade-form endosseous dental implant. The special controls for this device are:(i) The design characteristics of the device must ensure that the geometry and material composition are consistent with the intended use;
(ii) Mechanical performance (fatigue) testing under simulated physiological conditions to demonstrate maximum load (endurance limit) when the device is subjected to compressive and shear loads;
(iii) Corrosion testing under simulated physiological conditions to demonstrate corrosion potential of each metal or alloy, couple potential for an assembled dissimilar metal implant system, and corrosion rate for an assembled dissimilar metal implant system;
(iv) The device must be demonstrated to be biocompatible;
(v) Sterility testing must demonstrate the sterility of the device;
(vi) Performance testing to evaluate the compatibility of the device in a magnetic resonance (MR) environment;
(vii) Labeling must include a clear description of the technological features, how the device should be used in patients, detailed surgical protocol and restoration procedures, relevant precautions and warnings based on the clinical use of the device, and qualifications and training requirements for device users including technicians and clinicians;
(viii) Patient labeling must contain a description of how the device works, how the device is placed, how the patient needs to care for the implant, possible adverse events and how to report any complications; and
(ix) Documented clinical experience must demonstrate safe and effective use and capture any adverse events observed during clinical use.
Nobel Biocare Implant Orthodontic Anchor System (K000643)
Submission Summary (Full Text)
{0}------------------------------------------------
KO 33767
FEB 2 4 2004
### EXHIBIT 2
Jeil Medical Corporation 808. ACE Techno-Tower, 3rd 197-48, Kuro 3Dong Kuro-Ku, Seoul, KOREA Telephone : 82-2-2109-6781~9 Fax: 82-2-2109-6791 Revised February 17, 2004 Contact: N.K.Kim, R&D Director 510(k) Summary of Safety and Effectiveness
- Identification of the Device: 1. Proprietary-Trade Name: Dual Top Anchor System Screws (various models) Classification Name: Implant, Endosseous, Product Code DZE Common/Usual Name: Bone Screw
- Equivalent legally marketed devices: OsteoMed Orthodontic Screw System, (K031936), 2. Equivancen legary manett(K982509) the Nobel Biocare Inplant Orthodontic Anchor System (K000643) ..
- Indications for Use (intended use) . This device is intended to provide a fixed anchorage point 3. for attachment of orthodontic appliances to facilitate the orthodontic movement of teeth. It is used temporarily and is removed after orthodontic treatment has been completed. Screws are intended for single use only. For use in adults over the age of 12.
- 4. Description of the Device: This device is designed for orthodontic use. The screws are made Description of alloy (ASTM F 136-98). The head of the screw is dual head type. The upper head is a hex type. There is a hole in the screw head through which a wire can be passed to fix the mandible and maxilla in orthodontic treatment. Also, Dual head design of screw mandible and mannia in the screw with the orthodontic appliances (bracket, wire, and elastic band etc.) The tip of the screw is designed available to self-drilling and self-tapping. (Surgeon's option) This device is comprised of screws ranging in diameter of 1.4mm to 2.0mm and in length ranging from 6.0mm to 12mm. These screws are provided non-sterile. Therefore, this device must be sterilized prior to use. Steam sterilization is recommended. Therefore, and device mast Polyethylene package. The top of packaging bag is sealed with sealing strip.
- Potential Adverse Affects and Complications: (Common to all devices of this type) 5.
- Poor bone formation, Osteoporosis, Osteolysis, Osteomyelitis, inhibited revascularization, . or oon can cause loosening, bending, cracking or fracture of the device or premature loss of fixation with the bone, leading to nonunion.
- Migration, bending, fracture or loosening of the implant. .
- Metal sensitivity, or allergic reaction to a foreign body. .
- Pain, discomfort, or abnormal sensation due to the presence of the device. .
- Increased fibrous tissue response around the fracture site and/or the implant. .
- Necrosis of bone. .
- . Inadequate healing.
- madequate neaming.
Apart from these adverse effects there are always possible complications of any surgical
{1}------------------------------------------------
procedure such as, but not limited to, infection, nerve damage, and pain which may not be related to the implant
| Device Name | Nobel Biocare<br>Implant Orthodontic<br>Anchor System | OsteoMed<br>Orthodontic Screw<br>System, | Dual Top Anchor System<br>Screws (various models) |
|-------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Device Classification<br>Name | Implant, Endosseous,<br>Product Code DZE | Implant, Endosseous,<br>Product Code DZE | Implant, Endosseous, Product<br>Code DZE |
| Applicant | Nobel BioCare | Osteomed | Jeil Medical Corporation |
| 510(K) Number | (K000643). | (K031936) | (This submission) |
| Material | Titanium | Titanium Alloy | Titanium Alloy<br>(ASTM F 136-98) |
| Intended use | This device is intended<br>to provide a fixed<br>anchorage point for<br>attachment of<br>orthodontic appliances<br>to facilitate the<br>orthodontic movement<br>of teeth. It is used<br>temporarily and is<br>removed after<br>orthodontic treatment<br>has been completed.<br>Screws are intended for<br>single use only. | This device is used as an<br>anchorage for orthodontic<br>treatment in the mouth. | This device is intended to<br>provide a fixed anchorage<br>point for attachment of<br>orthodontic appliances to<br>facilitate the orthodontic<br>movement of teeth. It is used<br>temporarily and is removed<br>after orthodontic treatment has<br>been completed. Screws are<br>intended for single use only.<br>For use in adults over the age<br>of 12. |
| Single Use? | YES | YES | YES |
| Supplied Sterile? | No comment | Non sterile, steam<br>sterilize before use | Non sterile, steam sterilize<br>before use |
- 6. Safety and Effectiveness, comparisons to predicate devices:
- Conclusion: In all respects, the Dual Top Anchor System Screws (various models) Bone 7. Conclusion: In an respects, the Butan of currently marked devices. They are materials Screws are the equivalent of earling markets as the consection adverse effects are identical to allu have similar difficits and enazacts manufactured from material of titanium alloy (ASTM 136-98) that is used generally in this kind of bone screw. Similar devices made from (ASTM 136-98) that is used generally in this never are manufactured and sold around the world.
titanium alloy (ASTM 136-98) to this device are manufactured and sold around to this device is substantially equivalent in design, material, intended use and function to the This device is substantially orquiral with the certified by notified body for CE.
{2}------------------------------------------------
Image /page/2/Picture/1 description: The image shows the seal of the Department of Health & Human Services - USA. The seal is circular, with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. In the center of the seal is an abstract symbol that resembles an eagle or other bird-like figure. The symbol is composed of three curved lines that suggest wings or feathers, and the overall design is simple and modern.
#### Public Health Service
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
FEB 2 4 2004
Jeil Medical Corporation C/O Mr. Daniel Kamm Kamm & Associates P.O. Box 7007 Deerfield, Illinois 60015
Re: K033767
Trade/Device Name: Jeli Medical Corporation: Dual Top Anchor Systems Screws Regulation Number: 21 CFR 872.3640 Regulation Name: Endosseous Implant Regulatory Class: III Product Code: DZE Dated: November 20, 2003 Received: December 3, 2003
Dear Mr. Kamm:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially cquivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general approvisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting (1 Mr), it hay of badyoo in the Code of Federal Regulations, Title 21, Parts 800 to 898. In your device ear or roundsh further announcements concerning your device in the Federal Register.
{3}------------------------------------------------
#### Page 2 - Mr. Kamm
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice una libing (21 CF C 1 forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), rt you deente the Office of Compliance at (301) 594-4613. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html
Sincerely yours,
Chiu Lin, Ph., D.
Director Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
# KO33767
## Indications for Use
510(k) Number (if known): ___K033767_
Device Name:_Jeil Medical Corporation:___Dual Top Anchor System Screws_
Indications For Use:
This device is intended to provide a fixed anchorage point for attachment of othodontic This device is intended to provide a liked anchorage point of used temoved appliances to faciliate the othroublike movement of tooth. We single use only. For use on adults over the age of 12.
Susan Russer
(Division Sign-Off) Division of Anesthesiology, General Hospital, Infection Control, Dental Devices
510(k) Number:________________________________________________________________________________________________________________________________________________________________
Prescription Use __ (Part 21 CFR 801 Subpart D) AND/OR
Over-The-Counter Use _ (21 CFR 807 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Page 1 of 1
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.