K033737 · Philips Medical Systems North America, Inc. · OWB · Dec 9, 2003 · Radiology
Device Facts
Record ID
K033737
Device Name
ALLURA XPER FD20
Applicant
Philips Medical Systems North America, Inc.
Product Code
OWB · Radiology
Decision Date
Dec 9, 2003
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 892.1650
Device Class
Class 2
Attributes
3rd-Party Reviewed
Indications for Use
The Allura Xper FD20 is intended for: - Dedicated vascular and neurovascular imaging applications, including diagnostic and interventional procedures. This includes, e.g., peripheral, cerebral, thoracic and abdominal angiography, as well as PTCAs, stent placements, embolisations and thrombolysis. - Cardiac imaging applications including diagnostics, interventional procedures (such as PTCA, stent placing, atherectomies), pacemaker implantations, and electrophysiology (EP). - Non-vascular interventions such as drainages, biopsies and vertebroplasties procedures.
Device Story
Allura Xper FD20 is a stationary fluoroscopic angiographic X-ray system. It captures X-ray images for visualization during vascular, neurovascular, cardiac, and non-vascular interventional procedures. Operated by clinicians in clinical settings (e.g., interventional suites), the system provides real-time imaging to guide procedures like stent placement, biopsies, and electrophysiology. Output is displayed for the physician to monitor anatomy and device placement, facilitating minimally invasive interventions and improving procedural accuracy.
Clinical Evidence
No clinical data provided; device relies on bench testing and compliance with established radiological and safety standards.
Technological Characteristics
Stationary fluoroscopic X-ray system. Complies with 21 CFR Subchapter J (radiological health) and UL 2601-1 (safety). Supports ACR/NEMA DICOM standards for digital imaging communication.
Indications for Use
Indicated for patients requiring diagnostic or interventional vascular, neurovascular, cardiac, or non-vascular imaging procedures, including angiography, stent placement, embolization, thrombolysis, atherectomy, pacemaker implantation, electrophysiology, drainage, biopsy, and vertebroplasty.
Regulatory Classification
Identification
An image-intensified fluoroscopic x-ray system is a device intended to visualize anatomical structures by converting a pattern of x-radiation into a visible image through electronic amplification. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.
Special Controls
*Classification.* Class II (special controls). An anthrogram tray or radiology dental tray intended for use with an image-intensified fluoroscopic x-ray system only is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 892.9. In addition, when intended as an accessory to the device described in paragraph (a) of this section, the fluoroscopic compression device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 892.9.
{0}------------------------------------------------
DEC - 9 2003
Image /page/0/Picture/1 description: The image shows the Philips logo, which consists of the word "PHILIPS" at the top of a shield-shaped emblem. Below the text, there is a stylized globe with wavy lines representing water and curved lines representing land masses. The logo is in black and white.
Image /page/0/Picture/2 description: The image shows the word "PHILIPS" in a bold, sans-serif font. The letters are large and black, contrasting with the white background. The text appears to be a logo or brand name, likely for the electronics company Philips.
## Philips Medical Systems Nederland B.V.
# 510(k) Summary
Ko33737
The following information is being submitted in accordance with the requirements of 21 CFR 807.92.
| Company Name: | Philips Medical Systems North America Company |
|-----------------------------------|------------------------------------------------------------------------------|
| Address: | 22100 Bothell Everett Highway<br>P.O.Box 3003<br>Bothell, WA 98041-3003, USA |
| Registration No.: | 1217116 |
| Contact Person:<br>Telephone No.: | Lynn Harmer<br>(425) 487-7312 |
| Date Prepared: | November 18, 2003 |
| Device (Trade) Name: | Allura Xper FD20 |
| Classification Names: | Stationary X-ray system,<br>21CFR892.1680, Class II (code 90KPR) |
| | Angiographic X-ray system,<br>21CFR892.1600, Class II (code 90IZI) |
#### Predicate Device:
The Allura Xper FD20 is substantially equivalent to the Integris Allura, both manufactured by Philips Medical Systems. The Philips Integris Allura received a 510(k) substantially equivalent determination in K002016 on September 6, 2000.
#### Device description:
The Allura Xper FD20 is a stationary fluoroscopic angiographic X-ray system for vascular, neurovascular and cardiovascular procedures, as well as non-vascular procedures.
#### Intended use:
The Allura Xper FD20 is intended for:
- Dedicated vascular and neurovascular imaging applications, including diagnostic and . interventional procedures. This includes, e.g., peripheral, cerebral, thoracic and abdominal angiography, as well as PTCAs, stent placements, embolisations and thrombolysis.
- Cardiac imaging applications including diagnostics, interventional procedures (such . as PTCA, stent placing, atherectomies), pacemaker implantations, and electrophysiology (EP).
- Non-vascular interventions such as drainages, biopsies and vertebroplasties . procedures.
Document identification: XDB087-030820/RVI/vi
{1}------------------------------------------------
Image /page/1/Picture/0 description: The image shows the Philips logo. The logo consists of a shield-like emblem on the left and the word "PHILIPS" in bold, uppercase letters on the right. The emblem features a stylized representation of stars and waves, enclosed within the shield shape.
### General Safety and Effectiveness Information:
The device and their labeling will comply with the applicable requirements of:
- 21CFR, Subchapter J Radiological Health, parts 1010, 1020.30, 1020.32 & . 1040.10.
- Underwriters Laboratories Standard for Safety UL 2601-1 and be classified by � Underwriters Laboratories.
- ACR/NEMA DICOM digital imaging communication standard. .
#### Conclusion:
The Allura Xper FD20 does not introduce any new indications for use, nor does the use of the device result in any new potential hazard. Philips Medical Systems considers the Allura Xper FD20 to be substantially equivalent with the predicate device.
{2}------------------------------------------------
Image /page/2/Picture/0 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized caduceus symbol, which features a staff with a snake winding around it. The symbol is encircled by the text "DEPARTMENT OF HEALTH & HUMAN SERVICES USA". The text is arranged in a circular fashion around the caduceus symbol.
DEPARTMENT OF HEALTH & HUMAN SERVICES
Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
Philips Medical Systems North America Company c/o Mr. John C. So Senior Project Engineer Underwriters Laboratories, Inc. 6200 NW Lake Rd. Camas, WA 98607
JUL 1 9 2013
Re: K033737
Trade/Device Name: Allura Xper FD20 Regulation Number: 21 CFR 892.1650 Regulation Name: Intensified fluoroscopic x-ray system Regulatory Class: Class II Product Code: OWB. JAA, 1ZI Dated: November 26, 2003 Received: November 28, 2003
Dear Mr. So:
This letter corrects our substantially equivalent letter of December 9, 2003.
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA 's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must
{3}------------------------------------------------
Page 2 - J. So
comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act): 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Small Manufacturers, International and Consumer Assistance at its tollfree number (800) 638 2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safetv/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely vours.
Jaming M. Mense
Janine M. Morris Director Division of Radiological Health Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health
{4}------------------------------------------------
510(k) Number (if known): ___ バロタ:37 3 ア
Device name: Allura Xper FD20
Indications for use:
The Allura Xper FD20 is intended for:
- · Dedicated vascular and neurovascular imaging applications, including diagnostic and interventional procedures. This includes, " e.g., peripheral, cerebral, thoracic and abdominal angiography, as well as PTCAs, stent placements, embolisations and thrombolysis.
- · Cardiac imaging applications including diagnostics, interventional procedures (such as PTCA, stent placing, atherectomies), pacemaker implantations, and electrophysiology (EP),
- Non-vascular interventions such as drainages, biopsies and . vertebroplasties procedures.
Nancy C. Broughton
(Division Sign-Off)
Division of Reproductive, Abdominal,
and Radiological Devices
510(k) Number K033737
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
## Concurrence of CDRH, Office of Device Evaluation (ODE)
Prescription Use
(Per 21CFR801.109)
Use
OR
Over-The counter
Rel. nr. XDB087-030873/AVVrvi
3
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.